A Hybrid Approach of Simultaneous Extraction and Leadless Pacemaker Implantation in a Transvenous Lead Endocarditis Case

Equipment and Supplies/instrumentation; Pacemaker; Artificial/adverse effects; Leadless Pacemaker; Cardiac Pacing; Artificial/methods; Bradycardia/complications; Replantation; Method PISA

Introduction

Symptomatic bradycardia treatment is based on the transvenous method of endocardial pacing.11. Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J. 2021;42(35):3427-520. doi:10.1093/eurheartj/ehab364 However, cardiac electronic implantable devices (CEID) are associated with a potential risk of complications, including infections, with an estimated rate of 0.5% with primary implants and 1–7 % with secondary interventions.11. Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J. 2021;42(35):3427-520. doi:10.1093/eurheartj/ehab364 CIED infections are associated with longer hospital length of stay, higher clinical costs and mortality rates.22. Clémenty N, Carion PL, Léotoing L, Lamarsalle L, Bequet FW, Brown B, et al. Infections and associated costs following cardiovascular implantable electronic device implantations: a nationwide cohort study. Europace. 2018 ;20(12):1974-80. doi:10.1093/europace/eux387 , 33. Tarakji KG, Wazni OM, Harb S, Hsu A, Saliba W, Wilkoff BL. Risk factors for 1-year mortality among patients with cardiac implantable electronic device infection undergoing transvenous lead extraction: the impact of the infection type and the presence of vegetation on survival. Europace. 2014;16(10):1490-5. doi:10.1093/europace/euu147 According to guidelines, infective endocarditis related to CIED implicates complete removal of the system, followed by a period without intravascular therapy.11. Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J. 2021;42(35):3427-520. doi:10.1093/eurheartj/ehab364

2. Clémenty N, Carion PL, Léotoing L, Lamarsalle L, Bequet FW, Brown B, et al. Infections and associated costs following cardiovascular implantable electronic device implantations: a nationwide cohort study. Europace. 2018 ;20(12):1974-80. doi:10.1093/europace/eux387

3. Tarakji KG, Wazni OM, Harb S, Hsu A, Saliba W, Wilkoff BL. Risk factors for 1-year mortality among patients with cardiac implantable electronic device infection undergoing transvenous lead extraction: the impact of the infection type and the presence of vegetation on survival. Europace. 2014;16(10):1490-5. doi:10.1093/europace/euu147
- 44. Blomström-Lundqvist C, Traykov V, Erba PA, Burri H, Nielsen JC, Bongiorni MG, et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections—endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Europace. 2020;22(4):515-49. doi:10.1093/europace/euz246 Nevertheless, most of the patients require CIED reimplantation, which is known to be associated with a risk of reinfection between 2 and 11%, particularly in cases with only partial removal of the initial device.55. Boyle TA, Uslan DZ, Prutkin JM, Greenspon AJ, Baddour LM, Danik SB, et al. Reimplantation and repeat infection after cardiac-implantable electronic device infections: experience from the MEDIC (Multicenter Electrophysiologic Device Infection Cohort) database. Circ Arrhythm Electrophysiol. 2017;10(3):e004822 doi:10.1161/CIRCEP.116.004822

The PISA technique is a percutaneous procedure used with success in lead extraction of the CIED.66. Bongiorni MG, Giannola G, Arena G, Soldati E, Bartoli C, Lapira F, et al. Pacing and implantable cardioverter-defibrillator transvenous lead extraction. Ital Heart J. 2005;6(3):261-6. PMID: 15875518 This technique starts with the identification of the proximal portion of the lead. Then, a debridement is performed along the lead to achieve the site of venous insertion. Subsequently, a polypropylene dilator sheath is inserted and advanced externally up to the lead in rotational movements, while maintaining a light traction. Such movements will cause the release of adhesions around the lead, allowing, after the full advancement of the sheath, the entire lead removal.77. Bongiorni MG. Transvenous lead extraction: from simple traction internal transjugular approach. Milan:Springer; 2014. 176p. ISBN:978884705614 2014:

Clinical case

An 86-year-old female patient, a former smoker with a history of hypertension, hyperlipidemia, diabetes, obesity and hyperuricemia was admitted to the emergency room with atypical chest pain and fever ten months after a DDDR pacemaker implantation due to a third-degree atrioventricular block. The first clinical evaluation detected signs of local infection in the left hand, where the patient had a surgical intervention two months before. Blood tests suggested a systemic infection, despite negative blood cultures. A transthoracic echocardiogram showed normal left ventricular ejection fraction, without any significant valvular disease, and without the identification of vegetations or local complications suggesting endocarditis. Yet, considering the suspicion of endocarditis, a transesophageal echocardiogram was performed two weeks after the initial hospital admission, revealing two masses in the right atrium, attached to the ventricular lead, with maximum dimensions of 21x7 mm. Antibiotic therapy with vancomycin and ceftriaxone was initiated and continued for 35 days, while blood cultures remained sterile. She was then referred to our center for lead extraction using the PISA technique.

Moreover, the patient was totally dependent on the pacemaker rhythm. Therefore, to promote atrioventricular synchronization, we decided to implant the Micra AV (Medtronic Inc., Minneapolis, MN, USA) leadless pacemaker, enabling a VDD pacing mode. Pacemaker extraction and the Micra AV implantation were performed in a simultaneous procedure. An intracardiac leadless pacemaker was implanted initially according to the manufacturer’s training recommendation via the left femoral access. A MicraTMDelivery Catheter (105-cm-long) inserted in the MicraTMintroducer sheath (27-Fr outer diameter) was deflected into the right atrium without difficulty. One operator was responsible to maintain the delivery catheter in the right atrium, while two other operators started the pacemaker extraction. Complete removal of both leads was obtained with the PISA method ( Figure 1 ). At this stage, a suitable deployment of the intracardiac leadless pacemaker allowed a stable position to be reached in the mid-ventricular septum. The entire procedure occurred without complications.

Figure 1
– Images of the procedure, showing the extraction of the transvenous pacemaker and the implantation of the leadless pacemaker, positioned in the mid-ventricular septum (F).

On the next day, Micra AV stable parameters were confirmed and optimized, the femoral access was checked and did not show any complications. The patient received antibiotic therapy for 12 more days. A transesophageal echocardiogram was performed one week after the procedure, without any signs of endocarditis. At the one-month follow-up, the pacemaker parameters were stable, with 100% accurate atrial sensing and ventricular pacing, and the patient remained asymptomatic, without complications.

Discussion

Intracardiac leadless devices are now a safe and effective alternative to transvenous pacemakers, namely in patients with previous device-related infections, venous access issues, previous lead extraction, and comorbidities, including end-stage renal failure and diabetes.88. El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysio. 2019;30(4):569-74. doi:10.1111/jce.13851 , 99. Moltrasio M, Sicuso R, Tundo F, Tondo C. A new leadless pacemaker with atrioventricular synchronous pacing replacing a still working VVI leadless pacemaker: a case report. Eur Heart J Case Rep. 2021;5(8):ytab305. doi: 10.1093/ehjcr/ytab305 The absence of a subcutaneous pocket or transvenous leads prevent the potential risk of infection associated with these components.88. El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysio. 2019;30(4):569-74. doi:10.1111/jce.13851 , 1010. Kypta A, Blessberger H, Kammler J, Lichtenauer M, Lambert T, Silye R, et al. First autopsy description of changes 1 year after implantation of a leadless cardiac pacemaker: unexpected ingrowth and severe chronic inflammation. Can J Cardiol. 2016;32(12):1578.e1-e2. doi:10.1016/j.cjca.2015.12.028 The fact that the Micra system has a small surface area, is entirely endovascular (with encapsulation) and submitted to higher blood turbulence, velocity and pressure may also favor a lower infection risk.99. Moltrasio M, Sicuso R, Tundo F, Tondo C. A new leadless pacemaker with atrioventricular synchronous pacing replacing a still working VVI leadless pacemaker: a case report. Eur Heart J Case Rep. 2021;5(8):ytab305. doi: 10.1093/ehjcr/ytab305

