Abstracts
Objective:
despite the technologic advances in myocardial perfusion imaging, we keep using an uncomfortable and sometimes impracticable patient position - supine with arms raised above the head (U). The purpose of this study was to investigate whether perfusion and functional cardiac gated SPECT scan results of acquisition U are equivalent to another position modality: supine with arms down at the sides of the trunk (D).
Methods and Results:
we performed U acquisition and in sequence D acquisition in 120 patients (pts) using a one-day MPI (rest-gated/stress), with 99mTc-sestamibi (370 MBq and 1110 MBq). Images were processed by the iterative reconstruction method (OSEM). Rest (R) and stress (S) studies were scored using 17-segments model. Functional parameters (left ventricular ejection fraction, and volumes) were automatically obtained by the quantitative gated SPECT (QGS) program. According to the degree of stress defects observed in U study, the patients were categorized in two subgroups: normal (SSS ≤ 3 or < 5%) and abnormal (SSS>3 or ≥ 5%). Shoulder/back pain occurred in 23.3% of U patients and in 5% of D. No significant differences between U and D were found for SSS (p = 0.82) and SRS (p = 0.74) in normal group. In abnormal group, good correlation was found between U and D modes for SSS (Rho = 0.95, p = 0.0001) and SRS (Rho = 0.96 p = 0.0001), but the mean SSS (12.53 ± 7.54) and SRS (10.60 ± 7.08) values of D were significantly lower (p < 0.05) than SSS (13.43 ± 6.81) and SRS (11.33 ± 6.97) of U mode. Function measurements presented good correlations, except for end-diastolic volume (p = 0.0001).
Conclusion:
although D mode appears to be more comfortable and presented a good correlation with U values of SSS and SRS, in abnormal pts, the extent and severity of defects can be underestimated. Considering clinical implications of an accurate perfusion measurement, the acquisition with the arms down should be avoided.
artifacts; image interpretation; computer-assisted; radiopharmaceuticals/diagnostic use; technetium Tc 99m sestamibi/diagnostic use; tomography; emission-computed; single-photon/methods
Objetivo:
no estudo de perfusão miocárdica, apesar dos avanços tecnológicos, continuamos utilizando o desconfortável e às vezes impraticável posicionamento em decúbito dorsal com braços acima da cabeça (U). O objetivo do estudo foi investigar se os resultados da perfusão e da função cardíacas obtidos com a aquisição U são equivalentes ao de outra modalidade de posição: com braços para baixo nas laterais do tronco (D).
Métodos e resultados:
adquirida a aquisição U, na sequência foi realizada a aquisição D em 120 pacientes (pts) em um único dia (repouso e estresse sincronizado ao ECG), com 99mTc-sestamibi (370 MBq e 1110 MBq). As imagens foram processadas usando reconstrução iterativa (OSEM). Cada estudo foi quantificado usando o modelo de 17-segmentos ao repouso (R) e ao estresse (S). Os parâmetros funcionais (fração de ejeção do ventrículo esquerdo e volumes) foram obtidas automaticamente pelo programa gated SPECT quantitativa (QGS). De acordo com a dimensão da alteração perfusional no estresse na aquisição U, os pacientes foram classificados em dois subgrupos: 80 pts normais (SSS ≤ 3 ou < 5%) e 40 pts anormais (SSS > 3 ou ≥ 5%). Dor no ombro e/ou nas costas ocorreu em 23,3% dos pacientes de U e em 5% de D. Não houve diferenças significativas entre U e D quanto ao SSS (p = 0,82) e SRS (p = 0,74) no grupo normal. No grupo anormal, boa correlação foi encontrada entre U e D para SSS (Rho = 0,95, p = 0,0001) e SRS (Rho = 0,96 p = 0,0001), entretanto a média dos valores de SSS (12,53 ± 7,54) e SRS (10,60 ± 7,08 ) de D foram significativamente menores (p < 0,05) que o SSS (13,43 ± 6,81) e SRS (11,33 ± 6,97) de U. As medidas de função apresentaram boas correlações, exceto para o volume diastólico final (p = 0,0001).
