A simple cell for assembling commercial ion-selective electrodes in continuous flow analyzers

A simple flow cell for potentiometric detection is described. It was assembled by making use of two perspex pieces fixed together by means of four screws, and allow the connection of plane membrane conventional electrodes to flow system. Details about its construction are presented. The device performance was evaluated by making use of a cyanide ion-selective electrode. The relative standard deviation was about 0.5% with a detection limit of 8.0 x 10-6 mol CN- dm-3. Under experimental conditions, the linear range was 10-5 to10-2 mol dm-3.

flow cell; potentiometric detector; continuous flow analysis