Repeatability of contrast sensitivity testing in patients with age-related macular degeneration, glaucoma, and cataract

Reprodutibilidade dos escores de sensibilidade ao contraste em pacientes com degeneração macular relacionadas à idade, glaucoma e catarata

ABSTRACT

Purpose:

To analyze the intrasession and intersession repeatability of contrast sensitivity (CS) measurements in patients with glaucoma, cataract, or age-related macular degeneration (AMD) and healthy controls.

Methods:

CS measurements were performed using the OPTEC-Functional Vision Analyzer (FVA), which uses a standardized and closed (view-in) system. Measurements for patients with glaucoma, cataract, or AMD and healthy controls were repeated within 30 minutes (intrasession) and during two sessions (intersession), separated by one week to one month. Test-retest reliability and correlation were measured using the intraclass correlation coefficient (ICC) and coefficient of repeatability (COR).

Results:

Ninety subjects (90 eyes) with visual acuity of 0.17 logMAR or higher in the cataract group or 0.00 logMAR in the other groups were included. During the first session, the ICC values were 0.87, 0.90, 0.76, and 0.69, and COR values were 0.24, 0.20, 0.38, and 0.25 for the control, glaucoma, cataract, and AMD groups, respectively. The reliability scores significantly improved during the second session, except in the glaucoma group. There was an acceptable floor effect and no ceiling effect at higher frequencies in the glaucoma and AMD groups.

Conclusion:

In subjects with good visual acuity, the FVA system is useful for evaluating CS and demonstrates good repeatability, as shown by ICC and COR. Because there is no ceiling effect, this system is beneficial for evaluation of early changes in CS, particularly in patients with glaucoma or AMD.

Keywords:
Macular degeneration; Contrast sensitivity; Glaucoma; Age effect; Reproducibility of results

RESUMO

Objetivo:

Analisar a reprodutibilidade intrassessão e intersessão das medidas de sensibilidade ao contraste (CS) em pacientes com degeneração macular relacionada à idade (AMD), glaucoma e catarata.

Método:

As medidas de CS foram feitas pelo OPTEC-Funcional Visão Analyzer (FVA), que utiliza um sistema padronizado e fechado de avaliação da acuidade visual. Medidas em pacientes com AMD, glaucoma, catarata e nos controles saudáveis foram repetidas no prazo de 30 minutos (intrassessão) em duas visitas (intersessão), separadas por uma semana a um mês. A confiabilidade e correlação teste-reteste foram calculados por meio do coeficiente de correlação intraclasse (ICC) e coeficiente de reprodutibilidade (COR).

Resultados:

Noventa olhos de 90 indivíduos foram recrutados com acuidade visual de 0,17 logMAR ou melhor em catarata e 0,00 logMAR nos outros grupos. A confiabilidade da CS na primeira visita dos grupos normal, glaucoma, catarata e AMD foram, respectivamente, ICC 0,87; 0,90; 0,76; 0,69, e COR 0,24; 0,20; 0,38; 0,25. Os índices de confiabilidade foram significativamente melhorados nas segundas visitas, exceto no grupo glaucoma. Houve um efeito chão aceitável e nenhum efeito teto em frequências mais altas nos grupos glaucoma e AMD.

Conclusões:

Em indivíduos com boa acuidade visual, o sistema FVA de avaliação da CS é útil e apresenta boa confiabilidade, como mostrado pelas análises de ICC e COR. Por não apresentar efeito teto, este sistema parece ser benéfico para a avaliação das alterações precoces de CS, especialmente no glaucoma e AMD.

Descritores:
Degeneração macular; Sensibilidade de contraste; Glaucoma; Fator idade; Reprodutibilidade dos testes

INTRODUCTION

Contrast sensitivity (CS) is the ability to recognize small differences in luminance or differentiate two objects from each other and the background(11 Miller D, Schor P, Magnante P. Optics of the normal eye. In: Yanoff M, Duker JS, editor. Ophthalmology. China: Elsevier; 2014. p. 38-45.,22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.). CS is an important part of functional vision that is related to many activities of daily living and measuring it is one of the best ways to assess vision quality(33 Hong YT, Kim SW, Kim EK, Kim T. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010; 36(4):547-52.

4 Nielsen E, Hjortdal J. Visual acuity and contrast sensitivity after posterior lamellar keratoplasty. Acta Ophthalmol. 2012;90(8):756-60.
-55 Ginsburg AP. Contrast sensitivity: determining the visual quality and function of cataract, intraocular lenses and refractive surgery. Curr Opin Ophthalmol. 2006;17(1):19-26.). Despite having normal visual acuity measured by a Snellen chart, the satisfaction with the results of refractive surgery may vary relative to CS, and some eye diseases cause isolated loss of CS(44 Nielsen E, Hjortdal J. Visual acuity and contrast sensitivity after posterior lamellar keratoplasty. Acta Ophthalmol. 2012;90(8):756-60.,66 Aslam T, Mahmood S, Balaskas K, Patton N, Tanawade R, Tan S, et al. Repeatability of visual function measures in age-related macular degeneration. Graefes Arch Clin Exp Ophthalmol. 2014;252(2):201-6.