Conclusão:
embora o modo D pareça ser mais confortável e apresente boa correlação com valores de SSS e SRS do modo U, em pacientes anormais, a extensão e a gravidade dos defeitos podem ser subestimadas. Considerando as implicações clínicas da precisa avaliação da extensão de alterações na perfusão miocárdica, a aquisição com os braços para baixo deve ser evitada.
artefatos; interpretação de imagem assistida por computador; compostos radiofarmacêuticos/uso diagnóstico; tecnécio Tc 99m sestamibi/uso diagnóstico; tomografia computadorizada de emissão de fóton único/métodos
Introduction
Myocardial perfusion imaging (MPI) with single photon emission computed tomography
(SPECT) is part of routine clinical cardiology because of its value for the diagnosis
and risk stratification of coronary artery disease (CAD). 11 Underwood SR, Anagnostopoulos C, Cerqueira M, Ell PJ, et al. British
Cardiac Society; British Nuclear Cardiology Society; British Nuclear Medicine
Society; Royal College of Physicians of London; Royal College of Radiologists.
Myocardial perfusion scintigraphy: the evidence. Eur J Nucl Med Mol Imaging. 2004
Feb;31(2):261-91. Review. The ability of MPI for predicting future cardiac events has been
strongly demonstrated.22 Clark AN, Beller GA. The present role of nuclear cardiology in clinical
practice.Q J Nucl Med Mol Imaging. 2005 Mar;49(1):43-58. Review. ,33 Russell RR 3rd, Zaret BL.Nuclear cardiology: present and future. Curr
Probl Cardiol. 2006 Sep;31(9):557-629. Review. The extent and severity of stress perfusion
defects and the left ventricular ejection fraction are used to stratify the likelihood
of future events and to help define which patients have a short-term benefit from
revascularization.44 Hachamovitch R, Berman DS, Shaw LJ, Kiat H, Cohen I, Cabico JA, Friedman
J, et al. Incremental prognostic value of myocardial perfusion single photon emission
computed tomography for the prediction of cardiac death: differential stratification
for risk of cardiac death and myocardial infarction. Circulation. 1998;97(6):535-43.
Erratum in: Circulation. 1998;98(2):190.
5 Hachamovitch R, Hayes SW, Friedman JD, Cohen I, Berman DS. Comparison of
the short-term survival benefit associated with revascularization compared with
medical therapy in patients with no prior coronary artery disease undergoing stress
myocardial perfusion single photon emission computed tomography.
Circulation.2003;107(23):2900-7.-66 Hachamovitch R, Berman DS. The use of nuclear cardiology in clinical
decision making. Semin Nucl Med. 2005 Jan;35(1):62-72. Review.
SPECT images are affected by a variety of image degrading processes. Many technologic
advances in equipment and in reconstruction algorithms have been made to increase image
quality and to reduce the noise in the reconstructed SPECT images.77 Groch MW, Erwin WD. SPECT in the year 2000: basic principles. J Nucl Med
Technol. 2000;28(4):233-44.
8 Beekman FJ, Kamphuis C, King MA, Van Rijk PP, Viergever MA. Improvement
of image resolution and quantitative accuracy in clinical Single Photon Emission
Computed Tomography. Comput Med Imaging Graph. 2001 Mar-Apr;25(2):135-46. Review.
9 Germano G. Technical aspects of myocardial SPECT imaging. J Nucl Med.
2001; 42(10):1499-507.-1010 Paul AK, Nabi HA. Gated myocardial perfusion SPECT: basic principles,
technical aspects, and clinical applications. J Nucl Med Technol. 2004
Dec;32(4):179-87. Review. However,
since the introduction of cardiac SPECT imaging decades ago, to avoid increasing
attenuation and minimize the distance between the thorax and the camera, the patient
remains routinely positioned in supine position with the arms raised above the head.
