Comparison of Locked Plates and Blade Plates for Varus Osteotomy of the Proximal Femur by the Finite Element Method

Abstract

Objective:

The present study compared the difference in load and pressure distribution behavior of the blade plate and locked plate for varus osteotomy of the proximal femur per the finite element method.

Methods:

Modeling was performed by scanning a medium-sized left femur with medial valgus deformity made of polyurethane.

Results:

The stiffness of the locked plate is higher compared with that of the blade plate. However, this difference was not significant. In addition, the locked plate has proximal locking screws to ensure that the bending moments on the screws are smaller during loading.

Conclusion:

In summary, both plates are well-established and effective. However, the study using the finite element method plays a fundamental role in understanding the load and pressure distribution of the implant. Moreover, it opens up new possibilities for further studies, including surgical proposals and customized implant materials.

Keywords
orthopedics; traumatology; femur; bone plates

Resumo

Objetivo:

Testar a diferença do comportamento de distribuição de cargas e pressões da placa lâmina comparativamente com a placa bloqueada para osteotomia varizante do fêmur proximal através do método de elementos finitos.

Métodos:

A modelagem foi realizada através do escaneamento de um fêmur esquerdo com deformidade medial em valgo fabricado em poliuretano de tamanho médio.

Resultados:

Como resultados, pode-se inferir que a rigidez da placa bloqueada é maior do que a da placa lâmina. No entanto, essa diferença não foi significativa e, além disso, a placa bloqueada possui parafusos de travamento proximal para garantir que os momentos de flexão que agem nos parafusos sejam ainda menores durante o carregamento.

Conclusão:

Em síntese, ambos os materiais são consagrados e eficazes para serem utilizados, porém o estudo pelo método de elementos finitos apresenta papel importante para compreendermos a situação de distribuição de cargas e pressões do implante e abre novas possibilidades para novos estudos, como, por exemplo, o estudo da proposta cirúrgica e materiais a serem implantados de forma individual e personalizada.

Palavras-chave
ortopedia; traumatologia; fêmur; placas ósseas

Introduction

Varus osteotomies of the proximal femur are pediatric reconstructive surgeries widely performed in patients with neurological abnormalities, congenital hip diseases, sequelae, and acquired conditions.11 Joeris A, Audigé L, Ziebarth K, Slongo T. The Locking Compression Paediatric Hip Plate: technical guide and critical analysis. Int Orthop 2012;36(11):2299–2306,22 Ruzbarsky JJ, Swarup I, Garner MR, Meyers KN, Edobor-Osula F, Widmann RF, et al. Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies. HSS J 2020; 16(01):81–85 Fixation of these osteotomies may use several implants, including blade plates, dynamic compression plates (DCP), locked plates for the proximal femur, unilateral and circular external fixators, Kirschner wires, and screws.11 Joeris A, Audigé L, Ziebarth K, Slongo T. The Locking Compression Paediatric Hip Plate: technical guide and critical analysis. Int Orthop 2012;36(11):2299–2306,22 Ruzbarsky JJ, Swarup I, Garner MR, Meyers KN, Edobor-Osula F, Widmann RF, et al. Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies. HSS J 2020; 16(01):81–85,33 Jain A, Thompson JM, Brooks JT, Ain MC, Sponseller PD. Implant-related Fractures in Children With Proximal Femoral Osteotomy: Blade Plate Versus Screw-side Plate Constructs. J Pediatr Orthop 2016;36(01):e1–e5 The surgical complexity of these procedures resulted in advanced synthesis materials to facilitate surgery and improve outcomes.11 Joeris A, Audigé L, Ziebarth K, Slongo T. The Locking Compression Paediatric Hip Plate: technical guide and critical analysis. Int Orthop 2012;36(11):2299–2306,22 Ruzbarsky JJ, Swarup I, Garner MR, Meyers KN, Edobor-Osula F, Widmann RF, et al. Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies. HSS J 2020; 16(01):81–85,44 Crist BD, Khalafi A, Hazelwood SJ, Lee MA. A biomechanical comparison of locked plate fixation with percutaneous insertion capability versus the angled blade plate in a subtrochanteric fracture gap model. J Orthop Trauma 2009;23(09):622–627

