An Online Solid Phase Extraction-Liquid Chromatography Tandem Mass Spectrometry Method for the Analysis of Parabens in Urban Waters

Abstract

In this work, we report a method to quantify methylparaben, ethylparaben, propylparaben and butylparaben in urban waters of São Carlos (São Paulo State, Brazil) by online solid phase extraction-liquid chromatography tandem mass spectrometry (SPE-LC-MS/MS) system. The method achieved ng L-1 limits with good accuracy and precision (internal standards for each paraben were used). The total run cycle time takes 9.5 min per sample including the extraction, cleanup and columns conditioning cycles, lower collection volumes of urban water and solvent usage. The SPE cartridge applied showed robustness allowing over 500 sample injections (800 µL each) with good chromatographic performance. This study findings included the detection of parabens in samples in São Carlos.

Keywords:
sample cleanup; antimicrobials; personal care; SPE-LC-MS/MS


Introduction

Parabens are a group of substances used for more than 100 years as a preservative antimicrobials in cosmetics, food and pharmaceuticals and have been detected frequently in aquatic and biological environments.11 Garner, N.; Siol, A.; Eilks, I.; Chem. Action 2014, 103, 38, available at https://25593d2a-1813-4818-aca3-d56dd316e2c9.filesusr.com/ugd/7faa15_9e074968b5ee427a8cdca17b0a0aaa87.pdf, accessed in December 2020.
https://25593d2a-1813-4818-aca3-d56dd316...

2 Yoom, H.; Shin, J.; Ra, J.; Son, H.; Ryu, D.; Kim, C.; Lee, Y.; Sci. Total Environ. 2018, 634, 677.

3 Saunders, L. J.; Mazumder, A.; Lowe, C. J.; Environ. Toxicol. Chem. 2016, 35, 919.

4 Daughton, C. G.; Ternes, T. A.; Environ. Health Perspect. 1999, 107, 907.

5 Błędzka, D.; Gromadzińska, J.; Wąsowicz, W.; Environ. Int. 2014, 67, 27.

6 Brausch, J. M.; Rand, G. M.; Chemosphere 2011, 82, 1518.

7 Ramaswamy, B. R.; Shanmugam, G.; Velu, G.; Rengarajan, B.; Larsson, D. G. J.; J. Hazard. Mater. 2011, 186, 1586.

8 Jonkers, N.; Kohler, H. P. E.; Dammshäuser, A.; Giger, W.; Environ. Pollut. 2009, 157, 714.

9 Kang, S.; Kim, S.; Park, J.; Kim, H. J.; Lee, J.; Choi, G.; Choi, S.; Kim, S.; Kim, S. Y.; Moon, H. B.; Kim, S.; Kho, Y. L.; Choi, K.; Sci. Total Environ. 2013, 461-462, 214.

10 Cabaleiro, N.; De la Calle, I.; Bendicho, C.; Lavilla, I.; TrAC, Trends Anal. Chem. 2014, 57, 34.

11 Li, X. J.; Tang, X.; Zhou, X. H.; Chen, M. S.; Dai, J. L.; Xiong, W. M.; Zou, Z. X.; Liu, H. J.; Lv, S. Y.; Lin, H.; Lian, M.; Appl. Mech. Mater. 2013, 333-335, 1993.

12 Terasaki, M.; Takemura, Y.; Makino, M.; Environ. Chem. Lett. 2012, 10, 401.

13 Madureira, T. V.; Barreiro, J. C.; Rocha, M. J.; Cass, Q. B.; Tiritan, M. E.; J. Chromatogr. A 2009, 1216, 7033.

14 Zhang, D.; Gersberg, R. M.; Ng, W. J.; Tan, S. K.; Environ. Pollut. 2014, 184, 620.

15 Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985.

16 Pompei, C. M. E.; Campos, L. C.; da Silva, B. F.; Fogo, J. C.; Vieira, E. M.; Chemosphere 2019, 226, 210.

17 Marta-Sanchez, A. V.; Caldas, S. S.; Schneider, A.; Cardoso, S. M. V. S.; Primel, E. G.; Environ. Sci. Pollut. Res. 2018, 25, 14460.

18 Chen, Y.; Deng, P.; Xie, P.; Shang, R.; Wang, Z.; Wang, S.; Chemosphere 2017, 168, 1628.
-1919 Kang, H. M.; Kim, M. S.; Hwang, U. K.; Jeong, C. B.; Lee, J. S.; Chemosphere 2019, 226, 388. The parabens commonly found in cosmetics are: methylparaben (MP), ethylparaben (EP), propylparaben (PP) and butylparaben (BP).1010 Cabaleiro, N.; De la Calle, I.; Bendicho, C.; Lavilla, I.; TrAC, Trends Anal. Chem. 2014, 57, 34.

11 Li, X. J.; Tang, X.; Zhou, X. H.; Chen, M. S.; Dai, J. L.; Xiong, W. M.; Zou, Z. X.; Liu, H. J.; Lv, S. Y.; Lin, H.; Lian, M.; Appl. Mech. Mater. 2013, 333-335, 1993.

12 Terasaki, M.; Takemura, Y.; Makino, M.; Environ. Chem. Lett. 2012, 10, 401.

13 Madureira, T. V.; Barreiro, J. C.; Rocha, M. J.; Cass, Q. B.; Tiritan, M. E.; J. Chromatogr. A 2009, 1216, 7033.

14 Zhang, D.; Gersberg, R. M.; Ng, W. J.; Tan, S. K.; Environ. Pollut. 2014, 184, 620.

15 Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985.

16 Pompei, C. M. E.; Campos, L. C.; da Silva, B. F.; Fogo, J. C.; Vieira, E. M.; Chemosphere 2019, 226, 210.
-1717 Marta-Sanchez, A. V.; Caldas, S. S.; Schneider, A.; Cardoso, S. M. V. S.; Primel, E. G.; Environ. Sci. Pollut. Res. 2018, 25, 14460. Commercially, parabens are produced by esterification of the p-hydroxybenzoic acid associated to alcohol in the presence of a catalyst (for example, concentrated sulfuric acid or p-toluenesulfonic acid).11 Garner, N.; Siol, A.; Eilks, I.; Chem. Action 2014, 103, 38, available at https://25593d2a-1813-4818-aca3-d56dd316e2c9.filesusr.com/ugd/7faa15_9e074968b5ee427a8cdca17b0a0aaa87.pdf, accessed in December 2020.
https://25593d2a-1813-4818-aca3-d56dd316...
,22 Yoom, H.; Shin, J.; Ra, J.; Son, H.; Ryu, D.; Kim, C.; Lee, Y.; Sci. Total Environ. 2018, 634, 677.,1515 Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985.