1111 Hesse B, Tägil K, Cuocolo A, Anagnostopoulos C, Bardiés M, Bax J, et al.
EANM/ESC Group. EANM/ESC procedural guidelines for myocardial perfusion imaging in
nuclear cardiology. Eur J Nucl Med Mol Imaging. 2005
Jul;32(7):855-97.,1212 Hansen CL, Goldstein RA, Akinboboye OO, Berman DS, et al. American
Society of Nuclear Cardiology. Myocardial perfusion and function: single photon
emission computed tomography. J Nucl Cardiol. 2007
Nov-Dec;14(6):e39-60.
In clinical practice, this is a very uncomfortable and sometimes impracticable position. About 20 to 33 % of population present shoulder pain, and it is more common in women than man and increases with age.1313 McBeth J, Jones K. Epidemiology of chronic musculoskeletal pain. Best Pract Res Clin Rheumatol. 2007;21(3):403-25. Review. Many patients referred to myocardial scintigraphy have post-stroke motor sequel with consequent great difficulties in arms up positioning. Despite advances in SPECT hardware and software, in patients with difficulty to remain static, motion artifacts are a potential cause of error. Prone imaging has been reported to produce less patient motion and less inferior wall attenuation than supine imaging,1414 Peterson PN, Parker JA, Tepper MR, Hauser TH, English J, Danias PG. Prone SPECT myocardial perfusion imaging is associated with less cardiac drift during the acquisition duration than imaging in the supine position. Nucl Med Commun. 2005 Feb;26(2):115-7. but even in prone position, the arms are held above the head.
Just few studies1515 Cantinho G, Pena H, Veiga A, Marques V, Godinho F. Left arm position -
Effect of attenuation correction on myocardial tracer distribution. J. Nucl Cardiol.
1999;6(suppl):S28 (Presented at 4th Internatinal Conference of Nuclear Cardiology;
1999 April 18-21; Athens, Greece. Abstracts
16 Toma DM, White MP, Mann A, Phillips JM, Pelchat DA, et al. Influence of
arm positioning on rest/stress technetium-99m labeled sestamibi tomographic
myocardial perfusion imaging. J Nucl Cardiol. 1999; 6(2):163-8.-1717 Prvulovich EM, Jarritt PH, Lonn AHR, Vorontsova E, Bomanji JB, Ell PJ.
Influence of arm positioning on tomographic thallium-201 myocardial perfusion imaging
and the effect of attenuation correction. Eur J Nucl Med. 2000;
27(9):1349-1355. with no large series, analyzed the value of the
MPI with SPECT acquired with the patient's arms positioned at their sides. The findings
of these papers comparing myocardial perfusion results with the arms in up and down
position are contradictory. Thus, no strong evidence is available about the impact of
changes in resolution caused by the presence of the arm between the detector and the
heart and if this position can be used in routine clinical practice.
Furthermore, no study compared the reproducibility of gated SPECT measurement of myocardial function between these positions.
Our objective was to evaluate whether the perfusion and function results of MPI with SPECT acquired with the patients' arms in down position are comparable to the results obtained with the patients' arms in up position.
Methods
Patient population
In this prospective study, which was approved by the local ethics committee, 120 patients (83 men and 37 women; mean age 59 ± 11 years; range 36 to 82 years) who were referred for myocardial perfusion imaging were enrolled. All patients have signed informed consent.
The patients underwent either physical stress testing or pharmacologic stress (dipyridamole, adenosine or dobutamine) testing, according to their physician request. At the end of the study, we checked for the occurrence of back or arm pain and asked for the most comfortable acquisition position.
The thoracic circumference was measured excluding the arms (arms in up position) and including the arms (arms in down position). Clinical characteristics of the patient population are given in Table 1.
Data acquisition
All patients underwent same-day rest-stress technetium-99m labeled sestamibi (99mTc-sestamibi). SPECT myocardial perfusion scintigraphy and electrocardiographic-gated acquisition was performed only in stress imaging. In both parts (rest and stress), the first acquisition was done with the patients in supine position with the arms raised above the head (U) and immediately following the standard U position, the patient was placed in supine position with arms down at the sides of the trunk (D), as illustrated in Figure 1.
Part A: shows an example of a patient in supine position with the arms raised above the head (U), part B: shows the patient in supine position with arms down at the sides of the trunk (D).
