Comparison between the HemoCue® and an automated counter for measuring hemoglobin

Comparação entre o HemoCue® e um contador automático para determinação de hemoglobina

To determine the precision and agreement of the hemoglobin (Hb) measurements in capillary and venous blood samples by the HemoCue® and an automated counter. Hb was determined by both equipaments in blood samples of 29 pregnant women. The HemoCue® showed low repeatability of Hb measurements in duplicate in capillary (CR=0.53 g/dL, CV=13.6%) and venous blood (CR=0.53 g/dL, CV=13.6%). Hb measurements in capillary blood were higher than those in venous blood (12.4 and 11.7 g/dL, respectively; p<0.05). There was high agreement between Hb in capillary blood by the HemoCue® and in venous blood by the counter (r icc=0.86; p<0.01), and also between the diagnosis of anemia by both equipments (k=0.81; p<0.01). The HemoCue® seems to be more appropriate for capillary blood and require training of the measurers.

Anemia; Anemia; Hemoglobin; Hemocue; Pregnancy