7 Onal S, Yenice O, Cakir S, Temel A. FACT contrast sensitivity as a diagnostic tool in glaucoma: FACT contrast sensitivity in glaucoma. Int Ophthalmol. 2008;28(6):407-12.
-88 Zhang J, Zhou YH, Li R, Tian L. Visual performance after conventional LASIK and wavefront-guided LASIK with iris-registration: results at 1 year. Int J Ophthalmol. 2013; 6(4):498-504.). Snellen acuity tests have a limited ability to detect fine changes in contrast because high contrast letters (black on white) are used(99 Ginsburg AP. Contrast sensitivity and functional vision. Int Ophthalmol Clin. 2003; 43(2):5-15.). Also, recent advances in knowledge about refractive surgery, optical tissues, and glaucoma and macular diseases revealed that the Snellen acuity test is inadequate for early detection of eye diseases and measurement of functional vision(66 Aslam T, Mahmood S, Balaskas K, Patton N, Tanawade R, Tan S, et al. Repeatability of visual function measures in age-related macular degeneration. Graefes Arch Clin Exp Ophthalmol. 2014;252(2):201-6.,1010 Lahav K, Levkovitch-Verbin H, Belkin M, Glovinsky Y, Polat U. Reduced mesopic and photopic foveal contrast sensitivity in glaucoma. Arch Ophthalmol. 2011;129(1):16-22.

11 Yenice O, Onal S, Incili B, Temel A, Afşar N, Tanrıdag. Assessment of spatial-contrast function and short-wavelength sensitivity deficits in patients with migraine. Eye (Lond). 2007;21(2):218-23.
-1212 Richman J, Lorenzana LL, Lankaranian D, Dugar J, Mayer J, Wizov SS, et al. Importance of visual acuity and contrast sensitivity in patients with glaucoma. Arch Ophthalmol. 2010;128(12):1576-82.). CS decreases in all stages of glaucoma and age-related macular degeneration (AMD) have been documented using different testing methods(77 Onal S, Yenice O, Cakir S, Temel A. FACT contrast sensitivity as a diagnostic tool in glaucoma: FACT contrast sensitivity in glaucoma. Int Ophthalmol. 2008;28(6):407-12.,1212 Richman J, Lorenzana LL, Lankaranian D, Dugar J, Mayer J, Wizov SS, et al. Importance of visual acuity and contrast sensitivity in patients with glaucoma. Arch Ophthalmol. 2010;128(12):1576-82.

13 Hawkins AS, Szlyk JP, Ardickas Z, Alexander KR, Wilensky JT. Comparison of contrast sensitivity, visual acuity, and Humphrey visual field testing in patients with glaucoma. J Glaucoma. 2003;12(2):134-8.

14 Faria BM, Duman F, Zheng CX, Waisbourd M, Gupta L, Ali M, et al. Evaluating contrast sensitivity in Age-related macular degeneration using a novel computer-based test, the Spaeth/Richman contrast sensitivity test. Retina. 2015;35(7):1465-73.
-1515 Stangos N, Voutas S, Topouzis F, Karampatakis V. Contrast sensitivity evaluation in eyes predisposed to age-related macular degeneration and presenting normal visual acuity. Ophthalmologica. 1995;209(4):194-8.). Glaucoma, cataract, and AMD patients suffer from problems in vision-related activities of daily living and the inability to recognize targets in real world, which can be better identified by CS tests than by visual acuity tests(1414 Faria BM, Duman F, Zheng CX, Waisbourd M, Gupta L, Ali M, et al. Evaluating contrast sensitivity in Age-related macular degeneration using a novel computer-based test, the Spaeth/Richman contrast sensitivity test. Retina. 2015;35(7):1465-73.,1616 Puell MC, Barrio AR, Palomo-Alvarez C, Gómez-Sanz FJ, Clement-Corral A, Pérez-Carrasco MJ. Impaired mesopic visual acuity in eyes with early age-related macular degeneration. Invest Ophthalmol Vis Sci. 2012;53(11):7310-4.

17 Klein J, Pierscionek BK, Lauritzen J, Derntl K, Grzybowski A, Zlatkova MB. The effect of cataract on early stage glaucoma detection using spatial and temporal contrast sensitivity tests. PLoS One. 2015;10(6):e0128681.
-1818 Bansback N, Czoski-Murray C, Carlton J, Lewis G, Hughes L, Espallargues M, et al. Determinants of health related quality of life and health state utility in patients with age related macular degeneration: the association of contrast sensitivity and visual acuity. Qual Life Res. 2007;16(3):533-43.). The increasing importance of visual quality and the need for accurate measurement of visual acuity has led to more interest in CS tests(1212 Richman J, Lorenzana LL, Lankaranian D, Dugar J, Mayer J, Wizov SS, et al. Importance of visual acuity and contrast sensitivity in patients with glaucoma. Arch Ophthalmol. 2010;128(12):1576-82.,1919 Söğütlü Sari E, Kubaloğlu A, Unal M, Liorens DP, Koytak A,Ofluoglu AN, et al. Penetrating keratoplasty versus deep anterior lamellar keratoplasty: comparison of optical and visual quality outcomes. Br J Ophthalmol. 2012;96(8):1063-7.

20 Karakus SH, Basarir B, Pinarci EY, Kirandi EU, Demirok A. Long-term results of half-dose photodynamic therapy for chronic central serous chorioretinopathy with contrast sensitivity changes. Eye (Lond). 2013;27(5):612-20.
-2121 Gertnere J, Solomatin I, Sekundo W. Refractive lenticule extraction (ReLEx flex) and wavefront-optimized Femto-LASIK: comparison of contrast sensitivity and high-order aberrations at 1 year. Graefes Arch Clin Exp Ophthalmol. 2013;251(5):1437-42.).

There are several clinical tests for measuring CS with letters or gratings(55 Ginsburg AP. Contrast sensitivity: determining the visual quality and function of cataract, intraocular lenses and refractive surgery. Curr Opin Ophthalmol. 2006;17(1):19-26.,2222 Thayaparan K, Crossland MD, Michael D, Rubin GS. Clinical assessment of two new contrast sensitivity charts. Br J Ophthalmol. 2007;91(6):749-52.