A dose of 370 MBq of 99mTc-sestamibi (Cardiolite, Bristol-Myers Squibb Medical Imaging) was injected at rest and the acquisitions were performed 60 to 90 minutes later. Sixty-four patients performed a symptom-limited exercise treadmill test with standard Ellestad protocol. At near-maximal exercise 1110 MBq of 99mTc-sestamibi was injected. Fifty-five patients received the 1110 MBq of 99mTc-sestamibi during pharmacologic stress test with adenosine (140 mg/kg/min for 6 minutes) or four minutes after dipyridamole infusion (0.56 mg/kg/min for 4 minutes). One patient performed dobutamine stress test and 1110 MBq of 99mTc-sestamibi was injected during the infusion of 40 mg/kg/min of dobutamine. Stress perfusion scintigraphy acquisitions started 30 to 60 minutes later.
Paired studies were performed on commercially available SPECT systems: 37 Sophy camera-DST (GE Medical Systems), 38 on Cardio-MD (Adac-Phillips) and 45 on Forte (Adac-Phillips). The gamma cameras were equipped with low-energy, high-resolution, parallel-hole collimators. The data were acquired in 64 x 64 matrix for 64 projections over a 180-degree, non-circular orbit extending from 45º right anterior oblique to left posterior oblique. Each rest projection was acquired for 25 seconds and each stress gated-projection for 30 seconds, at 8 frames per R-R interval. The acquisition was repeated when patient motion occurred or if prominent activity adjacent to the heart was present. Attenuation correction, scatter correction and collimator response compensation were not applied.
All data were processed on the Pegasys 4.2 system (ADAC-Philips), Pegasys 5.0 system (ADAC-Philips) or Vision PowerStation system (GE Medical Systems) and transaxial tomograms were reconstructed by the use of iterative reconstruction method (OSEM) with an initial starting point obtained by filtered back-projection method. Vertical long-axis, horizontal long-axis, and short-axis tomograms were derived from transaxial images.
Perfusion and function analysis
The interpretation of U and D images were based on short-axis and vertical long-axis tomograms divided into 17 segments, totalizing 2040 segments. Rest and stress segments of U and D were interpreted blindly by three experienced nuclear medicine physicians separately, using a 5-point scoring system (0, normal; 1, mildly abnormal; 2, moderate abnormal; 3, severe abnormal; and 4, absence of segmental uptake). In both studies (U and D), summed scores for stress (SSS) and for rest (SRS) were calculated by summing of respective segmental scores. The SSS and SRS values were converted to a percentage (%) of the total myocardium involved dividing the summed scores by 68, the maximum potential score (4 × 17), and multiplying by 100.
The degree of stress defects was defined as: normal or minimally abnormal if involving less than 5% of the myocardium, mildly abnormal if involving 5 to 9%, moderately abnormal if involving 10 to 14% and severely abnormal if are greater than 15% of the myocardium area.
According to the degree of stress defects observed in U study, the patients were categorized in two subgroups: normal (SSS ≤ 3 or < 5%) and abnormal (SSS > 3 or ≥ 5%)
Left ventricular ejection fraction (LVEF), end-diastolic (EDV) and end-systolic volumes (ESV) were determined automatically by the use of QGS software (Cedars Sinai Medical Center, Los Angeles, Calif). We considered the lower normal limit of the EF 51% in women and 43% in men, the upper normal limit of EDV 102 mL for women and 149 mL for men and the upper normal limit of ESV 46 mL for women and 75 mL for men.
Statistical analysis
Data were expressed as mean, standard deviation (SD)
Overall agreement for perfusion findings between U and D modes was calculated on a per-segment basis, SSS and SRS values. Correlation of SSS and SRS values between U and D measurements were evaluated with the Spearman (Rho) value. Comparisons of variables were made using Wilcoxon signed-ranks test.
Comparisons between volumes and LVEFs of measurements made by repeated U and D SPECT were evaluated with the paired t test.
A P value of <0.05 was considered to indicate statistical significance.
