Acessibilidade / Reportar erro

Influence of biomaterials on scintigraphic diagnosis of periprosthetic infections. Ceftizoxime-99m technetium model 1 1 Research performed at Laboratory of Experimental Surgery, School of Medicine, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte-MG, Brazil.

Abstract

Purpose:

To compare the influence of two metallic implants in the diagnosis of periprosthetic infection using 99m technetium-labeled ceftizoxime.

Methods:

Twenty rats were randomly divided into four groups, which received sterile and contaminated titanium and stainless steel implants. After 3 weeks, scintilographic images were obtained using a gamma chamber. Radioactivity counts were obtained for the region of interest (ROI) on the operated and non-operated paws.

Results:

Groups A, B, and C showed homogenous distribution of the radiopharmaceutical. Hyper uptake was observed in the operated paw from group D. The ROI target count was higher in the two groups with stainless steel implants. Among the control groups, the count was higher in the stainless steel group. Furthermore, among the contaminated groups, the uptake was higher in the stainless steel group, with a significant difference. The target: non-target ratio was significantly lower in the control and contaminated groups with both titanium and stainless steel, but the comparison between control groups and contaminated groups was only significant in the former. The cpm/g observed after a decay of 48h showed statistically significant differences between groups.

Conclusion:

Different biomaterials used in implants have an influence on the results of scintigraphy with 99mTc-CFT.

Key words:
Biocompatible Materials; Radionuclide Imaging; Prosthesis-Related Infections; Rats

Introduction

In recent decades, there has been an increase in the use of implants and biomaterials in different areas of medicine. At the beginning of the 1990s, in the United States of America (USA), approximately 4.4 million people had received some type of implant for fracture fixation and 1.3 million had an articular prosthesis11. Praemer A, Furner S, Rice DP. Musculoskeletal conditions in the United States. American Academy of Orthopaedic Surgeons. 1ed. Park Ridge; 1992.. Although these biomaterials represented an advancement in the treatment of articular diseases and fractures, the following two factors limit their application: biocompatibility and infections.

Infections represent the second most common complication related to the use of orthopedic implants, with an incidence of 1%-5.6% in arthroplasties22. Duff GP, Lachiewicz PF, Kelley SS. Aspiration of the knee joint before revision arthroplasty. Clin Orthop Relat Res. 1996;331:132-9. PMID: 8895629.

3. Hanssen AD, Rand JA. Evaluation and treatment of infection at the site of a total hip or knee arthroplasty. Instr Course Lect. 1999;48:111-22. PMID: 10098033.
-44. Peersman G, Laskin R, Davis J, Peterson M. Infection in total knee replacement: a retrospective review of 6489 total knee replacements. Clin Orthop Relat Res. 2001;392:15-23. PMID: 11716377.. With respect to fractures, implants removed from patients presented a colonization rate of 50%, regardless of the clinical symptoms55. Moussa FW, Anglen JO, Gehrke JC, Christensen G, Simpson WA. The significance of positive cultures from orthopedic fixation devices in the absence of clinical infection. Am J Orthop. 1997;26:617-20. PMID: 9316724.. Bacterial adherence and formation of biofilms on the surface of the prostheses are key steps in the pathogenesis of these infections, making them resistant to the host defense mechanisms and the action of antimicrobial agents66. Bauer TW, Parvizi J, Kobayashi N, Krebs V. Diagnosis of periprosthetic infection. J Bone Joint Surg Am. 2006;88:869-82. PMID: 16595481.

7. Ehrlich GD, Stooley P, Kathju S, Zhao Y, McLeod BR, Balaban N, Hu FZ, Sotereanos NG, Costerton JW, Stewart PS, Post JC, Lin Q. Engineering approach for the detection and control of biofilm orthopedic infections. Clin Orthop Rel Res. 2005;437:59-66. PMID: 16056027.
-88. Gristina AG, Costerton JW. Bacterial adherence to biomaterials and tissue: The significance of its role in clinical sepsis. J Bone Joint Surg. 1985;67:264-73. PMID: 3881449..

In addition to treatment resistance, the diagnosis of periprosthetic infections also becomes difficult. Studies have shown that the formation of biofilms reduces the sensitivity and specificity of routine microbiological methods, but no standard test has been established that can be applied in clinical practice66. Bauer TW, Parvizi J, Kobayashi N, Krebs V. Diagnosis of periprosthetic infection. J Bone Joint Surg Am. 2006;88:869-82. PMID: 16595481.,99. Tunney MM, Patrick S, Curran MD, Ramage G, Hanna D, Nixon JR, Gorman SP, Davis RI, Anderson N. Detection of prosthetic hip infection at revision arthroplasty by immunofluorescence microscopy and PCR amplification of the bacterial 16S rRNA gene. J Clin Microbiol. 1999;37:3281-90. PMID: 10488193..

