Acessibilidade / Reportar erro

FAST SIMULTANEOUS DETERMINATION OF PROPIFENAZONE, PARACETAMOL AND CAFFEINE USING BATCH INJECTION ANALYSIS WITH AMPEROMETRIC DETECTION

A fast and simple procedure for simultaneous determination of propyphenazone (PRO), paracetamol (PAR) and caffeine (CAF) was performed using batch injection analysis with multiple pulse amperometric detection (BIA-MPA). Three potentials pulses were selected for the analysis: +0.90 V/50 ms: only PAR was oxidized; +1.20 V/50 ms: oxidation of PAR and PRO without the interference of CAF; +1.60 V/50 ms: oxidation of the three compounds (PAR, PRO and CAF). The sample aliquot (150 µL) was directly injected onto the boron-doped diamond (BDD) electrode immersed in a BIA cell. The analytical characteristics of the proposed method include high analytical frequency (75 injections per hour), good stability (RSD < 3.9%; n = 10), and low detection limits (0.3, 0.2 and 0.1 mg L−1 for PRO, PAR and CAF, respectively). The proposed method yielded similar results to those obtained by liquid chromatography at a 95% confidence level.

Keywords:
boron doped diamond electrode (BDDE); multiple-pulse amperometric detection; portable electrochemical method; pharmaceutical


Sociedade Brasileira de Química Secretaria Executiva, Av. Prof. Lineu Prestes, 748 - bloco 3 - Superior, 05508-000 São Paulo SP - Brazil, C.P. 26.037 - 05599-970, Tel.: +55 11 3032.2299, Fax: +55 11 3814.3602 - São Paulo - SP - Brazil
E-mail: quimicanova@sbq.org.br