Validity of a new contact mat system for evaluating vertical jump

The purpose of this study was to verify the validity of the flight and ground contact time measurement of a contact mat, SaltoBras (SB), comparing it to an oscilloscope (OS) and to a force plate (FP). For that, the SB was placed upon a FP. At first, four male and two female performed 15 jumps on the SB and the flight times were obtained. After, they performed six consecutive jumps and the ground contact times were obtained. SB software was valid in comparison with the OS, but underestimated the flight (~0.30 %) and ground contact (~1.01 %) time measurements when comparing it to the FP. Despite the differences, the error found between SB and FP was systematic, and two prediction equations were defined and added to the software for correction. The results suggest that SB is a valid instrument for the evaluation of vertical jump.

Instrumentation; Validity of tests; Biomechanics