Scintigraphic methods are often used. Gomes-Barreto et al.10 described the use of a third-generation cephalosporin called ceftizoxime (Cefizox®), labeled with 99m-technetium (99mTc-CFT), which showed good results in detecting soft tissue and bone infections1010. Gomes-Barreto V, Iglesias F, Roca M, Tubau F, Martin-Comin J. Labelling of ceftizoxime with 99mTC. Rev Esp Med Nuclear. 2000;19:479-83. PMID: 11171504.,1111. Diniz SOF, Rezende CMF, Serakides R, Ferreira RLB, Ribeiro TG, Martin-Comin J, Cardoso VN. Scintigraphic imaging using technetium-99m-labeled ceftizoxime in an experimental modelo of acute osteomyelitis in rats. Nucl Med Comm. 2008;29:830-6. doi: 10.1097/MNM.0b013e3283000514.
https://doi.org/10.1097/MNM.0b013e328300...
. However, for peri-implant infections, the intrinsic characteristics of the different biomaterials used can influence the effectiveness of scintigraphy because they directly interfere with bacterial adherence.

This study aims to compare the influence of the composition of two metallic implants (titanium and stainless steel) in the scintigraphic diagnosis of periprosthetic infections, using 99mTc-CFT as a radiopharmaceutical.

Methods

This study was approved by the ethics committee in animal experimentation of the Universidade Federal de Minas Gerais (protocol 150/2011).

Twenty 3-month-old Wistar rats, weighing an average of 322.5 g (230-390g), were used. Animals were housed in the vivarium at the UFMG Medical Faculty, with five animals per cage. The cages were kept in a temperature-controlled, ventilated area, with water and food ad libitum and natural day-night cycle and were monitored daily.

The sample size for the four groups was calculated using the software Minitab® version 14.1, by the ANOVA method, considering alpha = 0.05, test power > 0.90, and maximum deviation between the results equal to the value of the average. This resulted in five animals per group. The mean ± standard deviation (SD) values were established after completion of the pilot project.

Experimental design and technical procedures

Animals were randomly divided into four groups: group A (n = 5), animals with sterile titanium implants; group B (n = 5), those with contaminated titanium implants; group C (n = 5), those with sterile stainless steel implants; and group D (n = 5), those with contaminated stainless steel implants.

Animals were anesthetized and kept under sedation by administration of an intra-peritoneal injection of xylazine/ketamine solution at a dose of 60mg/kg - 15 mg/kg. After the trichotomy of the left posterior paw was completed, it was disinfected with polyvinylpyrrolidone-iodine (Riodine®) in an antiseptic solution, followed by an alcohol-based solution, and protection with sterile fields. A longitudinal incision was formed at the thigh, dissecting the layers to expose the posterior diaphyseal region of the femur.

The metal implant was aseptically placed in direct contact with the bone in groups A and C and was placed after contamination in groups B and D. The suture was performed using 4-0 monofilament nonabsorbable nylon wire (mononylon®). After 3 weeks, the animals were transported to the Laboratory of radioisotopes for diagnostic examinations.

Biomaterials

Metallic implants of titanium (Baumer®, São Paulo) and stainless steel (Baumer®, São Paulo) were used. The materials were constructed as cylinders (10 mm in length and 1.5 mm in diameter), with a smooth and polished surface, sterilized by ethylene oxide, and inserted using aseptic techniques.

Implant contamination

For contamination of the implants, multi-sensitive strains of Staphylococcus aureus (ATCC6538P) were used. Bacterial colonies were cultured in blood agar, suspended in saline solution, and a concentration of 1×1099. Tunney MM, Patrick S, Curran MD, Ramage G, Hanna D, Nixon JR, Gorman SP, Davis RI, Anderson N. Detection of prosthetic hip infection at revision arthroplasty by immunofluorescence microscopy and PCR amplification of the bacterial 16S rRNA gene. J Clin Microbiol. 1999;37:3281-90. PMID: 10488193. CFU/mL was adjusted by nephelometry. The implants used in the animals of groups B and D were immersed in this suspension for 5 min before being implanted.

