Acessibilidade / Reportar erro

Aquios CL flow cytometer performance in the automated quantification of lymphocyte subpopulations

ABSTRACT

Flow cytometry (FC) is an essential tool for diagnosis, prognosis and therapeutic follow-up of several hematologic malignancies. In addition, it performs the quantification of lymphocytes subpopulations for diagnosis and monitoring of primary and acquired immunodeficiencies through the antigenic expressions of CD19 and CD20 for B lymphocytes; CD2, CD3, CD4, CD8 for T lymphocytes; and CD56 and CD16 for the identification of natural killer (NK) cells. The cytometry technique has revolutionized the way that the cells are identified, and over the years this platform has progressed with several advances in hardware and software that aim to improve workflow resulting in higher productivity, quality and cost savings. The Aquios CL - Beckman Coulter (BC) is an example of this advance because it is a complete automation instrument in flow cytometry called “Load & Go flow cytometer” for quantification of lymphocyte subpopulations in the routine diagnosis. In this study, the Aquios CL was validated, and quantification in frequency and absolute numbers of the lymphocyte subpopulations had an excellent correlation with the results obtained by the dual platform quantification performed in the Cytomics FC500 (BC) and automated Sysmex XE-2100 cell analyzer.

Key words:
Aquios CL; flow cytometry; subpopulation of lymphocytes; CD4/CD8

Sociedade Brasileira de Patologia Clínica, Rua Dois de Dezembro,78/909 - Catete, CEP: 22220-040v - Rio de Janeiro - RJ, Tel.: +55 21 - 3077-1400 / 3077-1408, Fax.: +55 21 - 2205-3386 - Rio de Janeiro - RJ - Brazil
E-mail: jbpml@sbpc.org.br