Cardiac troponin T (cTnT) assessment using incompletely filled K2EDTA blood tubes is reliable

Dear Editor,

Cardiac troponin testing, which includes the measurement of cardiac troponin I (cTnI) or T (cTnT), is the mainstay of diagnosing acute coronary syndrome, especially in non-ST-elevation myocardial infarction (NSTEMI), in which elevated cTnI or cTnT is the only objective parameter. ( 11. Thygesen K, Alpert JS, Jaffe AS, Chaitman BR, Bax JJ, Morrow DA, White HD; Executive Group on behalf of the Joint European Society of Cardiology (ESC)/American College of Cardiology (ACC)/American Heart Association (AHA)/World Heart Federation (WHF) Task Force for the Universal Definition of Myocardial Infarction. Fourth Universal Definition of Myocardial Infarction (2018). J Am Coll Cardiol. 2018;72(18):2231-64. Review. ) Ensuring quality throughout the testing process is essential for maintaining the diagnosis integrity. The receipt of underfilled samples is common in clinical laboratories, ( 22. Lippi G, von Meyer A, Cadamuro J, Simundic AM. Blood sample quality. Diagnosis (Berl). 2019;6(1):25-31. Review. ) and this could impair the results of some immunoassays. ( 33. Toprak B, Yalcin H, Arı E, Colak A. EDTA interference in electrochemiluminescence ACTH assay. Ann Clin Biochem. 2016;53(6):699-701. ) Therefore, this study aimed to investigate the potential impact of underfilled K 2 EDTA blood tubes on cTnT assay results.

We studied 17 subjects (mean age 46±9 years, 65% females); recruited from the personnel of the clinical laboratory of the University Hospital of Verona (Italy). Blood was drawn through phlebotomy using a 10mL syringe (Plastipak Luer-Lok 10mL Syringe, Becton Dickinson, Madrid, Spain) and a 21-gauge disposable needle (KDL, Nanchang, Jiangxi, China). It was immediately dispensed into four 3.0mL primary blood tubes with spray-dried K 2 EDTA (5.4mg) (Vacutest, Kima, Arzergrande, Padova, Italy; Lot: XZ3062) at different volumes: 0.5mL (17% filling), 1.0mL (33% filling), 2.0mL (67% filling), and 3.0mL (100%; correct filling), as described earlier. ( 44. Lippi G, Dima F, Tosi M, Carpenè G, Celegon G, Favaloro EJ, et al. Incomplete filling of spray-dried K2EDTA evacuated blood tubes: impact on measuring routine hematological parameters on Sysmex XN-10. Diagnosis (Berl). 2023 Jul 24. doi: 10.1515/dx-2023-0064. Epub ahead of print.
https://doi.org/10.1515/dx-2023-0064...
) The empty tubes were premarked at different fill volumes through direct comparison with tubes filled with blood using a calibrated pipette. The samples were centrifuged, and the level of cTnT in the K 2 EDTA plasma was measured using a Roche Cobas E 601 Module (Roche Diagnostics, Basel, Switzerland), following the manufacturer’s instructions. A comprehensive description of this method and its performance has been reported. ( 55. von Meyer A, Albert G, Kunzelmann S, Rank C, Zerback R, Imdahl R. Evaluating the performance of an updated high-sensitivity troponin T assay with increased tolerance to biotin. Clin Chem Lab Med. 2020;59(3):591-7. ) All volunteers provided informed consent for participation in this study, which was performed in accordance with the Declaration of Helsinki and all relevant local legislation. This study was approved by the Ethics Committee of the University Hospital of Verona (970CESC; July 20, 2016).

The results are presented in figure 1 . No significant difference in the cTnT levels was observed among the three underfilled samples. Compared with the clinically significant threshold of variation reported as the reference change value, the variation in cTnT values remained well below the acceptable threshold of ±46%. ( 66. Frankenstein L, Wu AH, Hallermayer K, Wians FH Jr, Giannitsis E, Katus HA. Biological variation and reference change value of high-sensitivity troponin T in healthy individuals during short and intermediate follow-up periods. Clin Chem. 2011;57(7):1068-71. )

Figure 1
Influence of the filling level of K 2 EDTA blood tubes on cardiac troponin T (cTnT) testing. Differences from the reference (appropriately filled 3.0mL blood tube) are provided above the upper limit of the standard deviation, for comparison with the reference change value. Data are presented as mean±standard deviation

The collection of suboptimal whole blood volumes in K 2 EDTA blood tubes does not affect the results of the cTnT assays in ostensibly healthy individuals. Therefore, samples underfilled by up to 83% may still be acceptable for cTnT testing using the Roche Cobas E 601 module.

REFERENCES

  • 1
    Thygesen K, Alpert JS, Jaffe AS, Chaitman BR, Bax JJ, Morrow DA, White HD; Executive Group on behalf of the Joint European Society of Cardiology (ESC)/American College of Cardiology (ACC)/American Heart Association (AHA)/World Heart Federation (WHF) Task Force for the Universal Definition of Myocardial Infarction. Fourth Universal Definition of Myocardial Infarction (2018). J Am Coll Cardiol. 2018;72(18):2231-64. Review.
  • 2
    Lippi G, von Meyer A, Cadamuro J, Simundic AM. Blood sample quality. Diagnosis (Berl). 2019;6(1):25-31. Review.
  • 3
    Toprak B, Yalcin H, Arı E, Colak A. EDTA interference in electrochemiluminescence ACTH assay. Ann Clin Biochem. 2016;53(6):699-701.
  • 4
    Lippi G, Dima F, Tosi M, Carpenè G, Celegon G, Favaloro EJ, et al. Incomplete filling of spray-dried K2EDTA evacuated blood tubes: impact on measuring routine hematological parameters on Sysmex XN-10. Diagnosis (Berl). 2023 Jul 24. doi: 10.1515/dx-2023-0064. Epub ahead of print.
    » https://doi.org/10.1515/dx-2023-0064
  • 5
    von Meyer A, Albert G, Kunzelmann S, Rank C, Zerback R, Imdahl R. Evaluating the performance of an updated high-sensitivity troponin T assay with increased tolerance to biotin. Clin Chem Lab Med. 2020;59(3):591-7.
  • 6
    Frankenstein L, Wu AH, Hallermayer K, Wians FH Jr, Giannitsis E, Katus HA. Biological variation and reference change value of high-sensitivity troponin T in healthy individuals during short and intermediate follow-up periods. Clin Chem. 2011;57(7):1068-71.

Publication Dates

  • Publication in this collection
    05 Feb 2024
  • Date of issue
    2024

History

  • Received
    05 June 2023
  • Accepted
    11 July 2023
Instituto Israelita de Ensino e Pesquisa Albert Einstein Avenida Albert Einstein, 627/701 , 05651-901 São Paulo - SP, Tel.: (55 11) 2151 0904 - São Paulo - SP - Brazil
E-mail: revista@einstein.br