Abstract

This ex vivo study evaluated the quality of digital radiographic images obtained with the photostimulable phosphor plate system (Digora) according to the processing delay and maintenance of optical plates in either opaque (supplied with the system) or transparent protective plastic cases during this period. Five radiographs were obtained from the mandibular molar region of a dry human mandible using optical plates. These plates were placed in the protective plastic cases before obtaining the radiographs and were processed immediately or after processing delays of 5, 60 and 120 min, when the case was removed. The results revealed a reduction in image quality when processing was delay 120 min compared to the other times. The opaque case provided better protection to the sensor than the transparent case. In conclusion, a 120-min processing delay for the Digora system caused a reduction in image quality, yet without interfering with the quality of diagnosis. The opaque case supplied by the system's manufacturer provided better protection to the optical plate than the transparent case.

Radiography; Digital images; Phosphor storage plates; Digora


ORIGINAL ARTICLES

Evaluation of the effects of processing delays and protective plastic cases on image quality of a photostimulable phosphor plate system

Clóvis Monteiro BramanteI; Alexandre Silva BramanteII; Rogério Emílio de SouzaIII; Ivaldo Gomes MoraesIV; Norberti BernardineliI; Roberto Brandão GarciaV

IDDS, MSc, PhD, Professor, Department of Operative Dentistry, Endodontics and Dental Materials, Bauru School of Dentistry, University of São Paulo, Bauru, SP, Brazil

IIDDS, MSc, Brasiliense School of Dentistry, Brasília, DF, Brazil

IIIDDS, MSc, PhD, Dental School, FACIMP, Imperatriz, MA, Brazil

IVDDS, MSc, PhD, Associate Professor, Department of Operative Dentistry, Endodontics and Dental Materials, Bauru School of Dentistry, University of São Paulo, Bauru, SP, Brazil

VDDS, MSc, PhD, Assistant Professor, Department of Operative Dentistry, Endodontics and Dental Materials, Bauru School of Dentistry, University of São Paulo, Bauru, SP, Brazil

Corresponding address Corresponding address: Prof. Dr. Clovis Monteiro Bramante Faculdade de Odontologia de Bauru - USP Alameda Octávio Pinheiro Brisolla, Nº 9-75 17012-901 Bauru, SP, Brasil Phone +55-14-3235-8344 email: bramante@fob.usp.br

ABSTRACT

This ex vivo study evaluated the quality of digital radiographic images obtained with the photostimulable phosphor plate system (Digora) according to the processing delay and maintenance of optical plates in either opaque (supplied with the system) or transparent protective plastic cases during this period. Five radiographs were obtained from the mandibular molar region of a dry human mandible using optical plates. These plates were placed in the protective plastic cases before obtaining the radiographs and were processed immediately or after processing delays of 5, 60 and 120 min, when the case was removed. The results revealed a reduction in image quality when processing was delay 120 min compared to the other times. The opaque case provided better protection to the sensor than the transparent case. In conclusion, a 120-min processing delay for the Digora system caused a reduction in image quality, yet without interfering with the quality of diagnosis. The opaque case supplied by the system's manufacturer provided better protection to the optical plate than the transparent case.

Key words: Radiography. Digital images. Phosphor storage plates. Digora.

INTRODUCTION

Digital radiograph or radiovisiography was first described in 1987 and was commercially introduced by Trophy Radiologie in collaboration with Mouyen12 in 1993. Three systems are employed for radiovisiography: the CCD sensor (Charge-Coupled Device), CMOS sensor (Complementary Metallic Oxide Sensor), and PSPs (PSPs - Phosphor Storage Plate - optical plate); the latter includes the Digora, Digident, Denoptix and Den Ortix systems11. Some studies have found better outcomes of digitized images for endodontic diagnosis and treatment5-9,13,15-18.

However, some doubts remain in the use of radiovisiography based on the optical plate system as to the durability of the plate3, image quality according to the processing delay1,2,14, and storage conditions10. No study has yet investigated the influence of utilization of opaque cases on the quality of digital image.

This ex vivo study evaluated the quality of digital radiographic images obtained with the Digora system according to the processing delay and maintenance of optical plates in either opaque (supplied with the system) or transparent protective plastic cases during this period.

MATERIAL AND METHODS

The study was conducted on a dry human mandible fixed on an acrylic resin base. A plastic tube was fixed frontally to the mandible to standardize the position of the X-ray source, thus standardizing the vertical and horizontal angles (Figure 1). A groove was prepared on this resin base to standardize the position of the optical plate.