They have many advantages compared to other preservatives, such as stability and efficiency over a wide pH range, in addition to using water-soluble substances, they have no noticeable odor or taste, they do not affect color and are stable. Accepting these characteristics, parabens can be used effectively in personal care products, as preservatives even in the water phase and are also used as additives in food packaging.1010 Cabaleiro, N.; De la Calle, I.; Bendicho, C.; Lavilla, I.; TrAC, Trends Anal. Chem. 2014, 57, 34.

11 Li, X. J.; Tang, X.; Zhou, X. H.; Chen, M. S.; Dai, J. L.; Xiong, W. M.; Zou, Z. X.; Liu, H. J.; Lv, S. Y.; Lin, H.; Lian, M.; Appl. Mech. Mater. 2013, 333-335, 1993.

12 Terasaki, M.; Takemura, Y.; Makino, M.; Environ. Chem. Lett. 2012, 10, 401.

13 Madureira, T. V.; Barreiro, J. C.; Rocha, M. J.; Cass, Q. B.; Tiritan, M. E.; J. Chromatogr. A 2009, 1216, 7033.

14 Zhang, D.; Gersberg, R. M.; Ng, W. J.; Tan, S. K.; Environ. Pollut. 2014, 184, 620.
-1515 Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985. The antibacterial properties of parabens are directly proportional to the length of the ester group chain, and so, for example, butylparaben is 4 times more capable of inhibiting the bacterial growth than ethyl paraben. The increases on the length of the chain, however, rises partition coefficient problems resulting in lower water solubility.1111 Li, X. J.; Tang, X.; Zhou, X. H.; Chen, M. S.; Dai, J. L.; Xiong, W. M.; Zou, Z. X.; Liu, H. J.; Lv, S. Y.; Lin, H.; Lian, M.; Appl. Mech. Mater. 2013, 333-335, 1993.,1212 Terasaki, M.; Takemura, Y.; Makino, M.; Environ. Chem. Lett. 2012, 10, 401.

Approximately 80% of personal care products contain parabens.1212 Terasaki, M.; Takemura, Y.; Makino, M.; Environ. Chem. Lett. 2012, 10, 401.

13 Madureira, T. V.; Barreiro, J. C.; Rocha, M. J.; Cass, Q. B.; Tiritan, M. E.; J. Chromatogr. A 2009, 1216, 7033.

14 Zhang, D.; Gersberg, R. M.; Ng, W. J.; Tan, S. K.; Environ. Pollut. 2014, 184, 620.
-1515 Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985. In chlorinated waters, phenolic compounds containing hydroxyl groups such as parabens have favorable kinetics for chlorination, thus they are easily chlorinated by sodium hypochlorite, which is one of the most used techniques for water bleaching.22 Yoom, H.; Shin, J.; Ra, J.; Son, H.; Ryu, D.; Kim, C.; Lee, Y.; Sci. Total Environ. 2018, 634, 677.,1212 Terasaki, M.; Takemura, Y.; Makino, M.; Environ. Chem. Lett. 2012, 10, 401.,1414 Zhang, D.; Gersberg, R. M.; Ng, W. J.; Tan, S. K.; Environ. Pollut. 2014, 184, 620.,1515 Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985.

Parabens have been detected in effluents at concentrations up to 13 µg L-1, this shows that daily pharmaceutical and personal care products are continuously released into aquatic environments.66 Brausch, J. M.; Rand, G. M.; Chemosphere 2011, 82, 1518.,77 Ramaswamy, B. R.; Shanmugam, G.; Velu, G.; Rengarajan, B.; Larsson, D. G. J.; J. Hazard. Mater. 2011, 186, 1586.

Other studies22 Yoom, H.; Shin, J.; Ra, J.; Son, H.; Ryu, D.; Kim, C.; Lee, Y.; Sci. Total Environ. 2018, 634, 677.

3 Saunders, L. J.; Mazumder, A.; Lowe, C. J.; Environ. Toxicol. Chem. 2016, 35, 919.

4 Daughton, C. G.; Ternes, T. A.; Environ. Health Perspect. 1999, 107, 907.

5 Błędzka, D.; Gromadzińska, J.; Wąsowicz, W.; Environ. Int. 2014, 67, 27.
-66 Brausch, J. M.; Rand, G. M.; Chemosphere 2011, 82, 1518. indicate that bioaccumulation of parabens can occur in the environment. Results presented in scientific articles33 Saunders, L. J.; Mazumder, A.; Lowe, C. J.; Environ. Toxicol. Chem. 2016, 35, 919.

4 Daughton, C. G.; Ternes, T. A.; Environ. Health Perspect. 1999, 107, 907.
-55 Błędzka, D.; Gromadzińska, J.; Wąsowicz, W.; Environ. Int. 2014, 67, 27. mention that human exposure daily is up to 8 µg kg-1 from personal care products, 417 µg kg-1 from pharmaceutical products, 0.00253 µg kg-1 from food products, among other contacts coming from dust and air. The pathway of this human exposure is domestic sewage. Studies22 Yoom, H.; Shin, J.; Ra, J.; Son, H.; Ryu, D.; Kim, C.; Lee, Y.; Sci. Total Environ. 2018, 634, 677.

3 Saunders, L. J.; Mazumder, A.; Lowe, C. J.; Environ. Toxicol. Chem. 2016, 35, 919.

4 Daughton, C. G.; Ternes, T. A.; Environ. Health Perspect. 1999, 107, 907.

5 Błędzka, D.; Gromadzińska, J.; Wąsowicz, W.; Environ. Int. 2014, 67, 27.
-66 Brausch, J. M.; Rand, G. M.; Chemosphere 2011, 82, 1518. have shown the concentration in influent of 79.6 µg L-1, effluent 3.83 µg L-1, plant 8 µg g-1, sludge 0.202 µg g-1, soil 0.008 µg g-1, sediment 0.377 µg g-1, surface waters 3 µg L-1 and in fish 3.6 µg g-1.