Results
General findings
The thoracic circumference with the arms in down position was about 20% greater than the circumference with the arms in up position (mean 120 ± 9.8 cm and 100 ± 9.2 cm, respectively).
Fifty-three patients (44%) choose arms up position as more comfortable and sixty-seven ones (56%) preferred arms down position. Left arm, shoulder or back pain symptoms were reported for 28 patients (23%) in U mode against six patients (5%) in D mode.
Perfusion findings
The comparison of the 2040 segments obtained from the 120 studies demonstrated significant differences in the scores of U and D, at rest scans (P = 0.001; Wilcoxon signed-ranks test) and at stress scans (P = 0,0001; Wilcoxon signed-ranks test).
According U mode findings, 80 patients were classified as normal and 40 patients as abnormal (SSS ranging from 4 to 30).
No significant differences were found between U and D in normal group, at stress (P = 0.82) and rest (P = 0.74) scans.
In abnormal group, good correlation was observed between U and D values of SSS (Rho = 0.95; P = 0.0001) and SRS (Rho = 0.96; P = 0.0001) (Figure 2).
Correlation between SSSs and SRSs from U position and D position in the abnormal patients group.
However, SSS and SRS values of the matched scans differed significantly (p = 0.0083 and p = 0.0249; respectively). The mean value of SSS (12.53±7.54) and SRS (10.6±7.08) found in D acquisition were significantly lower than the mean values of SSS (13.43±6.81) and SRS (11.33±6.97) observed in U acquisition.
Among abnormal scans patients, 17 patients presented mildly abnormal, 8 ones moderately abnormal and 15 ones severely abnormal scans. Compared to U imaging, D position is associated with lower amount of perfusion abnormalities in 26 patients (22%) and in 14 patients (12%) an underestimation in the degree of the disease occurred.
Function findings
Two patients had non-regular heart rhythm, so the studies could not be performed with ECG triggering.
Among 118 patients evaluated, no statistical differences were found between U and D SPECT images in terms of LVEF and ESV (P > .05), but EDV differed significantly (P = 0.0001). The results of LVEF, ESV and EDV obtained in U and D mode are shown in Table 2.
Results of LVEF (left ventricular ejection fraction), EDV (end-diastolic volume) and ESV (end- systolic volume) values according to arm position
In standard U position, 95 patients presented normal LVEF values and 23 patients presented abnormal values according to gender. Comparing U and D modes, just seven patients changed the values of LVEF from normal to abnormal or from abnormal to normal in the repeated acquisitions.
Discussion
Just few data are available about how much the arm position really affects myocardial
perfusion images,1515 Cantinho G, Pena H, Veiga A, Marques V, Godinho F. Left arm position -
Effect of attenuation correction on myocardial tracer distribution. J. Nucl Cardiol.
1999;6(suppl):S28 (Presented at 4th Internatinal Conference of Nuclear Cardiology;
1999 April 18-21; Athens, Greece. Abstracts
16 Toma DM, White MP, Mann A, Phillips JM, Pelchat DA, et al. Influence of
arm positioning on rest/stress technetium-99m labeled sestamibi tomographic
myocardial perfusion imaging. J Nucl Cardiol. 1999; 6(2):163-8.-1717 Prvulovich EM, Jarritt PH, Lonn AHR, Vorontsova E, Bomanji JB, Ell PJ.
Influence of arm positioning on tomographic thallium-201 myocardial perfusion imaging
and the effect of attenuation correction. Eur J Nucl Med. 2000;
27(9):1349-1355. and this is the first study to analyze the
influence of arm position in the left ventricular function evaluation.
As expected, the arms down position appears to be more comfortable and cause less pain.
However, in our study the comparison of perfusion results was not so favorable.
In normal patients, no differences were found, but in abnormal patients, the SSS and SRS values were not equivalents. In abnormal patients, using arms down position, the amount of perfusion abnormalities can be underestimated. The underestimation of the myocardial lesion may interfere with clinical evaluation and management, since the severity and extent of cardiac SPECT perfusion abnormalities are clinically used in prognostic assessment and therapeutic management of patients with suspected or known coronary artery disease.