The most used materials in varus osteotomy of the proximal femur include locked and blade plates.11 Joeris A, Audigé L, Ziebarth K, Slongo T. The Locking Compression Paediatric Hip Plate: technical guide and critical analysis. Int Orthop 2012;36(11):2299–2306,22 Ruzbarsky JJ, Swarup I, Garner MR, Meyers KN, Edobor-Osula F, Widmann RF, et al. Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies. HSS J 2020; 16(01):81–85,33 Jain A, Thompson JM, Brooks JT, Ain MC, Sponseller PD. Implant-related Fractures in Children With Proximal Femoral Osteotomy: Blade Plate Versus Screw-side Plate Constructs. J Pediatr Orthop 2016;36(01):e1–e5,44 Crist BD, Khalafi A, Hazelwood SJ, Lee MA. A biomechanical comparison of locked plate fixation with percutaneous insertion capability versus the angled blade plate in a subtrochanteric fracture gap model. J Orthop Trauma 2009;23(09):622–627 Clinically, blade and locked plates have no statistical difference in the risk of failure (breakage),33 Jain A, Thompson JM, Brooks JT, Ain MC, Sponseller PD. Implant-related Fractures in Children With Proximal Femoral Osteotomy: Blade Plate Versus Screw-side Plate Constructs. J Pediatr Orthop 2016;36(01):e1–e5 which is the worst complication related to the choice of implant. Biomechanical tests with load application in experimental models show that locked plates with a support screw have higher axial resistance, lower resistance to torsion, and irreversible equivalent strain to deforming cycles compared with blade plates.44 Crist BD, Khalafi A, Hazelwood SJ, Lee MA. A biomechanical comparison of locked plate fixation with percutaneous insertion capability versus the angled blade plate in a subtrochanteric fracture gap model. J Orthop Trauma 2009;23(09):622–627

The finite element method (FEM) is a mathematical tool used to solve problems in engineering because it explores the effects of load application on the bone and its biomechanical behavior. One of its main advantages is its potential use in solids with irregular geometry presenting heterogeneous material properties. The introduction of FEM in orthopedic biomechanics occurred in the 1970s; since then, the number of publications on load analysis in bones, arthroplasty, and osteosynthesis has been increasing.55 Soni JF, Santili C, Lancellotti CLP, Hecke MB, Almeida FR, Karam LZ. Análise comparativa em modelo computadorizado bidimensional com simulação do emprego de hastes flexíveis de aço e titânio na fratura do fêmur da criança utilizando o método dos elementos finitos. Rev Bras Ortop 2008;43(05):183–192

The present study aims to analyze, using FEM, the load and pressure distribution behavior when implanting blade or locked plates for varus osteotomy of the proximal femur.

Methods

Model Generation

Modeling was based on a computed tomography scanning of a left femur made from medium-sized polyurethane (Nacional Ossos, Brazil, reference number 2025 DMVL) and presenting a medial valgus deformity. The software used for plate scanning and modeling was SolidWorks (Dassault Systems SolidWorks Corp., Waltham, MA, USA) (►Fig. 1a).

Fig. 1
Valgus femur from Nacional Ossos for scanning. (A) Simplified femur and (B) varus osteotomy. Source: Data constructed by the authors using SolidWorks software.

For femur simplification, we sectioned the diaphysis and discarded the distal part since it would not be studied. A varus osteotomy was performed for 20° correction. This simplification provided a gain in calculation processing by FEM without result distortion.66 Nag P, Chanda S. Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study. Med Biol Eng Comput 2021;59(02):271–285 The environment (femur) was the same for both studies, not distorting or favoring any data (►Fig. 1b).

