Open-access Biomechanical evaluation of a new femoral stem design for total hip replacement in a canine model

ABSTRACT

Purpose  To evaluate the biomechanical properties of a novel total hip replacement femoral stem.

Methods  Eight pairs of femurs from dog cadavers were used. The femurs were separated into different groups. A novel femoral stem with a convex proximal portion (Stem B) was biomechanically evaluated and compared to awell-known veterinary collared stem (Stem A). Femoral stems were inserted into the contralateral femurs from the same dog, forming 16 constructs. A flexo-compression load was applied on the axial axis of each sample. Maximum strength, deflection, stiffness, and energy absorption were analysed.

Results  Group B constructs showed significantly higher values (p ? 0.05) for the variables, except stiffness. The mean maximum strength was 1,347 ± 357 N for Group A and 1,805 ± 123 N for Group B (p ? 0.0069). The mean deflection was5.54 ± 2.63 mm for Group A and 10.03 ± 3.99 mm for Group B (p ? 0.0056). For the energy variable, the force was 6,203 ± 3,488 N/mm for Group A and 12,885 ± 5,056 N/mm for Group B (p ? 0.0054). Stem B had greater maximum strength, deflection, and energy.

Conclusions  The new stem was effective in neutralizing the impact of axial flexion-compression stresses during biomechanical tests in cadaveric models.

Key words Hip Prosthesis; Dogs

Introduction

Complication rates after canine total hip replacement (THR) range from 8 to 22% for both cemented and cementless techniques1,2. Recently, increasing numbers of cases require revision procedures3. The most common complications after THR in dogs are infection, aseptic loosening of implants, luxation, femoral fractures, and femoral stem subsidence4. Cementless systems are widely used in human patients and have been used in veterinary patients for the past two decades4,5. One of the reasons for their development was to eliminate the complications associated with the cement6. However, in spite of the good performance of canine cementless THR, many studies have described specific complications, especially those associated with the fixation of the femoral component such as catastrophic subsidence, peri-prosthetic fissures and femoral fractures1,2,7.

The longevity of cementless THR depends mainly on effective biological fixation. Anatomical fitting and correct insertion of the femoral stem are critical for bone ingrowth to occur8. Two basic types of cementless femoral stems are currently being used in dogs, press-fit and interlock system4,9,10. They have proved to be effective in allowing bone ingrowth11,12, but subsidence and femoral fractures are still reported10,12.

A number of femoral stem designs have been described and investigated for use in human THR. They promote better bone ingrowth and, thus, reduce complications. However, there is little information on the different canine femoral THR stem types available13. In an attempt to reduce the chance of subsidence or rotation of the femoral stem, a restricting medial collar has been used1,11. Recently, BioMedtrix has customized the BFX femoral component, including a lateral bolt, to provide additional resistance against subsidence11. In order to improve the problems that the stems have been suffering for some decades, new models have been tested to improve the resistance of the implant, improving the mechanical characteristics and the clinical result. Stems with lateral enlargement of the proximal portion were tested, and the results obtained show that these models enabled better filling of the medullary canal, greater resistance to axial forces and greater stability14-17. As these new designs of veterinary femoral stem are released, aiming to improve anatomical fit and bone ingrowth with reduced complication rates, they require further investigation to determine clinical effectiveness.

The objective of this study was to evaluate the biomechanical properties of a novel total hip replacement femoral stem (Stem B), which was compared to a well-known collared femoral stem (Stem A). The new stem model (Stem B) has as main features a convex proximal portion and a titanium plasma coating. Our hypothesis was that the novel femoral stem would provide more resistance to subsidence and better mechanical performance when compared to the collared stem.

Methods

Overall study design and specimen collection

The methodology adopted in this study was approved by the Ethics Committee on Animal Experimentation of Universidade Estadual Paulista “Júlio de Mesquita Filho” (UNESP), under the protocol no. 16,938/16.

