Open-access Determination of the melting temperature, heat of fusion, and purity analysis of different samples of zidovudine (AZT) using DSC

Abstracts

The determination of chemical purity, melting range, and variation of enthalpy in the process of characterizing medicines is one of the principal requirements evaluated in quality control of the pharmaceutical industry. In this study, the method of purity determination using DSC was outlined, as well as the application of this technique for the evaluation of commercial samples of zidovudine (AZT) (raw material) supplied by different laboratories. To this end, samples from six different laboratories (A, B, C, D, E, and F) and the standard reference (R) from the United States Pharmacopeia (USP) were analyzed. The DSC curves were obtained in the temperature range of 25 to 200 ºC under the dynamic atmosphere of N2 (50 mL min-1), heating rate of β=2 ºC min-1, using an Al capsule containing approximately 2 mg of sample material. The results demonstrated that the standard reference presented a proportion of 99.83% whereas the AZT samples presented a variation ranging from 97.59 to 99.54%. In addition, the standard reference was found to present a temperature of onset of melting point of 122.80 °C. Regarding the samples of active agents provided by the different laboratories, a variation ranging from 118.70 to 122.87 °C was measured. In terms of ΔHm, the samples presented an average value of 31.12 kJ mol-1.

Zidovudine; Differential scanning calorimetry; Thermal analysis; Medicines; Medicines


A determinação da pureza química, a faixa de fusão e a variação de entalpia envolvida no processo de caracterização de fármacos é um dos principais requisitos avaliados no controle de qualidade em indústrias farmacêuticas. Neste trabalho é feita uma breve abordagem sobre o método de determinação de pureza utilizando DSC, assim como a aplicação desta técnica para avaliação de amostras comerciais de zidovudina (AZT) (matéria-prima) fornecida por diferentes laboratórios. Para tal, foram analisadas amostras de seis diferentes laboratórios (A,B,C,D,E e F) e a substância química de referência (R) da United States Pharmacopeia (USP). As curvas DSC foram obtidas na faixa de temperatura entre 25 a 200 ºC, sob atmosfera dinâmica de N2 (50 mL min-1), β=2 ºC min-1, utilizando cápsula de Al contendo aproximadamente 2 mg de amostra. De acordo com os resultados, pode-se observar que a substância química de referência apresentou teor igual a 99,83% e que as amostras de AZT apresentaram uma faixa de variação entre 97,59 e 99,54%. Pode-se verificar, ainda, que a substância química de referência apresentou uma temperatura onset de fusão igual a 122,80 °C. Para as amostras dos princípios ativos fornecidos pelos diferentes laboratórios, pode-se verificar uma faixa de variação entre 118,70 e 122,87 °C. No que se refere ao ΔHm, as amostras apresentaram valor médio de 31,12 kJ.mol-1.

Zidovudina; Calorimetria exploratória diferencial; Análise térmica; Medicamentos; Medicamentos


ARTICLE

Determination of the melting temperature, heat of fusion, and purity analysis of different samples of zidovudine (AZT) using DSC

Adriano Antunes Souza AraújoI,*; Marília dos Santos BezerraI; Sílvia StorpirtisII; Jivaldo do Rosário MatosIII

IDepartament of Physiology, Federal University of Sergipe

IIDepartament of Pharmacy, Faculty of Pharmaceutical Sciences, University of São Paulo

IIIDepartament of Fundamental Chemistry, Institute of Chemistry, University of São Paulo