Fixation of the first femur employed a locked plate for the proximal femur with a 10-mm step, 100°, and three holes (Techimport, Rio Claro, SP, Brazil, reference number TI030.1003.100). Proximal fragment fixation used 3.5 × 50-mm diameter locked screws in two holes and a 3.5 × 40-mm locked screw in the third hole. Distal fragment fixation used 1 3.5 × 30-mm and 2 3.5 × 30-mm locked screws (►Fig. 2a). Fixation of the second femur employed a blade plate for the proximal femur with a 10-mm step, 100°, 3 holes, 50-mm blade (Techimport, Rio Claro, SP, Brazil, reference number ref. TI030.1010.350). The proximal hole received 3.5 × 50 mm-diameter locked screws, while the distal non-locked plate holes received two 3.5 × 30-mm non-locked compression screws and a 3.5 × 30-mm locked screw for distal fixation (►Fig. 2b).

Fig. 2
Locked plate fixed to the femur with screws (A). Blade plate fixed to the femur with screws (B). Source: Data constructed by the authors using SolidWorks software.

Plate and screw assembly through visual positioning used the SolidWorks software (Dassault Systems SolidWorks Corp., Waltham, MA, USA). We created a 27-mm distance restriction between the proximal axis of the locked plate screw and the central axis of the blade plate and a 71-mm distance restriction between the medial face of the proximal femur and the inner face of the plates. These restrictions standardized the flexor moment generated by applying the axial force and the displacement stress assessment (►Fig. 2).

Material Properties

Plates, screws, cortical, and cancellous bone models were homogeneous, linear, elastic, and isotropic, based on the properties described in the literature (►Table 1). The yield stress was set at 795 MPa, the limit stress of the elastic region of the titanium alloy.77 Maurer P, Holweg S, Schubert J. Finite-element-analysis of different screw-diameters in the sagittal split osteotomy of the mandible. J Craniomaxillofac Surg 1999;27(06):365–372,88 Janeček M, Novy F, Harcuba P, Stráský J, Trško L, Mhaede M, et al. The Very High Cycle Fatigue Behaviour of Ti-6Al-4V Alloy. Acta Phys Pol A 2015;128:497–503

Table 1
Elastic modulus, Poisson ratio, and number of fatigue cycles for all materials (data from Maurer et al., 199955 Soni JF, Santili C, Lancellotti CLP, Hecke MB, Almeida FR, Karam LZ. Análise comparativa em modelo computadorizado bidimensional com simulação do emprego de hastes flexíveis de aço e titânio na fratura do fêmur da criança utilizando o método dos elementos finitos. Rev Bras Ortop 2008;43(05):183–192 and Janecek et al., 201566 Nag P, Chanda S. Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study. Med Biol Eng Comput 2021;59(02):271–285)

Simulation Parameters (Load, Mesh, and Contact Conditions)

Fig. 2 shows the applied forces of 450, 500, 550, and 600 N considered a normal loading position, which assumes that the load vector has an angle of 8° of adduction with the hip longitudinal axis in the plane.66 Nag P, Chanda S. Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study. Med Biol Eng Comput 2021;59(02):271–285,99 Marco M, Giner E, Larraínzar R, Caeiro J, Miguélez H. Analysis of the variation of the mechanical behavior of the proximal extremity of the femur using the XFEM method (eXtended Finite Element Method). J Osteoporos Min Metabol 2016;8:61–69,1010 Perry J. Gait Analysis: Normal and Pathological Function. Thorofare, NJ: SLACK, Inc.; 1992:11–12,1111 Wang J, Ma JX, Lu B, Bai HH, Wang Y, Ma XL. Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS). Orthop Traumatol Surg Res 2020;106(01):95–101,1212 Taddei F, Viceconti M, Manfrini M, Toni A. Mechanical strength of a femoral reconstruction in paediatric oncology: a finite element study. Proc Inst Mech Eng H 2003;217(02):111–119,1313 Kainz H, Killen BA, Wesseling M, Perez-Boerema F, Pitto L, Aznar JMG, et al. A multi-scale modelling framework combining musculoskeletal rigid-body simulations with adaptive finite element analyses, to evaluate the impact of femoral geometry on hip joint contact forces and femoral bone growth. PLoS One 2020;15(07): e0235966,1414 Lewis GS, Mischler D, Wee H, Reid JS, Varga P. Finite Element Analysis of Fracture Fixation. Curr Osteoporos Rep 2021;19(04):403–416,1515 Zhang QY, Zhou Y, Li XX, Wang L, Cui XH, Wang SQ. Accelerated Formation of Tribo-oxide Layer and Its Effect on Sliding Wear of a Titanium Alloy. Tribol Lett 2016;63(01):. Doi: 10.1007/s11249-016-0694-7
https://doi.org/10.1007/s11249-016-0694-...
,1616 Adamus J, Lacki P, Motyka M. Kubiak KInvestigation of sheettitanium drawability. In: Ti 2011 Proceedings of the 12th World Conference on Titanium, Bei- jing, China, june 19 to ] une 24, 2011:337–341,1717 Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res 2009;467(08):1972–1980 In a clinical situation, both values are greater than those produced by touch support with crutches and should provide sufficient postoperative stability.1010 Perry J. Gait Analysis: Normal and Pathological Function. Thorofare, NJ: SLACK, Inc.; 1992:11–12,1111 Wang J, Ma JX, Lu B, Bai HH, Wang Y, Ma XL. Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS). Orthop Traumatol Surg Res 2020;106(01):95–101

