Abstract
Introduction:
The sphygmomanometer test is an alternative and inexpensive method for assessment of muscle strength. This instrument was considered valid and reliable to measure the isometric strength in different health conditions, however, in individuals with limb amputations the properties of this instrument for this purpose, have not been investigated.
Objective:
To investigate the concurrent criterion validity, test-retest and inter-rater reliabilities of the aneroid sphygmomanometer test (AST) without modification, for assessment of the strength of subjects with lower limb amputations.
Methods:
Twenty-two subjects (57.6 ± 15.2 years) with lower limb amputations were included in this study. Maximum isometric force was assessed with a handheld dynamometer (microFet2®) and the AST. To identify differences between trials, one-way ANOVA was applied. To assess the concurrent criterion validity, test-retest and inter-rater reliabilities of the AST, Pearson’s correlation coefficients of determination and intra-class correlation coefficient (ICC) were calculated.
Results:
For all muscle groups, no differences were observed between the trials (0.00001 ≤ F ≤ 0.10; 0.90 ≤ p ≤ 0.99). Significant, positive, and high to very high correlations were found between the HHD and the AST measures for the different numbers of trials for all assessed muscles (0.76 ≤ r ≤ 0.93; p ≤ 0.02). Test-retest (0.67 ≤ ICC ≤ 0.97) and inter-rater reliabilities (0.78 ≤ ICC ≤ 0.97) were adequate. The values obtained with the AST were good predictors of those obtained with HHD (0.58 ≤ r2 ≤ 0.85).
Conclusion:
For individuals with lower limb amputation, the AST showed adequate concurrent criterion validity, test-retest reliability and inter-rater reliability for the assessment of lower limbs muscle strength.
Keywords:
Sphygmomanometer; Amputation; Data Accuracy; Muscle Strength; Validity of Tests
Resumo
Introdução:
O teste do esfigmomanômetro é um método alternativo e de baixo custo para se avaliar a força muscular. Esse instrumento já foi considerado válido e confiável para mensurar a força isométrica em diversas condições de saúde. No entanto, em indivíduos com amputações de membros as propriedades desse instrumento para esse fim, ainda não foram investigadas.
Objetivo:
Investigar a validade de critério concorrente, a confiabilidade teste-reteste e entre examinadores, do teste do esfigmomanômetro aneroide (TEA) sem modificação, para a avaliação da força muscular de indivíduos com amputações de membros inferiores (MMII).
Métodos:
Participaram 22 indivíduos (57,6 ± 15,2 anos) com amputações de MMII. A força isométrica máxima foi avaliada através de um dinamômetro portátil (microFet2®) e do TEA. Para identificar diferenças entre as repetições, ANOVA one-way foi utilizada. Para avaliar a validade do critério concorrente, confiabilidade teste-reteste e entre examinadores do TEA, os coeficientes de correlação de Pearson, de determinação e coeficiente de correlação intra-classe (CCI) foram calculados.
Resultados:
Para todos os grupos musculares, não foram observadas diferenças nas repetições (0,00001 ≤ F ≤ 0,10; 0,90 ≤ p ≤ 0,99). Os resultados apresentaram correlação positiva elevada e muito elevada para as diferentes medidas e grupos musculares dos MMII (0,76 ≤ r ≤ 0,93; p ≤ 0,02). E confiabilidade teste-reteste (0,67 ≤ ICC ≤ 0,97) e entre examinadores (0,78 ≤ ICC ≤ 0,97) consideráveis. Os valores obtidos com o TEA foram bons preditores daqueles obtidos com o dinamômetro manual (0,58 ≤ r2 ≤ 0,85).
Conclusão:
Para indivíduos com amputações de membros, o TEA apresentou valores adequados de validade de critério concorrente, confiabilidade teste-reteste e entre examinadores, para a avaliação de força muscular de MMII.
Palavras-chave:
Esfigmomanômetro; Amputação; Confiabilidade dos Dados; Força Muscular; Validade dos Testes
Introduction
Individuals with lower limb amputations of traumatic or elective origin11 Reis G, Casa Jr AJ, Campos RS. Perfil epidemiológico de amputados de membros superiores e inferiores atendidos em um centro de referência. RESC. 2012;2(2):52-62. present diverse physical, psychological and social changes22 Highsmith MJ, Kahle JT, Miro RM, Orendurff MS, Lewandowski AL, Orriola JJ, et al. Prosthetic interventions for people with transtibial amputation: Systematic review and meta-analysis of high-quality prospective literature and systematic reviews. J Rehabil Res Dev. 2016;53(2):157-84.), (33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (44 Raya MA, Gailey RS, Fiebert IM, Roach KE. Impairment variables predicting activity limitation in individuals with lower limb amputation. Prosthet Orthot Int. 2010;34(1):73-84.), (55 Aytar A, Pekyavas NO, Ergun N, Karatas M. Is there a relationship between core stability, balance and strength in amputee soccer players? A pilot study. Prosthet Orthot Int. 2012;36(3):332-8.), (66 van Velzen JM, van Bennekom CA, Polomski W, Slootman JR, van der Woude LH, Houdijk H. Physical capacity and walking ability after lower limb amputation: a systematic review. Clin Rehabil. 2006;20(11):999-1016.. Therefore, emphasizing the complex process of functionality/disability, amputation is related to a disarrangement in all the factors that guide this process44 Raya MA, Gailey RS, Fiebert IM, Roach KE. Impairment variables predicting activity limitation in individuals with lower limb amputation. Prosthet Orthot Int. 2010;34(1):73-84.. Regarding the physical factors, limb loss generates changes in muscle strength and balance, increasing energy expenditure22 Highsmith MJ, Kahle JT, Miro RM, Orendurff MS, Lewandowski AL, Orriola JJ, et al. Prosthetic interventions for people with transtibial amputation: Systematic review and meta-analysis of high-quality prospective literature and systematic reviews. J Rehabil Res Dev. 2016;53(2):157-84.), (66 van Velzen JM, van Bennekom CA, Polomski W, Slootman JR, van der Woude LH, Houdijk H. Physical capacity and walking ability after lower limb amputation: a systematic review. Clin Rehabil. 2006;20(11):999-1016.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205., causing functional limitations and changes in the participation in individual and social activities33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (66 van Velzen JM, van Bennekom CA, Polomski W, Slootman JR, van der Woude LH, Houdijk H. Physical capacity and walking ability after lower limb amputation: a systematic review. Clin Rehabil. 2006;20(11):999-1016..
