Monte Carlo derivation of filtered tungsten anode X-ray spectra for dose computation in digital mammography* * Study developed at Centro de Desenvolvimento da Tecnologia Nuclear / Comissão Nacional de Energia Nuclear (CDTN/CNEN), Belo Horizonte, MG, Brazil.

Derivação Monte Carlo de espectros filtrados de raios X de ânodo de tungstênio para determinação de dose em mamografia digital

Abstract

Objective:

Derive filtered tungsten X-ray spectra used in digital mammography systems by means of Monte Carlo simulations.

Materials and Methods:

Filtered spectra for rhodium filter were obtained for tube potentials between 26 and 32 kV. The half-value layer (HVL) of simulated filtered spectra were compared with those obtained experimentally with a solid state detector Unfors model 8202031-H Xi R/F & MAM Detector Platinum and 8201023-C Xi Base unit Platinum Plus w mAs in a Hologic Selenia Dimensions system using a direct radiography mode.

Results:

Calculated HVL values showed good agreement as compared with those obtained experimentally. The greatest relative difference between the Monte Carlo calculated HVL values and experimental HVL values was 4%.

Conclusion:

The results show that the filtered tungsten anode X-ray spectra and the EGSnrc Monte Carlo code can be used for mean glandular dose determination in mammography.

Keywords:
Mammography; X-ray spectra; HVL; Monte Carlo