The force application frequency was 1Hz, considering walking 1 step per second.1111 Wang J, Ma JX, Lu B, Bai HH, Wang Y, Ma XL. Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS). Orthop Traumatol Surg Res 2020;106(01):95–101,1212 Taddei F, Viceconti M, Manfrini M, Toni A. Mechanical strength of a femoral reconstruction in paediatric oncology: a finite element study. Proc Inst Mech Eng H 2003;217(02):111–119 Since this is a temporary fixation device, the plate must withstand at least 1066 Nag P, Chanda S. Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study. Med Biol Eng Comput 2021;59(02):271–285 cycles, equivalent to ~ 1 year, with this frequency.1313 Kainz H, Killen BA, Wesseling M, Perez-Boerema F, Pitto L, Aznar JMG, et al. A multi-scale modelling framework combining musculoskeletal rigid-body simulations with adaptive finite element analyses, to evaluate the impact of femoral geometry on hip joint contact forces and femoral bone growth. PLoS One 2020;15(07): e0235966 We constrained the transverse face of the femoral shaft in all translational degrees of freedom (►Fig. 2).

We merged the mesh models with three-dimensional quadratic tetrahedral elements in SolidWorks software of 1 mm for plates, 0.5 mm for screws, and 5 mm for bone.1414 Lewis GS, Mischler D, Wee H, Reid JS, Varga P. Finite Element Analysis of Fracture Fixation. Curr Osteoporos Rep 2021;19(04):403–416 Bonding contact occurred between bone tissue and implants, except for osteotomy interfaces (►Fig. 3).

Fig. 3
Mesh applied to three-dimensional locked plate models (A). Mesh applied to the three-dimensional blade plate models (B). Source: Data constructed by the authors using SolidWorks software.

Plate and screw contact surfaces had a friction coefficient of 0.34.1515 Zhang QY, Zhou Y, Li XX, Wang L, Cui XH, Wang SQ. Accelerated Formation of Tribo-oxide Layer and Its Effect on Sliding Wear of a Titanium Alloy. Tribol Lett 2016;63(01):. Doi: 10.1007/s11249-016-0694-7
https://doi.org/10.1007/s11249-016-0694-...
,1717 Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res 2009;467(08):1972–1980 The screw and plate contacted at the surface of the screw head and the surface of the countersunk in the plate holes, all with a restriction to not allow penetration between them. As for the contacts, locked screws in plate holes were deemed connected and fixed.1717 Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res 2009;467(08):1972–1980

Results

Von Mises stress and displacement for the blade plate were higher compared with the locked plate (►Tables 2, 3, 4, 5). ►Fig. 4 demonstrates the displacement, in millimeters, of locked and blade plates after applying 450, 500, 550, and 600N forces. ►Fig. 5 shows the von Mises stress in MPa of locked and blade plates after applying 450, 500, 550, and 600N forces. ►Fig. 6 shows the von Mises stress in MPa on the bone fixed with locked or blade plates after applying 450, 500, 550, and 600N forces.