ABSTRACT

The determination of chemical purity, melting range, and variation of enthalpy in the process of characterizing medicines is one of the principal requirements evaluated in quality control of the pharmaceutical industry. In this study, the method of purity determination using DSC was outlined, as well as the application of this technique for the evaluation of commercial samples of zidovudine (AZT) (raw material) supplied by different laboratories. To this end, samples from six different laboratories (A, B, C, D, E, and F) and the standard reference (R) from the United States Pharmacopeia (USP) were analyzed. The DSC curves were obtained in the temperature range of 25 to 200 ºC under the dynamic atmosphere of N2 (50 mL min-1), heating rate of β=2 ºC min-1, using an Al capsule containing approximately 2 mg of sample material. The results demonstrated that the standard reference presented a proportion of 99.83% whereas the AZT samples presented a variation ranging from 97.59 to 99.54%. In addition, the standard reference was found to present a temperature of onset of melting point of 122.80 °C. Regarding the samples of active agents provided by the different laboratories, a variation ranging from 118.70 to 122.87 °C was measured. In terms of ΔHm, the samples presented an average value of 31.12 kJ mol-1.

Uniterms: Zidovudine/purity determination. Differential scanning calorimetry. Thermal analysis. Medicines/quality control. Medicines/qualitative analysis.

RESUMO

A determinação da pureza química, a faixa de fusão e a variação de entalpia envolvida no processo de caracterização de fármacos é um dos principais requisitos avaliados no controle de qualidade em indústrias farmacêuticas. Neste trabalho é feita uma breve abordagem sobre o método de determinação de pureza utilizando DSC, assim como a aplicação desta técnica para avaliação de amostras comerciais de zidovudina (AZT) (matéria-prima) fornecida por diferentes laboratórios. Para tal, foram analisadas amostras de seis diferentes laboratórios (A,B,C,D,E e F) e a substância química de referência (R) da United States Pharmacopeia (USP). As curvas DSC foram obtidas na faixa de temperatura entre 25 a 200 ºC, sob atmosfera dinâmica de N2 (50 mL min-1), β=2 ºC min-1, utilizando cápsula de Al contendo aproximadamente 2 mg de amostra. De acordo com os resultados, pode-se observar que a substância química de referência apresentou teor igual a 99,83% e que as amostras de AZT apresentaram uma faixa de variação entre 97,59 e 99,54%. Pode-se verificar, ainda, que a substância química de referência apresentou uma temperatura onset de fusão igual a 122,80 °C. Para as amostras dos princípios ativos fornecidos pelos diferentes laboratórios, pode-se verificar uma faixa de variação entre 118,70 e 122,87 °C. No que se refere ao ΔHm, as amostras apresentaram valor médio de 31,12 kJ.mol-1.

Unitermos: Zidovudina/determinação da pureza. Calorimetria exploratória diferencial. Análise térmica. Medicamentos/controle de qualidade. Medicamentos/análise qualitativa.

INTRODUCTION

Applications in the pharmaceutical field exploit thermal analysis as an important tool in the solution of problems involving the development, production, and quality control of medicines (Buckton et al., 1991). The principal applications in this area have included determining the degree of purity of drugs (Windmann, Scherrer, 1991), the conducting of the analyses of thermal stability and the kinetics of decomposition, the study of possible drug-drug and/or drug-excipient interactions in the pre-formulation phase (Iglesias et al., 1998; Saesma et al., 1991; Araújo et al., 2003), and in the characterization of raw materials and finished products (Ford, Timmins, 1989; Canotilho et al., 1992; Cammenga, Epple, 1995; Giron, 1998, 1999; Brown, Glass, 1999; Thompson, 2000).

Canotilho and collaborators (1992) reported that the resources and potential of thermal analysis in the pharmaceutical field are considerable not only for the study of polymorphism, but also for their versatility in other applications. These encompass characterization of polymorphs (Barbas et al., 2007), inclusion complexes and solid dispersants, studies of drug/drug compatibility and drug/excipients at the pre-formulation level (Kiss et al., 2006), determination of chemical purity, study of reactions in solid state (thermal stability and kinetic parameters) (Araújo et al., 2005), analysis of solid dosage forms, quality control of drugs and medicines (as in the determination of the contents of ash and humidity, among others) (Araújo et al., 2006), the evaluation of isomers and polymorphs, the analysis of raw materials (drugs and excipients) as well as finished dosage forms (emulsions, suppositories, microspheres, pills, liposomes, powders, creams, and gels) (Canotilho et al., 1992; Lvova et al., 1993).