The femoral stems (A and B) used in this study were made of surgical stainless steel and coated with a Cr-Co-Mo alloy (ASTM F75) (Cao Medica, Campinas, SP, Brazil). Femoral Stem A has a medial proximal collar and is covered with a titanium plasma coating on its proximal portion. Femoral Stem B has a more domed proximal region and is fully covered with a titanium plasma coating (Fig. 1). Both stems are available in two lengths (Stem A = 65.9 and 71.63 mm and Stem B = 58.54 and 57.34 mm). For this study, a 17-mm-diameter femoral head (neck +0) was used for both stems.

Figure 1
Design of Stem A and Stem B. The dashed lines (a and b) show that the Stem A has a titanium coating only on its proximal portion, while Stem B is fully coated. Additionally, there is a collar in Stem A. Stem B has a convexity with oblique chamfers on its proximal portion.

Sixteen femurs from skeletally mature male dog cadavers (25 ± 1 kg), euthanized for reasons unrelated to this study, were used. A randomized comparative experimental design was used, placing implant A in one of the femurs and implant B in the contralateral femur. Each femur was removed from its adjacent soft tissues and wrapped with surgical gauze soaked in a 0.9% NaCl solution. They were then placed in sealed bags and stored at -20ºC.

Radiographic projections (M1) (100 mA, 55 kV, 4 mAs) were taken with a calibration marker of 100 mm. A specific acetate template was used on digital X-rays to determine the size of stem to be implanted in each femur. All samples were subjected to densitometric analysis to ensure uniform bone mineral composition (BMC) and bone mineral density (BMD). The values of BMC and BMD were measured in the X-ray absorptiometry device in two dual energy X-ray absorptiometry (DXA) hologic energies (DXA, Hologic, Marlborough, Massachusetts, United States). This device emits photons that are collimated in a beam that cross the analyzed structure until reaching the detector, in which the intensity of the transferred beam is registered. The mechanism moves from side to side, forming the scan lines that will compose the image. Measurements were made by emitting X-ray beams at different energy levels, creating photoelectric peaks between 80 and 140 kV and 3 mA/s.

In order to check the quality control of the equipment, before scanning, the device was calibrated using a model (hologic DXA quality control spine phantom, hologic, Marlborough, Massachusetts, United States) provided by the manufacturer. This object has an area of 54.4 cm2, 51.1 g of BMC and 0.94 g/cm2 of BMD. The specimen was placed on the table, proximal to the flow, and a scan was performed. The data values were automatically computed, and the value obtained, compared with the expected value (previously determined by the manufacturer, at the time of installation). For small samples, less than 50 cm, according to the manufacturer, a second calibration using another model (Rat Step Phanton) is required. Smaller and in a density scale, the object is used to determine the scan area of the sample. If the obtained value differs by more than 1.5% from the expected value, the measurement must be repeated, and maintaining the difference, it is recommended to discontinue the use. The images were analyzed with the aid of the Hologic Discovery Wi10 software (Hologic, Marlborough, Massachusetts, United States). Once the regions of interest are obtained, in this case, the entire sample body of the femur (area) is selected.

It should be noted that all samples were previously weighed and measured, with respect to the length and area of the femur. Weighing was performed using a digital scale and, to measure the length of the femur, a 12-inch digital caliper was used, both properly calibrated. Then, the captured measurements were inserted in the software developed by the equipment manufacturer.

Stem implantation and instrumentation

The (left and right) femurs were thawed at room temperature. A femoral head and neck ostectomy was made ± 0.5 cm proximal to the lesser trochanter, using a cutting guide and oscillating saw. Using a specific set of rasps and drills, the medullary canal was prepared following a previously described technique11. The femoral component was inserted by holding the neck of the implant, without touching the porous parts. The stem was inserted up to 3/4 of the total length of the stem and filled 85 to 90% of the medullary canal. Finally, a specialized impactor was used to push the stem down into the femur to achieve a press-fit fixation.

All specimens were radiographed (M2) before the biomechanical evaluation, to verify correct positioning of the implants.