Because amputation is an aggravating physical condition, the evaluation of muscle strength by means of quantitative methods is essential for the success of the therapeutic management33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (44 Raya MA, Gailey RS, Fiebert IM, Roach KE. Impairment variables predicting activity limitation in individuals with lower limb amputation. Prosthet Orthot Int. 2010;34(1):73-84.), (55 Aytar A, Pekyavas NO, Ergun N, Karatas M. Is there a relationship between core stability, balance and strength in amputee soccer players? A pilot study. Prosthet Orthot Int. 2012;36(3):332-8.), (88 Nolan L. Lower limb strength in sports-active transtibial amputees. Prosthet Orthot Int. 2009;33(3):230-41.), (99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45.), (1212 Toohey LA, Noronha M, Taylor C, Thomas J. Is a sphygmomanometer a valid and reliable tool to measure the isometric strength of hip muscles? A systematic review. Physiother Theory Pract. 2015;31(2):114-9.), (1313 Kegel B, Burgess EM, Starr TW, Daly WK. Effects of isometric muscle training on residual limb volume, strength, and gait of below-knee amputees. Phys Ther. 1981;61(10):1419-26.. The primary objective of rehabilitation after amputation is the fitting of a prosthesis22 Highsmith MJ, Kahle JT, Miro RM, Orendurff MS, Lewandowski AL, Orriola JJ, et al. Prosthetic interventions for people with transtibial amputation: Systematic review and meta-analysis of high-quality prospective literature and systematic reviews. J Rehabil Res Dev. 2016;53(2):157-84.), (1313 Kegel B, Burgess EM, Starr TW, Daly WK. Effects of isometric muscle training on residual limb volume, strength, and gait of below-knee amputees. Phys Ther. 1981;61(10):1419-26.), (1414 Boutwell E, Stine R, Gard S. Shock absorption during transtibial amputee gait: Does longitudinal prosthetic stiffness play a role? Prosthet Orthot Int. 2017;41(2):178-85., with this condition being dependent on good muscle strength33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (44 Raya MA, Gailey RS, Fiebert IM, Roach KE. Impairment variables predicting activity limitation in individuals with lower limb amputation. Prosthet Orthot Int. 2010;34(1):73-84.), (55 Aytar A, Pekyavas NO, Ergun N, Karatas M. Is there a relationship between core stability, balance and strength in amputee soccer players? A pilot study. Prosthet Orthot Int. 2012;36(3):332-8.), (1414 Boutwell E, Stine R, Gard S. Shock absorption during transtibial amputee gait: Does longitudinal prosthetic stiffness play a role? Prosthet Orthot Int. 2017;41(2):178-85..
There are many ways to measure muscle strength in individuals with amputation33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.), (1616 Moirenfeld I, Ayalon M, Bem-Sira D, Isakov E. Isokinetic strength and endurance of the knee extensors and flexors in trans-tibial amputees. Prosthet Orthot Int. 2000;24(3):221-5., however most of them present high costs, making them unusable in the clinical practice. Therefore, the physiotherapist uses subjective resources of limited reliability to quantify muscle strength1313 Kegel B, Burgess EM, Starr TW, Daly WK. Effects of isometric muscle training on residual limb volume, strength, and gait of below-knee amputees. Phys Ther. 1981;61(10):1419-26.), (1717 Bohannon RW. Manual muscle testing: does it meet the standards of an adequate screening test? Clin Rehabil. 2005;19(6):662-7.), (1818 Dvir Z. Differentiation of submaximal from maximal trunk extension effort: an isokinetic study using a new testing protocol. Spine (Phila Pa 1976). 1997;22(22):2672-6.. Based on this assumption, many researchers have investigated the validity and reliability of equipment such as the sphygmomanometer, an instrument already established to measure blood pressure1919 Shahbabu B, Dasgupta A, Sarkar K, Sahoo SK. Which is More Accurate in Measuring the Blood Pressure? A Digital or an Aneroid Sphygmomanometer. J Clin Diagn Res. 2016;10(3):LC11-4., to evaluate the isometric muscle strength of individuals with different health conditions99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45.), (1212 Toohey LA, Noronha M, Taylor C, Thomas J. Is a sphygmomanometer a valid and reliable tool to measure the isometric strength of hip muscles? A systematic review. Physiother Theory Pract. 2015;31(2):114-9.), (2020 Agnew PJ, Maas F. Jamar Dynamometer and Adapted Sphygmomanometer for Measuring Grip Strength in Patients with Rheumatoid Arthritis. OTJR (Thorofare N J). 1991;11(5):259-70.), (2121 Chiu TT, Sing KL. Evaluation of cervical range of motion and isometric neck muscle strength: reliability and validity. Clin Rehabil. 2002;16(8):851-8.. Considering the positive results of these investigations, the hypothesis of the present study was that the aneroid sphygmomanometer test (AST) could be a potential instrument to evaluate the muscle strength of individuals with limb amputations.
The sphygmomanometer represents a favorable alternative due to being portable and basic equipment of the health professional1919 Shahbabu B, Dasgupta A, Sarkar K, Sahoo SK. Which is More Accurate in Measuring the Blood Pressure? A Digital or an Aneroid Sphygmomanometer. J Clin Diagn Res. 2016;10(3):LC11-4. and is highlighted when the cost is taken into account. In addition, satisfactory results of concurrent criterion validity (0.75 ≤ r ≤ 0.90; p ≤ 0.001), test-retest reliability (0.77 ≤ ICC ≤ 0.98; p ≤ 0.001) and inter-rater reliability (0.53 ≤ ICC ≤ 0.91, p ≤ 0.001) have guaranteed the use of the equipment for measuring strength directed at various muscle groups1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.. The aneroid sphygmomanometer test (AST) has also been used to evaluate the strength of healthy individuals2222 Mahashabde R, Fernandez R, Sabnis S. Validity and reliability of the aneroid sphygmomanometer using a paediatric size cuff for craniocervical flexion test. Int J Evid Based Healthc. 2013;11(4):285-90.), (1212 Toohey LA, Noronha M, Taylor C, Thomas J. Is a sphygmomanometer a valid and reliable tool to measure the isometric strength of hip muscles? A systematic review. Physiother Theory Pract. 2015;31(2):114-9. and those with various health conditions, such as Parkinson’s disease1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45., stroke99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8., neck pain2121 Chiu TT, Sing KL. Evaluation of cervical range of motion and isometric neck muscle strength: reliability and validity. Clin Rehabil. 2002;16(8):851-8., hip fracture2323 Sherrington C, Lord SR. Reliability of simple portable tests of physical performance in older people after hip fracture. Clin Rehabil. 2005;19(5):496-504. and rheumatoid arthritis2020 Agnew PJ, Maas F. Jamar Dynamometer and Adapted Sphygmomanometer for Measuring Grip Strength in Patients with Rheumatoid Arthritis. OTJR (Thorofare N J). 1991;11(5):259-70..
