Phosphorescence lifetime in polymer relaxation studies: phase resolution method

We describe in this work a simple experimental set up to perform time dependent luminescence experiments in time scales from mili-seconds to seconds, based on the phase resolution of the emission signal. This system is composed by modulation of a continuous light source with an external chopper controlled by a lock-in amplifier. We exemplified the utility of the system in studies of phosphorecence emission using benzophenone dissolved in polystyrene and since the phosphorescence intensity and lifetime are temperature-dependent processes, we also studied polymer relaxation processes in the temperature range from 20 to 400K. A software that drives the spectrofluorimeter and controls both the cryosystem and the lock-in amplifier was developed.

phase resolution lifetime determinations; benzophenone; polymer relaxation processes


Sociedade Brasileira de Química Instituto de Química, Universidade Estadual de Campinas (Unicamp), CP6154, 13083-0970 - Campinas - SP - Brazil