Toma et al.1616 Toma DM, White MP, Mann A, Phillips JM, Pelchat DA, et al. Influence of arm positioning on rest/stress technetium-99m labeled sestamibi tomographic myocardial perfusion imaging. J Nucl Cardiol. 1999; 6(2):163-8. assessed the influence of arm positioning in MPI using 99mTc-sestamibi in 41 patients. These authors concluded that arm positioning does not influence the interpretation of MPI with regard to image quality or defect location and extent. Our present study demonstrates results which are not consistent to this previous study. However, in this previous study, more than half of the population was interpreted as normal in both positions. In our study, also no significant differences were found in normal patients. The perfusion differences were observed exactly in the abnormal patients group. More comparable with our study findings, although using 201-tallium as tracer, prior investigation1717 Prvulovich EM, Jarritt PH, Lonn AHR, Vorontsova E, Bomanji JB, Ell PJ. Influence of arm positioning on tomographic thallium-201 myocardial perfusion imaging and the effect of attenuation correction. Eur J Nucl Med. 2000; 27(9):1349-1355. has reported that mean segmental relative uptake values for arms up position studies were higher than for arms down position studies across the majority of the myocardium, especially in anterolateral region. Another research,1515 Cantinho G, Pena H, Veiga A, Marques V, Godinho F. Left arm position - Effect of attenuation correction on myocardial tracer distribution. J. Nucl Cardiol. 1999;6(suppl):S28 (Presented at 4th Internatinal Conference of Nuclear Cardiology; 1999 April 18-21; Athens, Greece. Abstracts reported in abstract form, concluded that only the inferior wall of the heart was unaffected by left arm position. All of these studies evaluated small populations and none of them analyzed normal and abnormal groups distinctly.
SPECT images are affected by a variety of image degrading processes such as tissue attenuation. Superimposition of left arm over the lateral part of the heart may result in areas of relatively decreased activity in MPI. It might be one of the factors to explain the differences found, but certainly it's not the unique. If left arm attenuation were a main factor, it was expected that differences were noticed as much in normal as in abnormal groups. Furthermore, arm attenuation probably would cause an overestimation of the defects, not an underestimation as observed in our results.
The resolution of a gamma camera is depth-dependent, so increase in source-to-detector distance may result in variation of spatial resolution and loss of fine details in SPECT images. In our population, the thoracic circumference with the arms in down position was about 20 % greater than the circumference with the arms in up position. So increased source-to-detector distance in arms down acquisition, may explain the differences observed between arms down and up scan results, especially in abnormal patients where fine details are important in the interpretation of the images.
In our study, we performed a non-circular orbit, which has the advantage of minimizing the distance between the patient and the camera throughout the scan. However, due to increased variation of source-to-detector distance, resulting in variation of spatial resolution, some artifacts may be observed.
The acquisition projections that are tangential to a particular segment of the left ventricular wall are of primary importance in defining the shape and count distribution in that segment of the reconstructed slice. Abufadel et al.1818 Abufadel A, Eisner RL, Schafer RW. Differences due to collimator blurring in cardiac images with use of circular and elliptic camera orbits. J Nucl Cardiol. 2001 Jul-Aug;8(4):458-65. demonstrated that there was a significant loss of counts at the apex relative to other regions of the left ventricle. It results in more blurring in the projection of the apex and became larger as the orbit became more and more elliptic.
Physiological process may be evolved too. Hemodynamic variations occur during postural changes. Compared to supine imaging, the prone position is associated with a change in the location of the heart in the chest. Pump et al.1919 Pump B, Talleruphuus U, Christensen NJ, Warberg J, Norsk P.Effects of supine, prone, and lateral positions on cardiovascular and renal variables in humans.Am J Physiol Regul Integr Comp Physiol. 2002 Jul;283(1):R174-80. analyzing the effects of horizontal posture in cardiovascular variables demonstrated that prone position decreases stroke volume and increases sympathetic nervous activity, heart rate and total peripheral vascular resistance. Mckeough et al.2020 McKeough ZJ, Alison JA, Bye PT. Arm positioning alters lung volumes in subjects with COPD and healthy subjects. Aust J Physiother. 2003;49(2):133-7. observed that arm positioning above the head places the chest wall in an inflated position. Therefore, arm position alters lung volumes, so probably the heart position in the chest also is altered.