Specimen preparation and data processing

A force of 250 N was applied to all specimens in the static flexo-compression test, corresponding to 50% of the maximum stress in the pilot test used to determine the strength limit of the system (failure of the construct). An EMIC universal testing machine (maximum capacity of 10,000 kgf, Instron Brasil Equipamentos Científicos, São Jose dos Pinhais, PR, Brazil) was used to deliver a load cell of 1,961.33 N (200 kg). The rate of load was promoted displacement of the stem by 5 mm/min (Fig. 2).

Figure 2
The static flexion-compression test, in which a force of 250 N (red arrow) was used, corresponding to 50% of the maximum force observed during the pilot test. The load cell used was 1,961.33 N (200 kg) and the speed used to promote displacement was 5 mm/min, in the two sample models (A and B).

A polymethyl methacrylate support base attached to the distal end of each femur was created to keep the specimens in place. In addition, the acetabular dome (28 mm) was adapted to maintain the specimen in a vertical position, while the flexo-compression stress was applied in the long axis of the samples. The convex surface of the acetabular implant was filled with polymethyl methacrylate through which a hex screw with fine, central threads on its vertex was inserted. The objective was to create a femoral axis at an angle of 15º in the coronal-vertical plane. All tests were performed with the samples at room temperature, average of 22ºC. After the flexion-compression tests, the machine software calculated the maximum force supported by the femur / implant set, the deflection, the relative stiffness, and the energy absorbed during the test for each sample. Failure in the axial compression test was defined as fracture of the specimen.

Statistical analysis

A randomized block design was used for the comparison of groups (A and B), with two treatments and 16 repetitions (eight for each treatment). Statistics were performed using R software (R Foundation for Statistical Computing, Vienna, Austria). Initially, the homoscedasticity of variances (Bartlett test) and the normal distribution (Shapiro-Wilk test) were tested. The variables from the bone densitometry analysis were compared between groups with the Student’s t-test (the independent samples t test) to verify the homogeneity of the sample units. The variables obtained from the static axial compression test were subsequently compared between groups through analysis of variance (one-way ANOVA) in a randomized blocks (animal) design. The significance level was set at 95% (p < 0.05) for all tests, and the data was presented as mean ± standard deviation (SD). When the test revealed a statistically significant difference, the Tukey test was also used.

Results

The pre-test radiographic examinations revealed homogeneity of the samples. The implants were inserted in accordance with the methodological description. The statistical analyses of the variables for BMD and BMC revealed similarity between the selected samples for both Group A and Group B—the samples were homogeneous, and there was no significant difference between groups (p = 0.05) regarding BMD and BMC.

The following variables were obtained during the biomechanical axial static flexo-compression test in the femurs with stems A and B: maximum strength (MS) supported by the implants, deflection (D), stiffness (S) of the construct, and energy (En) applied until failure. Group B had significantly higher values (p ≤ 0.05) for the variables (MS, D, En) resulting from this study, with the exception of S, which was statistically similar between groups (p = 0.05; Table 1).

Table 1
Comparison of biomechanical axial flexo-compression test variables between two cementless implant models, Groups A and B. The corresponding values of the resulting variables are expressed as mean (±) and standard deviation (SD).

Discussion

Canine cementless total hip replacement provides excellent clinical outcomes. However, complications such as subsidence, periprosthetic fractures, stress shielding and infections still occur18,19. Despite improvements in existing systems over recent years1,2,11,12, the development of new implants should reduce the currently reported complication rate20,21. This study aimed to test and compare the biomechanics of a new model of femoral stem with specific characteristics, designed to promote better mechanical performance and adequate adjustment to the femur. The initial hypothesis was that the new femoral stem model (Stem B) would perform well, especially in relation to greater resistance and stress distribution. In fact, Stem B performed better than Stem A.