The determination of the melting point using the differential scanning calorimetry (DSC) method has been satisfactorily used as a method of evaluating the degree of purity of a compound (Widmann, Scherrer, 1991). By means of DSC, the melting range can be determined for a substance, and based on the equation of Van't Hoff (Canotilho et al., 1992, Bezjak et al., 1992) (Equation 1) it is possible to determine the molar fraction of impurities x2 (number of mols of impurities by the total mol number) contained in this material in relation to the melting range. However, this method can be applied only if x2 lies between 0.95 and 0.999 (or 95 and 99.99 mol %).

where Tm represents the melting temperature of impurities in the process of melting, since the greatest component of remaining crystalline solids is in equilibrium with the phase already melted; To is the melting temperature of the main component expressed in Kelvin; R is the constant of gases (8.3143 J K-1 mol-1); x2 is the molar fraction of impurities in liquid phase; ΔHm is the heat of melting of the main component expressed in J mol-1; ΔHm (R.To2) and is the denominated cryoscopic constant.

When the substance is heated, the collection of impurities is melted, forming a liquid phase. Above this temperature, the solid phase only consists of a pure substance. As the temperature decreases with the formation of the eutectic phase of impurities (Figure 1), the molar fraction of impurities in liquid phase x2 is constantly diminished since the pure substance dissolves in eutectic solution within this temperature range (Giron, Goldbronn, 1995; Staub, Perron, 1974). This is expressed by Equation 2:

in which x2* is a fraction of the impurities in the liquid phase of the completely melted substance or in the original substance, and F is the melted fraction.


When Equation 2 is substituted in Equation 1 it gives Equation 3. This equation describes the linear correlation between Tm and 1/F with an intercept of To and slope of x2*.R.To2 /ΔHm :

x2*- mol fraction of impurity in the original sample,

To - melting temperature of the pure main component, in K,

ΔHm - melting heat of the pure main component in J mol-1,

F- fraction of sample melted at Tm,

R- gas constant, 8.314 J mol-1 deg.

To obtain the graph of Tm x 1/F from the DSC curve, some points, normally within the region between 10 and 50% of the peak, are evaluated. The fraction melted is the ratio of the partial area ΔHi (measured from the beginning of the peak to temperature Tm) to the total area ΔHtot. The range between 10 and 50% of the peak is also used to exclude the highest concentration of impurities in liquid phase and to avoid errors in the region where the rate of melting is also greater. If the calculated values show correlation among each other then the curve produced by Tm x 1/F will typically be represented by a straight line (Figure 2). The purity determination by DSC is explained by the theory of non-linearity, that is, the area where melting is normally initiated occurs due to a series of impurities melting before the main compound, thereby explaining a non-linear process. Thus, the accepted procedure for linearization of these data includes correction for this area (Blaine, Schoff, 1983). The melting temperature, melting heat, and purity were determined using Shimadzu TASYS software.


The determination of degree of purity utilizing DSC cannot be performed in the following situations: a) when the compounds decompose at the stage subsequent to melting; b) when the impurities form a eutectic with the principal component, that is, the impurities are soluble in the liquid phase of the principal component; c) for amorphous materials, since the melting peak is related to the variation of enthalpy in function of crystallinity; d) when the melting heat is independent of temperature; e) when other events take place close to melting point such as volatization or polymorphic transition (Blaine, Schoff, 1983).

As thermal analysis has stimulated great interest from investigators and scientists in the last few years regarding studies involving the quality control of drugs and medicines, the objective of this study was the use of DSC for determining the purity of zidovudine (AZT) of samples provided by different pharmaceutical laboratories. In this study, the following parameters were determined: purity, temperature of onset of melting (Tonset), variation of enthalpy occurring during melting, besides the effects of variation of mass and heating rate on the results.

