OBJECTIVE: To evaluate the correlation and concordance between the measurements of echocardiographic analysis of cardiac dimensions obtained through the Echo off-line applicative (software for obtaining digitized image measurements in a dedicated workstation) available to public domain, and those obtained through the conventional method. METHOD: Transversal contemporary study, of 56 randomized patients whose images were obtained during examina-tions. The measurements of the M mode and 2D, were done in the ventricles, left atrium, and aorta by the Echo off-line program. These measurements were compared to those obtained by another professional, through Pearson's correlation test (r), alpha = 0.05 and by concordance analysis (Bland and Altman). RESULT: The measurements carried out by the Echo off-line system showed r that varied from 0.85 to 0.98. The analysis of concordance showed that for most measu-rements, the mean difference between the methods was approximately zero. The variation of absolute values did not show, in average, a clinical significance. The Echo off-line applicative allows a reduction of approximately 30% in the time spent to obtain the measurements. CONCLUSION: This study demonstrated the accuracy of the Echo off-line program to measure cardiac dimensions in a dedicated workstation, showing that it can be routinely used in echocardiography labs.