Radiographs were obtained from the mandibular molar region by utilization of optical plates with an x-ray unit (Gnatus model XR 6010; Gnatus, Ribeirão Preto, SP, Brazil) set at 60 kVp and 10 mA, with open tube, total filtration of 1.5 mm Al, exposure time of 0.3 s, and optical plate of the Digora system (Soredex, Orion Corporation, Finland). Two groups were constituted for x-ray exposure: in group A, the optical plate was inserted and sealed in the opaque plastic case supplied by the manufacturer and kept as such until processing; in group B, the optical plate was stored in a transparent plastic case (Figure 2). Five optical plates were exposed for each period of processing delay.


The optical plates were processed in the Digora scanner immediately and 5, 60 and 120 min after X-ray exposure, and were transferred to the computer for analysis of digitized images. Analysis was performed directly on the computer screen by three examiners using the Digora for Windows software. The images were analyzed as to brightness, contrast and resolution of tooth structures (enamel, dentin, root and pulp cavity) and periapical structures (periodontal space, cortical bone, alveolar bone), which received scores 0 to 2, as follows: 0- Poor image quality; 1- Good image quality; 2- Excellent image quality.

Data were submitted to the Kruskal-Wallis nonparametric test, and comparison of results between opaque and transparent cases were submitted by the Mann-Whitney test. Significance level was set at 5%.

RESULTS

Table 1 presents the scores assigned by examiners to the images, according to processing delays and utilization of opaque or transparent plastic cases.

There was statistically significant difference (p<0.05) regarding the processing delays, namely immediate, 5 and 60 min with utilization of opaque case (Table 2). This difference was observed between the 120-min delay and the other periods (Table 3). No statistically significant difference (p>0.05) was observed among processing delays with utilization of transparent case (Table 4). Comparison of results between opaque and transparent cases did not reveal significant differences (p>0.05) (Table 5).

DISCUSSION

Digital radiography was introduced to dental practice with a view to replace the conventional radiography due to the shorter exposure to radiation, achievement of high-quality images, possibility of adjustments with aid of software, and comparable reliability to conventional radiography, for both endodontic diagnosis and treatment3,5-7,19.

The main disadvantage of the optical plate system is the need of additional time for processing and possible loss of quality when the sensitized optical plate is exposed to light. Some authors have emphasized that, after exposure to x-ray, the optical plates should be processed within 1 h in order to avoid loss of image quality1,2. Other authors believe that the plates can be processed within 6 h if stored in appropriate cases10.

In the present study, a processing delay of 120 min caused a mild reduction in image quality, yet without compromising the quality of diagnosis (Table 1). The three examiners were able to interpret radiographic details in all study periods. It should be highlighted that the images were not altered as to the brightness and contrast provided by the system, being analyzed as produced on the computer screen.

Together with the Digora system, the manufacturer supplies opaque plastic cases that offer protection against contamination and the deleterious effect of light. Images of optical plates maintained in the original opaque cases after exposure to x-ray maintained their image quality up to 120 min (Table 2), with significant difference between the 120min processing delay and the other periods (Table 3). When the optical plates were maintained in transparent cases, there was also a progressive reduction of image quality (Table 1), which was more accentuated at 120 min (Table 4), though without statistical difference among periods. This is due to the fact that the optical plate protected with transparent cases is subjected to the continuous action of light, with progressive loss of image quality. However, images obtained with optical plates protected with both opaque and transparent cases and with processing delays of up to 120 min provided good quality for diagnosis.

CONCLUSIONS

The findings of the present study revealed that the processing delay of 120 min for the Digora system caused a reduction in image quality, yet without interfering with the quality of diagnosis. The opaque case supplied by the system's manufacturer provided better protection to the optical plate than the transparent case.

