Acessibilidade / Reportar erro

Comparison: real and simulated ear insertion gain

The development of hearing aid (HA) software programming does not replace the analysis of real ears with probe microphones. AIM: To compare simulated insertion gain in HA software programming and real ear insertion gain. METHOD: A prospective study of 62 patients (aged from 29 to 93 years; 30 male and 32 female). All patients presented unilateral (n=14) or bilateral (n=48) and mild to profound sensorineural hearing impairment. 110 ears assessed. Data was gathered from medical records and the insertion gain was obtained in real ears for comparison with the simulated insertion gain in HA software programming. Statistical tests were applied to analyze the correlation of data - difference of real ear and simulated insertion gain. RESULTS: HA software programming simulated insertion gain was higher than real ear insertion gain obtained with probe microphone measures. There were statistically significant frequency differences. Age did not correlate with the difference of real ear and simulated measures. CONCLUSION: The use of real ear measures is important during verification of HA.

hearing aids; hearing loss; simulation


Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico-Facial. Sede da Associação Brasileira de Otorrinolaringologia e Cirurgia Cérvico Facial, Av. Indianópolia, 1287, 04063-002 São Paulo/SP Brasil, Tel.: (0xx11) 5053-7500, Fax: (0xx11) 5053-7512 - São Paulo - SP - Brazil
E-mail: revista@aborlccf.org.br