The concurrent criterion validity indicates that the instrument is adequate to measure the strength and can be compared to portable instruments that are gold standard2424 Bohannon RW. Reference values for extremity muscle strength obtained by hand-held dynamometry from adults aged 20 to 79 years. Arch Phys Med Rehabil. 1997;78(1):26-32.), (2525 Bohannon RW, Andrews AW, Thomas MW. Normative values for isometric muscle force measurements obtained with hand-held dynamometers. Phys Ther. 1996;76(3):248-59.), (2626 Souza LAC, Martins JC, Moura JB, Teixeira-Salmela LF, Paula FVR, Faria CDCM. Avaliação da força muscular com o teste do esfigmomanômetro modificado: qual o melhor método e forma de operacionalização? Braz J Phys Ther. 2014;18(2):191-200.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.. The use of the AST to evaluate the muscle strength of individuals with amputations may be an alternative of choice to complement the evaluations. Both validity and reliability of this instrument for measuring the muscle strength of these individuals are fundamental for it to become a reliable and valid option for use in studies and mainly in the clinical practice.
As mentioned, measuring muscle strength gain in individuals with limb amputations is a strategic determinant for the clinical evaluation33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (44 Raya MA, Gailey RS, Fiebert IM, Roach KE. Impairment variables predicting activity limitation in individuals with lower limb amputation. Prosthet Orthot Int. 2010;34(1):73-84.), (55 Aytar A, Pekyavas NO, Ergun N, Karatas M. Is there a relationship between core stability, balance and strength in amputee soccer players? A pilot study. Prosthet Orthot Int. 2012;36(3):332-8.), (88 Nolan L. Lower limb strength in sports-active transtibial amputees. Prosthet Orthot Int. 2009;33(3):230-41., because after the fitting of a prosthesis, muscle mass gain attenuates the gait asymmetries66 van Velzen JM, van Bennekom CA, Polomski W, Slootman JR, van der Woude LH, Houdijk H. Physical capacity and walking ability after lower limb amputation: a systematic review. Clin Rehabil. 2006;20(11):999-1016.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.. Due to increased energy expenditure during walking, patients opt for a more economical velocity77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205., however, the more muscle strength individuals gain, the more they increase the walking ability22 Highsmith MJ, Kahle JT, Miro RM, Orendurff MS, Lewandowski AL, Orriola JJ, et al. Prosthetic interventions for people with transtibial amputation: Systematic review and meta-analysis of high-quality prospective literature and systematic reviews. J Rehabil Res Dev. 2016;53(2):157-84.), (66 van Velzen JM, van Bennekom CA, Polomski W, Slootman JR, van der Woude LH, Houdijk H. Physical capacity and walking ability after lower limb amputation: a systematic review. Clin Rehabil. 2006;20(11):999-1016.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9..
Thus, the aim of the present study was to investigate the concurrent criterion validity, test-retest reliability and inter-rater reliability, of the unmodified AST in the evaluation of the muscle strength of individuals with lower limb amputations.
Methods
The present study was evaluated and approved by the Research Ethics Committee of the University of Itaúna (No. 1.333.943), considering the requirements of Resolution No. 510/2016 of the National Health Council.
Sample
The study was performed with volunteers, with unilateral or bilateral lower limb amputations, who were recruited from the Amputee Sports Association of Minas Gerais (AMDA - Belo Horizonte), the Integrated Physiotherapy Clinics of the University of Itaúna and the Primary Health Units of the municipalities of Itaúna and Itatiaiuçu of Minas Gerais state. After providing consent, the evaluations were carried out at the centers involved or in the home of the volunteer. All individuals who participated in the study signed a consent form.
Female and male volunteers who presented lower limb amputation of traumatic or elective origin, with time since the amputation of four months or more and aged over 20 years were included in the study. The individuals who presented the following were excluded; cognitive alterations or the inability to understand the proposed tests1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.; contractures in the residual limb; pain in the joints involved; residual pain or phantom pain2727 Nikolasjsen L, Jensen TS. Phantom limb pain. Br J Anaesth. 2001;87:107-16. greater than three on the Visual Analogue Scale (VAS)2828 Alves AMB, Natour J, Assis MR, Feldman D. Avaliação de instrumentos de medida usados em pacientes com fibromialgia. Rev Bras Reumatol. 2012;52(4):496-506.; residual limb of less than 10cm in length and other health conditions that could alter the strength of the lower limb33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.), (2929 Isakov E, Burger H, Gregoric M, Marincek C. Stump length as related to atrophy and strength of the thigh muscles in trans-tibial amputees. Prosthet Orthot Int. 1996;20(2):96-100..
Equipment for Muscle Strength Measurement
For the measurement of the isometric muscular force, the Tyros® brand aneroid sphygmomanometer (AS) (WelchAllyn Inc., NY, USA, Model DS-44) was used, unmodified, with the units of measurement in mmHg. For its use in the study, the equipment was inflated to 100 mmHg to remove possible folds, and then deflated to 20 mmHg, keeping the valve closed, establishing a measurement range of 20 to 304 mmHg99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (2020 Agnew PJ, Maas F. Jamar Dynamometer and Adapted Sphygmomanometer for Measuring Grip Strength in Patients with Rheumatoid Arthritis. OTJR (Thorofare N J). 1991;11(5):259-70.. For methodological control, this procedure was repeated with each test1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8..
