Sutureless Aortic Valve Replacement vs. Transcatheter Aortic Valve Implantation in Patients with Small Aortic Annulus: Clinical and Hemodynamic Outcomes from a Multi-Institutional Study

ABSTRACT

Objective:

This study aimed to compare hemodynamic performances and clinical outcomes of patients with small aortic annulus (SAA) who underwent aortic valve replacement by means of sutureless aortic valve replacement (SUAVR) or transcatheter aortic valve implantation (TAVI).

Methods:

From 2015 to 2020, 622 consecutive patients with SAA underwent either SUAVR or TAVI. Through a 1:1 propensity score matching analysis, two homogeneous groups of 146 patients were formed. Primary endpoint: all cause-death at 36 months. Secondary endpoints: incidence of moderate to severe patient-prosthesis mismatch (PPM) and incidence of major adverse cardiovascular and cerebrovascular events (MACCEs)

Results:

All-cause death at three years was higher in the TAVI group (SUAVR 12.2% vs. TAVI 21.0%, P=0.058). Perioperatively, comparable hemodynamic performances were recorded in terms of indexed effective orifice area (SUAVR 1.12 ± 0.23 cm2/m2 vs. TAVI 1.17 ± 0.28 cm2/m2, P=0.265), mean transvalvular gradients (SUAVR 12.9 ± 5.3 mmHg vs. TAVI 12.2 ± 6.2 mmHg, P=0.332), and moderate-to-severe PPM (SUAVR 4.1% vs. TAVI 8.9%, P=0.096). TAVI group showed a higher cumulative incidence of MACCEs at 36 months (SUAVR 18.1% vs. TAVI 32.6%, P<0.001). Pacemaker implantation (PMI) and perivalvular leak ≥ 2 were significantly higher in TAVI group and identified as independent predictors of mortality (PMI: hazard ratio [HR] 3.05, 95% confidence interval [CI] 1.34-6.94, P=0.008; PPM: HR 2.72, 95% CI 1.25-5.94, P=0.012).

Conclusion:

In patients with SAA, SUAVR and TAVI showed comparable hemodynamic performances. Moreover, all-cause death and incidence of MACCEs at follow-up were significantly higher in TAVI group.

Keywords:
Transcatheter Aortic Valve Replacement; Aorta Valve; Hemodynamics; Prostheses and Implants; Propensity Scores