23 Keane PA, Patel PJ, Ouyang Y, Chen FK, Ikeji F, Walsh AC, et al. Effects of retinal morphology on contrast sensitivity and reading ability in neovascular age-related macular degeneration. Invest Ophthalmol Vis Sci. 2010;51(11):5431-7.

24 Haughom B, Strand T-E. Sine wave mesopic contrast sensitivity - defining the normal range in a young population. Acta Ophthalmol. 2013;91(2):176-82.

25 Li JH, Feng YF, Zhao YE, Zhao YY, Lin L. Contrast visual acuity after multifocal intraocular lens implantation: aspheric versus spherical design. Int J Ophthalmol. 2014; 7(1):100-3.
-2626 Haymes SA, Roberts KF, Cruess AF, Cruess AF, Nicolela MT, LeBlanc RP, et al. The letter contrast sensitivity test: clinical evaluation of a new design. Invest Ophthalmol Vis Sci. 2006;47(6):2739-45.). A limitation of CS tests is low reliability associated with varying environmental conditions and the subjective nature of the tests. In particular, CS tests that use charts may be affected by inadequate or uneven lighting, fading, the subject's learning curve, and reflections(22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.,33 Hong YT, Kim SW, Kim EK, Kim T. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010; 36(4):547-52.). The Functional Acuity Contrast Test (FACT) assesses five spatial frequencies (1.5, 3, 6, 12, and 18 cycles per degree (cpd)) and 9 levels of contrast for each frequency. FACT uses a small increment (0.15 log unit) and a forced-choice method with three options(22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.). Although the FACT test has greater reliability compared with other CS testing methods, administration of the CS test in a closed system, with charts on the screen inside the device, with a Functional Visual Acuity (FVA) device has additional advantages. These advantages include the ability to display contrast charts in a random order and the administration of the test with standardized lighting, which is difficult to maintain under other conditions(33 Hong YT, Kim SW, Kim EK, Kim T. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010; 36(4):547-52.,2727 Hohberger B, Laemmer R, Adler W, Juenemann AG, Horn FK. Measuring contrast sensitivity in normal subjects with OPTEC 6500: influence of age and glare. Graefes Arch Clin Exp Ophthalmol. 2007;245(12):1805-14.). A CS test is expected to have a good reliability to differentiate a small loss or detect real progression, which can be better achieved by a closed test system, which has inherently stable testing conditions.

To the best of our knowledge, there has been little research published about the comparative reliability of these devices for healthy controls and patients with eye disease. This study investigates the repeatability of the closed system FVA CS test in control subjects and in patients with glaucoma, cataracts and good Snellen visual acuity, or age-related macular degeneration (AMD).

METHODS

This prospective, observational study included patients with glaucoma, cataracts, or AMD and healthy control subjects. All subjects were recruited from the Eye Clinic at Canakkale Onsekiz Mart University. Written informed consent was obtained from all participants following a comprehensive explanation of the study. The research was approved by the local clinical research ethics committee and complied with the tenets of the Helsinki Declaration.

The study included subjects between 40 and 70 years old to facilitate enrollment of subjects with best corrected visual acuity (BCVA) of 0.00 logMAR for the groups without cataracts. One eye of each patient was included in the study. Subjects were not eligible to be included if they had ocular diseases other than glaucoma, cataracts, or AMD, refractive error <-3 or >3 diopters (D) sphere or >2.00 D cylinder, or a history of previous eye surgery or laser treatment. Patients with psychiatric or neurologic disorders, poor compliance during the CS test, or who used systemic or ocular medications could affect CS were excluded. All patients underwent an ophthalmic examination, including auto refractometry, BCVA evaluation, anterior segment biomicroscopy, funduscopy, and IOP, measured with a Goldmann applanation tonometer (GAT).

The glaucoma group included subjects who were diagnosed with primary open-angle glaucoma (POAG) or normotensive glaucoma (NTG). The inclusion criteria for patients with POAG were the characteristic visual field loss, glaucomatous optic neuropathy (having rim thinning or notching, retinal nerve fiber layer defects, or disc hemorrhage and disc asymmetry between the eyes ≥0.2), and intraocular pressure (IOP) >21 mmHg. The inclusion criteria for patients with NTG were visual field loss, glaucomatous optic neuropathy, and IOP ≤21 mmHg. Glaucomatous visual field defects, which were repeatable in at least two standard automated perimetry evaluations, included at least one of the following: nasal step, arcuate scotoma, paracentral scotoma, or temporal wedge. All glaucoma patients had a grade 3 or 4 open angle according to the Shaffer Classification System(2929 Shaffer RN. Primary glaucomas. Gonioscopy, ophthalmoscopy and perimetry. Trans Am Acad Ophthalmol Otolaryngol. 1960;64:112-27.) and a mean deviation no worse than -6 dB.

In the cataract group, only eyes with a BCVA of 0.17 logMAR or better and a nuclear sclerosis of less than grade 2 were included. Nuclear sclerosis was graded by the slit-lamp method according to the Lens Opacities Classification System III(2828 Chylack LT Jr, Wolfe JK, Singer DM, Leske MC, Bullimore MA, Bailey IL, et al. The Lens Opacities Classification System III. The Longitudinal Study of Cataract Study Group. Arch Ophthalmol. 1993;111(6):831-6.). The inclusion criteria for the AMD group were retinal pigment epithelial abnormalities in the form of hypopigmentation or hyperpigmentation, without evidence of active choroidal neovascularization, and no prior treatment with any anti-vascular endothelial growth factor injections. The control subjects had normal findings on a comprehensive eye examination (including anterior segment biomicroscopy and fundoscopy of the macula and optic nerve) and IOP <21 mmHg.

