Assessing peri-implant bone microarchitecture: conventional vs. osseodensification drilling - ex vivo analysis

Abstract:

The aim was to evaluate primary implant stability and bone microarchitecture in two drilling situations, by comparing the conventional technique (CT) and osseodensification (OD) (Versah Burs - Jackson - Mississippi - USA). The implant insertion torque (IT), implant stability quotient (ISQ), and the peri-implant trabecular microstructure were assessed on bone fragments obtained from pig’s tibia (n=12), divided between CT (n=6) and OD (n=6). After the drilling procedure, the implants were installed (3.5x8.5 mm, Epikut - SIN - São Paulo - Brazil). The IT and ISQ were measured using a digital torque wrench and resonance frequency analysis. Then, the bone fragments containing the implants were removed with a trephine and analyzed by Microtomography (µCT, 8.0 µm). The comparison between groups was performed using the unpaired t-test (α=0.05). The results revealed that OD promotes a higher insertion torque (CT: 7.67±2.44 Ncm; OD: 19.78±5.26 Ncm) (p=0.0005), although ISQ was not different (CT: 61.33±4.66; OD: 63.25±4.58) (p=0.48). There was a significant increase in peri-implant bone volume (CT: 23.17±3.39 mm3; OD: 32.01±5.75 mm3) (p=0.008), and trabecular parameters: separation (CT: 0.4357±0.03 mm; OD: 0.3865±0.04 mm) (p=0.0449), number (CT: 1.626±0.18 1/mm; OD: 1.946±0.13 1/mm) (p=0.007), and thickness (CT: 0.1130±0.009 mm; OD: 0.1328±0.015 mm) (p=0.02). Structure model index (SMI) data demonstrate no significant differences between groups (CT: 1.7±0.2; OD: 1.4±0.4) (p=0.12). In conclusion, OD increases the insertion torque values and promotes beneficial changes regarding bone microarchitecture compared with CT, revealing more peri-implant bone volume with consequent higher primary stability.

Key Words:
Osteotomy; X-Ray Microtomography; Dental Implants; Bone and Bones

Fundação Odontológica de Ribeirão Preto Av. do Café, S/N, 14040-904 Ribeirão Preto SP Brasil, Tel.: (55 16) 3602-3982, Fax: (55 16) 3633-0999 - Ribeirão Preto - SP - Brazil
E-mail: bdj@forp.usp.br