MATERIAL AND METHODS

Material

The standard reference of AZT (C10H13N5O4 ; 3´-azido-2,3´-dideoxythymidine) was obtained from the United States Pharmacopeia (lot: F, N° cat. 72450). DSC analyses of six AZT samples supplied by Laboratório Farmacêutico do Estado de Pernambuco-LAFEPE, batch AZN001, Fundação para o Remédio Popular-FURP, batch 22459710, Indústria Química do Estado de Goiás-IQUEGO, batch 00100/001, Sanval Indústria Farmacêutica LTDA, batch AB934, Fundação Oswaldo Cruz-Farmanguinhos, batch AZL027, and Glaxo Wellcome Indústria Farmacêutica, lote 9054001, were carried out. These samples were randomly labeled A, B, C, D, E and F. The standard reference of AZT was called the R product.

Methods

DSC curves were recorded in a DSC-50 cell (Shimadzu) using aluminum crucibles containing ~ 2 mg of samples, under dynamic nitrogen atmosphere (50 mL min-1) and a heating rate of 2 °C min-1 in the temperature range of 25 to 600 °C. The DSC cell was calibrated with indium (m.p. 156.6 °C; ΔHm = 28.54 J g-1). An empty pan sealed with a cover pan was used as a reference sample. The analyses were done in triplicate.

RESULTS AND DISCUSSION

The endotherm of melting corresponds to the portion of the DSC curve that is far from the baseline, and, later, returns to it. The melting temperature, Tonset, is defined by the extrapolated beginning of the curve, being defined by the point of intersection of the tangent with the point of maximum slope, on the principal side of the peak with the base line extrapolated. Figure 3 shows the endotherm of melting of AZT (reference standard) obtained at the heating rate of 2 °C min-1. The graph shows that the sample presents an onset temperature for melting of 122.80 °C with ΔHm= 31.12 kJ mol-1 and a purity of 99.83%. The 1/F plot shows the melting temperature Tm measured at several ΔHi as a function of the reciprocal of the liquid fraction 1/F before and after the required linearization (Figure 4). This figure shows the melting temperature, melting heat, and purity obtained by Shimadzu TASYS software.



The effect of heating rate on the purity determination, melting point, and enthalpy of melting was conducted using a sample mass of about 2 mg. To perform the analysis, sample B was randomly chosen for evaluation of the effects of heating rate and mass. Figure 5 shows the DSC curves of AZT (substance A) at heating rates of 1, 2, 5, 10, and 15 °C min-1.This study demonstrated that the heating rate had a significant effect on the values obtained for purity, since the values of purity diminish as the heating rate increases (Table I). This observation can be explained by the fact that the Van´t Hoff equation is valid for conditions of equilibrium and therefore lower heating rates are recommended.


Figure 6 depicts the DSC curve of AZT (substance A) when the mass of the sample is varied by 0.58 to 7.44 mg, for a heating rate of 2 °C min-1. In this case, the degree of purity, employing different values of sample mass, was above 99.53%. These results show clearly that the purity, melting heat, and temperature for the onset of melting undergo small variations when the sample mass is changed, as shown in Table II.


The representative DSC curves for the six AZT samples provided by laboratories A, B, C, D, E, and F (Figure 7). The determination of chemical purity of the samples was done based on the correction factor (corrected value) shown for each in Table III. Consequently, the smaller the correction factor used, the greater the purity of the material. The degree of purity, the melting temperature, and the ΔHm found in the analyses by DSC for the different samples of AZT are described in Table III.