A closed-system CS testing device, the OPTEC-FVA (Stereo Optical Co, Inc., Chicago, IL, USA), was used to examine CS values under daytime without glare. Figure 1 shows the FVA CS testing device. The FVA has rotating slide packages for other visual functions, including visual acuity, disability glare, and stereo and color vision. The CS testing device contains the FACT chart, which uses three orientations for the gratings: oriented vertically or tilted 15º to the right or left. Subjects identified the orientation of each grating and were required to guess when they could not determine the orientation of gratings.

Figure 1
The functional vision analyzer contrast sensitivity test device.

The CS testing device controls light levels with sensors to achieve consistent conditions for each test. This device supplies a target luminance level of 85 cd/m2 for daytime conditions in compliance with the American National Standards Institute (ANSI). It evaluates spatial frequencies using sine-wave grating charts of 1.5, 3, 6, 12, and 18 cpd. Figure 2 presents the sine-wave gratings charts for the frequencies that the CS testing device used. Contrast was defined according to the Michelson formula (C = [Lmaximum − Lminimum]/[Lmaximum + Lminimum], C: Contrast, L: Luminance)(22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.). CS was measured during two sessions separated by at least one week and up to one month (intersession). At each session, two measurements taken 30 minutes apart (intrasession). All tests were performed with the optimum refractive correction and a natural pupil. All the subjects performed the CS test for the first time during our study, and each test lasted approximately 10 minutes. The CS measurements were obtained by a single operator who was blinded to the patient's group.

Figure 2
Sine-wave grating charts in the view-in contrast sensitivity test device.

All the CS values were converted to logarithmic values. The test-retest reliability of the CS test was examined using the intraclass correlation coefficient (ICC) and the coefficient of repeatability (COR) and the corresponding 95% confidence interval (CI). The ICC was calculated with absolute agreement. The COR is calculated as 1.96 times the standard deviation of the difference between the test and retest scores. An ICC value of 1.0 and a COR value of 0.0 represent perfect test-retest reliability(22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.). Reliability analyses were performed within individual sessions and between sessions for all groups. Possible floor and ceiling effects were evaluated by calculating the percentage of subjects with lowest and highest test scores, respectively, for each of the four tests at each spatial frequency and within each group.

Spearman rank-order correlations were used to investigate the possible correlation between age and changes in CS score. The Wilcoxon and Mann-Whitney U-tests were used to compare CS scores between groups. The Bonferroni correction, which applies a statistical significance level of 1/n times the "p" value (n= the number of comparisons), was used to test for multiple comparisons to ensure that an appropriate level of significance was applied to the individual tests(3030 Armstrong RA, Davies LN, Dunne MC, Gilmartin B. Statistical guidelines for clinical studies of human vision. Ophthalmic Physiol Opt. 2011;31(2):123-36.). Statistical analysis was performed using SPSS software (Windows version 19.0; SPSS, Inc., Chicago, IL, USA), and statistical significance was defined as p<0.05.

RESULTS

The analysis was included 90 eyes of 90 Caucasian subjects, including healthy control (n=30; mean ± standard deviation age 47.1 ± 6.7 years; 15 males), glaucoma (n=13; age 59.6 ± 4.1 years; 8 males), cataract (n=29; age 56.3 ± 5.2 years; 14 males), and AMD (n=18; age 59.9 ± 4.6 years; 9 males) groups.

The test-retest reliability of the CS test, including ICC and COR, is shown in table 1. The reliability of the CS test in the control, glaucoma, cataract, and AMD groups were good at the first session (ICC: 0.87 [95% CI 0.84 -0.9], 0.90 [CI 0.84 -0.96], 0.76 [CI 0.65 -0.87], and 0.69 [CI 0.59 -0.79], respectively; and COR 0.24, 0.20, 0.38, and 0.25, respectively). However, at the second session, statistically significant increases were found in the reliability of the CS test, except in the glaucoma group (ICC: 0.92 [CI 0.88 -0.96], p=0.042; 0.93 [CI 0.88 -0.98], p=0.461; 0.90 [CI 0.86 -0.94], p=0.043; and 0.86 [CI 0.82 -0.9], p=0.043; and COR of 0.16, 0.16, 0.24, and 0.21 in the control, glaucoma, cataract, and AMD groups, respectively).

Table 1
Test-retest reliability of the contrast sensitivity test in the groups of control, glaucoma, cataract, and age-related macular degeneration groups

Figure 3 illustrates the mean CS score for all groups as a function of spatial frequency. The CS scores for the AMD group were significantly lower than for those in the control group at all frequencies (significance of Bonferroni correction 0.012, p<0.001). In the glaucoma group, except at a frequency of 1.5 cpd, the CS scores were significantly lower compared with control group (significance of Bonferroni correction 0.012, p<0.012). There was also a statistically significant correlation between age and changes in CS scores of subjects at all spatial frequencies (1st session p<0.001, r=-0.399, -0.484, -0.484, -0.386, and -0.395 at 1.5, 3, 6, 12, and 18 cpd, respectively; 2nd session p<0.001, r=-0.424, -0.450, -0.495, -0.432, and -0.444 at 1.5, 3, 6, 12, and 18 cpd, respectively).

Figure 3
Contrast sensitivity for the control, glaucoma, cataract, and age-related macular degeneration (AMD) groups as a function of spatial frequency.

Table 2 summarizes the results of the analysis of possible floor and ceiling effects at each spatial frequency and for each group. There was a prominent floor effect at the highest frequency (18 cpd) in all groups. In healthy eyes, a floor effect, except in the second test, only occurred at 18 cpd. However, in the glaucoma group, the floor effect was greatest for the first two tests and decreased for the last CS tests. There was no ceiling effect in the glaucoma or AMD groups at any of the tested frequencies. In healthy and cataract eyes, a low ceiling effect (3.3%-10%), with a slight decrease during the last two tests, was detected.