Table 2
von Mises stress and strain for each plate and bone at a 450 N force
Table 3
von Mises stress and strain for each plate and bone at a 500 N force
Table 4
von Mises stress and strain for each plate and bone at a 550 N force
Table 5
von Mises stress and strain for each plate and bone at a 600 N force

Fig. 4
Locked and blade plate displacement versus applied forces. Source: Data constructed by the authors using SolidWorks software.

Fig. 5
Locked and blade plate von Mises stress versus applied forces. Source: Data constructed by the authors using SolidWorks software.

Fig. 6
Von Mises stress in the bone versus applied forces. Source: Data constructed by the authors using SolidWorks software.

The blade plate caused the highest displacement (►Table 2). The highest displacement site for locked and blade plates was at the point of vertical force application at the femoral head (►Fig. 7). The blade plate had the highest von Mises stress (►Table 2). For the locked plate, the highest stress concentration was in the locking area between the proximal screw and the plate. For the blade plate, the region with the highest stress concentration was at the beginning of the advancement, at the osteotomy level (►Fig. 8).

Fig. 7
Locked plate total displacement, F = 450 N (A), blade plate total displacement, F = 450 N (B), locked plate total displacement, F = 500 N (C), blade plate total displacement, F = 500 N (D), locked plate total displacement, F = 550 N (E), blade plate total displacement, F = 550 N (F), locked plate total displacement, F = 600 N (G), blade plate total displacement, F = 600 N (H). Source: Data constructed by the authors using SolidWorks software.

Fig. 8
Von Mises stress for the locked plate, F = 450 N (A), von Mises stress for the blade plate, F = 450 N (B), von Mises stress for the locked plate, F = 500 N (C), von Mises stress for the blade plate, F = 500 N (D), von Mises stress for the locked plate, F = 550 N (E), von Mises stress for the blade plate, F = 550 N (F), von Mises stress for the locked plate, F = 600 N (G), von Mises stress for the blade plate, F = 600 N (H). Source: Data constructed by the authors using SolidWorks software.

The bone fixed with the blade plate showed the highest von Mises stress (►Table 2). The region with the highest bone stress concentration was the osteotomy, in the corner of the proximal part with the spongy bone surface in the distal part. The stress occurred at the same regions for all applied forces (►Fig. 9).

Fig. 9
Von Mises stress for locked plate, F = 450 N (A), von Mises stress for blade plate, F = 450 N (B). Source: Data constructed by the authors using SolidWorks software.

Fixation with the locked plate withstand more cycles compared with the blade plate for loads of 450, 500, 550, and 600N in a 1Hz frequency (►Table 6).

Table 6
Number of cycles for locked and blade plates at a 1 Hz frequency

Fig. 10 shows the number of cycles for locked and blade plates under 450, 500, 550, and 600N loads. Furthermore, it demonstrates that after applying a load >600 N, locked and blade plates tend to fail in the first cycle.

Fig. 10
Number of cycles versus applied forces. Source: Data constructed by the authors using SolidWorks software.

Discussion

Conduction of clinical investigations on the forces required to stimulate bone healing is complex. The finite element method is widely used in the medical and orthopedic field because it provides a comprehensive view of vector dissolution in undermined structures and allows for accurate failure detection. Moreover, it may avoid unnecessary costs when failure is identifiable only after structural design or manufacturing. The finite element method also reduces the time from the first conceptual design to production, as the creation of a large number of experimental specimens becomes unnecessary. Finite element method analysis provides access to information difficult to obtain under laboratory conditions, such as predicted stress distribution and material strength, which are fundamental to assessing fatigue strength.1818 Faria FF, Gruhl CEM, Ferro RR, Rached RN, Soni JF, Trevilatto P. Finite Element Analysis of a Controlled Dynamization Device for External Circular Fixation. Rev Bras Ortop 2021;56(01):36–41

In the present comparative biomechanical study, we investigated the flexural stiffness of the locked and blade plates for proximal femoral varus osteotomy. Our data suggest the locked plate presents higher stiffness compared with the blade plate. However, this difference was not significant (►Tables 2, 3, 4, 5). In addition, the locked plate has proximal locking screws to ensure that the bending moments acting on screws are lower during loading.