The results were obtained using a low heating rate (2 °C min-1) in order to better define the event. The data obtained evidences that the standard reference presented a degree of purity of 99.83%, and that the AZT samples presented a variation ranging from 97.59 to 99.54%. Given that the standard deviation is generally used to define intervals around the average, the results obtained for the tests in triplicate by DSC were shown to be sufficiently consistent and with a low dispersion of values. The results in Table I show that the standard reference presented a temperature for the onset of melting of 122.80 °C. Concerning the samples of active ingredients provided by the different laboratories, a variation of the range of melting of between 118.70 and 122.87°C is evident. In terms of ΔHm,, the samples presented an average value of 31.12 kJ mol-1.

The results obtained for product D show a greater proportion of impurity (2.41%) than did the remaining products, while a widening and shifting of the temperature onset of melting to a lower temperature was also observed, in this case to 118.70 °C.

CONCLUSION

The determination of the purity of substances by DSC is considered a fast and reliable technique for eutectic systems, in which melting is not accompanied by thermal decomposition of the material. The main advantages of the technique over chromatography and spectroscopy are that analysis of all the impurities normally does not require standards except for the main compound, and permits simultaneous measurement of the melting range and variation of enthalpy. In this study, the comparison of six commercial samples of AZT medicine and the analysis of the chemical reference substance provided by the United States Pharmacopeia was conducted.

The results showed that the standard reference presented a proportion of 99.83% while the AZT samples presented a range of variation between 97.59 and 99.54%. Given that standard deviation is generally used to define intervals around the average, the obtained results for the tests in triplicate by DSC show sufficient agreement and a low dispersion of values. This can be verified even though the standard reference presented a temperature for the onset of melting of 122.80 °C. For the samples of active ingredients provided by the different laboratories, a variation ranging from 118.70 to 122.87 °C was ascertained. With regards to ΔHm, the samples presented an average value of 31.12 kJ.mol-1. In conclusion, the number of studies published in the last few years, the speed in obtaining results, and the development of computer calculus programs have stimulated this technique and shown it to be promising.

This study demonstrated that the heating rate had a significant effect on the values obtained for purity, since the values of purity diminish as the heating rate increases. The results show clearly that the purity, melting heat, and temperature for the onset of melting undergo small variations when the sample mass is changed.

ACKNOWLEDGEMENTS

The authors would like to thank the Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) and the Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP) for their financial support.