The analytical method for determining drug/pollutants traces in complex environmental matrices involves a considerable number of methods already described in the literature,2020 Ribeiro, C.; Ribeiro, A. R.; Maia, A. S.; Gonçalves, V. M. F.; Tiritan, M. E.; Crit. Rev. Anal. Chem. 2014, 44, 142.

21 Caban, M.; Lis, E.; Kumirska, J.; Stepnowski, P.; Sci. Total Environ. 2015, 538, 402.
-2222 Liu, Z.-h; Lu, G.-n; Yin, H.; Dang, Z.; Littier, H.; Liu, Y.; TrAC, Trends Anal. Chem. 2015, 64, 149. and this number has increased considerably. More than twenty years ago, gas chromatography technique coupled with mass spectrometry (GC-MS) and extraction of compounds using 1 L of samples and in some cases derivatization were the most used approach to quantify small organic molecules in low concentrations, such as µg L-1 and ng L-1.2323 Hodgeson, J. W.; Collins, J.; Barth, R. E.; Method 552: Determination of Haloacetic Acids in Drinking Water by Liquid-Liquid Extraction, Derivatization, and Gas Chromatography with Electron Capture Detection; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1990, available at https://www.o2si.com/docs/epa-method-552.pdf, accessed in December 2020.
https://www.o2si.com/docs/epa-method-552...

24 Munch, J. W.; Method 525.2: Determination of Organic Compounds in Drinking Water by Liquid-Solid Extraction and Capillary Column Gas Chromatography/Mass Spectrometry; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.epa.gov/sites/production/files/2015-06/documents/epa-551.1.pdf, accessed in December 2020.
https://www.epa.gov/sites/production/fil...

25 Munch, J. W.; Hautman, D. P.; Method 551.1: Determination of Chlorination Disinfection Byproducts, Chlorinated Solvents, and Halogenated Pesticides/Herbicides in Drinking Water by Liquid-Liquid Extraction and Gas Chromatography with Electron-Capture Detection; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.epa.gov/sites/production/files/2015-06/documents/epa-525.2.pdf, accessed in December 2020.
https://www.epa.gov/sites/production/fil...

26 Munch, J. W.; Method 515.2: Determination of Chlorinated Acids in Water Using Liquid-Solid Extraction and Gas Chromatography with an Electron Capture Detector; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.o2si.com/docs/epa-method-515.2.pdf, accessed in December 2020.
https://www.o2si.com/docs/epa-method-515...

27 Munch, J. W.; Method 508: Determination of Chlorinated Pesticides in Water by Gas Chromatography with An Electron Capture Detector; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.o2si.com/docs/epa-method-508.pdf, accessed in December 2020.
https://www.o2si.com/docs/epa-method-508...
-2828 Munch, J. W.; Method 506: Determination of Phthalate and Adipate Esters in Drinking Water by Liquid-Liquid Extraction or Liquid-Solid Extraction and Gas Chromatography with Photoionization Detection; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at http://www.cromlab.es/Articulos/Metodos/EPA/500/506.pdf, accessed in December 2020.
http://www.cromlab.es/Articulos/Metodos/...

Sample collection, storage and preservation, transport and pre-treatment steps are as important as the type of sample preparation selected to guarantee the quality of the results,2929 Majors, E. R. In Sample Preparation Fundamentals for Chromatography; Agilent: Delaware, USA, 2013, available at https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf, accessed in December 2020.
https://www.agilent.com/cs/library/prime...

30 Raynie, E. D.; LCGC North America 2016, 34, 174, available at https://www.chromatographyonline.com/view/trends-sample-preparation-4, accessed in December 2020.
https://www.chromatographyonline.com/vie...

31 Chirila, E. S.; Draghici, C.; Dobrinas, S. In Chemicals as Intentional and Accidental Global Environmental Threats; Simeonov, L.; Chirila, E. eds.; Springer: Brasov, 2006, p. 7.
-3232 Petrovic, M.; Farré, M.; de Alda, M. L.; Perez, S.; Postigo, C.; Köck, M.; Radjenovic, J.; Gros, M.; Barcelo, D.; J. Chromatogr. A 2010, 1217, 4004. but higher volumes of water as 0.1-1 L involve larger rooms to process the samples, larger amount of extraction solvents and more sample preparation time.

Studies2929 Majors, E. R. In Sample Preparation Fundamentals for Chromatography; Agilent: Delaware, USA, 2013, available at https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf, accessed in December 2020.
https://www.agilent.com/cs/library/prime...
,3030 Raynie, E. D.; LCGC North America 2016, 34, 174, available at https://www.chromatographyonline.com/view/trends-sample-preparation-4, accessed in December 2020.
https://www.chromatographyonline.com/vie...
show that the time and labor intensity spent on sample preparation remains responsible for nearly 70% of the total time in the laboratory to achieve lower limit of quantification. The time spent in most cases involves time-consuming steps. For this reason, and with the increasing number of samples to be analyzed in monitoring laboratories, there is a continuous search for methods of high productivity, cost effective and easy handling. With this, there has been an increase in the use of automated instruments that integrate extraction, purification and detection.3333 Ramos, L.; J. Chromatogr. A 2012, 1221, 84.

34 Poole, C. F.; Gunatilleka, A. D.; Sethuraman, R.; J. Chromatogr. A 2000, 885, 17.

35 Hennio, M.-C.; J. Chromatogr. A 1999, 856, 3.
-3636 Gilart, N.; Borrull, F.; Fontanals, N.; Marcé, R. M.; Trends Environ. Anal. Chem. 2014, 1, e8. Also, surveys2929 Majors, E. R. In Sample Preparation Fundamentals for Chromatography; Agilent: Delaware, USA, 2013, available at https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf, accessed in December 2020.
https://www.agilent.com/cs/library/prime...
,3030 Raynie, E. D.; LCGC North America 2016, 34, 174, available at https://www.chromatographyonline.com/view/trends-sample-preparation-4, accessed in December 2020.
https://www.chromatographyonline.com/vie...
indicate that sample loads and sample per week will increase in the next years, which means the need of faster sample preparation and chromatography methods.