Table 2
The floor and ceiling effect for each spatial frequency and group (measured by the percentage of maximum and minimum CS scores in the control (n=30), glaucoma (n=13), cataract (n=29), and AMD (n=18) groups)

DISCUSSION

This study showed that the FVA CS test has good reliability, as shown by ICC and COR analysis. The CS test is useful for patients with good visual acuity who have healthy eyes without glaucoma, cataracts, or AMD. Although there are other commercially available CS tests using letters, symbols, or sine-wave gratings, these tests are not clearly standardized for performing a common CS evaluation such as visual field analyses(22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.,2323 Keane PA, Patel PJ, Ouyang Y, Chen FK, Ikeji F, Walsh AC, et al. Effects of retinal morphology on contrast sensitivity and reading ability in neovascular age-related macular degeneration. Invest Ophthalmol Vis Sci. 2010;51(11):5431-7.). Sine-wave gratings permit sensitive testing of individual visual channels and are part of spatial frequency for evaluating CS function in vision science(1212 Richman J, Lorenzana LL, Lankaranian D, Dugar J, Mayer J, Wizov SS, et al. Importance of visual acuity and contrast sensitivity in patients with glaucoma. Arch Ophthalmol. 2010;128(12):1576-82.). The FACT chart also uses sine-wave gratings, which have been chosen at given luminance and glare levels for the ANSI by the United States Food and Drug Administration(55 Ginsburg AP. Contrast sensitivity: determining the visual quality and function of cataract, intraocular lenses and refractive surgery. Curr Opin Ophthalmol. 2006;17(1):19-26.). Although the FACT chart is a modified version of the Vistech, it uses smaller increments (0.15 log units) and subjects must choose from three options. Previous CS studies with sine-wave gratings reported insufficient scores for reliability analyses (ICC 0.28-0.64 and COR 0.26-0.58)(22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.). Despite having poor reliability with FACT wall charts, closed systems using the standard FACT luminance and glare values were reported to exactly match the ANSI standards, thereby providing better reliability(33 Hong YT, Kim SW, Kim EK, Kim T. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010; 36(4):547-52., 2727 Hohberger B, Laemmer R, Adler W, Juenemann AG, Horn FK. Measuring contrast sensitivity in normal subjects with OPTEC 6500: influence of age and glare. Graefes Arch Clin Exp Ophthalmol. 2007;245(12):1805-14.). Hohberger et al.(2727 Hohberger B, Laemmer R, Adler W, Juenemann AG, Horn FK. Measuring contrast sensitivity in normal subjects with OPTEC 6500: influence of age and glare. Graefes Arch Clin Exp Ophthalmol. 2007;245(12):1805-14.) reported average reliability coefficients of 0.80-0.96 for each cpd with an OPTEC 6500 CS test. Using the same device, Hong et al.(33 Hong YT, Kim SW, Kim EK, Kim T. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010; 36(4):547-52.) reported ICC 0.85 and COR 0.20, which are comparable to our results with the FVA closed system CS test (intersession ICC 0.86-0.94 and COR 0.16-0.19).

Previous studies with FACT charts in closed systems omitted any comparison with earlier and latter CS test scores that might have explained possible improvements in test reliability. We think that such comparisons are important to establish a reliable clinical application of the CS test. In this study, the mean ICCs of the closed system CS test were better at the second session (0.86-0.94) than the first session (0.69-0.90) in all groups. Furthermore, the CORs improved at the second session (first session 0.20-0.38; second session 0.16-0.24). There may be a learning effect, which possibly arises from cognitive abilities or getting familiar with the test procedure during the repeated tests. It has been reported that using letter charts for CS tests result in learning effects according to variable abilities to recognize letters. The FACT charts that use sine-wave gratings eliminate these problems(22 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.,3131 Bühren J, Terzi E, Bach M, Wesemann W, Kohnen T. Measuring contrast sensitivity under different lighting conditions: comparison of three tests. Optom Vis Sci. 2006;83(5): 290-8.). However, subjects high probability of correct guessing with the FACT charts, so repetitive tests are needed to reduce this effect(2727 Hohberger B, Laemmer R, Adler W, Juenemann AG, Horn FK. Measuring contrast sensitivity in normal subjects with OPTEC 6500: influence of age and glare. Graefes Arch Clin Exp Ophthalmol. 2007;245(12):1805-14.,3131 Bühren J, Terzi E, Bach M, Wesemann W, Kohnen T. Measuring contrast sensitivity under different lighting conditions: comparison of three tests. Optom Vis Sci. 2006;83(5): 290-8.). Subjects' responses may become more reliable after learning the FVA CS test procedure so it is important to conduct multiple CS tests to obtain more reliable clinical results.

