Open-access Real-time changes in rib cage expansion and use of abdominal mechanical stimulation in newborns: a quasi-experimental study

Mudanças na expansibilidade torácica com o uso de estímulo mecânico na região abdominal em recém-nascidos: um estudo quase-experimental

ABSTRACT

Objective:  To assess the rib cage expansion and respiratory rate in newborns using an abdominal stabilization band.

Methods:  The study included 32 newborns of both genders, with gestational age between 35 and 41 weeks. The abdominal stabilization band was used for 15 minutes between the xiphoid process and the anterosuperior iliac crest, with an abdominal contention 0.5cm smaller than the abdominal circumference. The rib cage expansion was evaluated by a breathing transducer (Pneumotrace II™) three minutes before using the band, during the use (15 minutes), and ten minutes after removing the band. The Shapiro-Wilk test verified data normality, and the Wilcoxon test compared the variables considering rib cage expansion and respiratory rate. Significance was set to p<0.05.

Results:  There was an increase in respiratory rate when comparing before and ten minutes after removing (p=0.008) the abdominal stabilization band, as well as when comparing during its use and ten minutes after its removal (p=0.001). There was also an increase in rib cage expansion when comparing before and during the use of the abdominal stabilization band (p=0.005).

Conclusions:  The use of the abdominal stabilization band promoted an increase in the rib cage expansion and respiratory rate in the assessed newborns and may be a viable option to improve the respiratory kinematics of this population.

Keywords: Infant; newborn; Respiratory mechanics; Pulmonary ventilation

RESUMO

Objetivo:  Avaliar a expansibilidade torácica e a frequência respiratória em recém-nascidos que fizeram uso de uma faixa de estabilização abdominal.

Métodos:  O estudo incluiu 32 recém-nascidos de ambos os sexos, com idade gestacional entre 35 e 41 semanas. A faixa de estabilização abdominal foi mantida por 15 minutos entre o processo xifoide e a espinha ilíaca anterossuperior, com contenção abdominal 0,5 cm menor do que a circunferência abdominal. A expansibilidade torácica foi avaliada por um transdutor piezoelétrico (Pneumotrace II™) 3 minutos antes do uso da faixa, durante 15 minutos, e 10 minutos após sua retirada. A normalidade das variáveis foi testada pelo teste de Shapiro-Wilk e a análise comparativa da expansibilidade torácica e da frequência respiratória foi realizada por meio do teste t pareado, considerando-se p<0,05.

Resultados:  Houve aumento da frequência respiratória quando comparados os tempos antes do uso da faixa e 10 minutos após a retirada (p=0,008), bem como quando comparados os tempos durante o uso e 10 minutos após a retirada da faixa (p=0,001). Houve aumento da expansibilidade torácica quando comparados os tempos antes e durante o uso da faixa (p=0,005).

Conclusões:  O uso da faixa de estabilização abdominal conferiu aumento da expansibilidade torácica e da frequência respiratória nos recém-nascidos estudados, podendo ser uma opção viável para a melhora da cinemática respiratória dessa população.

Palavras-chave: Recém-nascido; Mecânica respiratória; Ventilação pulmonar

INTRODUCTION

The biomechanical immaturity and neural control of the respiratory system of newborns (NB) put them in a vulnerable position during situations of increased ventilatory demand.1 The horizontal position of the costal margins decreases the diaphragmatic zone of apposition and the efficiency of the intercostal muscles, and unlike the lung, which is less compliant, less mineralized ribs imply a more compliant and deformable rib cage.2 All of these characteristics offer less stability to the chest wall and predispose to the emergence of paradoxical chest wall movements, which increase the diaphragmatic work.3

Unlike in adults, rib cage expansion in the neonatal period obtains advantages from the rib cage response.4 Since NB’ ribs are horizontal, which alters the length-tension relationship of the muscles attached to the rib cage.3,5 Conditions that offer greater support for the rib cage can improve the force of contraction and increase thoracic expansion.4 Lanza et al.6 demonstrated that chest wall mobility in healthy individuals is related to respiratory muscle strength and lung function. However, the authors studied only adult individuals.