Scintigraphic method

Ceftizoxime (Cefizox®, Smith-kline, Lisbon) was labeled using a 1.2 ml sodium pertechnetate solution (Na99mTcO4-), with a radioactivity of 481 MBq (IPEN, Brazil). The radiopharmaceutical was incubated for 10 min in a boiling bath (100°C) and cooled under running water for 5 min. After cooling, the solution was filtered through a 0.22-μm cellulose ester membrane and collected in a vacuum flask. The complete CTF labeling procedure was performed according to the procedure followed by Diniz et al.33. Hanssen AD, Rand JA. Evaluation and treatment of infection at the site of a total hip or knee arthroplasty. Instr Course Lect. 1999;48:111-22. PMID: 10098033.

After labeling ceftizoxime with 99m-technetium (99mTc-CFT), ampules containing approximately 0.1ml (62.9 MBq) of the solution were subjected to radioactivity count using a well counter (Wizard®, Turku/Finland) and then administrated into the tail vein of anesthetized animals.

Scintigraphy was performed after 390 min, using a gamma chamber with a low-energy collimator for the acquisition of layered static images over a 10-min period, using a matrix of 256 × 256 pixels (Nuclide® TH 22, Mediso, Hungary).

Radioactivity counts were performed in the region of interest (ROI) of the operated paw (target area) and non-operated paw (non-target area) to determine the total concentration and target/not target ratio (Figure 1).

Figure 1
Radioactivity counts of the region of interest (ROI) and illustration of the counted area in the contralateral region (non-target).

After the radioactivity counts were performed, animals were euthanized by anesthetic overdose, and the implant was removed and incubated in Trypticase soy broth. The paw was disarticulated and kept under refrigeration for calculating the radioactivity count/minute per gram (cpm/g) of the tissue after 48 h of decay.

Results were analyzed by subjective evaluation of the images, separating them into homogeneous or heterogeneous. The quantitative variables considered were radioactivity count of ROI, ratio between the operated and non-operated paws (ROI target/non-target), and cpm/g of the tissue after 48h of decay of radioactivity. The comparison between groups with the same implant (A/B) and (C/D) determined the effectiveness of test in detecting the infection. The aim of the comparative analysis between control groups (A/C) and (B/D) was to evaluate the influence of the type of implant on the effectiveness of the method.

All quantitative results were presented as mean ± SD. Data were compared by the Kruskal-Wallis test, using SPSS® for Windows, version 17.0 (Chicago, United States). A confidence interval of 95% was used, and the differences were considered statistically significant when p value was less than or equal to 0.05 (p≤0.05).

Results

In the animals of the control groups, all implant cultures were negative after incubation for 5 days. All samples from the contaminated groups showed bacterial growth.

In the evaluation of the images, groups A, B, and C showed a homogeneous distribution of the radiopharmaceutical. Thus, it was not possible to differentiate between operated and non-operated groups (Figure 2). In all animals of group D, evident hyper-uptake was observed in the operated paw, both in the early and later images (Figure 3).

Figure 2
Homogeneous pattern of representative images of animals from groups A, B, and C 390 min after an injection of 99mTc-CFT.

Figure 3
Heterogeneous pattern of scintigraphic images of animals of group D, 390 min after an injection of the radiopharmaceutical.

The radioactivity mean counts of the operated paw (ROI target), non-target paw, and the target/non-target ratio are represented in Table 1.

Table 1
Average total radioactivity count of the region of interest (ROI), contralateral (non-target ROI) region, and target/non-target ratio 390 min after an injection of 99mTc-CFT.

The ROI target count was higher in the two groups that used stainless steel implants than that in the groups that used titanium implants. After 390 min, the counts between the control groups (A/C) were higher in group C, with a significant difference (p = 0.016). In the contaminated groups (B/D), group D had higher counts than those of group B (p = 0.041) (Figure 4).

Figure 4
Comparison between mean total count of the region of interest of the target paw, 390 min after injection of 99m Tc-CFT.

The target/non-target ratio was significantly lower in group A than in group B (p = 0.038). Similar results were observed in groups C and D that had stainless steel implants (p = 0.048). However, comparing the control (A and C) and contaminated groups (B and D), the difference was significant only in the control groups (p = 0.001). Similar results were not observed in the contaminated groups (p = 0.348). The target/non-target ratio averages for the groups studied are shown in Figure 5.

Figure 5
Mean ratio for target and non-target region counts by animal groups (target/non-target ROI).