Some key features are crucial for biomechanical strength of a stem, including the construction material, dimensions and design22. The new femoral stem model is coated with a Cr-Co-Mo alloy. This material is widely used in the manufacture of dental and orthopedic implants, mainly for its properties involving biocompatibility, resistance to fatigue and corrosion23,24. The novel femoral stem (Stem B) has specific characteristics, such as a conical form, a convex and wider proximal portion and fully blasted surface, that result in greater resistance factors. Stem designs focused on a lateral enlargement of the proximal portion have been described for human use, and these were created to maximize the proximal adjustment and the filling of the medullary canal, providing greater torsional and axial stability14-17. This design potentially increases the contact of the implant in the metaphyseal region, thus increasing the cross-sectional diameter of the stem, allowing the stem to have a broader base of metaphysical support25,26. As a result, there is a better distribution of endosteal compressive forces, which can potentially reduce complications such as subsidence and aseptic loosening25.

Walker et al.28 compared lateral extended and conventional straight human stems and concluded that the lateral extension resulted in better distribution of periprosthetic stress, allowing the use of smaller stems, facilitating implantation and causing less bone remodeling of the femur. Similarly, Leali and Fetto27 demonstrated that larger lateral proximal diameter stems produce an important effect on load transfer to the bone metaphysis. With these stems, the forces between the implant and the bone are concentrated around the level of the lesser trochanter, whereas in straight stems they are located more distally in the diaphysis. Long-term evaluations of these implants confirmed that the change in the proximal geometry provided greater stability and femoral bone preservation, which provides more resistance to migration of the implant. In addition, good femoral canal filling produces high circumferential tensile stresses, that can result in reduced loss of bone29.

Despite these positive features of the novel stem design, the lateral and proximal bulging can potentially create an over stress in the proximal femoral canal. Additionally, an over-reaming of the femur is needed, leading to a higher risk for intraoperative fissures and postoperative femoral fracture29.

The novel stem has a slightly conical shape, thus differing from the straight conventional stems30. Conical-shaped stems have been widely used in human patients, particularly for surgical revisions in which bone loss has occurred in the proximal femur31. These provide greater primary rotational stability through continuous pressure in the diaphyseal portion of the bone and better distribution of axial loads and accommodation of the implants in the femoral bone bed, thus reducing complications30,32,33. Additional advantages such as simpler surgical technique, easier intramedullary positioning, better adjustment to different femoral anatomies and simpler correction of retroversion have also been reported31,32,34,35. As previously reported, during the implantation of the new stem model (Stem B), and in results of the mechanical tests, there was good coaptation of the implant to the femur. It is possible that this was one of the factors that resulted in the good biomechanical properties.

The coating of implants with materials such as titanium spray plasma assists in the resistance to axial displacement of the stem, allowing a low index of distal micromovement36-38. Both implants had surface treatment with blasting of glass microspheres and a titanium plasma coating. This coating is thought to provide greater rigidity to the stems (A and B), promoting greater surface contact between the implant and the endosteum. Although the coatings were similar, a higher MS, D and En were observed for construct B. One potential explanation for this is that the differences between groups was due to the extent of the covered area, since Stem B had the coating along its entire length, while the coating of Stem A was restricted to the proximal portion. However, the differences between the proximal regions of each stem, represented by the presence of the collar in Group A, and by the bulging in Group B, might also have resulted in the statistical differences observed.

This research has a number of limitations. More targeted studies are needed to determine whether blasting of glass and titanium plasma microspheres are important for the rigidity and strength of the systems, and whether the extent of the coating is also relevant. Stems A and B in this study were slightly different in size, and it would be interesting to compare the results with a further study using identically sized stems. Additionally, this study was conducted in canine cadaveric models. Therefore, it was not possible to evaluate and compare the effect of the biological fixation obtained by bone growth after the insertion of the stems; further studies in living patients are necessary.

The number of different femoral stem models and designs has been gradually increasing as veterinary medicine advances. However, so far there is no reliable information on the effect of stem design on resistance to subsidence or other complications of total hip prosthesis surgery.

Since the collared femoral stem improves stability and promotes better transfer of stress to the calcar bone39, we believe that Stem B can, potentially, behave in the same manner as the collared stem. It is possible that femoral Stem B, due to its larger proximal portion (bulging), would adapt better to the anatomy of the proximal femur, preventing femoral displacement and contributing to adequate filling and osteointegration after implantation1,40,41.