Generic approaches have been developed to prepare water samples with the final analysis by LC-MS/MS (liquid chromatography coupled with sequential mass spectrometry) mainly involving techniques such as: LLE (liquid-liquid extraction), SPE (solid phase extraction), QuEChERS (quick, easy, cheap, effective, rugged, and safe), SPME (solid phase microextraction) and SBSE (stir bar sorptive extraction), yet other techniques such as: RAM (restricted access materials) and TFC (turbulent flow chromatography) have been explored.2828 Munch, J. W.; Method 506: Determination of Phthalate and Adipate Esters in Drinking Water by Liquid-Liquid Extraction or Liquid-Solid Extraction and Gas Chromatography with Photoionization Detection; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at http://www.cromlab.es/Articulos/Metodos/EPA/500/506.pdf, accessed in December 2020.
http://www.cromlab.es/Articulos/Metodos/...

29 Majors, E. R. In Sample Preparation Fundamentals for Chromatography; Agilent: Delaware, USA, 2013, available at https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf, accessed in December 2020.
https://www.agilent.com/cs/library/prime...

30 Raynie, E. D.; LCGC North America 2016, 34, 174, available at https://www.chromatographyonline.com/view/trends-sample-preparation-4, accessed in December 2020.
https://www.chromatographyonline.com/vie...

31 Chirila, E. S.; Draghici, C.; Dobrinas, S. In Chemicals as Intentional and Accidental Global Environmental Threats; Simeonov, L.; Chirila, E. eds.; Springer: Brasov, 2006, p. 7.

32 Petrovic, M.; Farré, M.; de Alda, M. L.; Perez, S.; Postigo, C.; Köck, M.; Radjenovic, J.; Gros, M.; Barcelo, D.; J. Chromatogr. A 2010, 1217, 4004.

33 Ramos, L.; J. Chromatogr. A 2012, 1221, 84.

34 Poole, C. F.; Gunatilleka, A. D.; Sethuraman, R.; J. Chromatogr. A 2000, 885, 17.

35 Hennio, M.-C.; J. Chromatogr. A 1999, 856, 3.

36 Gilart, N.; Borrull, F.; Fontanals, N.; Marcé, R. M.; Trends Environ. Anal. Chem. 2014, 1, e8.

37 Fumes, B. H.; Silva, M. R.; Andrade, F. N.; Nazario, C. E. D.; Lanças, F. M.; TrAC, Trends Anal. Chem. 2015, 71, 9.

38 Poole, C. F.; TrAC, Trends Anal. Chem. 2003, 22, 362.

39 Barreiro, J. C.; Vanzolini, K. L.; Madureira, T. V.; Tiritan, M. E.; Cass, Q. B.; Talanta 2010, 82, 384.

40 Valls-Cantenys, C.; Scheurer, M.; Iglesias, M.; Sacher, F.; Brauch, H. J.; Salvadó, V.; Anal. Bioanal. Chem. 2016, 408, 6189.

41 Cerqueira, M. B. R.; Caldas, S. S.; Primel, E. G.; J. Chromatogr. A 2014, 1336, 10.

42 Bylda, C.; Thiele, R.; Kobold, U.; Volmer, D. A; Analyst 2014, 139, 2265.

43 Lopes, B. R.; Barreiro, J. C.; Cass, Q. B.; J. Pharm. Biomed. Anal. 2016, 130, 318.

44 Barciela-Alonso, M. C.; Otero-Lavandeira, N.; Bermejo-Barrera, P.; Microchem. J. 2017, 132, 233.
-4545 Cassiano, N. M.; Barreiro, J. C.; Cass, Q. B.; J. Braz. Chem. Soc. 2014, 25, 9.

Currently, LC-MS/MS offers high sensitivity,2929 Majors, E. R. In Sample Preparation Fundamentals for Chromatography; Agilent: Delaware, USA, 2013, available at https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf, accessed in December 2020.
https://www.agilent.com/cs/library/prime...
,4040 Valls-Cantenys, C.; Scheurer, M.; Iglesias, M.; Sacher, F.; Brauch, H. J.; Salvadó, V.; Anal. Bioanal. Chem. 2016, 408, 6189. uses smaller sample volumes, allows lower transportation and storage prices, reduces the possibility of handling error, employs less amount of solvent providing double benefits by less acquisition and disposal cost, no evaporation step decreasing time-consuming and an integrated system of software rising productivity.

The SPE technique basically consists of four stages: (i) conditioning, to activate the silica-based materials avoiding hydrophobic-hydrophilic shielding and/or to ensure the cleaning of the polymer-based materials, (ii) loading of the sample for the extraction/enrichment of the analytes of interest, (iii) cleanup for purification of the SPE phase, since many interferents are extracted with the target compounds and (iv) elution to obtain the analyte.2929 Majors, E. R. In Sample Preparation Fundamentals for Chromatography; Agilent: Delaware, USA, 2013, available at https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf, accessed in December 2020.
https://www.agilent.com/cs/library/prime...
,3030 Raynie, E. D.; LCGC North America 2016, 34, 174, available at https://www.chromatographyonline.com/view/trends-sample-preparation-4, accessed in December 2020.
https://www.chromatographyonline.com/vie...
,3434 Poole, C. F.; Gunatilleka, A. D.; Sethuraman, R.; J. Chromatogr. A 2000, 885, 17.

35 Hennio, M.-C.; J. Chromatogr. A 1999, 856, 3.
-3636 Gilart, N.; Borrull, F.; Fontanals, N.; Marcé, R. M.; Trends Environ. Anal. Chem. 2014, 1, e8.

Silica-based phases materials due to their diversity have shown to be able to extract a variety classes of analytes. The most used in aqueous matrices has being the C18 type derivative, other types are also used such as C8, phenyl, NH2, SAX (strong anion exchange) and SCX (strong cation exchange), however, silica-based phases have limitations due to the narrow operating pH range. Thus, polymeric phases have been gaining popularity since they have similar C18 and weak ion exchange types selectivity, but high pH range applications.3535 Hennio, M.-C.; J. Chromatogr. A 1999, 856, 3.