In order to investigate the concurrent criterion validity, the force in Newtons (N) was measured using a microFet2® Handheld Dynamometer (HHD) (HogganHealth Industries, Salt Lake City, UT, USA), considered a gold standard for isometric strength measurement99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (2424 Bohannon RW. Reference values for extremity muscle strength obtained by hand-held dynamometry from adults aged 20 to 79 years. Arch Phys Med Rehabil. 1997;78(1):26-32.), (2525 Bohannon RW, Andrews AW, Thomas MW. Normative values for isometric muscle force measurements obtained with hand-held dynamometers. Phys Ther. 1996;76(3):248-59.. A digital timer (Casio HS-30W) was also used to measure the duration of the force pressures obtained with the AST.
Procedures
The test-retest reliability evaluates the consistency of the measurements performed under the same evaluation conditions at two different times. Inter-rater reliability, on the other hand, evaluates the consistency of the measurements performed by two different examiners3030 Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Practice. 3rd ed. New Jersey: Prentice-Hall Health; 2009.. Thus, according to established procedures, the participants were evaluated in two moments by two previously trained evaluators (Examiner 1 and 2), with an average interval of 10 days between the first and second evaluation, following the same criteria of evaluation, such as: administration, location, time, instruction and protocols.
On the first day, for investigation and collection of data, the volunteers responded to an evaluation form containing demographic data, anthropometric data and clinical data related to the amputation. In addition, they were informed of the objectives and methods of the study, and then underwent the first strength evaluation using the AST and HHD.
The order of use of the equipment (AST and HHD) was randomized before each evaluation through a simple draw1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.. An approximate interval of five minutes between the measurement of each piece of equipment was established99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.. The examiners applied manual resistance against the force exerted by the volunteer, keeping the limb in question static99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45.. To assist each examiner, two other researchers helped in the tests, controlling the time, reading the equipment99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (2424 Bohannon RW. Reference values for extremity muscle strength obtained by hand-held dynamometry from adults aged 20 to 79 years. Arch Phys Med Rehabil. 1997;78(1):26-32.), (2525 Bohannon RW, Andrews AW, Thomas MW. Normative values for isometric muscle force measurements obtained with hand-held dynamometers. Phys Ther. 1996;76(3):248-59. and recording the data, so that the examiners who were operating the equipment were not influenced by the transcription of the data1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (3030 Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Practice. 3rd ed. New Jersey: Prentice-Hall Health; 2009..
Previously, for familiarization training for the test, the volunteers performed submaximal isometric contraction. Only the affected side was evaluated and in cases of bilateral amputation, the residual limb of greater length was the side considered for the analysis33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (1212 Toohey LA, Noronha M, Taylor C, Thomas J. Is a sphygmomanometer a valid and reliable tool to measure the isometric strength of hip muscles? A systematic review. Physiother Theory Pract. 2015;31(2):114-9.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.), (2929 Isakov E, Burger H, Gregoric M, Marincek C. Stump length as related to atrophy and strength of the thigh muscles in trans-tibial amputees. Prosthet Orthot Int. 1996;20(2):96-100.. The muscle groups evaluated depended on the level of the amputation. For the transfemoral amputations, the muscle groups evaluated were: flexors, extensors and hip abductors. For evaluation of the flexor muscles of the hip the volunteer remained seated in a chair with a backrest, maintaining 90° flexion of the hip and contralateral knee33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (3131 Thorborg K, Petersen J, Magnusson SP, Holmich P. Clinical assessment of hip strength using a hand-held dynamometer is reliable. Scand J Med Sci Sports. 2010;20(3):493-501.. In order to perform this test, the volunteer was allowed to hold the edges of the chair for stabilization, and the isometric movement was performed, resisted by the examiner3131 Thorborg K, Petersen J, Magnusson SP, Holmich P. Clinical assessment of hip strength using a hand-held dynamometer is reliable. Scand J Med Sci Sports. 2010;20(3):493-501., with the equipment positioned on the anterior surface of the residual limb. For the evaluation of hip extensors, the volunteer was placed in the ventral decubitus (VD) position and the equipment was positioned on the posterior surface of the residual limb3131 Thorborg K, Petersen J, Magnusson SP, Holmich P. Clinical assessment of hip strength using a hand-held dynamometer is reliable. Scand J Med Sci Sports. 2010;20(3):493-501.. For the evaluation of the hip abductor muscles, the volunteer was lying in lateral decubitus (LD) position, with the contralateral limb kept flexed and the equipment positioned on the lateral side of the residual limb33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (3232 Kolk S, Fluit R, Luijten J, Heesterbeek PJ, Geurts AC, Verdonschot N, et al. Gait and lower limb muscle strength in women after triple innominate osteotomy. BMC Musculoskelet Disord. 2015;16:68..
In the transtibial amputations and disarticulations of the foot, the muscle groups evaluated were knee flexors and extensors. For the evaluation of the extensors, the volunteer remained seated in a chair with back, hips and knees flexed at 90°1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.), (3131 Thorborg K, Petersen J, Magnusson SP, Holmich P. Clinical assessment of hip strength using a hand-held dynamometer is reliable. Scand J Med Sci Sports. 2010;20(3):493-501.. For this movement, the volunteer was allowed to hold the edges of the chair3131 Thorborg K, Petersen J, Magnusson SP, Holmich P. Clinical assessment of hip strength using a hand-held dynamometer is reliable. Scand J Med Sci Sports. 2010;20(3):493-501.. The devices were positioned on the anterior surface of the residual limb. For the knee flexor muscles, the volunteer was lying in the VD position with the contralateral knee in extension and the amputated limb at 45° flexion3333 Oliveira LC, Saad MC, Felício LR, Grossi DB. Muscle strength analysis of hip and knee stabilizers in individuals with Patellofemoral Pain Syndrome. Fisioter Pesqui. 2014;21(4):327-32.. The equipment was positioned on the posterior surface of the residual limb.