Conclusions

The new stem model had some results similar to the collar model, mainly by effectively neutralizing the impact of axial flexion-compression stresses on dog femurs. Although the new stem model has achieved superior results to the collar model, and this hypothetically suggests greater mechanical strength, further studies are needed to determine whether the new model is superior mechanically or at least as effective as the model already known.

Acknowledgments

Not applicable.

  • Data availability statement
    Data will be available upon request.
  • Funding
    BioMedica Company.
  • Research performed at Laboratory of Bases of Surgical Techniques, Department of Clinical and Veterinary Surgery, Faculty of Agricultural and Veterinary Sciences, Universidade Estadual Paulista “Júlio de Mesquita Filho”, Jaboticabal-SP, Brazil. Part of PhD degree thesis, Postgraduate Program in Veterinary Surgery, UNESP. Tutor: Dr. Bruno Watanabe Minto.

References

  • 1 Liska WD, Doyle ND. Use of an electron beam melting manufactured titanium collared cementless femoral stem to resist subsidence after canine total hip replacement. Vet Surg. 2015;44(7):883-94. https://doi.org/10.1111/vsu.12353
    » https://doi.org/10.1111/vsu.12353
  • 2 Buks Y, Wendelburg KL, Stover SM, Garcia‐Nolen TC. The effects of interlocking a universal hip cementless stem on implant subsidence and mechanical properties of cadaveric canine femora. Vet Surg. 2016;45(2):155-64. https://doi.org/10.1111/vsu.12437
    » https://doi.org/10.1111/vsu.12437
  • 3 Johnson KA. Cementless total hip replacement complications. Vet Comp Orthop Traumatol. 2015;28(1):V-VI. https://doi.org/10.3415/VCOT-14-12-0182
    » https://doi.org/10.3415/VCOT-14-12-0182
  • 4 Hummel DW, Lanz OI, Were SR. Complications of cementless total rip replacement. A retrospective study of 163 cases. Vet Comp Orthop Traumatol. 2010;23(6):424-32. https://doi.org/10.3415/VCOT-09-07-0071
    » https://doi.org/10.3415/VCOT-09-07-0071
  • 5 Vezzoni L, Vezzoni A, Boudrieau RJ. Long‐term outcome of Zürich cementless total hip arthroplasty in 439 cases. Vet Surg. 2015;44(8):921-9. https://doi.org/10.1111/vsu.12371
    » https://doi.org/10.1111/vsu.12371
  • 6 Skurla CP, James SP. Assessing the dog as a model for human total hip replacement: analysis of 38 postmortem-retrieved canine cemented acetabular components. J Biomed Mater Res. 2005;73(2):260-70. https://doi.org/10.1002/jbm.b.30204
    » https://doi.org/10.1002/jbm.b.30204
  • 7 Korani HM, Marcellin-little DJ, Roe SC. Variability associated with assessing changes in position of a canine uncemented femoral stem prosthesis. Vet Comp Orthop Traumatol. 2015;28(6):409-16. https://doi.org/10.3415/VCOT-15-03-0044
    » https://doi.org/10.3415/VCOT-15-03-0044
  • 8 De Andrade CR, Minto BW, Dreibi RM, Diogo LM, Kawamoto FY, Franco GG.; Uscategui RAR, Dias LGGG. Accuracy in determining canal flare index using different radiographical positions for imaging canine femurs. Vet Comp Orthop Traumatol. 2019;32(3):234-40. https://doi.org/10.1055/s-0039-1683390
    » https://doi.org/10.1055/s-0039-1683390
  • 9 Ponzio DY, Shahi A, Park AG, Purtill JJ. Intraoperative proximal femoral fracture in primary cementless total hip arthroplasty. J Arthroplasty. 2015;30(8):1418-22. https://doi.org/10.1016/j.arth.2015.02.043