36 Gilart, N.; Borrull, F.; Fontanals, N.; Marcé, R. M.; Trends Environ. Anal. Chem. 2014, 1, e8.

37 Fumes, B. H.; Silva, M. R.; Andrade, F. N.; Nazario, C. E. D.; Lanças, F. M.; TrAC, Trends Anal. Chem. 2015, 71, 9.

38 Poole, C. F.; TrAC, Trends Anal. Chem. 2003, 22, 362.

39 Barreiro, J. C.; Vanzolini, K. L.; Madureira, T. V.; Tiritan, M. E.; Cass, Q. B.; Talanta 2010, 82, 384.

40 Valls-Cantenys, C.; Scheurer, M.; Iglesias, M.; Sacher, F.; Brauch, H. J.; Salvadó, V.; Anal. Bioanal. Chem. 2016, 408, 6189.

41 Cerqueira, M. B. R.; Caldas, S. S.; Primel, E. G.; J. Chromatogr. A 2014, 1336, 10.
-4242 Bylda, C.; Thiele, R.; Kobold, U.; Volmer, D. A; Analyst 2014, 139, 2265.,4646 Munch, J. W.; Method 528: Determination of Phenols in Drinking Water by Solid Phase Extraction and Capillary Column Gas Chromatography/Mass Spectrometry (GC/MS); Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 2000, available at https://www.epa.gov/sites/production/files/2015-09/documents/m_528.pdf, accessed in December 2020.
https://www.epa.gov/sites/production/fil...

The online mode presents as main advantages reusable column type cartridges, fast elution with less solvent consumption, minimal degradation and no loss of analyte by evaporation resulting in reduced analysis time. The disadvantages are fewer material phases availability and lower analyte concentration factor.4646 Munch, J. W.; Method 528: Determination of Phenols in Drinking Water by Solid Phase Extraction and Capillary Column Gas Chromatography/Mass Spectrometry (GC/MS); Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 2000, available at https://www.epa.gov/sites/production/files/2015-09/documents/m_528.pdf, accessed in December 2020.
https://www.epa.gov/sites/production/fil...

47 Moghimi, A.; Hosseini, S. A.; Far, A. M.; Orient. J. Chem. 2015, 31, 263.

48 Gołębiowski, M.; Stepnowski, P.; Leszczyńska, D.; Chem. Pap. 2017, 71, 831.

49 Buszewski, B.; Szultka, M.; Crit. Rev. Anal. Chem. 2012, 42, 198.
-5050 Liska, I.; J. Chromatogr. A 1993, 655, 163.

The term online SPE applies when switching valves are used to couple the SPE column to the analytical column. The chromatographic system can be configured with high flexibility on. The most selected are the 10-port system that present the same functionality as the 6-port valve but has the advantage of working with two SPE columns at the same time with higher-throughput.5151 Tlili, I.; Caria, G.; Ouddane, B.; Ghorbel-Abid, I.; Ternane, R.; Trabelsi-Ayadi, M.; Net, S.; Sci. Total Environ. 2016, 563-564, 424.,5252 Kuklenyik, Z.; Calafat, A. M.; Barr, J. R.; Pirkle, J. L.; J. Sep. Sci. 2011, 34, 3606.

Applications of online SPE systems have become increasingly popular and different types of applications have been published such as drugs, carbamates, pyrethroids, sulfonamides and pharmaceuticals in water. The common injection volumes usually are 0.5 mL up to 10 mL.5252 Kuklenyik, Z.; Calafat, A. M.; Barr, J. R.; Pirkle, J. L.; J. Sep. Sci. 2011, 34, 3606.

53 Heuett, N. V.; Ramirez, C. E.; Fernandez, A.; Gardinali, P. R.; Sci. Total Environ. 2015, 511, 319.

54 Fernández-Ramos, C.; Šatínský, D.; Solich, P.; Talanta 2014, 129, 579.

55 García-Galán, M. J.; Díaz-Cruz, M. S.; Barceló, D.; Talanta 2010, 81, 355.

56 Bones, J.; Thomas, K.; Nesterenko, P. N.; Paull, B.; Talanta 2006, 70, 1117.
-5757 Cass, Q. B.; Cassiano, N. M. In Cromatografia Líquida: Novas Tendências e Aplicações; Elsevier: Rio de Janeiro, Brazil, 2015.

This study aimed to investigate the occurrence of parabens in urban water and herein we report a qualified online SPE-LC-MS/MS method at ng L-1 limits. The results presented herein might be useful to support new researches on analysis of water contaminants with greener methods.

Experimental

Standards, materials and reagents

For the experiments, standards of target compounds were acquired from Sigma-Aldrich (Saint Louis, USA) including 1 g per flask of methylparaben (MP) part number PHR1012-1G, ethylparaben (EP) part number PHR1011-1G, propylparaben (PP) part number PHR1010-1G and butylparaben (BP) part number PHR1022-1G, and a mixture of internal standard (IS) at 10 mg L-1 in acetone of each paraben represented as 13C6 part number 32124-5ML. A fulvic acid solution IHSS SUWANNE batch 1S101 was donated by Analytical Instrumentation laboratory of Embrapa (São Carlos, Brazil) and humic acid of the Sigma-Aldrich (Saint Louis, USA) part number 53680 was donated by the Environmental Biogeochemistry Laboratory of UFSCar (São Carlos, Brazil).

Regenerated cellulose membrane Agilent (Santa Clara, USA) part number 5190-5109 (15 mm × 0.45 µm) were used for sample filtration, and Agilent 2 mL vials, pre-slit septa and screw caps (part number 5183-2076 and 5182-0715) were also used. A polymeric reversed phase for small molecules sorbent that is hydrophobic, with no bonded phase or alkyl ligands, Agilent PLRP-S (2.1 × 12.5 mm, 15 µm) were used as the online SPE cartridge while the analytical column was the Poroshell SB-C18 column (2.1 × 50 mm, 2.7 µm).

The solvents methanol and acetonitrile were LC-MS grade from Sigma-Aldrich or Honeywell (Charlotte, USA) and ultra-pure water from a Sartorius (Göttingen, Germany) system. Otherwise stated, the mobile phases were prepared as v/v relationship.