On comparing contaminated and control groups, Cpm/g observed after 48 h of decay showed statistically significant differences between the groups. Among the groups with titanium implants, group A presented an average of 18077.07 ± 4948.95 and group B presented an average of 44983.32 ± 15575.87 (p = 0.006). Within the group with stainless steel implants, group C had a mean of 55168 ± 2944.58 and group D had a mean of 93735.54 ± 35523.58 (p = 0.028).

Discussion

The main characteristic of periprosthetic infections is the formation of biofilms, which allows the confinement of organized bacterial colonies that are protected against the actions of host defense and antimicrobial agents33. Hanssen AD, Rand JA. Evaluation and treatment of infection at the site of a total hip or knee arthroplasty. Instr Course Lect. 1999;48:111-22. PMID: 10098033.,88. Gristina AG, Costerton JW. Bacterial adherence to biomaterials and tissue: The significance of its role in clinical sepsis. J Bone Joint Surg. 1985;67:264-73. PMID: 3881449.,1313. Vuong C, Gerke C, Somerville GA, Fischer ER, Otto M. Quorum-sensing control of biofilm factors in Staphylococcus epidermidis. J Infect Dis. 2003;188:706-18. PMID: 12934187.. This process forms a complex structure that makes the treatment, diagnosis of infections, and/or isolation of the pathogen by traditional microbiological methods difficult. Implant composition also changes the local microenvironment, leading to foreign body-type reactions that can influence diagnostic methods.

Scintigraphy is a widely used method that is easy to apply in clinical practice and has high sensitivity. Most authors agree that a scintigraphy without uptake excludes a periprosthetic infection. However, the significance of a local uptake is still uncertain because its ability to differentiate between inflammatory infections is low. In our study, we used a third-generation cephalosporin as the radiopharmaceutical that is active against gram-positive, gram-negative, and anaerobic bacteria. Moreover, its application as a radiopharmaceutical has shown to be effective for soft tissue infections, osteomyelitis, and prosthetic infections1010. Gomes-Barreto V, Iglesias F, Roca M, Tubau F, Martin-Comin J. Labelling of ceftizoxime with 99mTC. Rev Esp Med Nuclear. 2000;19:479-83. PMID: 11171504.,1111. Diniz SOF, Rezende CMF, Serakides R, Ferreira RLB, Ribeiro TG, Martin-Comin J, Cardoso VN. Scintigraphic imaging using technetium-99m-labeled ceftizoxime in an experimental modelo of acute osteomyelitis in rats. Nucl Med Comm. 2008;29:830-6. doi: 10.1097/MNM.0b013e3283000514.
https://doi.org/10.1097/MNM.0b013e328300...
,1212. Diniz SOF, Ferrari CS, Nelson DL, Martin-Comín J, Cardoso VN. 99mTechnetium-ceftizoxime kit preparation. Braz Arch BioTechnol. 2005;48:89-96. doi: 10.1590/S1516-89132005000700014.
https://doi.org/10.1590/S1516-8913200500...
,1414. Gomes Barreto V, Rabiller G, Iglesias F, Soroa V, Tubau F, Roca M, Martín-Comín J. 99mTc-ceftizoxime scintigraphy in normal rats and abscess induced rats. Rev Esp Med Nucl. 2005;24:312-8. PMID: 16194463.,1515. Teixeira LE, Soares GG, Teixeira HC, Takenaka IK, Diniz SO, de Andrade MAP, Cardoso VN, de Araújo ID. Efficacy of (99m)Tc-labeled ceftizoxime in the diagnosis of subclinical infections associated with titanium implants in rats. Surg Infect. 2015;16:352-7. PMID: 26046250..

Comparison between the control and contaminated groups showed that the method is useful for diagnosing infections, regardless of the composition of the periprosthetic metal used. However, we observed two patterns of results that were determined by the type of material used. In the group with titanium implants, the radioactive counts were lower, with less variability and less significant values between the contaminated and control groups. The images were also similar between these groups. In the groups that used stainless steel implants, the counts showed higher values, with a great variability in the counts, and all values were lower in the control group. The visual images also differed, which suggests the presence of an infection.

This difference can be explained by the interaction between the implant and host. Of the metals, titanium alloys are less prone to infections compared with stainless steel alloys and chromium cobalt1616. Cordero J, Manuera L, Folgueira MD. Influence of bacterial strains on bone infection. J Orthop Res. 1996;14:663-7. PMID: 8764878.

