Comment on “Measuring optic nerve sheath diameter using ultrasonography in patients with idiopathic intracranial hypertension”

Comentário a “Medición del diámetro de la vaina del nervio óptico mediante ultrasonografía en pacientes con hipertensión intracraneal idiopática”

Dear Editor,

We are writing to comment on Dağdelen et al.'s study on the measurement of the optic nerve sheath diameter (ONSD) in patients with idiopathic intracranial hypertension.11 Dağdelen K, EkiciM. Measuring optic nerve sheath diameter using ultrasonography in patients with idiopathic intracranial hypertension. Arq Neuropsiquiatr 2022;80(06):580–585

Ocular ultrasonography is considered to be the least invasive diagnostic method to detect intracranial hypertension, even in an emergency setting.

In their paper, the authors utilized the B-scan technique, with the probe placed in the superolateral of the globe, with the upper eyelid closed. The B-scan technique is very sensitive in detecting small optic nerve calcifications, such as in cases of optic nerve drusen,22 Rosa N, De Bernardo M, Abbinante G, Vecchio G, Cione F, Capasso L. Optic Nerve Drusen Evaluation: A Comparison between Ultrasound and OCT. J Clin Med 2022;11(13):3715 but it is not so reliable for measurements because the optic nerve sheath's appearance is influenced by the B scan gain: with decreased gain, the ONSD appears larger compared to the one acquired with an increased gain (blooming effect).33 Tenuta M, De Bernardo M, Rosa N. Neuromuscular ultrasonography of cranial nerves. J Clin Neurol 2017;13(02):212–213 44 Iaconetta G, De Bernardo M, Rosa N. Coronal Axis Measurement of the Optic Nerve Sheath Diameter. J Ultrasound Med 2017;36(05): 1073

Other problems are related to the performance of the examination through closed eyelids, which makes impossible the recognition of gaze position and leads to Bell's phenomenon, with an upward and outward movement of the eye when an attempt to close the eyes is made. The consequence is the acquisition of images of optic nerve in different positions and measurements of sheath diameter stretched differently, thus not comparable.

Moreover, with the eyelids closed, there is a decrease of the image quality score due to the sound attenuation produced by the lids, making the results even more unpredictable.

For this reason, we suggest performing ocular ultrasonography with open eyelids, using methylcellulose and anesthetic drops.

However, the best way to perform the examination is with a standardized A scan. It permits the unequivocal measurement of the ONSD with its hyperreflective spikes from the interface between the arachnoid and subarachnoid fluids, thus making the blooming effect harmless.55 Rosa N, De Bernardo M, Di Stasi M, Cione F, Capaldo I. A-Scan Ultrasonographic Evaluation of Patients with Idiopathic Intracranial Hypertension: Comparison of Optic Nerves. J Clin Med 2022; 11(20):6153

There are a few other points that, in our opinion, deserve to be discussed.

The case group is meaningfully heterogeneous: in fact, some of the patients (57.4%) were treated with acetazolamide, while the remaining (42.6%) were not.

The authors found a negative correlation between ONSD and age, and a positive correlation between ONSD and body mass index (BMI).

Unfortunately, both case and control groups are significantly different in age and BMI, making the comparison ineffective.

We are afraid that all these problems could make the results of this study non significative.

References

  • 1
    Dağdelen K, EkiciM. Measuring optic nerve sheath diameter using ultrasonography in patients with idiopathic intracranial hypertension. Arq Neuropsiquiatr 2022;80(06):580–585
  • 2
    Rosa N, De Bernardo M, Abbinante G, Vecchio G, Cione F, Capasso L. Optic Nerve Drusen Evaluation: A Comparison between Ultrasound and OCT. J Clin Med 2022;11(13):3715
  • 3
    Tenuta M, De Bernardo M, Rosa N. Neuromuscular ultrasonography of cranial nerves. J Clin Neurol 2017;13(02):212–213
  • 4
    Iaconetta G, De Bernardo M, Rosa N. Coronal Axis Measurement of the Optic Nerve Sheath Diameter. J Ultrasound Med 2017;36(05): 1073
  • 5
    Rosa N, De Bernardo M, Di Stasi M, Cione F, Capaldo I. A-Scan Ultrasonographic Evaluation of Patients with Idiopathic Intracranial Hypertension: Comparison of Optic Nerves. J Clin Med 2022; 11(20):6153

Publication Dates

  • Publication in this collection
    17 July 2023
  • Date of issue
    June 2023

History

  • Received
    30 Dec 2022
  • Accepted
    20 Feb 2023
Academia Brasileira de Neurologia - ABNEURO R. Vergueiro, 1353 sl.1404 - Ed. Top Towers Offices Torre Norte, 04101-000 São Paulo SP Brazil, Tel.: +55 11 5084-9463 | +55 11 5083-3876 - São Paulo - SP - Brazil
E-mail: revista.arquivos@abneuro.org