Some studies have tried to demonstrate that rib cage stabilization in NB can positively contribute to cardiorespiratory indications,4,7,8 promoting an increase in the response to oxygen saturation or a decrease in heart4 and respiratory rates,7 as well as thoracoabdominal asynchrony.8 However, little is known about the behavior of rib cage expansion after stabilization.

In the last years, significant attention has been directed to methods for non-invasive measurement of the chest wall movements and patterns.9,10 The use of respiratory transducers is a promising method to detect changes in chest expansion and an important variable for thoracoabdominal kinematics.11

Thus, this study hypothesized that the use of an abdominal band can stabilize the NB’ rib cage. We believe that restricting the movement of the floating ribs during spontaneous breathing can modulate respiratory behavior, supporting the thoracoabdominal synchrony and increasing rib cage expansion. Furthermore, we believe that the respiratory transducer may be an effective method to assess this outcome. In this context, the aim of this study was to assess rib cage expansion and respiratory rate in newborns using an abdominal stabilization band.

METHOD

This quasi-experimental study was conducted in the neonatal units of a Maternity School between July and December 2020. Parents were informed about the purposes, benefits, and risks before enrollment, and all agreed to participate and signed a consent form. The study was approved by a Research Ethics Committee (protocol number 3.302.184) in accordance with the Declaration of Helsinki.

The study included 32 NB, of both genders, with gestational age between 35 and 41 weeks, without ventilatory support or supplemental oxygen at the time of evaluation. Those with chest wall deformities or skin lesions, necrotizing enterocolitis, hyperthermia, hypothermia, abdominal diseases, congenital diaphragmatic hernia, or presenting a chest drain tube were excluded. Those who missed the study procedures due to intense inconsolable crying were also excluded from the analysis.

Data collection was performed in an adapted portable crib with the NB in the supine position. The rib cage perimeter was measured before placing the stabilization band using a non-extensible measuring tape (in centimeters), with the xiphoid process as a reference point.12

For rib cage stabilization, a hypoallergenic abdominal band (cotton, polyamide, and elastodiene), with velcro closure, and marked in centimeters in the horizontal direction (Figure 1) was used. At the end of the inspiratory phase, the band was adjusted between the xiphoid process and the anterosuperior iliac crest, with an abdominal contention 0.5cm smaller than the abdominal circumference measured without the band. The measure of abdominal circumference was obtained with a non stretchable tape wrapped around the xiphoid process level while NB were in a supine position.

Figure 1
Newborn using the abdominal stabilization band (beige color) and Pneumotrace II™ (black color) to assess the rib cage expansion.

The abdominal stabilization band was maintained for 15 minutes, and the rib cage expansion (mV) and respiratory rate (per minute) were evaluated using a respiratory transducer (Pneumotrace II™, UFI, Morro Bay, CA, United States) attached to the abdominal stabilization band and with amplification and filtering acquired by Powerlab® (ADInstruments, Australia). The respiratory belt transducer is designed to measure changes in chest diameter and produces a linear voltage proportional to changes in length resulting from breathing.

The respiratory transducer was positioned on the thoracic area (xiphoid process, in the intermammary cleft), and chest expansion was measured three minutes before the band use, during its use (after 15 minutes of use) and ten minutes after its removal (Figure 2), using a sampling frequency of 20 millivolts. A blinded researcher measured offline the results obtained by the Pneumotrace II (respiratory rate and rib cage expansion) at the three different moments assessed in this study.

Figure 2
Study procedures.