Site of sample collection and pre-treatment

Water samples were collected at discharge along the Monjolinho River in the region of São Carlos, SP, Brazil in November 2016. The location of samples sites are: (1) Monjolinho spring water: latitude 22°00’33” S and longitude 47°50’07” W and presented pH 7; (2) areas with possible industrial and urban contamination: latitude 21°98’58” S, longitude 47°88’22” W and presented pH 7.5, it is located at the Federal University of São Carlos (UFSCar); (3) close to the confluence with the Tijuco Preto stream, inserted in an urbanized area latitude 22°00’68” S, longitude 47°90’47” W and presented pH 7.5 and (4) is after an urban area, latitude 22°02’21” S, longitude 47°91’40” W and presented pH 7.5. The sewage treatment system in the city of São Carlos is operational at 100%, currently. Samples were transported to the laboratory, and to remove suspended particles they were filtered through regenerated cellulose membrane before being stored at 4 °C if not instantly analyzed after preparation.

Working solutions

Individual solutions of target parabens were prepared in the following concentrations: 280 mg L-1 MP, 740 mg L-1 EP, 1,050 mg L-1 PP and 1,020 mg L-1 BP. The preparation was carried out by dissolving 2.8 mg of MP, 7.4 mg of EP, 10.5 mg of PP and 10.2 mg of BP in 10 mL of solvent mixture of methanol and water (50:50). Thus, a solution of the mixture of parabens at 10 mg L-1 in water and methanol (95:5) was finally prepared.

A methanolic solution containing the four (13C66 Brausch, J. M.; Rand, G. M.; Chemosphere 2011, 82, 1518.) internal standards at 10 µg L-1 was prepared to be added to the blank and water samples, and calibration solutions.

Calibration curve, quality controls and samples analysis

The paraben mixture solution at 10,000,000 ng L-1 was diluted in analyte-free spring water from Monjolinho River to produce the working solution at 1,000 ng L-1 from which the calibrators were prepared at the following concentrations: 3.9, 7.8, 15.6, 31.25, 62.5, 125, 250, 500 and 1000 ng L-1. The quality controls (QCs) solutions evaluated were also prepared on an analyte-free spring water in the following concentrations 20, 400, 800 ng L-1 for MP and PP, 90, 400, 800 ng L-1 for EP and 10, 400, 800 ng L-1 for BP. A 5.0 µL solution of IS at 10,000 ng L-1 was added to each 1 mL of blanks, QCs, water samples and calibrators, corresponding to 50 ng L-1 of deuterated standard at final concentration.

Online SPE-LC-MS/MS method

The SPE-LC-MS/MS system was from Agilent and composed by an autosampler, an online SPE unit, and an ultra high pressure liquid chromatography (UHPLC) 1290 with a MS/MS 6460C model having an electrospray ionization source with jet stream technology (ESI-AJS). The MassHunter acquisition, qualitative, quantitative and optimizer software was used for data analysis.5858 MassHunter Optimizer, version B07; Agilent Inc., Santa Clara, CA, USA, 2014.

The samples load injections were of 800 µL at a flow rate of 900 µL per min. Then, a mixture of water: methanol (95:5) flushed the PLRP-S cartridges at 1 mL min-1 by 90 s to polar interferents removal (Figure 1a). For the analysis, the valve position was changed to connect the SPE cartridge to the Poroshell SB-C18 analytical column set at the UHPLC system (Figure 1b). Gradient elution with water (A) and acetonitrile (B) without any additives were used at the following conditions: 5% of B held on 1.76 min to 67% of B at 5.86 min and finally 100% of B at 6.26 min held until 8 min at 0.4 mL min-1. Oven temperature at 50 °C and MS/MS were used while one SPE cartridge were in running analyses, simultaneously the valve configuration allowed the second cartridge to be prepared for the next load injection. For that, methanol at 1 mL min-1 for 200 s was used aiming to fasten cleanup cartridge and minimizing the carryover, then, the conditioning solution of water: methanol (95:5) at 1 mL per min for 230 s was applied allowing the cartridge to be ready for the next analysis. The total run of analysis time was 9.5 min per sample.

Figure 1
(a) Online SPE system at position 1 with 10-port valve with two available positions; (b) online SPE system at position 2 with a 10-port valve with two available positions.

The conditions used for ESI-AJS in negative mode were gas temperature at 300 °C, gas flow at 10 L min-1, nebulizer at 50 psi, heat sheath gas temperature at 350 °C, sheath gas flow at 11 L min-1, capillary 3500 V, nozzle voltage 500 V and cell accelerator at 7 V. The SRM (selected reaction monitoring) for target analytes were MP 151 > 92 (collision energy (CE) 16 V), 151 > 136 (CE 8 V) and 85 V for fragmentor, EP 165 > 92 (CE 16 V), 165 > 137 (CE 8 V) and 65 V for fragmentor, PP 179 > 192 (CE 20 V), 179 > 137 (CE 7 V) and 80 V for fragmentor, BP 193 > 92 (CE 20 V), 193 > 136 (CE 23 V) and 60 V for fragmentor. The SRM for internal standards were MP 13C6 157 > 98 (CE 16 V), EP 13C6 171 > 98 (CE 18 V), PP 13C6 185 > 98 (CE 20 V), BP 13C6 199 > 98 (CE 20 V) and fragmentor voltage were the same as target analytes, respectively.

Matrix effect procedure and method validation

The qualification protocol of the method was based on the bioanalytical validation guide EMA (European Medicines Agency)5959 Smith, G.; Bioanalysis 2012, 4, 865. and following parameters were evaluated: memory effect, calibration curve, selectivity, recovery (extraction efficiency), precision and accuracy, limits of quantification and detection, dilution integrity, matrix effect and stability.

The evaluation of matrix effect (ME) was determined by calculating the peak area of internal standards compounds in spiked samples with the lowest and highest QCs from sites samples 2, 3 and 4, and then, the percentage of increase/decrease of signals in different lots of spring water samples (site 1) were calculated. Subsequently, for the investigation of ME, an injection of samples (2, 3 and 4) and standards solutions at 14.5 mg L-1 of fulvic acids and 2.9 mg L-1 humic acids prepared in ultra-purified water were analyzed by the same online SPE and UHPLC method conditions for target analytes, however, coupling to diode array detector (Agilent 1290 model) at 254 nm with the MS/MS valve on waste position.