With regard to measurement of myocardial function, great degree of reproducibility between ejection fractions and end-systolic volume were observed in two positions.
The functional evaluation seems to be less affected by the arm position.
In our study, sequential acquisitions (arms up and arms down) were performed. Therefore variation in the time course of actual change in volumes and ejection fraction was minimized. However, left ventricular end-diastolic volume differed significantly, maybe related to physiological modifications.
Limitations
The supine position with arms above the head is routinely used for SPECT imaging; therefore it was considered the standard position. The comparison was not based on coronary angiography or other imaging modalities results. However, the aim of the study was not to define the accuracy of the procedures. The present study examines the repeatability of myocardial perfusion imagines in different arm positions.
Attenuation correction, scatter correction and collimator response compensation were not applied. These recent innovations in hardware and software have provided the capabilities for improved spatial resolution and contrast, and it's not established how much the use of these innovations could improve arms down scan diagnostic performance. 2121 Links JM, Becker LC, Rigo P, Taillefer R, Hanelin L, Anstett F, et al. Combined corrections for attenuation, depth-dependent blur, and motion in cardiac SPECT: a multicenter trial. J Nucl Cardiol. 2000 Sep-Oct;7(5):414-25.,2222 El Fakhri G, Buvat I, Benali H, Todd-Pokropek A, Di Paola R. Relative impact of scatter, collimator response, attenuation, and finite spatial resolution corrections in cardiac SPECT. J Nucl Med. 2000 Aug;41(8):1400-8.
Conclusion
Arms down position seem to be more comfortable and cause less pain than arms up position.
MPI SPECT acquired with the patients' arms in down position presented perfusion findings comparable with the results of patients' arms in up position, just in normal patients.
In abnormal scans perfusion abnormalities may be underestimated.
The functional evaluation seems to be less affected by the arm position. Great degree of reproducibility between ejection fractions and end-systolic volume were observed in two positions. Significant differences were found in end-diastolic volume measurement.
Therefore, considering the important therapeutic and prognostic implications of an accurate perfusion measurement, the cardiac SPECT acquisition with the arms in down position should be avoided.
-
Study conducted at the Department of Nuclear Medicine and Molecular Imaging, Heart Institute, Hospital das Clínicas, Faculty of Medicine, University of São Paulo - HCFMUSP, São Paulo, SP
References
-
1Underwood SR, Anagnostopoulos C, Cerqueira M, Ell PJ, et al. British Cardiac Society; British Nuclear Cardiology Society; British Nuclear Medicine Society; Royal College of Physicians of London; Royal College of Radiologists. Myocardial perfusion scintigraphy: the evidence. Eur J Nucl Med Mol Imaging. 2004 Feb;31(2):261-91. Review.
-
2Clark AN, Beller GA. The present role of nuclear cardiology in clinical practice.Q J Nucl Med Mol Imaging. 2005 Mar;49(1):43-58. Review.
-
3Russell RR 3rd, Zaret BL.Nuclear cardiology: present and future. Curr Probl Cardiol. 2006 Sep;31(9):557-629. Review.
-
4Hachamovitch R, Berman DS, Shaw LJ, Kiat H, Cohen I, Cabico JA, Friedman J, et al. Incremental prognostic value of myocardial perfusion single photon emission computed tomography for the prediction of cardiac death: differential stratification for risk of cardiac death and myocardial infarction. Circulation. 1998;97(6):535-43. Erratum in: Circulation. 1998;98(2):190.
-
5Hachamovitch R, Hayes SW, Friedman JD, Cohen I, Berman DS. Comparison of the short-term survival benefit associated with revascularization compared with medical therapy in patients with no prior coronary artery disease undergoing stress myocardial perfusion single photon emission computed tomography. Circulation.2003;107(23):2900-7.
-
6Hachamovitch R, Berman DS. The use of nuclear cardiology in clinical decision making. Semin Nucl Med. 2005 Jan;35(1):62-72. Review.