Statistical analysis was performed using the Statistical Package for the Social Sciences software (SPSS Inc., Chicago, IL, USA), version 20.0. Data normality was tested using the Shapiro-Wilk test, showing a not normal distribution, and data were presented as median and interquartile range. The Wilcoxon test compared the variables considering rib cage expansion and respiratory rate at three different moments, independently: before using the abdominal band, immediately after the abdominal band, and ten minutes after the last measurement. For all statistical analyses, a significance level of <0.05 was adopted. A post-hoc analysis considering a Cohen’s d effect size of 0.95 between rib cage expansion before and ten minutes after the abdominal stabilization band and α error of 0.02 showed a statistical power of 53% for this study.

RESULTS

During the period of this study, a total of 90 NB were hospitalized and 35 met the inclusion criteria. Three NB were excluded from the study due to intense inconsolable crying. Out of the 32 NB who completed the protocol, 14 were female, and 21 were born by cesarian section. Table 1 presents the sample description.

Table 1
Sample characteristics (n=32).

No differences in respiratory rate were observed before and during the use of the abdominal stabilization band. However, there was a significant increase in respiratory rate when comparing before and ten minutes after removing (p=0.008) the abdominal stabilization band, as well as when comparing during its use and ten minutes after its removal (p=0.001).

When comparing rib cage expansion before and during the use of the abdominal stabilization band, a significant increase was observed (p=0.005). However, although there was also an increase in rib cage expansion when comparing before and ten minutes after using the abdominal stabilization band, this difference did not reach statistical significance. Values from the rib cage expansion and respiratory rate before, during and ten minutes after using the abdominal stabilization band, as well as the Wilcoxon test, are shown in Table 2.

Table 2
Rib cage expansion and respiratory rate before, during, and ten minutes after using the abdominal stabilization band (n=32).

DISCUSSION

The present study aimed to assess, in a short term, the response of rib cage expansion in NB during and after using an abdominal stabilization band. The results demonstrate that there was a significant increase in respiratory rate when comparing before and ten minutes after removing the abdominal stabilization band, as well as when comparing during its use and ten minutes after its removal. Rib cage expansion increased significantly while the band was being used. These findings can be attributed to the activation of chemoreceptors and baroreceptors, as well as the biomechanical effects induced by the utilization of the abdominal stabilization band.13,14

The changes that continually occur in ventilation are the result of the integration of afferent and efferent signals between chemoreceptors, baroreceptors, upper airways and lungs, which have influence in the control of breathing through the modulation of the neuronal networks that make up the respiratory center, a structure located in the brainstem.15 However, during the neonatal period, this communication between the respiratory center and lungs is still in an immature state, which could potentially explain the delayed response (observed after 10 minutes) to mechanical stimulation.13

Moreover, the increase in the respiratory rate could also be explained by the mechanical adjustment provided by the band, which restricts the abdominal wall, limiting the diaphragmatic excursion and leading to a compensatory activation of the respiratory center.16

Despite the observed increase in respiratory rate, no increase in respiratory distress was observed in the assessed infants. In the short term, this enhanced respiratory rate, along with improved stabilization of the thoracoabdominal compartment, may contribute to increased rib cage expansion. This finding is particularly noteworthy, as it aligns with the observed increase in rib cage expansion during the use of the abdominal stabilization band.

Neural control plasticity can favor interventions that minimize the mechanical disadvantages experienced by neonates, especially in premature newborns, who present worse thoracoabdominal synchrony response in situations of increased inspiratory resistive load.17 Considering the anatomical conditions of the respiratory mechanics in preterm NB, such as the horizontal position of the costal margins, less mineralized ribs and a more compliant and deformable rib cage,18,19 which are not favorable for the maintenance of a stable lung volume, the abdominal stabilization band utilized in this study, could support the respiratory physiotherapy sessions.

