Simultaneous determination of cations by capillary electrophoresis with contactless conductivity detection performed in a home-made equipment

This work describes the development of a home-made capillary electrophoresis (CE) system based on the capacitively coupled contactless conductivity detection (C4D) for the separation of the metallic species Zn2+, Cr3+, Pb2+, Cd2+, Co2+, Cu2+, Ni2+ e Tl+. A background electrolyte composed of MES/Histidine 0,02 mol L-1 (pH 5.0) was optimized for the separation of the metallic species by using organic solvents and complexing agents as additives. The system allowed the determination of the metallic species using MES/Histidine 0,02 mol L-1 and methanol 5% (pH 5.0) as a background electrolyte, 15 kV separation voltage and hydrodynamic injection by gravity.

capillary electrophoresis; contactless conductivity detection; cations