    » https://doi.org/10.1016/j.arth.2015.02.043
  • 10 Bayer K, Matiasovic M, Steger H, Böttcher P. Complications and long-term outcome in 16 canine cementless hybrid hip arthroplasties. Vet Comp Orthop Traumatol. 2019;32(1):73-8. https://doi.org/10.1055/s-0038-1676072
    » https://doi.org/10.1055/s-0038-1676072
  • 11 Schiller TD. BioMedtrix total hip replacement systems: an overview. Vet Clin North Am Small Anim Pract. 2017;47(4):899-916. https://doi.org/10.1016/j.cvsm.2017.03.005
    » https://doi.org/10.1016/j.cvsm.2017.03.005
  • 12 Vezzoni L, Bazzo S, Vezzoni A. Revision of a BFX total hip replacement stem using a Kyon stem and a head adaptor in two dogs. Vet Comp Orthop Traumatol. 2017;30(1):81-7. https://doi.org/10.3415/VCOT-16-05-0070
    » https://doi.org/10.3415/VCOT-16-05-0070
  • 13 Kawamoto FY, Minto BW, Diogo LM, De Faria LG, Shimano AC, Macedo AP, Camassa JAA, Dias LGGG. Photoelastic analysis of two different cementless femoral stems for total hip arthroplasty in a canine model. VCOT Open. 2019;2(1):e36-e42. https://doi.org/10.1055/s-0039-1691792
    » https://doi.org/10.1055/s-0039-1691792
  • 14 Zadeh HG, Hua J, Walker PS, Muirhead-Allwood SK. Uncemented total hip arthroplasty with subtrochanteric derotational osteotomy for severe femoral anteversion. J Arthroplasty. 1999;14 (6):682-8. https://doi.org/10.1016/s0883-5403(99)90223-1
    » https://doi.org/10.1016/s0883-5403(99)90223-1
  • 15 Renkawitz T, Santori FS, Grifka J, Valverde C, Morlock MM, Learmonth ID. A new short uncemented, proximally fixed anatomic femoral implant with a prominent lateral flare: design rationals and study design of an international clinical trial. BMC Musculoskelet Disord. 2008;9(1):1-6. https://doi.org/10.1186/1471-2474-9-147
    » https://doi.org/10.1186/1471-2474-9-147
  • 16 Ghera S, Pavan L. The DePuy Proxima™ Hip: a short stem for total hip arthroplasty. early experience and technical considerations. Hip Int. 2009;19(3):215-20. https://doi.org/10.1177/112070000901900305
    » https://doi.org/10.1177/112070000901900305
  • 17 Mirza SB, Dunlop DG, Panesar SS, Naqvi SG, Gangoo S, Salih S. Basic science considerations in primary total hip replacement arthroplasty. Open Orthop J. 2010;4:169-80. https://doi.org/10.2174/1874325001004010169
    » https://doi.org/10.2174/1874325001004010169
  • 18 Mitchell MM, Hudson CC, Beale BS. Comparison of femoral stem subsidence between three types of press‐fit cementless total hip replacement in dogs. Vet Surg. 2020;49(4):787-93. https://doi.org/10.1111/vsu.13391
    » https://doi.org/10.1111/vsu.13391
  • 19 Olsen RS, Lanz OI. Revision of a canine Zürich cementless total hip replacement using a ‘Cupless’ system. Vet Rec Case Rep. 2020;8(2): e001035. https://doi.org/10.1136/vetreccr-2019-001035
    » https://doi.org/10.1136/vetreccr-2019-001035
  • 20 Koyano G, Jinno T, Koga D, Yamauchi Y, Muneta T, Okawa A. Comparison of bone remodeling between an anatomic short stem and a straight stem in 1-stage bilateral total hip arthroplasty. J Arthroplasty. 2017;32(2):594-600. https://doi.org/10.1016/j.arth.2016.07.016
    » https://doi.org/10.1016/j.arth.2016.07.016