The chemical stability of parabens was evaluated by the results of solution prepared at 125 ng L-1 in ultra-pure water and QCs of spiked sample 4 at low and high concentration at room temperature and injected as freshly prepared, after 4 h, 24 h (autosampler stability) and finally after 8 days of refrigeration at 4 °C.

To determine linearity, six blank spring water samples were spiked in triplicate at six different concentration level and external calibration curve was constructed by plotting the peak area versus concentration and internal standard. For selectivity evaluation, collected analyte-free spring water samples were compared with spring water samples spiked with the mixture of the parabens at LOQ (limit of quantification) concentration. Samples were individually analyzed and evaluated for interference. Precision was expressed as relative standard deviation (RSD) of the replicate measurements and the accuracy of the method was evaluated as the percentage between the calculated and the nominal concentration of each analyte. The extraction efficiencies were calculated using the QC samples. The percentage of recovery was obtained comparing the processes data of QC samples with those prepared at the same concentration in ultrapure water. The LOQ values were determined using spiked spring water samples prepared in triplicate. The LOD (limit of detection) was calculated as the minimum detectable amount of compound, with a signal-to-noise (S/N) ratio of ≥ three. The LOQ was obtained as the lowest calibration level with the accepted criteria that the precision and the accuracy for three extracted samples had variability of less or equal to 20%. In accordance with EMA protocol, the obtained RSD were under 15% variability.

Results and Discussion

Online SPE systems as illustrated in Figures 1a and 1b have demonstrated to be an asset for environmental laboratories as it requires minimal time pre-treatment, sample preparation and, thereby increases throughput. Moreover, it reduces total volume of collected samples succeeding in requiring less storage space in a cooling chamber or refrigerator.3333 Ramos, L.; J. Chromatogr. A 2012, 1221, 84.

34 Poole, C. F.; Gunatilleka, A. D.; Sethuraman, R.; J. Chromatogr. A 2000, 885, 17.

35 Hennio, M.-C.; J. Chromatogr. A 1999, 856, 3.

36 Gilart, N.; Borrull, F.; Fontanals, N.; Marcé, R. M.; Trends Environ. Anal. Chem. 2014, 1, e8.

37 Fumes, B. H.; Silva, M. R.; Andrade, F. N.; Nazario, C. E. D.; Lanças, F. M.; TrAC, Trends Anal. Chem. 2015, 71, 9.

38 Poole, C. F.; TrAC, Trends Anal. Chem. 2003, 22, 362.

39 Barreiro, J. C.; Vanzolini, K. L.; Madureira, T. V.; Tiritan, M. E.; Cass, Q. B.; Talanta 2010, 82, 384.
-4040 Valls-Cantenys, C.; Scheurer, M.; Iglesias, M.; Sacher, F.; Brauch, H. J.; Salvadó, V.; Anal. Bioanal. Chem. 2016, 408, 6189.

For setting up the online SPE system, a sequence of events must be carefully examined. This involves the steps illustrated at Figure 2: (i) analytes ionization, including the selection of ionization mode, collision energy and capillary voltage; (ii) selection of mobile phases that involves experiments with different solvents, flow rate and additives within the analytical column with/without coupling to the SPE column; (iii) the final adjustment of the ionization source parameters involving temperature, nebulizing flow, sheath and drying gas, and finally, (iv) selection of suitable solvents for the conditioning of cartridge and cleanup of interferences from the samples via online SPE.

Figure 2
Sequence of events for method development and optimization followed by first (i) upper analytes ionization and SRM transitions by direct injection, (ii) selection of mobile phase for sensitivity and separation enhancement, (iii) evaluation of source optimization in equilibrium with previous steps and finally (iv) adjustments of online SPE steps.

The four steps (i-iv) designed in flow chart (Figure 2) were followed. In this respect, evaluation of the ionization conditions of the analytes on positive and negative ionization modes were the first parameters selected (i). The highest signal intensities were obtained at the negative polarity mode, and thus selected for the experiments. The parameters were adjusted by the Optimizer software for SRM for each paraben.5858 MassHunter Optimizer, version B07; Agilent Inc., Santa Clara, CA, USA, 2014. By optimization of the fragmentor voltage (V) for each ion, selection of the product ions and CE values of each transition were carried out for selecting the quantification and qualification ion transitions. The standard conditions ESI-AJS source for small molecule were evaluated and no significant changes in the sensitivity were observed under manual optimization, thus, the capillary voltage of 3500 V was used. The data acquisition rate was also manually estimated using the dwell time function.

The main fragments of the deprotonated compounds is the loss of alkyl chain linked to the ester followed by the loss of CO2.6060 González-Mariño, I.; Quintana, J. B.; Rodríguez, I.; Cela, R. S.; Rapid Commun. Mass Spectrom. 2009, 23, 1756. Thus, the nominated SRM transitions were [M - H - CH3 - CO2]- and [M - H - CH3]- for methylparaben, [M - H - CH3 - CO2]- and [M - H - CH2]- for ethylparaben, [M - H - CH3 - CO2]- and [M - H - CH2]- for propylparaben and [M - H - CH3 - CO2]- and [M - H - CH3]- for propylparaben.

As chromatographic conditions (step (ii)), isocratic elution using methanol (MeOH) or acetonitrile (ACN) as organic modifier in water were examined at the following proportions: 40:60, 50:50 and 70:30 but, they afforded high tail factors chromatographic bands when the SPE cartridge was coupled to the analytical column. The linear gradient elution mode (5 to 100% ACN) at 15% per min gradient step gave the highest resolution and minor tail factor and, thus, it was selected as the chromatographic conditions.

All selected parabens have approximately pKa of 8 and partition coefficient (log P) between 1.6-3.2. To improve ionization, experiments were carried out with and without additives such as: acetic and formic acid and ammonium formate in the mobile phase. No significant improvement in sensitivity for formic and acetic acid (less than 10%) was observed, and with ammonium formate there was a decrease in signal intensity. Based on these results, and to avoid favoring the ionization of matrices interferents no additives were used. Flow rate at 0.4, 0.5 and 0.6 mL min-1 were also examined, and due to the ESI mass analyzer, the flow at 0.4 mL min-1 was selected.