17. Petty W, Spanier S, Schuster JJ, Silverthorne C. The influence of skeletal implants on incidence of infection. J Bone Joint Surg. 1985;67:1236-44. PMID: 3902846.
-1818. Speranza G, Gottardi G, Pederzolli C, Lunelli L, Canteri R, Pasquardini I, Carli E, Lui A, Maniglio D, Brugnara M, Anderle M. Role of chemical interactions in bacterial adhesion to polymer surfaces. Biomaterials. 2004;25:2029-37. PMID: 14741617.. This is caused by better biointegration of titanium, which are more inert when implanted in the human tissue1919. Spangehl MJ, Younger ASE, Marsi BA, Duncan CP. Diagnosis of infection following total hip arthroplasty. J Bone Joint Surg. 1997;79:1578-88. PMID: 9571430.. Alloys, such as stainless steel, have low integration with host cells, resulting in more free energy on their surface, which allows greater bacterial colonization1919. Spangehl MJ, Younger ASE, Marsi BA, Duncan CP. Diagnosis of infection following total hip arthroplasty. J Bone Joint Surg. 1997;79:1578-88. PMID: 9571430.. This may also explain the higher uptake in the stainless steel control group than in titanium control group because stainless steel results in a local inflammatory reaction that could modify the uptake or promote greater bacterial adherence.

On evaluating the group with titanium implants, the images did not clearly detect the presence of an infection. According to the study conducted by Phillips2020. Phillips WT. Delivery of gamma-imaging agents by liposomes. Adv Drug Deliv Rev. 1999;37:13-32. PMID: 10837724., the target/non-target ratio must present minimum values of 1.5, i.e., 50% more activity in the target compared with that in the contralateral region (non-target), for the images to differentiate local uptake. As the value of the target/non-target ratio observed in the groups with titanium implants was only 1.6, it was difficult to produce clear images that would guarantee the presence of infection. Thus, in the case of negative images, the examination should be based on the local radioactivity count and its relationship with the contralateral side.

Therefore, in our study, an image with evident hyper-uptake confirms the presence of infection. However, if the images are negative, the target/non-target ratio should be considered. Studies in humans and with a larger sample size are needed to determine the cut-off value that indicates infection in examinations with homogeneous images. A recent study suggests that the association of 99mTc-CFT with liposomes can increase its effectiveness2121. Ferreira SM, Domingos GP, Ferreira Ddos S, Rocha TG, Serakides R, de Faria Rezende CM, Cardoso VN, Fernandes SO, Oliveira MC. Technetium-99m-labeled ceftizoxime loaded long-circulating and pH-sensitive liposomes used to identify osteomyelitis. Bioorg Med Chem Lett. 2012;22:4605-8. PMID: 22726927..

Conclusion

Scintigraphy with 99mTc-CFT was effective in the diagnosis of periprosthetic infections; however, different biomaterials used in implants have an influence on the outcome.