Received for publication on 17th February 2009

Accepted for publication on 03th June 2009

References

  • ARAÚJO, A. A. S., STORPIRTIS, S., MERCURI, L. P., CARVALHO, F. M. S. SANTOS FILHO, M., MATOS, J. R. Thermal analysis of the antiretroviral zidovudine (AZT) and evaluation of the compatibility with excipients used in solid dosage forms. Int. J. Pharm., v.260, p.303-314,2003.
  • ARAÚJO, A. A. S., CIDES, L. C. S., STORPIRTIS, S., MATOS, J. R., BRUNS, R. E. Effects of experimental conditions on the estimation of kinetic parameters of the thermal decomposition of AZT using factorial design. J. Therm. Anal. Cal, v.79, p.6187-6192, 2005.
  • ARAÚJO, A. A. S., MERCURI, L. P., SEIXAS, S. R. S., STORPIRTIS, S., MATOS, J. R. Determinação dos teores de umidade e cinzas de amostras comerciais de guaraná utilizando métodos convencionais e análise térmica. Rev. Bras. Ciênc. Farm., v.42, p.269-277, 2006.
  • BARBAS, R. R. P., PUIGJANER, C. A new polymorph of norfloxacin. Complete characterization and relative stability of its trimorphic system. J. Therm. Anal. Cal., v.89, p.687-692, 2007.
  • BEZJAK, A., GRDINIC, V., CETINA-CIZMEK., B. Refinement of parameters in differential scanning calorimetric purity determinations. Acta Pharm., v.42, p.239-250, 1992.
  • BLAINE, R. L. & SCHOFF, C. K. Purity determinations by thermal methods Baltimore: ASTM Publications Code Number, 1983. p.151.
  • BOWMAN, P. B., ROGERS, L. B. Purity determinations by differential thermal analysis. Talanta, v.14, p.377-411, 1967.
  • BUCKTON, G., RUSSEL S. J., BEEZER, A. E. Pharmaceutical calorimetry: a selective review. Thermochim. Acta, v.193, p.195-214, 1991.
  • CAMMENGA, H. K., EPPLE, M. Basic principles of thermoanalytical techniques and their applications in preparative chemistry. Angew. Chem., Int. Ed. Engl, v.34, p.1171-1187, 1995.
  • CANOTILHO, J., SOUSA, AT. B., PINTO, J. A. M. C. Análise térmica: interesse em tecnologia farmacêutica. Rev. Port. Farm, v.42, n.4, p.5-12, 1992.
  • FORD, J. L., TIMMINS, P. Pharmaceutical thermal analysis: technique and application. New York: Willey-Interscience, 1986. 313 p.
  • GIRON, D. Contribution of thermal methods and related techniques to the rational development of pharmaceuticals. Part 1. Pharm. Sci. Technol Today, v.1, p.191-199, 1998.
  • GIRON, D. Contribution of thermal methods and related techniques to the rational development of pharmaceuticals. Part 2. Pharm. Sci. Technol Today, v.6, p.262-268, 1998.
  • GIRON, D., GOLDBRONN, C. Place of DSC purity analysis in pharmaceutical development. J. Therm. Anal., v.44, p.217-251, 1995.
  • GIRON, D. Thermal analysis, microcalorimetry and combined techniques for the study of pharmaceuticals. J. Therm. Anal. Cal, v.56, p.1285-1304, 1999.
  • HAINES, P. J. Thermal methods of analysis: principles, application and problems. London: Chapman & Hall, 1995. 286 p.
  • HARDY, M. J. Pharmaceutical applications of thermal analysis. In. CHARSLEY, E. L., WARRINGTON, S. B. (Eds.). Thermal Analysis: techniques & applications. Cambridge: Royal Society of Chemistry, 1992. p.180-197.
  • KISSI, D., ZELKÓ, R., NOVÁK, C. S., ÉHEN, Z. S. Application of DSC and NIRS to study the compatibility of metronidazole with different pharmaceutical excipients. J. Therm. Anal. Cal, v.84, p.447-451, 2006.
  • LVOVA, M. Sh.; KOZLOV, E. I.; WISHNEVEZKAYA, I. A. Thermal analysis in the quality control and standardization of some drugs. J. Therm. Anal, v.40, p.405-411, 1993.
  • STAUB, H., PERRON, W. New method of purity determination by means of calorimetric differential thermal analysis. Anal. Chem., v.46, p.128-130, 1974.
  • THOMPSON, K. C. Pharmaceutical applications of calorimetric measurements in the new millennium. Thermochim. Acta, v.355, p.83-87, 2000.
  • WENDLANDT, W. W. Thermal analysis 3.ed. New York: Willey, 1986. 814 p.
  • WIDMANN, G., SCHERRER, O. A new program for DSC purity analysis. J. Therm. Anal., v.37, p.1957-1964, 1991.
  • *
    Correspondence: A. A. S. Araújo. Departamento de Fisiologia, Universidade Federal de Sergipe-UFS. Av. Marechal Rondom, s/n -49000-100 - São Cristóvão - Sergipe-Brazil. E-mail:
  • Publication Dates

    • Publication in this collection
      27 May 2010
    • Date of issue
      Mar 2010

    History

    • Accepted
      03 June 2009
    • Received
      17 Feb 2009
    location_on
    Universidade de São Paulo, Faculdade de Ciências Farmacêuticas Av. Prof. Lineu Prestes, n. 580, 05508-000 S. Paulo/SP Brasil, Tel.: (55 11) 3091-3824 - São Paulo - SP - Brazil
    E-mail: bjps@usp.br
    rss_feed Acompanhe os números deste periódico no seu leitor de RSS
    Acessibilidade / Reportar erro