Comparative study on imaging quality in PET acquisition modes: validation of a protocol for reducing the radiation dose

OBJECTIVE: The present study is aimed at validating a 3D acquisition protocol for positron emission tomography as a replacement for the 2D mode, to reduce the radiation dose delivered to patients, without any loss in the quality of images. MATERIALS AND METHODS: The study comprised 27 simulations in a Discovery ST equipment with four-minute 2D acquisitions, and two-minute and four-minute 3D acquisitions, utilizing a chest phantom according to the National Electrical Manufacturers Association protocol. Six spheres with different diameters were inserted into this phantom as a means for determining the images quality. The images were blindly reviewed by three experienced nuclear physicians who did not know the acquisition modes. Each observer attributed a score 1 when one of the spheres was not identified, or 2 for visible spheres. RESULTS: The qualitative analysis based on generalized kappa coefficient demonstrated that the frequency of visible spheres was higher with four-minute 3D acquisitions (85%), with an also higher interobserver agreement (88.9%), generalized kappa = 0.725 [0.507;0.942]. CONCLUSION: The protocol with four-minute 3D acquisition with lower 18F-FDG activity can be utilized for patients with a biotype similar to the phantom, without any loss in the imaging quality.

PET/CT; Imaging quality; 2D mode; 3D mode; 18F-FDG; Dose reduction


Publicação do Colégio Brasileiro de Radiologia e Diagnóstico por Imagem Av. Paulista, 37 - 7º andar - conjunto 71, 01311-902 - São Paulo - SP, Tel.: +55 11 3372-4541, Fax: 3285-1690, Fax: +55 11 3285-1690 - São Paulo - SP - Brazil
E-mail: radiologiabrasileira@cbr.org.br