Locked plate failure occurs with significantly higher forces compared with the blade plate. The locked plate has significantly higher stiffness and load to failure values due to the nature of its design.At19mm,the locked plate is 8mmwider in the proximal area, thus withstanding more stress than the blade plate. Both plates have approximately the same thickness of 3mm. This results in a higher moment of inertia on the locked plate, which reflects in the proximal force results.

Consistent with our data, a comparative analysis by Radtke et al.1919 Radtke K, Goede F, Schwarze M, Paes P, Ettinger M, Welke B. Fixation stability and stiffness of two implant systems for proximal femoral varization osteotomy. Appl Sci (Basel) 2020; •••;. Doi: 10.5867.10.3390/app10175867
https://doi.org/10.5867.10.3390/app10175...
found mean values of 554N for locked plates and 399N for blade plates. These authors used synthetic bone and plates. Forward et al.2020 Forward DP, Doro CJ, O'Toole RV, Kim H, Floyd JCP, Sciadini MF, et al. A biomechanical comparison of a locking plate, a nail, and a 95° angled blade plate for fixation of subtrochanteric femoral fractures. J Orthop Trauma 2012;26(06):334-340 reported mean values of 620N for locked plates and 450N for blade plates in a study performed with cadavers.

Femoral stress distribution was consistent with a study by Sim et al.2121 Sim E, Freimüller W, Reiter TJ. Finite element analysis of the stress distributions in the proximal end of the femur after stabilization of a pertrochanteric model fracture: a comparison of two implants. Injury 1995;26(07):445-449 reporting a higher stress concentration between the proximal and distal parts at the point of separation osteotomy.

Absolute displacement values were higher for the blade plate, which had two areas of load application with higher variation (femur head). However, there is a biomechanical advantage associated with the stress areas of the system. In locked plates, the stress area of the joint between the plate and the bone is at the proximal screw. On the other hand, in blade plates, this area is at the blade advance region. Thus, in blade plates, all the stress concentrates in the osteotomy region, which usually constitutes an obstacle to bone consolidation resulting from the Wolff law.2222 Wolff J. Das Gesetz der Transformation der Knochen. Berlin: Hirschwald; 1892 The success of biological bone healing depends on a favorable mechanical environment. In addition,the Wolff law and the Perren strain theory allow using several osteosynthesis systems to promote adequate stabilization and the differentiation of various cell types at the bone healing site.2222 Wolff J. Das Gesetz der Transformation der Knochen. Berlin: Hirschwald; 1892,2323 Perren S, Boitzy A. Cellular differentiation and bone biomechanics during the consolidation of a fracture. Clin Anat 1978;1: 13-28

The relative stability indicated for comminuted diaphyseal or extra-articular fractures allows for some controlled mobility at the fracture site and exuberant formation of bone callus, which characterizes an indirect or endochondral ossification. Direct or intramembranous ossification is recommended to avoid developing bulky bone calluses in joint fractures, following absolute fixation with greater rigidity.2424 Glatt V, Matthys R. Adjustable stiffness, external fixator for the rat femur osteotomy and segmental bone defect models. J Vis Exp 2014;9(92):e51558

The locked plate system also presents a larger stress area at the osteotomy region but produced lower absolute values than the blade plate. Thus, we inferred that locked plates create a more favorable biomechanical situation for bone consolidation.

Our study has limitations regarding FEM, which considers structures as a gathering of small particles of finite quantity, the so-called finite elements, connected to a finite number of points, the nodes, or nodal points. These particles represent the approximate result of every discretized system.2525 Zienkiewicz OC. The finite element method. 3rd expanded and rev. ed. London; New York: McGraw-Hill; c1977 The finite element method allows the evaluation of the approximate stress distribution in a structure, observing the element strain through visualization and image interpretation on a color chart.55 Soni JF, Santili C, Lancellotti CLP, Hecke MB, Almeida FR, Karam LZ. Análise comparativa em modelo computadorizado bidimensional com simulação do emprego de hastes flexíveis de aço e titânio na fratura do fêmur da criança utilizando o método dos elementos finitos. Rev Bras Ortop 2008;43(05):183–192

The present study demonstrated that the blade plate resisted fewer cycles when the loading forces were lower. However, locked and blade plates tend to fail under the highest applied force. Thus, despite being exposed to controlled load situations, the locked plate was more resistant to implant failure. In exacerbated load situations, locked and blade plates tend to fail.