Received : October 03, 2007

Modification : January 28, 2008

Accepted : May 07, 2008

References

  • 1- Akdeniz BG, Grondahl HG, Kose T. Effect of delayed scannig of storage phosphor plates. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2005;99:603-7.
  • 2- Akdeniz BC, Grondahl HG. Degradation of storage phosphor images due to scanning delay. Dentomaxillofacial Radiol. 2006;35:74-7
  • 3- Bedard A, Davis TD, HG. Angelopoulos C. How durable are they as a digital dental radiographic system? J Contemp Dent Pract. 2004;5:57-69.
  • 4- Bramante AS, Bramante CM, Bernardineli N, Moraes I, Garcia RB. Diagnóstico de defeitos ósseos por meio da radiografia convencional, digital e tomografia helicoidal. Rev Port Estom Medicina Dentária Cir Maxilofac. 2007;48:15-21.
  • 5- Brettle DS, Workman A, Ellwood RP, Launders JH, Horner K, Davies RM. The imaging performance of a storage phosphor system for dental radiography. Brit J Radiol. 1996;69:256-61.
  • 6- Hedrick RT, Dove SB, Peters DD, McDavid WD. Radiographic determinations of canal length: direct digital radiography versus conventional radiography. J Endod. 1994;20:320-6.
  • 7- Holtzmann DJ, Johnson WT, Southard TE, Khademi JA, Chang PJ, et al. Storage-phosphor computed radiography versus film radiography in the detection of pathologic periradicular bone loss in cadavers. Oral Surg Oral Med Oral Pathol. 1998;86:90-7.
  • 8-Huda W, Rill LN, Benn DK, Pettigrew JC. Comparison of a photostimulable phosphor system with film for dental radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1997;83:725-31.
  • 9- Kaeppler G, Vogel A, Axmann-Krcmar D. Intra-oral storage phosphor and conventional radiography in the assessment of alveolar bone structures. Dentomaxillofacial radiol. 2000;29:362-7.
  • 10- Martins MGMQ, Haiter F Neto, Whaites EJ. Analysis of digital images acquired using different phosphor storage plates (PSPs) subjected to varying reading times and storage conditions. Dentomaxilofacial Rad. 2003;32:186-90.
  • 11- Miles DA, The deal on digital: the status of radiographic imaging. Compedium 2001;22:1057-64.
  • 12- Mouyen F, Benz C, Sonnabend E, Lodter JP. Presentation and physical evaluation of radiovisiography. Oral Surg Oral Med Oral Pathol. 1989;68:238-42.
  • 13- Nair MK, Nair UP. Digital and advanced imaging in endodontics: a review. J Endod. 2007;33:1-6.
  • 14- Oliveira AE, Almeida SM, Paganini GA, Haiter F Neto, Bóscolo FN. Comparative study of two digital radiographic storage phosphor systems. Braz Dent J. 2000;11:11-6.
  • 15- Parissis N, Kondylidou-Sidira A, Tsirlis A, Patias P. Conventional radiographs vs digitized radiographs: image quality assessment. Dentomaxillofacial Radiol. 2005;34:353-6.
  • 16- Shearer AC, Mullane E, Macfarlane TV, Grondahl HG, Horner K. Three phosphor plate systems and film compared for imaging root canals. Int Endod J. 2001;34:275-9.
  • 17- Yoshiura K, Kawazu T, Chikui T, Tatsumi M, Tokumori K, Tanaka T, et al. Assessment of image quality in dental radiography, part 2: optimum exposure conditions for detection of small mass changes in 6 intraoral radiography systems. Oral Surg. Oral Med. Oral Pathol. Oral Radiol Endod. 1999;87:123-9.
  • 18- Westphalen VPD, Moraes IG, Westphalen FH, Martins WD, Souza PHC. Conventional and digital radiographic methods in the detection of simulated external root resorptions: a comparative study. Dentomaxillofacial Radiol. 2004;33:233-5.
  • Corresponding address:
    Prof. Dr. Clovis Monteiro Bramante
    Faculdade de Odontologia de Bauru - USP
    Alameda Octávio Pinheiro Brisolla, Nº 9-75
    17012-901 Bauru, SP, Brasil
    Phone +55-14-3235-8344
    email:
  • Publication Dates

    • Publication in this collection
      10 Oct 2008
    • Date of issue
      Oct 2008

    History

    • Received
      03 Oct 2007
    • Accepted
      07 May 2008
    location_on
    Faculdade De Odontologia De Bauru - USP Serviço de Biblioteca e Documentação FOB-USP, Alameda Dr. Octávio Pinheiro Brisolla 9-75, 17012-901 Bauru SP Brasil, Tel.: +55 14 3235-8373 - Bauru - SP - Brazil
    E-mail: jaos@usp.br
    rss_feed Acompanhe os números deste periódico no seu leitor de RSS
    Acessibilidade / Reportar erro