-
7Groch MW, Erwin WD. SPECT in the year 2000: basic principles. J Nucl Med Technol. 2000;28(4):233-44.
-
8Beekman FJ, Kamphuis C, King MA, Van Rijk PP, Viergever MA. Improvement of image resolution and quantitative accuracy in clinical Single Photon Emission Computed Tomography. Comput Med Imaging Graph. 2001 Mar-Apr;25(2):135-46. Review.
-
9Germano G. Technical aspects of myocardial SPECT imaging. J Nucl Med. 2001; 42(10):1499-507.
-
10Paul AK, Nabi HA. Gated myocardial perfusion SPECT: basic principles, technical aspects, and clinical applications. J Nucl Med Technol. 2004 Dec;32(4):179-87. Review.
-
11Hesse B, Tägil K, Cuocolo A, Anagnostopoulos C, Bardiés M, Bax J, et al. EANM/ESC Group. EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging. 2005 Jul;32(7):855-97.
-
12Hansen CL, Goldstein RA, Akinboboye OO, Berman DS, et al. American Society of Nuclear Cardiology. Myocardial perfusion and function: single photon emission computed tomography. J Nucl Cardiol. 2007 Nov-Dec;14(6):e39-60.
-
13McBeth J, Jones K. Epidemiology of chronic musculoskeletal pain. Best Pract Res Clin Rheumatol. 2007;21(3):403-25. Review.
-
14Peterson PN, Parker JA, Tepper MR, Hauser TH, English J, Danias PG. Prone SPECT myocardial perfusion imaging is associated with less cardiac drift during the acquisition duration than imaging in the supine position. Nucl Med Commun. 2005 Feb;26(2):115-7.
-
15Cantinho G, Pena H, Veiga A, Marques V, Godinho F. Left arm position - Effect of attenuation correction on myocardial tracer distribution. J. Nucl Cardiol. 1999;6(suppl):S28 (Presented at 4th Internatinal Conference of Nuclear Cardiology; 1999 April 18-21; Athens, Greece. Abstracts
-
16Toma DM, White MP, Mann A, Phillips JM, Pelchat DA, et al. Influence of arm positioning on rest/stress technetium-99m labeled sestamibi tomographic myocardial perfusion imaging. J Nucl Cardiol. 1999; 6(2):163-8.
-
17Prvulovich EM, Jarritt PH, Lonn AHR, Vorontsova E, Bomanji JB, Ell PJ. Influence of arm positioning on tomographic thallium-201 myocardial perfusion imaging and the effect of attenuation correction. Eur J Nucl Med. 2000; 27(9):1349-1355.
-
18Abufadel A, Eisner RL, Schafer RW. Differences due to collimator blurring in cardiac images with use of circular and elliptic camera orbits. J Nucl Cardiol. 2001 Jul-Aug;8(4):458-65.
-
19Pump B, Talleruphuus U, Christensen NJ, Warberg J, Norsk P.Effects of supine, prone, and lateral positions on cardiovascular and renal variables in humans.Am J Physiol Regul Integr Comp Physiol. 2002 Jul;283(1):R174-80.
-
20McKeough ZJ, Alison JA, Bye PT. Arm positioning alters lung volumes in subjects with COPD and healthy subjects. Aust J Physiother. 2003;49(2):133-7.
-
21Links JM, Becker LC, Rigo P, Taillefer R, Hanelin L, Anstett F, et al. Combined corrections for attenuation, depth-dependent blur, and motion in cardiac SPECT: a multicenter trial. J Nucl Cardiol. 2000 Sep-Oct;7(5):414-25.
-
22El Fakhri G, Buvat I, Benali H, Todd-Pokropek A, Di Paola R. Relative impact of scatter, collimator response, attenuation, and finite spatial resolution corrections in cardiac SPECT. J Nucl Med. 2000 Aug;41(8):1400-8.
Publication Dates
-
Publication in this collection
Jul-Aug 2014
History
-
Received
17 July 2012 -
Accepted
30 Jan 2013