Comparison of refractometric measurements using 2win® Photoscreener and manual retinoscopy in asymptomatic preschoolers

ABSTRACT

Objective:

Evaluate the sensitivity, specificity e accuracy of the ocular refraction measured by the 2WIN® photoscreener as a screening method to identify children in need of spectacles prescription according to the criteria published by the Brazilian Society of Pediatric Ophthalmology (BSPO) in asymptomatic children, 6 to 36 months old, and determine the impact of cycloplegia in the sensitivity, specificity and accuracy of this method.

Methods:

One hundred seventy-eight (178) eyes of asymptomatic children between the ages of 6 and 36 months have been submitted to ocular refraction measurements by the gold-standard method, the manual retinoscopy under cycloplegia, and the method been tested, the 2WIN® photoscreening, both before and under cycloplegia.

Results:

The 2WIN® photoscreener before cycloplegia was able to identify those patients in need of spectacles prescription according to the criteria published by the BSPO with 100% sensitivity, 93.18% specificity and 93.26% accuracy, when compared to the manual retinoscopy under cycloplegia. The 2WIN® photoscreener under cycloplegia maintained a sensitivity of 100%, but increased specificity to 96.59% and accuracy to 96.63%.

Conclusion:

The 2WIN® photoscreener before cycloplegia showed high sensitivity, specificity, and accuracy in detection of patients in need of spectacles prescription according to the criteria published by the BSPO in the tested population, with minor increase in specificity and accuracy when the measurements were performed under cycloplegia.

Keywords:
Refraction, ocular; Retinoscopy; Amblyopia; Screening; Child, preschool