The measuring instruments, AST and HHD, were positioned 5 cm from the distal end of the residual limb to resist the isometric motion. During the tests, the volunteers were verbally encouraged to perform the maximum isometric contraction, maintaining it for 5 seconds (s)99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.), (3131 Thorborg K, Petersen J, Magnusson SP, Holmich P. Clinical assessment of hip strength using a hand-held dynamometer is reliable. Scand J Med Sci Sports. 2010;20(3):493-501.. The examiners applied manual resistance, contrary to the force exerted by the volunteer, keeping the limb static. Six measurements were performed, three with HHD and three with AST99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.. Peak force values were recorded. A rest interval of 15s was given between each measurement1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.. All the care with testing and operations related to the muscular force test, such as positioning, equipment calibration, and pre-test care, were observed and performed following all the observations suggested by the manufacturers. If a prosthesis was used, it was removed for the tests.
Sample Calculation
To determine the sample size, a previous pilot study was performed with 10 individuals with lower limb amputations3030 Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Practice. 3rd ed. New Jersey: Prentice-Hall Health; 2009.. For the determination of the calculation, the lowest value of the coefficient of determination obtained for the different muscle groups was considered: r 2 = 0.57. Using a Power of 0.90 and an α = 0.05 significance level, n = 11 was found. As individuals of different age groups would be investigated, two groups, 11 adults (20 ≤ n < 60 years) and 11 older adults (≥ 60 years), were determined to attenuate the assumptions of sample variability3030 Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Practice. 3rd ed. New Jersey: Prentice-Hall Health; 2009.. All data analyzes were performed with all the subjects pooled. For the calculation of the sample size, the statistical program G-Power©, version 3.1.9.2 (Franz Faul, Universität Kiel, Germany, 1992-2014) was used.
Statistical Analysis
Descriptive statistics and normality tests were performed. In order to compare the results obtained with the AST, one-way ANOVA was used considering the measurements obtained by examiners 1 and 2, on the first day, using different repetitions (first repetition, mean of 1st and 2nd repetition and mean of the three repetitions) for all muscle groups 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.. For the analysis of the concurrent criterion validity of the AST, Pearson’s correlation coefficient was used to determine the correlation between muscle strength measurements obtained with the HHD (N) 2424 Bohannon RW. Reference values for extremity muscle strength obtained by hand-held dynamometry from adults aged 20 to 79 years. Arch Phys Med Rehabil. 1997;78(1):26-32.), (2525 Bohannon RW, Andrews AW, Thomas MW. Normative values for isometric muscle force measurements obtained with hand-held dynamometers. Phys Ther. 1996;76(3):248-59. and with the AST (mmHg), for each of the muscle groups evaluated. Linear regression analyzes were performed to identify the best model, which could explain the relationship between the measurements obtained with the two pieces of equipment and thus present the regression equations to predict the strength values, in N, through those obtained in mmHg by the AST.
For the analysis of the reliability of the AST, the intra-class correlation coefficient (ICC) was used, with 95% confidence interval (CI), for the inter-rater measurements obtained and for the analysis of the test-retest reliability. For the ICC values that showed significant results, the magnitude of the correlation was classified as previously described 3434 Souza LAC, Martins JC, Teixeira-Salmela LF, Godoy MR, Aguiar LT, Faria CDCM. Avaliação da força muscular pelo teste do esfigmomanômetro modificado: uma revisão da literatura. Fisioter Mov. 2013;26(2):437-52.. For the correlations that presented significant results, the magnitude of the correlation was classified as follows: 0-0.25 very low; 0.26-0.49 low; 0.50-0.69 moderate; 0.70-0.89 high; 0.90-1.00 very high. Systematic differences between the two sessions, test-retest or inter-rater reliability were verified through paired t-tests, presenting 95% CI of the mean of the difference. All the analyzes were performed using the SPSS statistical package for Windows (version 22.0). The level of significance was α = 0.05.
Results
For the analysis of the of the concurrent criterion validity of the AST, 22 individuals with lower limb amputations were evaluated (Table 1). For the tests of test-retest and inter-rater reliability 18 individuals were evaluated, with a mean age of 58.6 (15.9) years, mean time since the amputation of 10.73 (14.5) years. Four volunteers did not participate in the second evaluation because they were unable to attend. The majority of the individuals underwent elective amputation as a result of peripheral vascular disease or diabetes.
Table 2 presents the values obtained from the three repetitions considered for analysis of strength measurement. As no differences were observed (0.00001 ≤ F ≤ 0.10; 0.90 ≤ p ≤ 0.99), analysis of the validity and reliability data were performed considering the three repetitions.
Validity
Table 3 presents the correlation values and regression analysis between the AST and HHD. There was a high and very high positive correlation for the lower limb muscle groups (0.76 ≤ r ≤ 0.93; p ≤ 0.02).
The results of the linear regression analysis (r 2), obtained in the first session (repetition 1), indicated that the values presented by the AST were good predictors of those obtained by the HHD. According to the results of the determination coefficient, at least 58% of the values obtained with the HHD, in N, were explained by the values obtained with the AST, in mmHg, for all muscle groups. The equations shown in Table 3 can be used to predict the strength values in N, through the measurements obtained by the AST in mmHg.
Reliability
According to the results of the test-retest reliability of the AST, considering the three measurements and the two examiners, ICC ≥ 0.81 and ICC ≥ 0.67 were observed for the knee and hip flexor and extensor musculature and hip abductors, respectively (Table 4). High and very high values were presented for the knee and hip flexors and extensors and moderate or high for the hip abductors.
Intra-Class Correlation Coefficient (ICC) values for the test-retest reliability test using the AST to evaluate the strength of the hip abductor muscles; flexor and extensor muscles of the hip and knee, considering the measurements in the two distinct sessions, including data from examiner 1 and 2, n = 18
Evaluating the results of the inter-rater reliability (Table 5), high or very high values (0.78 ≤ ICC ≤ 0.97) were observed for all lower limb muscle groups using the AST. No differences were observed between the evaluators or between the two days of evaluation.
Intra-Class Correlation Coefficient (ICC) values for the inter-examiner reliability test using the AST to evaluate the strength of the hip abductor muscles; flexor and extensor muscles of the hip and knee, considering the repetitions in the two distinct sessions, including data from examiner 1 and 2
In general, the 95% CI of the test-retest (Table 4) and inter-rater reliabilities (Table 5) presented a wide variation between the different repetitions. For the hip abductors, related to the second evaluator, in the test-retest reliability test the ICC was below 0.70, this being moderate. The 95% CI of the ICC were broad and the ranges included zero.