Manual techniques or devices that promote improvement in oxygenation and ventilation levels have a physiological rationale highlighted by the action of baroreceptors.20,21 Studies suggest that NB are more sensitive to peripheral stimuli on the regulation of neural respiratory control and that baroreceptors may be functional even in the neonatal period, but they go through a process of structure sophistication over time.14,22

Other studies corroborate with our hypothesis and have shown that mechanical adjustments performed by physiotherapists7 or adapted accessories4,23 for stabilization of the costal margins may improve some respiratory parameters, such as oxygen saturation,4 thoracoabdominal asynchrony,8 respiratory rate, and respiratory distress.7 Therefore, we hypothesize that concomitantly with the use of the abdominal stabilization band, the physiotherapist’s hands could be employed to enhance another aspect of respiratory function,19 potentially improving respiratory outcomes within a single session. However, the currently existing literature lacks sufficient data regarding the impact of stabilization on rib cage expansion in newborns.

Associated with this mechanical condition, the development of respiratory control comprises complex and dynamic interactions13 and may be subject to structural or functional changes during critical maturation periods.5 Critical periods represent a moment of greater flexibility in response to respiratory control, and it is supposed that at this moment, in which there is a greater susceptibility to alterations, there is also a greater capacity for reversibility (e.g., increased variability in oxygen levels, which are inherent to neonatal intensive care).24

Although different hypotheses were presented here to justify the variability in rib cage expansion and respiratory rate in the evaluated NB, we recognize that this study has some limitations, such as the fact that the protocol start order was not randomized. In addition, the absence of a thoracic belt for a more comprehensive assessment of the thoracoabdominal synchrony is also a limitation that could be explored in future studies. On the other hand, this was the first study assessing the rib cage expansion in NB using an abdominal stabilization band, a low-cost intervention which may help in understanding the responses of respiratory biomechanics in this population.

It is important to highlight that the increase in rib cage expansion was not sustained after the removal of the abdominal stabilization band, which leads to the need to investigate the lasting effects of the manual and instrumental techniques used in respiratory physiotherapy. Besides, studying chest maneuvers to help professionals to optimize health care in a neonatal care is crucial,19 since the literature is still uncertain about respiratory physiotherapy techniques which can minimize the impacts of mechanical ventilation in NB.2527 Therefore, introducing the abdominal stabilization band, as an alternative procedure, could be a start for new clinical trials with preterm NB or NB with complex respiratory conditions, such as respiratory distress syndrome and bronchopulmonary dysplasia.

Based on this, it was possible to observe that the use of the mechanical stimulus with the abdominal stabilization band in the studied NB increased rib cage expansion during its use, as well as increased the respiratory rate during and after the use of the band.

  • Funding
    This study did not receive any funding.
  • Declaration
    The database that originated the article is available with the corresponding author.