  • 21 Schmidutz F, Woiczinski M, Kistler M, Schröder C, Jansson V, Fottner A. Influence of different sizes of composite femora on the biomechanical behavior of cementless hip prosthesis. Clin Biomech. 2017;41:60-5. https://doi.org/10.1016/j.clinbiomech.2016.12.003
    » https://doi.org/10.1016/j.clinbiomech.2016.12.003
  • 22 Zhou XM, Walker PS, Robertson DD. Effect of press-fit femoral stems on strains in the femur: a photoelastic coating study. J Arthroplasty. 1990;5(1):71-82. https://doi.org/10.1016/S0883-5403(06)80013-6
    » https://doi.org/10.1016/S0883-5403(06)80013-6
  • 23 Hedberg YS, Qian B, Shen Z, Virtanen S, Wallinder IO. In vitro biocompatibility of CoCrMo dental alloys fabricated by selective laser melting. Dent Mater. 2014;30(5):525-34. https://doi.org/10.1016/j.dental.2014.02.008
    » https://doi.org/10.1016/j.dental.2014.02.008
  • 24 Lucchetti MC, Fratto G, Valeriani F, De Vittori E, Giampaoli S, Papetti P, Spica VR, Manzon L. Cobalt-chromium alloys in dentistry: an evaluation of metal ion release. J Prosthet Dent. 2015;114(4):602-8. https://doi.org/10.1016/j.prosdent.2015.03.002
    » https://doi.org/10.1016/j.prosdent.2015.03.002
  • 25 Leali A, Fetto J, Insler H. Favorable midterm results of total hip arthroplasties with a lateral flare uncemented stem. Clin Orthop Relat Res. 2006;450:138-44. https://doi.org/10.1097/01.blo.0000214410.60722.b1
    » https://doi.org/10.1097/01.blo.0000214410.60722.b1
  • 26 Leali A, Fetto J. Promising mid-term results of total hip arthroplasties using an uncemented lateral-flare hip prosthesis: a clinical and radiographic study. Int Orthop. 2007;31(6):845-9. https://doi.org/10.1007/s00264-006-0267-8
    » https://doi.org/10.1007/s00264-006-0267-8
  • 27 Leali A, Fetto J. Preservation of femoral bone mass after total hip replacements with a lateral flare stem. Int Orthop. 2004;28:151-4. https://doi.org/10.1007/s00264-004-0554-1
    » https://doi.org/10.1007/s00264-004-0554-1
  • 28 Walker PS, Culligan SG, Hua J, Muirhead-Allwood SK, Bentley G. The effect of a lateral flare feature on uncemented hip stems. Hip Int. 1999;9(2):71-80. https://doi.org/10.1177/112070009900900210
    » https://doi.org/10.1177/112070009900900210
  • 29 Walker PS, Schneeweis D, Murphy S, Nelson P. Strains and micromotions of press-fit femoral stem prostheses. J Biomech. 1987;20(7):693-702. https://doi.org/10.1016/0021-9290(87)90035-2
    » https://doi.org/10.1016/0021-9290(87)90035-2
  • 30 Jakubowitz E, Bitsch RG, Heisel C, Lee C, Kretzer JP, Thomsen MN. Primary rotational stability of cylindrical and conical revision hip stems as a function of femoral bone defects: an in vitro comparison. J Biomech. 2008;41(14):3078-84. https://doi.org/10.1016/j.jbiomech.2008.06.002
    » https://doi.org/10.1016/j.jbiomech.2008.06.002
  • 31 Da Assunção RE, Pollard TCB, Hrycaiczuk A, Curry J, Glyn-Jones S, Taylor A. Revision arthroplasty for periprosthetic femoral fracture using an uncemented modular tapered conical stem. Bone Joint J. 2015;97(8):1031-7. https://doi.org/10.1302/0301-620X.97B8.34431
    » https://doi.org/10.1302/0301-620X.97B8.34431
  • 32 Zhu J, Shen C, Chen X, Cui Y, Peng J, Cai G. Total hip arthroplasty with a non-modular conical stem and transverse subtrochanteric osteotomy in treatment of high dislocated hips. J Arthroplasty. 2015;30(4):611-4. https://doi.org/10.1016/j.arth.2014.11.002