Discussion
Validity
In the present study, the regression analyzes were able to explain how the association between HHD and AST measurements occurred and could infer judicious data regarding the variations of HHD values obtained by AST. The coefficient of determination (r 2 ) of the various muscle groups was categorical, varying from 0.58 to 0.85, which means that at least 58% of the changes in the DPM measurements can be explained by the variations in the AST measurements, indicating that the data obtained through the AST are good predictors of those obtained by the HHD. In the study by Aguiar et al. 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-., r 2 ranged from 0.54 to 0.77 for the hip and knee musculature. In the study by Souza et al. 1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8., r 2 ranged from 0.57 to 0.79.
The equations presented in this study may help professionals to convert force measurement data, or that of other strength measurements, from mmHg to N, which facilitates the comparison of data from the literature, following the clinical evolution of individuals based on scientific evidence.
In the present study, the three repetitions were considered for analysis 1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45.), (2020 Agnew PJ, Maas F. Jamar Dynamometer and Adapted Sphygmomanometer for Measuring Grip Strength in Patients with Rheumatoid Arthritis. OTJR (Thorofare N J). 1991;11(5):259-70., with no differences being observed between them (p ≥ 0.90). Other studies also did not identify differences 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.. In the study by Agnew 2020 Agnew PJ, Maas F. Jamar Dynamometer and Adapted Sphygmomanometer for Measuring Grip Strength in Patients with Rheumatoid Arthritis. OTJR (Thorofare N J). 1991;11(5):259-70., there were significant variations between the measurements, however, the researchers used the AST to evaluate manual grip strength (p < 0.01).
Test-Retest Reliability
The test-retest reliability of the AST, considering the three repetitions and the two examiners, revealed ICC ≥ 0.81 for the knee and hip flexor and extensor musculature and ICC ≥ 0.67 for the hip abductors. In the study by Aguiar et al. 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-., the test-retest reliability in subjects with hemiparesis ranged from 0.75-0.97 (ICC) for hip flexors and extensors; 0.86-0.97 (ICC) for abductors; 0.65-0.87 and 0.76-0.91 (ICC) for knee flexors and extensors. Moderate to very high correlations for hip flexors (ICC 0.66-0.93) and abductors (ICC 0.75-0.93) were also observed in other studies 2323 Sherrington C, Lord SR. Reliability of simple portable tests of physical performance in older people after hip fracture. Clin Rehabil. 2005;19(5):496-504.), (3535 Perossa DR, Dziak M, Vernon HT, Hayashita K. The intra-examiner reliability of manual muscle testing of the hip and shoulder with a modified sphygmomanometer: A preliminary study of normal subjects. J Can Chiropr Assoc. 1998;42(2):73-82.. A very high correlation was also observed for the hip extensors (ICC 0.90-0.91) 3535 Perossa DR, Dziak M, Vernon HT, Hayashita K. The intra-examiner reliability of manual muscle testing of the hip and shoulder with a modified sphygmomanometer: A preliminary study of normal subjects. J Can Chiropr Assoc. 1998;42(2):73-82.. This indicates concordance between the present study and the aforementioned studies. Even with levers of different lengths the results of the ICC were similar.
Observing the ICC values in the test-retest evaluations for the hip abductors, moderate results were identified regarding the second evaluator (0.67 - 0.69), with a wide 95% CI, containing zero. The variability can be explained by the force generation being dependent on the length of the lever arm 1212 Toohey LA, Noronha M, Taylor C, Thomas J. Is a sphygmomanometer a valid and reliable tool to measure the isometric strength of hip muscles? A systematic review. Physiother Theory Pract. 2015;31(2):114-9.. Sherrington et al. 2323 Sherrington C, Lord SR. Reliability of simple portable tests of physical performance in older people after hip fracture. Clin Rehabil. 2005;19(5):496-504. also observed a 12% lower difference in the ICC in the abduction movement using the sphygmomanometer without modification, when compared to the HHD.
Inter-examiner Reliability
In this study, the inter-rater reliability of the AST generated results considered high or very high. For knee extensors and flexors, the ICC scores ranged from 0.88 - 0.96. For the hip abductors the ICC was ≥ 0.78. For the hip flexors and extensors the inter-rater ICC was ≥ 0.86. No differences between the evaluators and the two days of evaluation were observed.
According to previous studies 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.), (1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (3636 Kaegi C, Thibault MC, Giroux F, Bourbonnais D. The interrater reliability of force measurements using a modified sphygmomanometer in elderly subjects. Phys Ther. 1998;78(10):1095-103., inter-rater reliability values have been found to present variability according to the sample investigated. While the ICC values observed for the hip extensors in the present study were very high, moderate inter-rater correlations (ICC = 0.67) were found in a cohort of older adults 3636 Kaegi C, Thibault MC, Giroux F, Bourbonnais D. The interrater reliability of force measurements using a modified sphygmomanometer in elderly subjects. Phys Ther. 1998;78(10):1095-103.. In individuals with subacute 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-., and chronic 1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8. hemiparesis, the ICC were moderate, high and very high, (0.62 - 0.94 and 0.57-0.97) respectively, for the muscle groups evaluated in this study. Bohannon et al. 2525 Bohannon RW, Andrews AW, Thomas MW. Normative values for isometric muscle force measurements obtained with hand-held dynamometers. Phys Ther. 1996;76(3):248-59. observed moderate to low values for inter-rater reliability in the hip abduction and ankle dorsiflexion movements, however, using the HHD.