Patients with systemic infections have worse short- and long-term prognosis, and, even when complete device extraction has been accomplished and the systemic infection was resolved, high mortality rates are observed at one year, between 20% and 35%,33. Tarakji KG, Wazni OM, Harb S, Hsu A, Saliba W, Wilkoff BL. Risk factors for 1-year mortality among patients with cardiac implantable electronic device infection undergoing transvenous lead extraction: the impact of the infection type and the presence of vegetation on survival. Europace. 2014;16(10):1490-5. doi:10.1093/europace/euu147 several times associated with an even higher risk of reinfection. It is particularly important to select the most suitable procedure, and that is why choosing an intracardiac leadless device seems to guarantee an important alternative approach for dependent-pacemaker patients. El-Chami, et al.88. El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysio. 2019;30(4):569-74. doi:10.1111/jce.13851 showed that Micra implantation is safe after a previous pacemaker infection, since no Micra infection was observed and systemic infections requiring device removal were not detected during the follow-up. Also, the ability of the intracardiac leadless device to provide atrioventricular synchrony is advantageous, since it may prevent some of the deleterious effects associated with a single-chamber pacing.1111. Chinitz L, Ritter P, Khelae SK, Iacopino S, Garweg C, Grazia-Bongiorni M, et al. Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: results from the Micra atrioventricular feasibility studies. Heart Rhythm. 2018;15(9):1363-71. doi:10.1016/j.hrthm.2018.05.004

The optimal timing to perform the reimplantation is unknown and controversial. Previous literature88. El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysio. 2019;30(4):569-74. doi:10.1111/jce.13851 , 1212. Chang D, Gabriels JK, Kim BS, Ismail H, Willner J, Beldner SJ, et al. Concomitant leadless pacemaker implantation and lead extraction during an active infection. J Cardiovasc Electrophysiol. 2020;31(4):860-7.doi: 10.1111/jce.14390
https://doi.org/10.1111/jce.14390...
described just a few cases of a simultaneous procedure, since most of the operators prefer to complete the extraction and perform the reimplantation a few days later. Nonetheless, the recommendations suggest that reimplantation should occur at least 72 hours after the extraction and documentation of negative blood cultures.1313. Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14(12):e503-51. doi:10.1016/j.hrthm.2017.09.001 Curiously, a recent meta-analysis concluded that the reimplantation after 72 hours was associated with a higher risk of reinfection of the new cardiac system.1414. Chew D, Somayaji R, Conly J, Exner D, Rennert-May E. Timing of device reimplantation and reinfection rates following cardiac implantable electronic device infection: a systematic review and meta-analysis. BMJ Open. 2019;9(9):e029537. doi:10.1136/bmjopen-2019-029537 Reimplantation in a simultaneous procedure of a new pacing system and lead extraction has already proved to be feasible, without increasing complication rates.1313. Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14(12):e503-51. doi:10.1016/j.hrthm.2017.09.001 Our decision to make both techniques simultaneously was made according to a risk-benefit balance evaluation. Considering the clinical evolution during the hospitalization, age and the presence of several comorbidities, the intrinsic risk of infection associated with repeated procedures and all the potential complications, a single procedure seemed to be the best alternative approach. Moreover, literature evidence88. El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysio. 2019;30(4):569-74. doi:10.1111/jce.13851 of the safety of intracardiac leadless device in patients with pre-existing infections reinforces this assumption and, therefore, the approach used by our team.