It has been well documented that the spatial and temporal types of CS decrease in people with glaucoma(1111 Yenice O, Onal S, Incili B, Temel A, Afşar N, Tanrıdag. Assessment of spatial-contrast function and short-wavelength sensitivity deficits in patients with migraine. Eye (Lond). 2007;21(2):218-23.,3232 Lundh BL. Central and peripheral contrast sensitivity for static and dynamic sinusoidal gratings in glaucoma. Acta Ophthalmol. 1985;63(5):487-92.,3333 McKendrick AM, Sampson GP, Walland MJ, Badcock DR. Contrast sensitivity changes due to glaucoma and normal aging: low-spatial-frequency losses in both magnocellular and parvocellular pathways. Invest Ophthalmol Vis Sci. 2007;48(5):2115-22.). Grating CS tests are promising for the detection of early glaucoma and its progression(3232 Lundh BL. Central and peripheral contrast sensitivity for static and dynamic sinusoidal gratings in glaucoma. Acta Ophthalmol. 1985;63(5):487-92.,3434 Ansari EA, Morgan JE, Snowden RJ. Psychophysical characterisation of early functional loss in glaucoma and ocular hypertension. Br J Ophthalmol. 2002;86(10):1131-5.). Klein et al.(1717 Klein J, Pierscionek BK, Lauritzen J, Derntl K, Grzybowski A, Zlatkova MB. The effect of cataract on early stage glaucoma detection using spatial and temporal contrast sensitivity tests. PLoS One. 2015;10(6):e0128681.) suggested that spatial CS were a sensitive indicator of early glaucomatous loss in the presence of cataracts. Although visual acuity was good, the CS of the eyes with glaucoma was significantly less than among healthy subjects except at 1.5 cpd (p<0.012). This result was compatible with that previously reported by Onal et al.(77 Onal S, Yenice O, Cakir S, Temel A. FACT contrast sensitivity as a diagnostic tool in glaucoma: FACT contrast sensitivity in glaucoma. Int Ophthalmol. 2008;28(6):407-12.) using the FACT wall chart. The authors suggested that the FACT chart was useful for early diagnosis of patients with glaucoma, accompanied by a short-wavelength automated perimetry. This may be associated with early degeneration in the magnocellular ganglion cells, which are important for CS, in glaucoma(3535 Sun H, Swanson WH, Arvidson B, Dul MW. Assessment of contrast gain signature in inferred magnocellular and parvocellular pathways in patients with glaucoma. Vision Res. 2008;48(26):2633-41.).

The CS of eyes with AMD was worse than among healthy subjects at all frequencies (p<0.001), similar to previous studies(66 Aslam T, Mahmood S, Balaskas K, Patton N, Tanawade R, Tan S, et al. Repeatability of visual function measures in age-related macular degeneration. Graefes Arch Clin Exp Ophthalmol. 2014;252(2):201-6.,3636 Patel PJ, Chen FK, Rubin GS, Tufail A. Intersession repeatability of contrast sensitivity scores in age-related macular degeneration. Invest Ophthalmol Vis Sci. 2009;50(6): 2621-5.). As there are limited options for treatment of AMD, early detection of the disease and prevention of progression are very important. Visual discomfort in performing daily activities often occurs in patients with AMD even though they have a BCVA of 0.00 logMAR on Snellen charts. This vision quality loss is seen in early AMD before any detectable retinal changes occur(1818 Bansback N, Czoski-Murray C, Carlton J, Lewis G, Hughes L, Espallargues M, et al. Determinants of health related quality of life and health state utility in patients with age related macular degeneration: the association of contrast sensitivity and visual acuity. Qual Life Res. 2007;16(3):533-43.,3737 Scilley K, Jackson GR, Cideciyan AV, Maguire MG, Jacobson SG, Owsley C. Early age-related maculopathy and self-reported visual difficulty in daily life. Ophthalmology. 2002;109(7):1235-42.). CS reduction in AMD patients becomes more evident and uncomfortable under conditions of low illumination, which provides a low contrast environment(1616 Puell MC, Barrio AR, Palomo-Alvarez C, Gómez-Sanz FJ, Clement-Corral A, Pérez-Carrasco MJ. Impaired mesopic visual acuity in eyes with early age-related macular degeneration. Invest Ophthalmol Vis Sci. 2012;53(11):7310-4.).

It was reported previously that the closed system FACT chart had a lower ceiling effect than the wall chart test, which might increase its ability to detect small changes. In our study, no ceiling effect was observed in eyes with glaucoma or AMD using the FVA closed system CS test, and the ceiling effect was acceptable in healthy and cataract eyes(33 Hong YT, Kim SW, Kim EK, Kim T. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010; 36(4):547-52.). Even after a gap of at least one week between the first and second CS tests, the ceiling effect in normal and cataract eyes tended to decrease. These results suggest that the FACT chart in a closed system has the potential to detect small decreases in high CS scores in healthy, cataract, and AMD eyes and otherwise healthy eyes with early stages of glaucoma. A more pronounced floor-effect emerged in glaucoma and AMD eyes at higher frequencies, which may represent the effects of early retinal or ganglion cell defects on the loss of CS.

It is critical to know if the patients are affected by glaucoma, cataracts, or AMD in their daily lives when considering treatment options. CS has become more important for quality of life assessments, which have been receiving increase attention in recent years(3737 Scilley K, Jackson GR, Cideciyan AV, Maguire MG, Jacobson SG, Owsley C. Early age-related maculopathy and self-reported visual difficulty in daily life. Ophthalmology. 2002;109(7):1235-42.

38 Browne C, Brazier J, Carlton J, Alavi Y, Jofre-Bonet M. Estimating quality-adjusted life years from patient-reported visual functioning. Eye (Lond). 2012;26(10):1295-301.
-3939 Wei H, Sawchyn AK, Myers JS, Katz LJ, Moster MR, Wizov SS, Steele M, Lo D, Spaeth GL. A clinical method to assess the effect of visual loss on the ability to perform activities of daily living. Br J Ophthalmol. 2012;96(5):735-41.). CS is involved with differentiating low-contrast objects, such as human faces, and recognizing movement(3737 Scilley K, Jackson GR, Cideciyan AV, Maguire MG, Jacobson SG, Owsley C. Early age-related maculopathy and self-reported visual difficulty in daily life. Ophthalmology. 2002;109(7):1235-42.,4040 Rubin GS, Bandeen-Roche K, Huang GH, Muñoz B, Schein OD, Fried LP, West SK. The association of multiple visual impairments with self-reported visual disability: SEE project. Invest Ophthalmol Vis Sci. 2001;42(1):64-72.). Because of the close relationship between CS and the ability to perform activities of daily living, despite normal BCVA, CS testing is a useful clinical method to assess how ocular disease affects what patients can do. The present study found that FVA closed system CS evaluation provided reliable results with an easy method and standardized luminance.