Clinical Considerations
Studies that evaluate reproducibility (reliability and concordance) aim to evaluate the variability of a method or an instrument, to attenuate poor interpretations of data before and after interventions or observations 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.) - (1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45.), (3636 Kaegi C, Thibault MC, Giroux F, Bourbonnais D. The interrater reliability of force measurements using a modified sphygmomanometer in elderly subjects. Phys Ther. 1998;78(10):1095-103.. With muscle groups of considerable functional magnitude, such as those evaluated in the present study, the importance of evaluating hip abductors and extensors and knee extensors in individuals with lower limb amputation, is emphasized. Individuals who have a short residual limb (< 15.1 cm) deserve special attention because, when weak, these muscles alter the socket system, limiting acuity in basic tasks such as walking and standing up 77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (2929 Isakov E, Burger H, Gregoric M, Marincek C. Stump length as related to atrophy and strength of the thigh muscles in trans-tibial amputees. Prosthet Orthot Int. 1996;20(2):96-100.. In the present study, since the level of amputation was different, presenting residual limbs of different sizes (mean 30.4 cm [10.4 SD]), the positioning of the equipment presented variability, however, always considering 5cm to the distal end of the stump. According to biomechanics, the production of force is dependent on the size of the lever arm and the position of the joint 77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (2929 Isakov E, Burger H, Gregoric M, Marincek C. Stump length as related to atrophy and strength of the thigh muscles in trans-tibial amputees. Prosthet Orthot Int. 1996;20(2):96-100.. This could present differences when comparing studies on the reliability of the AST. As no studies with amputees were found in the literature review, the comparison with the present results was performed with studies presenting different samples 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.) - (1212 Toohey LA, Noronha M, Taylor C, Thomas J. Is a sphygmomanometer a valid and reliable tool to measure the isometric strength of hip muscles? A systematic review. Physiother Theory Pract. 2015;31(2):114-9.), (2323 Sherrington C, Lord SR. Reliability of simple portable tests of physical performance in older people after hip fracture. Clin Rehabil. 2005;19(5):496-504.), (3535 Perossa DR, Dziak M, Vernon HT, Hayashita K. The intra-examiner reliability of manual muscle testing of the hip and shoulder with a modified sphygmomanometer: A preliminary study of normal subjects. J Can Chiropr Assoc. 1998;42(2):73-82.), (3636 Kaegi C, Thibault MC, Giroux F, Bourbonnais D. The interrater reliability of force measurements using a modified sphygmomanometer in elderly subjects. Phys Ther. 1998;78(10):1095-103.. According to Pedrinelli et al. 77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205., the residual limb length does not cause interference when evaluations are performed with the use of prostheses. However, as the evaluations were performed without the use of prostheses in the present study, the force length ratio was preponderant.
In this study, the non-amputated side was not evaluated, since it is already evident that the variability of strength measurements is lower in the non-amputated limb, in which the strength and power are greater 77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205..
The researchers performed the test-retest with two examiners, reinforcing the methodological criteria. The majority of the studies investigated 99 Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.) - (1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45. evaluated the reproducibility, intra-examiner reliability, with only one examiner. In addition to reliability, validity was also a subject of the study.
Limitations
Although valid and reliable, the AST can be limited when the individual has a force that exceeds 300 mmHg, the maximum limit of measurement of the equipment. In the present study, the participants presented different levels of lower limb amputation, with levers of varying lengths. Unfortunately, this is a process inherent and particular to the studies with amputations of corporeal segments 33 Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.) , (2929 Isakov E, Burger H, Gregoric M, Marincek C. Stump length as related to atrophy and strength of the thigh muscles in trans-tibial amputees. Prosthet Orthot Int. 1996;20(2):96-100.. Regarding the number of individuals evaluated, the ideal would be 11 individuals for each age group 1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.. However, as in the majority of the studies with similar samples 22 Highsmith MJ, Kahle JT, Miro RM, Orendurff MS, Lewandowski AL, Orriola JJ, et al. Prosthetic interventions for people with transtibial amputation: Systematic review and meta-analysis of high-quality prospective literature and systematic reviews. J Rehabil Res Dev. 2016;53(2):157-84.), (77 Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.), (1414 Boutwell E, Stine R, Gard S. Shock absorption during transtibial amputee gait: Does longitudinal prosthetic stiffness play a role? Prosthet Orthot Int. 2017;41(2):178-85.), (1515 Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.), (2929 Isakov E, Burger H, Gregoric M, Marincek C. Stump length as related to atrophy and strength of the thigh muscles in trans-tibial amputees. Prosthet Orthot Int. 1996;20(2):96-100., the number of individuals evaluated was restricted. This is a common situation when using a sample of amputees, as well as in other health conditions that present different situations 1010 Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.), (1111 Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45..
Conclusion
The AST is a simple measurement method, with valid and reliable measurement capacity. In individuals with lower limb amputations evaluated by two independent examiners, the AST presented adequate concurrent criterion validity, test-retest reliability and inter-rater reliability. After familiarization training, only one repetition was sufficient to reproduce valid and reliable results.
References
-
1Reis G, Casa Jr AJ, Campos RS. Perfil epidemiológico de amputados de membros superiores e inferiores atendidos em um centro de referência. RESC. 2012;2(2):52-62.
-
2Highsmith MJ, Kahle JT, Miro RM, Orendurff MS, Lewandowski AL, Orriola JJ, et al. Prosthetic interventions for people with transtibial amputation: Systematic review and meta-analysis of high-quality prospective literature and systematic reviews. J Rehabil Res Dev. 2016;53(2):157-84.
-
3Pauley T, Devlin M, Madan-Sharma P. A single-blind, cross-over trial of hip abductor strength training to improve Timed Up & Go performance in patients with unilateral, transfemoral amputation. J Rehabil Med. 2014;46(3):264-70.
-
4Raya MA, Gailey RS, Fiebert IM, Roach KE. Impairment variables predicting activity limitation in individuals with lower limb amputation. Prosthet Orthot Int. 2010;34(1):73-84.
-
5Aytar A, Pekyavas NO, Ergun N, Karatas M. Is there a relationship between core stability, balance and strength in amputee soccer players? A pilot study. Prosthet Orthot Int. 2012;36(3):332-8.
-
6van Velzen JM, van Bennekom CA, Polomski W, Slootman JR, van der Woude LH, Houdijk H. Physical capacity and walking ability after lower limb amputation: a systematic review. Clin Rehabil. 2006;20(11):999-1016.
-
7Pedrinelli A, Saito M, Coelho RF, Fontes RB, Guarniero R. Comparative study of the strength of the flexor and extensor muscles of the knee through isokinetic evaluation in normal subjects and patients subjected to trans-tibial amputation. Prosthet Orthot Int. 2002;26(3):195-205.
-
8Nolan L. Lower limb strength in sports-active transtibial amputees. Prosthet Orthot Int. 2009;33(3):230-41.