Both osteosynthesis implants are consecrated and effective. Nevertheless, our study using FEM shows a fundamental role in understanding the biomechanical situation of the implant. It also opens up new possibilities for further studies, including surgical proposals and customized implant materials. Therefore, our study corroborates a hypothesis raised by common sense, that is, the superiority of the locked plate compared with blade plates for varus osteotomy of the proximal femur. Still, there is no unanimity in the literature, especially regarding clinical outcomes.

Our study may yield future models with laboratory biomechanical tests to prove the differences between locked and blade plate fixation systems.

Conclusion

Both osteosynthesis implants are well-established and effective in clinical practice. However, our study applying FEM demonstrated the biomechanical superiority of the locked plate compared with the blade plate for proximal femoral varus osteotomy in the proposed model.

  • Financial Support
    The present study received no specific financial support from public, commercial, or not-for-profit sources.

References

  • 1
    Joeris A, Audigé L, Ziebarth K, Slongo T. The Locking Compression Paediatric Hip Plate: technical guide and critical analysis. Int Orthop 2012;36(11):2299–2306
  • 2
    Ruzbarsky JJ, Swarup I, Garner MR, Meyers KN, Edobor-Osula F, Widmann RF, et al. Biomechanical Comparison of Two Pediatric Blade Plate Designs in Proximal Femoral Osteotomies. HSS J 2020; 16(01):81–85
  • 3
    Jain A, Thompson JM, Brooks JT, Ain MC, Sponseller PD. Implant-related Fractures in Children With Proximal Femoral Osteotomy: Blade Plate Versus Screw-side Plate Constructs. J Pediatr Orthop 2016;36(01):e1–e5
  • 4
    Crist BD, Khalafi A, Hazelwood SJ, Lee MA. A biomechanical comparison of locked plate fixation with percutaneous insertion capability versus the angled blade plate in a subtrochanteric fracture gap model. J Orthop Trauma 2009;23(09):622–627
  • 5
    Soni JF, Santili C, Lancellotti CLP, Hecke MB, Almeida FR, Karam LZ. Análise comparativa em modelo computadorizado bidimensional com simulação do emprego de hastes flexíveis de aço e titânio na fratura do fêmur da criança utilizando o método dos elementos finitos. Rev Bras Ortop 2008;43(05):183–192
  • 6
    Nag P, Chanda S. Biomechanical design prognosis of two extramedullary fixation devices for subtrochanteric femur fracture: a finite element study. Med Biol Eng Comput 2021;59(02):271–285
  • 7
    Maurer P, Holweg S, Schubert J. Finite-element-analysis of different screw-diameters in the sagittal split osteotomy of the mandible. J Craniomaxillofac Surg 1999;27(06):365–372
  • 8
    Janeček M, Novy F, Harcuba P, Stráský J, Trško L, Mhaede M, et al. The Very High Cycle Fatigue Behaviour of Ti-6Al-4V Alloy. Acta Phys Pol A 2015;128:497–503
  • 9
    Marco M, Giner E, Larraínzar R, Caeiro J, Miguélez H. Analysis of the variation of the mechanical behavior of the proximal extremity of the femur using the XFEM method (eXtended Finite Element Method). J Osteoporos Min Metabol 2016;8:61–69
  • 10
    Perry J. Gait Analysis: Normal and Pathological Function. Thorofare, NJ: SLACK, Inc.; 1992:11–12
  • 11