It should be noted that there are several limitations of this study. The numbers of patients with eye disease and healthy controls were small. Reliability was evaluated by only two sessions with CS test each, so more sessions are required to strengthen the results. In addition, the interval between sessions varied from one week to one month.

CS measures visual quality and can provide detailed information by spatial frequency assessment. Thus, it can be used to detect early subclinical findings or changes during treatment for visual-system-related eye diseases. However, to differentiate healthy from abnormal characteristics and to detect real disease progression, the CS test must have strong clinical reliability. In this study, we found good reliability of the FVA closed system CS test in all groups and reliability improved during the second session. Therefore, to get more reliable results, repeated measurements are needed. Although, there was an acceptable floor-effect with the closed system test in patients with glaucoma and AMD, we found no ceiling effect; consequently, the view-in test has potential to detect fine changes at high frequencies.

In summary, to the best of our knowledge, this is the first study reporting the repeatability of FACT test scores using a closed system in patients with AMD and glaucoma. Hopefully, the reliability of this FACT will encourage clinicians to incorporate contrast sensitivity into their thorough evaluation of the early changes that accompany glaucoma, cataract, and AMD.

  • Funding: No specific financial support was available for this study.
  • Approved by the following research ethics committee: Canakkale Onsekiz Mart University (#050. 99-151).

