Acessibilidade / Reportar erro

Simple flow injection amperometric system for simultaneous determination of dipyrone and paracetamol in pharmaceutical formulations

In this work a simple flow injection methodology with pulsed amperometric detection for simultaneous determination of dipyrone (DI) and paracetamol (PCT) in pharmaceutical formulations is described. The compounds are detected by applying four sequential potential pulses as function of the time. Dipyrone is directly detected at +0.40 V and paracetamol indirectly at 0.00 V through the reduction of the oxidation product (N-acetyl-p-benzoquinoneimine) electrochemically generated at +0.65 V. The fourth potential pulse (-0.05 V) is applied for the cleaning of the electrode surface (glassy carbon). The linear response range was optimized between 9 and 45 mg L-1 for dipyrone and 6 and 30 mg L-1 for paracetamol. Linear regression of these two series of experiments leads to excellent correlation coefficients (R = 0.999) for both analytes and recoveries of around 100%. The proposed methodology allows an analytical frequency of 45 injections h-1 with a consumption of 1.5 mL of solution per minute. The relative standard deviation for 24 successive injections of solutions containing DI and PCT was calculated as 0.4% and 0.2%, respectively. The developed methodology was successfully used for the determination of dipyrone and paracetamol in pharmaceutical samples.

simultaneous determination; dipyrone; paracetamol; flow injection analysis (FIA); multiple pulse amperometric detection


Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br