References

  • 1
    Praemer A, Furner S, Rice DP. Musculoskeletal conditions in the United States. American Academy of Orthopaedic Surgeons. 1ed. Park Ridge; 1992.
  • 2
    Duff GP, Lachiewicz PF, Kelley SS. Aspiration of the knee joint before revision arthroplasty. Clin Orthop Relat Res. 1996;331:132-9. PMID: 8895629.
  • 3
    Hanssen AD, Rand JA. Evaluation and treatment of infection at the site of a total hip or knee arthroplasty. Instr Course Lect. 1999;48:111-22. PMID: 10098033.
  • 4
    Peersman G, Laskin R, Davis J, Peterson M. Infection in total knee replacement: a retrospective review of 6489 total knee replacements. Clin Orthop Relat Res. 2001;392:15-23. PMID: 11716377.
  • 5
    Moussa FW, Anglen JO, Gehrke JC, Christensen G, Simpson WA. The significance of positive cultures from orthopedic fixation devices in the absence of clinical infection. Am J Orthop. 1997;26:617-20. PMID: 9316724.
  • 6
    Bauer TW, Parvizi J, Kobayashi N, Krebs V. Diagnosis of periprosthetic infection. J Bone Joint Surg Am. 2006;88:869-82. PMID: 16595481.
  • 7
    Ehrlich GD, Stooley P, Kathju S, Zhao Y, McLeod BR, Balaban N, Hu FZ, Sotereanos NG, Costerton JW, Stewart PS, Post JC, Lin Q. Engineering approach for the detection and control of biofilm orthopedic infections. Clin Orthop Rel Res. 2005;437:59-66. PMID: 16056027.
  • 8
    Gristina AG, Costerton JW. Bacterial adherence to biomaterials and tissue: The significance of its role in clinical sepsis. J Bone Joint Surg. 1985;67:264-73. PMID: 3881449.
  • 9
    Tunney MM, Patrick S, Curran MD, Ramage G, Hanna D, Nixon JR, Gorman SP, Davis RI, Anderson N. Detection of prosthetic hip infection at revision arthroplasty by immunofluorescence microscopy and PCR amplification of the bacterial 16S rRNA gene. J Clin Microbiol. 1999;37:3281-90. PMID: 10488193.
  • 10
    Gomes-Barreto V, Iglesias F, Roca M, Tubau F, Martin-Comin J. Labelling of ceftizoxime with 99mTC. Rev Esp Med Nuclear. 2000;19:479-83. PMID: 11171504.
  • 11
    Diniz SOF, Rezende CMF, Serakides R, Ferreira RLB, Ribeiro TG, Martin-Comin J, Cardoso VN. Scintigraphic imaging using technetium-99m-labeled ceftizoxime in an experimental modelo of acute osteomyelitis in rats. Nucl Med Comm. 2008;29:830-6. doi: 10.1097/MNM.0b013e3283000514.
    » https://doi.org/10.1097/MNM.0b013e3283000514
  • 12
    Diniz SOF, Ferrari CS, Nelson DL, Martin-Comín J, Cardoso VN. 99mTechnetium-ceftizoxime kit preparation. Braz Arch BioTechnol. 2005;48:89-96. doi: 10.1590/S1516-89132005000700014.
    » https://doi.org/10.1590/S1516-89132005000700014
  • 13
    Vuong C, Gerke C, Somerville GA, Fischer ER, Otto M. Quorum-sensing control of biofilm factors in Staphylococcus epidermidis. J Infect Dis. 2003;188:706-18. PMID: 12934187.
  • 14
    Gomes Barreto V, Rabiller G, Iglesias F, Soroa V, Tubau F, Roca M, Martín-Comín J. 99mTc-ceftizoxime scintigraphy in normal rats and abscess induced rats. Rev Esp Med Nucl. 2005;24:312-8. PMID: 16194463.
  • 15
    Teixeira LE, Soares GG, Teixeira HC, Takenaka IK, Diniz SO, de Andrade MAP, Cardoso VN, de Araújo ID. Efficacy of (99m)Tc-labeled ceftizoxime in the diagnosis of subclinical infections associated with titanium implants in rats. Surg Infect. 2015;16:352-7. PMID: 26046250.
  • 16
    Cordero J, Manuera L, Folgueira MD. Influence of bacterial strains on bone infection. J Orthop Res. 1996;14:663-7. PMID: 8764878.
  • 17
    Petty W, Spanier S, Schuster JJ, Silverthorne C. The influence of skeletal implants on incidence of infection. J Bone Joint Surg. 1985;67:1236-44. PMID: 3902846.
  • 18
    Speranza G, Gottardi G, Pederzolli C, Lunelli L, Canteri R, Pasquardini I, Carli E, Lui A, Maniglio D, Brugnara M, Anderle M. Role of chemical interactions in bacterial adhesion to polymer surfaces. Biomaterials. 2004;25:2029-37. PMID: 14741617.
  • 19
    Spangehl MJ, Younger ASE, Marsi BA, Duncan CP. Diagnosis of infection following total hip arthroplasty. J Bone Joint Surg. 1997;79:1578-88. PMID: 9571430.
  • 20
    Phillips WT. Delivery of gamma-imaging agents by liposomes. Adv Drug Deliv Rev. 1999;37:13-32. PMID: 10837724.
  • 21
    Ferreira SM, Domingos GP, Ferreira Ddos S, Rocha TG, Serakides R, de Faria Rezende CM, Cardoso VN, Fernandes SO, Oliveira MC. Technetium-99m-labeled ceftizoxime loaded long-circulating and pH-sensitive liposomes used to identify osteomyelitis. Bioorg Med Chem Lett. 2012;22:4605-8. PMID: 22726927.
  • Financial source:

    none
  • 1
    Research performed at Laboratory of Experimental Surgery, School of Medicine, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte-MG, Brazil.

Publication Dates

  • Publication in this collection
    Jan 2018

History

  • Received
    28 Sept 2017
  • Reviewed
    26 Nov 2017
  • Accepted
    29 Dec 2017
Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia https://actacirbras.com.br/ - São Paulo - SP - Brazil
E-mail: actacirbras@gmail.com