REFERENCES

  • 1 Davis GM, Bureau MA. Pulmonary and chest wall mechanics in the control of respiration in the newborn. Clin Perinatol. 1987;14:551-79. PMID: 3311539.
  • 2 Papastamelos C, Panitch HB, England SE, Allen JL. Developmental changes in chest wall compliance in infancy and early childhood. J Appl Physiol (1985). 1995;78:179-84. https://doi.org/10.1152/jappl.1995.78.1.179
    » https://doi.org/10.1152/jappl.1995.78.1.179
  • 3 Heldt GP. Development of stability of the respiratory system in preterm infants. J Appl Physiol (1985). 1988;65:441-4. https://doi.org/10.1152/jappl.1988.65.1.441
    » https://doi.org/10.1152/jappl.1988.65.1.441
  • 4 Brunherotti MA, Martinez FE. Response of oxygen saturation in preterm infants receiving rib cage stabilization with an elastic band in two body positions: a randomized clinical trial. Braz J Phys Ther. 2013;17:105-11. https://doi.org/10.1590/S1413-35552012005000082
    » https://doi.org/10.1590/S1413-35552012005000082
  • 5 Mitchell GS, Johnson SM. Neuroplasticity in respiratory motor control. J Appl Physiol (1985). 2003;94:358-74. https://doi.org/10.1152/japplphysiol.00523.2002
    » https://doi.org/10.1152/japplphysiol.00523.2002
  • 6 Lanza FC, Camargo AA, Archija LR, Selman JP, Malaguti C, Dal Corso S. Chest wall mobility is related to respiratory muscle strength and lung volumes in healthy subjects. Respir Care. 2013;58:2107-12. https://doi.org/10.4187/respcare.02415
    » https://doi.org/10.4187/respcare.02415
  • 7 Roussenq KR, Scalco JC, Rosa GJ, Honório GJ, Schivinski CI. Rebalancing thoracoabdominal movements in preterms infants: effects on cardiorespiratory parameters, in behavior, in pain and in the respiratory effort. Acta Fisiatr. 2013;20:118-23. https://doi.org/10.5935/0104-7795.20130019
    » https://doi.org/10.5935/0104-7795.20130019
  • 8 Oliveira TG, Rego MA, Pereira NC, Vaz LO, França DC, Vieira DS, et al. Prone position and reduced thoracoabdominal asynchrony in preterm newborns. J Pediatr (Rio J). 2009;85:443-8. https://doi.org/10.2223/JPED.1932
    » https://doi.org/10.2223/JPED.1932
  • 9 Brüllmann G, Fritsch K, Thurnheer R, Bloch KE. Respiratory monitoring by inductive plethysmography in unrestrained subjects using position sensor-adjusted calibration. Respiration. 2010;79:112-20. https://doi.org/10.1159/000212117
    » https://doi.org/10.1159/000212117
  • 10 Banovcin P, Seidenberg J, von der Hardt H. Pressure sensor plethysmography: a method for assessment of respiratory motion in children. Eur Respir J. 1995;8:167-71. https://doi.org/10.1183/09031936.95.08010167
    » https://doi.org/10.1183/09031936.95.08010167
  • 11 Piuzzi E, Pisa S, Pittella E, Podestà L, Sangiovanni S. Low-cost and portable impedance plethysmography system for the simultaneous detection of respiratory and heart activities. IEEE Sens J. 2019;19:2735-46. https://doi.org/10.1109/JSEN.2018.2887303
    » https://doi.org/10.1109/JSEN.2018.2887303
  • 12 Sociedade Brasileira de Pediatria. Avaliação nutricional da criança e do adolescente: manual de orientação. São Paulo: Sociedade Brasileira de Pediatria. Departamento de Nutrologia; 2009.
  • 13 Carroll JL. Developmental plasticity in respiratory control. J Appl Physiol (1985). 2003;94:375-89. https://doi.org/10.1152/japplphysiol.00809.2002
    » https://doi.org/10.1152/japplphysiol.00809.2002
  • 14 Myers MM, Gomez-Gribben E, Smith KS, Tseng A, Fifer WP. Developmental changes in infant heart rate responses to head-up tilting. Acta Paediatr. 2007;95:77-81. https://doi.org/10.1080/08035250500325074
    » https://doi.org/10.1080/08035250500325074
  • 15 Gauda EB, McLemore GL, Tolosa J, Marston-Nelson J, Kwak D. Maturation of peripheral arterial chemoreceptors in relation to neonatal apnoea. Semin Neonatol. 2004;9:181-94. https://doi.org/10.1016/j.siny.2003.11.002
    » https://doi.org/10.1016/j.siny.2003.11.002