    » https://doi.org/10.1016/j.arth.2014.11.002
  • 33 Schuroff AA, Deeke M, Pedroni MA, Lupselo FS, Kunz RE, Lima AM. Avaliação radiográfica da técnica de cimentação com haste femoral cônica polida e tripla cunha em artroplastia do quadril. Rev Bras Ortop. 2017;52:40-5. https://doi.org/10.1016/j.rboe.2017.08.019
    » https://doi.org/10.1016/j.rboe.2017.08.019
  • 34 Wagner H, Wagner M. Cone prosthesis for the hip joint. Arch Orthop Trauma Surg. 2000;120(1-2):88-95. https://doi.org/10.1007/pl00021223
    » https://doi.org/10.1007/pl00021223
  • 35 Viceconti M, Pancanti A, Varini E, Traina F, Cristofolini L. On the biomechanical stability of cementless straight conical hip stems. Proc Inst Mech Eng H. 2006;220(3):473-80. https://doi.org/10.1243/09544119H06904
    » https://doi.org/10.1243/09544119H06904
  • 36 Pernell RT, Gross RS, Milton JL, Montgomery RD, Wenzel JG, Savory CG, Aberman HM. Femoral strain distribution and subsidence after physiological loading of a cementless canine femoral prosthesis: the effects of implant orientation, canal fill, and implant fit. Vet Surg. 1994;23(6):503-18. https://doi.org/10.1111/j.1532-950x.1994.tb00512.x
    » https://doi.org/10.1111/j.1532-950x.1994.tb00512.x
  • 37 Yoon BJV, Xavier F, Walker BR, Grinberg S, Cammisa FP, Abjornson C. Optimizing surface characteristics for cell adhesion and proliferation on titanium plasma spray coatings on polyetheretherketone. Spine J. 2016;16(10):1238-43. https://doi.org/10.1016/j.spinee.2016.05.017
    » https://doi.org/10.1016/j.spinee.2016.05.017
  • 38 Pap K, Vasarhelyi G, Gal T, Nemeth G, Abonyi B, Hangody LR, Hangody GM, Hangody L. Evaluation of clinical outcomes of cemented vs uncemented knee prostheses covered with titanium plasma spray and hydroxyapatite: a minimum two years follow-up. Joint Dis Rel Surg. 2018;29(2):65-70. https://doi.org/10.5606/ehc.2018.61076
    » https://doi.org/10.5606/ehc.2018.61076
  • 39 Jeon I, Bae JY, Park JH, Yoon TR, Todo M, Mawatari M, Hotokebuchi T. The biomechanical effect of the collar of a femoral stem on total hip arthroplasty. Comput Methods Biomech Biomed Engin. 2011;14(1):103-12. https://doi.org/10.1080/10255842.2010.49351341
    » https://doi.org/10.1080/10255842.2010.49351341
  • 40 Demey G, Fary C, Lustig S, Neyret P, Si Selmi TA. Does a collar improve the immediate stability of uncemented femoral hip stems in total hip arthroplasty? A bilateral comparative cadaver study. J Arthroplasty. 2011;26(8):1549-55. https://doi.org/10.1016/j.arth.2011.03.030
    » https://doi.org/10.1016/j.arth.2011.03.030
  • 41 Bonin N, Gedouin JE, Pibarot V, Bejui-Hughues J, Bothorel H, Saffarini M, Batailler C. Proximal femoral anatomy and collared stems in hip arthroplasty: is a single collar size sufficient? J Exp Orthop. 2017;4(1):32. https://doi.org/10.1186/s40634-017-0107-3
    » https://doi.org/10.1186/s40634-017-0107-3

Publication Dates

  • Publication in this collection
    18 June 2021
  • Date of issue
    2021

History

  • Received
    16 Jan 2021
  • Reviewed
    12 Mar 2021
  • Accepted
    14 Apr 2021
location_on
Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia https://actacirbras.com.br/ - São Paulo - SP - Brazil
E-mail: actacirbras@gmail.com
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Acessibilidade / Reportar erro