Referências

  • 1
    Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricchio A, Barbash IM, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: Developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J. 2021;42(35):3427-520. doi:10.1093/eurheartj/ehab364
  • 2
    Clémenty N, Carion PL, Léotoing L, Lamarsalle L, Bequet FW, Brown B, et al. Infections and associated costs following cardiovascular implantable electronic device implantations: a nationwide cohort study. Europace. 2018 ;20(12):1974-80. doi:10.1093/europace/eux387
  • 3
    Tarakji KG, Wazni OM, Harb S, Hsu A, Saliba W, Wilkoff BL. Risk factors for 1-year mortality among patients with cardiac implantable electronic device infection undergoing transvenous lead extraction: the impact of the infection type and the presence of vegetation on survival. Europace. 2014;16(10):1490-5. doi:10.1093/europace/euu147
  • 4
    Blomström-Lundqvist C, Traykov V, Erba PA, Burri H, Nielsen JC, Bongiorni MG, et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections—endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Europace. 2020;22(4):515-49. doi:10.1093/europace/euz246
  • 5
    Boyle TA, Uslan DZ, Prutkin JM, Greenspon AJ, Baddour LM, Danik SB, et al. Reimplantation and repeat infection after cardiac-implantable electronic device infections: experience from the MEDIC (Multicenter Electrophysiologic Device Infection Cohort) database. Circ Arrhythm Electrophysiol. 2017;10(3):e004822 doi:10.1161/CIRCEP.116.004822
  • 6
    Bongiorni MG, Giannola G, Arena G, Soldati E, Bartoli C, Lapira F, et al. Pacing and implantable cardioverter-defibrillator transvenous lead extraction. Ital Heart J. 2005;6(3):261-6. PMID: 15875518
  • 7
    Bongiorni MG. Transvenous lead extraction: from simple traction internal transjugular approach. Milan:Springer; 2014. 176p. ISBN:978884705614 2014:
  • 8
    El-Chami MF, Johansen JB, Zaidi A, Faerestrand S, Reynolds D, Garcia-Seara J, et al. Leadless pacemaker implant in patients with pre-existing infections: Results from the Micra postapproval registry. J Cardiovasc Electrophysio. 2019;30(4):569-74. doi:10.1111/jce.13851
  • 9
    Moltrasio M, Sicuso R, Tundo F, Tondo C. A new leadless pacemaker with atrioventricular synchronous pacing replacing a still working VVI leadless pacemaker: a case report. Eur Heart J Case Rep. 2021;5(8):ytab305. doi: 10.1093/ehjcr/ytab305
  • 10
    Kypta A, Blessberger H, Kammler J, Lichtenauer M, Lambert T, Silye R, et al. First autopsy description of changes 1 year after implantation of a leadless cardiac pacemaker: unexpected ingrowth and severe chronic inflammation. Can J Cardiol. 2016;32(12):1578.e1-e2. doi:10.1016/j.cjca.2015.12.028
  • 11
    Chinitz L, Ritter P, Khelae SK, Iacopino S, Garweg C, Grazia-Bongiorni M, et al. Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: results from the Micra atrioventricular feasibility studies. Heart Rhythm. 2018;15(9):1363-71. doi:10.1016/j.hrthm.2018.05.004
  • 12
    Chang D, Gabriels JK, Kim BS, Ismail H, Willner J, Beldner SJ, et al. Concomitant leadless pacemaker implantation and lead extraction during an active infection. J Cardiovasc Electrophysiol. 2020;31(4):860-7.doi: 10.1111/jce.14390
    » https://doi.org/10.1111/jce.14390
  • 13
    Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14(12):e503-51. doi:10.1016/j.hrthm.2017.09.001
  • 14
    Chew D, Somayaji R, Conly J, Exner D, Rennert-May E. Timing of device reimplantation and reinfection rates following cardiac implantable electronic device infection: a systematic review and meta-analysis. BMJ Open. 2019;9(9):e029537. doi:10.1136/bmjopen-2019-029537
  • Study association
    This study is not associated with any thesis or dissertation work.
  • Ethics approval and consent to participate
    This article does not contain any studies with human participants or animals performed by any of the authors.
  • Sources of funding: There were no external funding sources for this study.