  • 16 Bodegård G. Control of respiration in newborn babies. IV. Rib cage stability and respiratory regulation. Acta Paediatr Scand. 1976;65:257-66. https://doi.org/10.1111/j.1651-2227.1976.tb16547.x
    » https://doi.org/10.1111/j.1651-2227.1976.tb16547.x
  • 17 Deoras KS, Greenspan JS, Wolfson MR, Keklikian EN, Shaffer TH, Allen JL. Effects of inspiratory resistive loading on chest wall motion and ventilation: differences between preterm and full-term infants. Pediatr Res. 1992;32:589-94. https://doi.org/10.1203/00006450-199211000-00022
    » https://doi.org/10.1203/00006450-199211000-00022
  • 18 Ribeiro SN, Lourenço LB, Sena GS, Sampaio SS, Pereira SA. Thoracoabdominal interaction and its relationship with biological risk factors in premature newborns. Fisioter Pesqui. 2020;27:155-60. https://doi.org/10.1590/1809-2950/19011927022020
    » https://doi.org/10.1590/1809-2950/19011927022020
  • 19 Souza BH, Sampaio SS, Moura JR, Holanda H, Torres VB, Azevedo IG, et al. Thoracoabdominal rebalancing is not superior to manual hyperinflation to increase the amount of pulmonary secretion removed in preterm newborns: a randomized crossover trial. Health Sci Rep. 2023;6:e1367. https://doi.org/10.1002/hsr2.1367
    » https://doi.org/10.1002/hsr2.1367
  • 20 Armstrong M, Kerndt CC, Moore RA. Physiology, baroreceptors. In: StatPearls [Internet]. Treasure Island: StatPearls Publishing; 2023. PMID: 30844199.
  • 21 Cooper VL, Pearson SB, Bowker CM, Elliott MW, Hainsworth R. Interaction of chemoreceptor and baroreceptor reflexes by hypoxia and hypercapnia – a mechanism for promoting hypertension in obstructive sleep apnoea. J Physiol. 2005;568:677-87. https://doi.org/10.1113/jphysiol.2005.094151
    » https://doi.org/10.1113/jphysiol.2005.094151
  • 22 Grieve PG, Stark RI, Isler JR, Housman SL, Fifer WP, Myers MM. Electrocortical functional connectivity in infancy: response to body tilt. Pediatr Neurol. 2007;37:91-8. https://doi.org/10.1016/j.pediatrneurol.2007.04.004
    » https://doi.org/10.1016/j.pediatrneurol.2007.04.004
  • 23 Petersen CG. The effect of abdominal binding on weaning a patient with a large ventral abdominal wall hernia. Chest. 2003;124:312S. https://doi.org/10.1378/chest.124.4_MeetingAbstracts.312S-a
    » https://doi.org/10.1378/chest.124.4_MeetingAbstracts.312S-a
  • 24 Bavis RW, MacFarlane PM. Developmental plasticity in the neural control of breathing. Exp Neurol. 2017;287:176-91. https://doi.org/10.1016/j.expneurol.2016.05.032
    » https://doi.org/10.1016/j.expneurol.2016.05.032
  • 25 Gomes DC, Fonseca Filho GG, Araujo AG, Gomes VL, Medeiros Júnior NB, Cavalcanti BE, et al. Evaluation of thoracoabdominal mobility by biophotogrammetry in newborns after respiratory physical therapy. Fisioter Bras. 2018;19:28-34.
  • 26 Matilde IN, Eid RA, Nunes AF, Ambrozin AR, Moura RH, Carnieli-Cazati D, et al. Manobras de higiene brônquica em pacientes em ventilação mecânica: quais e por que são usadas? Einstein (São Paulo). 2018;16:1-7. https://doi.org/10.1590/S1679-45082018AO3856
    » https://doi.org/10.1590/S1679-45082018AO3856
  • 27 Nicolau CM, Falcão MC. Effects of respiratory physiotherapy on the cardiopulmonary function of very low birth weight newborn infants. Rev Paul Pediatr. 2010;28:170-5. https://doi.org/10.1590/S0103-05822010000200007
    » https://doi.org/10.1590/S0103-05822010000200007

Publication Dates

  • Publication in this collection
    22 Dec 2023
  • Date of issue
    2024

History

  • Received
    13 Feb 2023
  • Accepted
    10 Sept 2023
location_on
Sociedade de Pediatria de São Paulo R. Maria Figueiredo, 595 - 10o andar, 04002-003 São Paulo - SP - Brasil, Tel./Fax: (11 55) 3284-0308; 3289-9809; 3284-0051 - São Paulo - SP - Brazil
E-mail: rpp@spsp.org.br
rss_feed Acompanhe os números deste periódico no seu leitor de RSS
Acessibilidade / Reportar erro