REFERENCES

  • 1
    Miller D, Schor P, Magnante P. Optics of the normal eye. In: Yanoff M, Duker JS, editor. Ophthalmology. China: Elsevier; 2014. p. 38-45.
  • 2
    Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.
  • 3
    Hong YT, Kim SW, Kim EK, Kim T. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010; 36(4):547-52.
  • 4
    Nielsen E, Hjortdal J. Visual acuity and contrast sensitivity after posterior lamellar keratoplasty. Acta Ophthalmol. 2012;90(8):756-60.
  • 5
    Ginsburg AP. Contrast sensitivity: determining the visual quality and function of cataract, intraocular lenses and refractive surgery. Curr Opin Ophthalmol. 2006;17(1):19-26.
  • 6
    Aslam T, Mahmood S, Balaskas K, Patton N, Tanawade R, Tan S, et al. Repeatability of visual function measures in age-related macular degeneration. Graefes Arch Clin Exp Ophthalmol. 2014;252(2):201-6.
  • 7
    Onal S, Yenice O, Cakir S, Temel A. FACT contrast sensitivity as a diagnostic tool in glaucoma: FACT contrast sensitivity in glaucoma. Int Ophthalmol. 2008;28(6):407-12.
  • 8
    Zhang J, Zhou YH, Li R, Tian L. Visual performance after conventional LASIK and wavefront-guided LASIK with iris-registration: results at 1 year. Int J Ophthalmol. 2013; 6(4):498-504.
  • 9
    Ginsburg AP. Contrast sensitivity and functional vision. Int Ophthalmol Clin. 2003; 43(2):5-15.
  • 10
    Lahav K, Levkovitch-Verbin H, Belkin M, Glovinsky Y, Polat U. Reduced mesopic and photopic foveal contrast sensitivity in glaucoma. Arch Ophthalmol. 2011;129(1):16-22.
  • 11
    Yenice O, Onal S, Incili B, Temel A, Afşar N, Tanrıdag. Assessment of spatial-contrast function and short-wavelength sensitivity deficits in patients with migraine. Eye (Lond). 2007;21(2):218-23.
  • 12
    Richman J, Lorenzana LL, Lankaranian D, Dugar J, Mayer J, Wizov SS, et al. Importance of visual acuity and contrast sensitivity in patients with glaucoma. Arch Ophthalmol. 2010;128(12):1576-82.
  • 13
    Hawkins AS, Szlyk JP, Ardickas Z, Alexander KR, Wilensky JT. Comparison of contrast sensitivity, visual acuity, and Humphrey visual field testing in patients with glaucoma. J Glaucoma. 2003;12(2):134-8.
  • 14
    Faria BM, Duman F, Zheng CX, Waisbourd M, Gupta L, Ali M, et al. Evaluating contrast sensitivity in Age-related macular degeneration using a novel computer-based test, the Spaeth/Richman contrast sensitivity test. Retina. 2015;35(7):1465-73.
  • 15
    Stangos N, Voutas S, Topouzis F, Karampatakis V. Contrast sensitivity evaluation in eyes predisposed to age-related macular degeneration and presenting normal visual acuity. Ophthalmologica. 1995;209(4):194-8.
  • 16
    Puell MC, Barrio AR, Palomo-Alvarez C, Gómez-Sanz FJ, Clement-Corral A, Pérez-Carrasco MJ. Impaired mesopic visual acuity in eyes with early age-related macular degeneration. Invest Ophthalmol Vis Sci. 2012;53(11):7310-4.
  • 17
    Klein J, Pierscionek BK, Lauritzen J, Derntl K, Grzybowski A, Zlatkova MB. The effect of cataract on early stage glaucoma detection using spatial and temporal contrast sensitivity tests. PLoS One. 2015;10(6):e0128681.
  • 18
    Bansback N, Czoski-Murray C, Carlton J, Lewis G, Hughes L, Espallargues M, et al. Determinants of health related quality of life and health state utility in patients with age related macular degeneration: the association of contrast sensitivity and visual acuity. Qual Life Res. 2007;16(3):533-43.
  • 19
    Söğütlü Sari E, Kubaloğlu A, Unal M, Liorens DP, Koytak A,Ofluoglu AN, et al. Penetrating keratoplasty versus deep anterior lamellar keratoplasty: comparison of optical and visual quality outcomes. Br J Ophthalmol. 2012;96(8):1063-7.
  • 20
    Karakus SH, Basarir B, Pinarci EY, Kirandi EU, Demirok A. Long-term results of half-dose photodynamic therapy for chronic central serous chorioretinopathy with contrast sensitivity changes. Eye (Lond). 2013;27(5):612-20.
  • 21
    Gertnere J, Solomatin I, Sekundo W. Refractive lenticule extraction (ReLEx flex) and wavefront-optimized Femto-LASIK: comparison of contrast sensitivity and high-order aberrations at 1 year. Graefes Arch Clin Exp Ophthalmol. 2013;251(5):1437-42.
  • 22
    Thayaparan K, Crossland MD, Michael D, Rubin GS. Clinical assessment of two new contrast sensitivity charts. Br J Ophthalmol. 2007;91(6):749-52.
  • 23
    Keane PA, Patel PJ, Ouyang Y, Chen FK, Ikeji F, Walsh AC, et al. Effects of retinal morphology on contrast sensitivity and reading ability in neovascular age-related macular degeneration. Invest Ophthalmol Vis Sci. 2010;51(11):5431-7.
  • 24
    Haughom B, Strand T-E. Sine wave mesopic contrast sensitivity - defining the normal range in a young population. Acta Ophthalmol. 2013;91(2):176-82.
  • 25
    Li JH, Feng YF, Zhao YE, Zhao YY, Lin L. Contrast visual acuity after multifocal intraocular lens implantation: aspheric versus spherical design. Int J Ophthalmol. 2014; 7(1):100-3.
  • 26
    Haymes SA, Roberts KF, Cruess AF, Cruess AF, Nicolela MT, LeBlanc RP, et al. The letter contrast sensitivity test: clinical evaluation of a new design. Invest Ophthalmol Vis Sci. 2006;47(6):2739-45.
  • 27
    Hohberger B, Laemmer R, Adler W, Juenemann AG, Horn FK. Measuring contrast sensitivity in normal subjects with OPTEC 6500: influence of age and glare. Graefes Arch Clin Exp Ophthalmol. 2007;245(12):1805-14.
  • 28
    Chylack LT Jr, Wolfe JK, Singer DM, Leske MC, Bullimore MA, Bailey IL, et al. The Lens Opacities Classification System III. The Longitudinal Study of Cataract Study Group. Arch Ophthalmol. 1993;111(6):831-6.
  • 29
    Shaffer RN. Primary glaucomas. Gonioscopy, ophthalmoscopy and perimetry. Trans Am Acad Ophthalmol Otolaryngol. 1960;64:112-27.
  • 30
    Armstrong RA, Davies LN, Dunne MC, Gilmartin B. Statistical guidelines for clinical studies of human vision. Ophthalmic Physiol Opt. 2011;31(2):123-36.
  • 31
    Bühren J, Terzi E, Bach M, Wesemann W, Kohnen T. Measuring contrast sensitivity under different lighting conditions: comparison of three tests. Optom Vis Sci. 2006;83(5): 290-8.
  • 32
    Lundh BL. Central and peripheral contrast sensitivity for static and dynamic sinusoidal gratings in glaucoma. Acta Ophthalmol. 1985;63(5):487-92.
  • 33
    McKendrick AM, Sampson GP, Walland MJ, Badcock DR. Contrast sensitivity changes due to glaucoma and normal aging: low-spatial-frequency losses in both magnocellular and parvocellular pathways. Invest Ophthalmol Vis Sci. 2007;48(5):2115-22.
  • 34
    Ansari EA, Morgan JE, Snowden RJ. Psychophysical characterisation of early functional loss in glaucoma and ocular hypertension. Br J Ophthalmol. 2002;86(10):1131-5.
  • 35
    Sun H, Swanson WH, Arvidson B, Dul MW. Assessment of contrast gain signature in inferred magnocellular and parvocellular pathways in patients with glaucoma. Vision Res. 2008;48(26):2633-41.
  • 36
    Patel PJ, Chen FK, Rubin GS, Tufail A. Intersession repeatability of contrast sensitivity scores in age-related macular degeneration. Invest Ophthalmol Vis Sci. 2009;50(6): 2621-5.
  • 37
    Scilley K, Jackson GR, Cideciyan AV, Maguire MG, Jacobson SG, Owsley C. Early age-related maculopathy and self-reported visual difficulty in daily life. Ophthalmology. 2002;109(7):1235-42.
  • 38
    Browne C, Brazier J, Carlton J, Alavi Y, Jofre-Bonet M. Estimating quality-adjusted life years from patient-reported visual functioning. Eye (Lond). 2012;26(10):1295-301.
  • 39
    Wei H, Sawchyn AK, Myers JS, Katz LJ, Moster MR, Wizov SS, Steele M, Lo D, Spaeth GL. A clinical method to assess the effect of visual loss on the ability to perform activities of daily living. Br J Ophthalmol. 2012;96(5):735-41.
  • 40
    Rubin GS, Bandeen-Roche K, Huang GH, Muñoz B, Schein OD, Fried LP, West SK. The association of multiple visual impairments with self-reported visual disability: SEE project. Invest Ophthalmol Vis Sci. 2001;42(1):64-72.

Publication Dates

  • Publication in this collection
    Sep-Oct 2016

History

  • Received
    21 May 2015
  • Accepted
    20 May 2016
Conselho Brasileiro de Oftalmologia Rua Casa do Ator, 1117 - cj.21, 04546-004 São Paulo SP Brazil, Tel: 55 11 - 3266-4000, Fax: 55 11- 3171-0953 - São Paulo - SP - Brazil
E-mail: abo@cbo.com.br