Some technical aspects should be taken into consideration in order to guarantee the reliability of the assessment of pulmonary emphysema with lung computed tomography densitometry. Changes in lung density associated with variations in lungs inspiratory and expiratory levels, computed tomography slice thickness, reconstruction algorithm and type of computed tomography apparatus make tomographic comparisons more difficult in follow up studies of pulmonary emphysema. Nevertheless, quantitative computed tomography has replaced the visual assessment competing with pulmonary function tests as a sensitive method to measure pulmonary emphysema. The present review discusses technical variables of lung computed tomography and their influence on measurements of pulmonary emphysema.