    Wang J, Ma JX, Lu B, Bai HH, Wang Y, Ma XL. Comparative finite element analysis of three implants fixing stable and unstable subtrochanteric femoral fractures: Proximal Femoral Nail Antirotation (PFNA), Proximal Femoral Locking Plate (PFLP), and Reverse Less Invasive Stabilization System (LISS). Orthop Traumatol Surg Res 2020;106(01):95–101
  • 12
    Taddei F, Viceconti M, Manfrini M, Toni A. Mechanical strength of a femoral reconstruction in paediatric oncology: a finite element study. Proc Inst Mech Eng H 2003;217(02):111–119
  • 13
    Kainz H, Killen BA, Wesseling M, Perez-Boerema F, Pitto L, Aznar JMG, et al. A multi-scale modelling framework combining musculoskeletal rigid-body simulations with adaptive finite element analyses, to evaluate the impact of femoral geometry on hip joint contact forces and femoral bone growth. PLoS One 2020;15(07): e0235966
  • 14
    Lewis GS, Mischler D, Wee H, Reid JS, Varga P. Finite Element Analysis of Fracture Fixation. Curr Osteoporos Rep 2021;19(04):403–416
  • 15
    Zhang QY, Zhou Y, Li XX, Wang L, Cui XH, Wang SQ. Accelerated Formation of Tribo-oxide Layer and Its Effect on Sliding Wear of a Titanium Alloy. Tribol Lett 2016;63(01):. Doi: 10.1007/s11249-016-0694-7
    » https://doi.org/10.1007/s11249-016-0694-7
  • 16
    Adamus J, Lacki P, Motyka M. Kubiak KInvestigation of sheettitanium drawability. In: Ti 2011 Proceedings of the 12th World Conference on Titanium, Bei- jing, China, june 19 to ] une 24, 2011:337–341
  • 17
    Eberle S, Gerber C, von Oldenburg G, Hungerer S, Augat P. Type of hip fracture determines load share in intramedullary osteosynthesis. Clin Orthop Relat Res 2009;467(08):1972–1980
  • 18
    Faria FF, Gruhl CEM, Ferro RR, Rached RN, Soni JF, Trevilatto P. Finite Element Analysis of a Controlled Dynamization Device for External Circular Fixation. Rev Bras Ortop 2021;56(01):36–41
  • 19
    Radtke K, Goede F, Schwarze M, Paes P, Ettinger M, Welke B. Fixation stability and stiffness of two implant systems for proximal femoral varization osteotomy. Appl Sci (Basel) 2020; •••;. Doi: 10.5867.10.3390/app10175867
    » https://doi.org/10.5867.10.3390/app10175867
  • 20
    Forward DP, Doro CJ, O'Toole RV, Kim H, Floyd JCP, Sciadini MF, et al. A biomechanical comparison of a locking plate, a nail, and a 95° angled blade plate for fixation of subtrochanteric femoral fractures. J Orthop Trauma 2012;26(06):334-340
  • 21
    Sim E, Freimüller W, Reiter TJ. Finite element analysis of the stress distributions in the proximal end of the femur after stabilization of a pertrochanteric model fracture: a comparison of two implants. Injury 1995;26(07):445-449
  • 22
    Wolff J. Das Gesetz der Transformation der Knochen. Berlin: Hirschwald; 1892
  • 23
    Perren S, Boitzy A. Cellular differentiation and bone biomechanics during the consolidation of a fracture. Clin Anat 1978;1: 13-28
  • 24
    Glatt V, Matthys R. Adjustable stiffness, external fixator for the rat femur osteotomy and segmental bone defect models. J Vis Exp 2014;9(92):e51558
  • 25
    Zienkiewicz OC. The finite element method. 3rd expanded and rev. ed. London; New York: McGraw-Hill; c1977

Publication Dates

  • Publication in this collection
    20 May 2024
  • Date of issue
    Jan-Feb 2024

History

  • Received
    28 Nov 2022
  • Accepted
    24 Jan 2023
Sociedade Brasileira de Ortopedia e Traumatologia Al. Lorena, 427 14º andar, 01424-000 São Paulo - SP - Brasil, Tel.: 55 11 2137-5400 - São Paulo - SP - Brazil
E-mail: rbo@sbot.org.br