-
9Aguiar LT, Lara EM, Martins JC, Teixeira-Salmela LF, Quintino LF, Christo PP, et al. Modified sphygmomanometer test for the assessment of strength of the trunk, upper and lower limbs muscles in subjects with subacute stroke: reliability and validity. Eur J Phys Rehabil Med. 2016 Oct;52(5):637-.
-
10Souza LA, Martins JC, Teixeira-Salmela LF, Lara EM, Moura JB, Aguiar LT, et al. Validity and reliability of the modified sphygmomanometer test to assess strength of the lower limbs and trunk muscles after stroke. J Rehabil Med. 2014;46(7):620-8.
-
11Silva SM, Correa FI, Silva PF, Silva DF, Lucareli PR, Correa JC. Validation and reliability of a modified sphygmomanometer for the assessment of handgrip strength in Parkinson's disease. Braz J Phys Ther. 2015;19(2):137-45.
-
12Toohey LA, Noronha M, Taylor C, Thomas J. Is a sphygmomanometer a valid and reliable tool to measure the isometric strength of hip muscles? A systematic review. Physiother Theory Pract. 2015;31(2):114-9.
-
13Kegel B, Burgess EM, Starr TW, Daly WK. Effects of isometric muscle training on residual limb volume, strength, and gait of below-knee amputees. Phys Ther. 1981;61(10):1419-26.
-
14Boutwell E, Stine R, Gard S. Shock absorption during transtibial amputee gait: Does longitudinal prosthetic stiffness play a role? Prosthet Orthot Int. 2017;41(2):178-85.
-
15Demirbuken I, Ozyurek S, Angin S. Altered movement strategies in sit-to-stand task in persons with transtibial amputation. Prosthet Orthot Int. 2014;38(4):303-9.
-
16Moirenfeld I, Ayalon M, Bem-Sira D, Isakov E. Isokinetic strength and endurance of the knee extensors and flexors in trans-tibial amputees. Prosthet Orthot Int. 2000;24(3):221-5.
-
17Bohannon RW. Manual muscle testing: does it meet the standards of an adequate screening test? Clin Rehabil. 2005;19(6):662-7.
-
18Dvir Z. Differentiation of submaximal from maximal trunk extension effort: an isokinetic study using a new testing protocol. Spine (Phila Pa 1976). 1997;22(22):2672-6.
-
19Shahbabu B, Dasgupta A, Sarkar K, Sahoo SK. Which is More Accurate in Measuring the Blood Pressure? A Digital or an Aneroid Sphygmomanometer. J Clin Diagn Res. 2016;10(3):LC11-4.
-
20Agnew PJ, Maas F. Jamar Dynamometer and Adapted Sphygmomanometer for Measuring Grip Strength in Patients with Rheumatoid Arthritis. OTJR (Thorofare N J). 1991;11(5):259-70.
-
21Chiu TT, Sing KL. Evaluation of cervical range of motion and isometric neck muscle strength: reliability and validity. Clin Rehabil. 2002;16(8):851-8.
-
22Mahashabde R, Fernandez R, Sabnis S. Validity and reliability of the aneroid sphygmomanometer using a paediatric size cuff for craniocervical flexion test. Int J Evid Based Healthc. 2013;11(4):285-90.
-
23Sherrington C, Lord SR. Reliability of simple portable tests of physical performance in older people after hip fracture. Clin Rehabil. 2005;19(5):496-504.
-
24Bohannon RW. Reference values for extremity muscle strength obtained by hand-held dynamometry from adults aged 20 to 79 years. Arch Phys Med Rehabil. 1997;78(1):26-32.
-
25Bohannon RW, Andrews AW, Thomas MW. Normative values for isometric muscle force measurements obtained with hand-held dynamometers. Phys Ther. 1996;76(3):248-59.
-
26Souza LAC, Martins JC, Moura JB, Teixeira-Salmela LF, Paula FVR, Faria CDCM. Avaliação da força muscular com o teste do esfigmomanômetro modificado: qual o melhor método e forma de operacionalização? Braz J Phys Ther. 2014;18(2):191-200.
-
27Nikolasjsen L, Jensen TS. Phantom limb pain. Br J Anaesth. 2001;87:107-16.
-
28Alves AMB, Natour J, Assis MR, Feldman D. Avaliação de instrumentos de medida usados em pacientes com fibromialgia. Rev Bras Reumatol. 2012;52(4):496-506.
-
29Isakov E, Burger H, Gregoric M, Marincek C. Stump length as related to atrophy and strength of the thigh muscles in trans-tibial amputees. Prosthet Orthot Int. 1996;20(2):96-100.
-
30Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Practice. 3rd ed. New Jersey: Prentice-Hall Health; 2009.
-
31Thorborg K, Petersen J, Magnusson SP, Holmich P. Clinical assessment of hip strength using a hand-held dynamometer is reliable. Scand J Med Sci Sports. 2010;20(3):493-501.
-
32Kolk S, Fluit R, Luijten J, Heesterbeek PJ, Geurts AC, Verdonschot N, et al. Gait and lower limb muscle strength in women after triple innominate osteotomy. BMC Musculoskelet Disord. 2015;16:68.
-
33Oliveira LC, Saad MC, Felício LR, Grossi DB. Muscle strength analysis of hip and knee stabilizers in individuals with Patellofemoral Pain Syndrome. Fisioter Pesqui. 2014;21(4):327-32.
-
34Souza LAC, Martins JC, Teixeira-Salmela LF, Godoy MR, Aguiar LT, Faria CDCM. Avaliação da força muscular pelo teste do esfigmomanômetro modificado: uma revisão da literatura. Fisioter Mov. 2013;26(2):437-52.
-
35Perossa DR, Dziak M, Vernon HT, Hayashita K. The intra-examiner reliability of manual muscle testing of the hip and shoulder with a modified sphygmomanometer: A preliminary study of normal subjects. J Can Chiropr Assoc. 1998;42(2):73-82.
-
36Kaegi C, Thibault MC, Giroux F, Bourbonnais D. The interrater reliability of force measurements using a modified sphygmomanometer in elderly subjects. Phys Ther. 1998;78(10):1095-103.
Publication Dates
-
Publication in this collection
2017
History
-
Received
27 June 2016 -
Accepted
12 May 2017