For the final adjustment of the ionization source parameters (step (iii)) nebulizing flow, sheath, drying gas temperature and nitrogen flow parameters were all adjusted via the Source optimizer software.5858 MassHunter Optimizer, version B07; Agilent Inc., Santa Clara, CA, USA, 2014. This might be done manually, but it is just much easier to use the software.

For the analysis cycle, parameters such as mobile phase for sample injection and SPE cartridges cleanups as well as SPE conditioning must be all evaluated (step (iv)). For selecting the appropriated conditions, one needs to consider the physicochemical characteristics of the SPE stationary phases. In this study, the polymeric-based SPE cartridge allowed the use of MeOH and ACN as organic modifiers. For sample cleanup, after injection, these modifiers in water at the proportion of 5 to 20% were examined. Due to the higher ACN elution strength, it affected the extraction of MP resulting in signal losses of at least 10% at total ion chromatogram (TIC). In this regard, 5% of MeOH in water was selected. For the SPE cartridge additional cleanup, 100% MeOH was used, previously to conditioning at 5% MeOH for the next injection. Since the analyses were carried out with ACN as mobile phase, the use of MeOH at the cleanup stage, resulted in higher selectivity.

To evaluate the cleanup time, the SPE cartridge was connected directly to the diode array detector (DAD) and the elution monitored at 254 nm. The cleanup time was evaluated for the samples collected from four different locations from the urban waters in São Carlos-SP. The 1.75 min elution time using 5% of MeOH in water, at a flow rate of 1 mL min-1 accomplished the polar interferences removal. The use of a 10 port and 2 positions SPE valve allowed that while running the analysis, the second SPE cartridge to be cleaned/conditioned for the following injection.

The analytical method was qualified following the EMA5959 Smith, G.; Bioanalysis 2012, 4, 865. validation guide and had the following parameters evaluated: memory effect, selectivity, calibration curve, precision and accuracy, recovery (extraction efficiency), limit of quantification and method detection, dilution integrity, stability and matrix effect. The SRM chromatogram of final method is illustrated in Figure 3.

Figure 3
Final SRM chromatograms at 1,000 ng L-1 presented being the sequential of retention time methyl, ethyl, propyl and butyl parabens with their internal standard 13C6.

The analytical calibration curve was plotted by linear regression in triplicate at six concentration level. The linear coefficient determination (R2) was above 0.99 for all analytes, the RSD of the triplicates of the curve was lower than 10% showing suitability. The accuracy showed a deviation less than 15% of the nominal value in all concentrations of the curves, except for the LOQ of butylparaben where the deviation found was of 16% and some of these details are presented in Table 1. The precision and accuracy of the method were evaluated through inter and intra-batch repeatability by low, medium and high concentration as quality control checks, and the results for accuracy and intermediate precision values obtained were 92-115% of the nominal value with RSD between 1.3-7.8%, respectively.

Table 1
LOD (limit of detection), LOQ (limit of quantification), R2 (linear regression coefficient), equation of calibration curve and linear range are presented

The inter-batch accuracy and intermediate precision by two analysts in three different days still showed accuracy between 95-107% and RSD less than 10%. The LOQ is the first concentration of the calibration curve with accuracy and precision that varied from 86-111% and 1-18% respectively, for the four parabens. For the LOD, verified experimentally, the standard solution with the lowest reliable concentration that differed from the memory effect varied between 2.0-15.6 ng L-1. The integrity of the dilution was given by diluting Monjolinho spring water 20 times in the concentration of 2,000 ng L-1 (100 ng L-1) in ultrapure water. The accuracy was 93-103% and the RSD was 3.8%.

Stability assessment was carried out by comparing the average areas of the standards injected as soon as samples were prepared with 4 and 24 h in the auto injector and finally after 8 days of refrigeration. For that, the sample of urban water from site 4 at low and high concentrations was used since it was the most challenging matrix presenting the highest ionic suppression, these concentrations were evaluated in comparison to the standard solution of 125 ng L-1. Regarding stability, the first batch results were compared to 4 and 24 h after sample preparation, the accuracy values were between 88-106% and precision less than 15%, showing satisfactory results up to 24 h period. Also, the results of the standard solution refrigerated for 8 days had an accuracy between 114-118%, however, the sample analyzed in the same condition showed an accuracy of 12-20% for MP, 74-77% for EP, 202-210% for PP and 140-144% for BP. For the proposed condition, these results were considered unsatisfactory for sample storage period.

The method application was demonstrated applying it to samples of urban waters collected in the city of São Carlos-SP. The parabens quantifications in these samples showed the following results < LOQ (site 2), 20.0 ng L-1 of MP (site 3), 18.0 ng L-1 of MP and 4.1 ng L-1 of BP (site 4).

The use of IS was important for correcting the matrix effect of the samples from the collecting sites 2, 3 and 4. Naturally, water presents different levels of humic and fulvic substances, scientific research6161 Carder, K. L.; Steward, R. G.; Harvey, G. R.; Ortner, P. B.; Limnol. Oceanogr. 1989, 34, 68.

62 Weigel, S.; Kallenborn, R.; Hühnerfuss, H.; J. Chromatogr. A 2004, 1023, 183.

63 Guenu, S.; Hennion, M. C.; J. Chromatogr. A 1996, 737, 15.
-6464 Lin, W. C.; Chen, H. C.; Ding, W. H.; J. Chromatogr. A 2005, 1065, 279. has shown that extracting organic contaminants at a pH close to neutral limits influence the extraction of these interferents by SPE. Based on the chromatographic profile of humic substances standards compared to sample profiles obtained on DAD, as shown in Figures 4 and 5, it was obvious that the matrix effect was due to the amount of humic and fulvic acids in these samples. Although the method has determined conditions to minimize this effect, it was not possible to eliminate it as characteristics of the cartridge phase used for extraction and limitation of the configuration of online SPE. Without IS the results could had shown negative values based on ion suppression observed which was from (-69.8) to (-16.8). Nevertheless, due to useful use of IS, all urban waters exhibited precision within RSD below 15% and accuracy between 85-115%.

Figure 4
DAD signals at 254 nm, being sample from site 2, 3 and 4 represented by the colors black, red and green, respectively.
Figure 5
DAD signals at 254 nm, being sample from site 4, humic and fulvic acids standards represented by the colors green, red and black, respectively.

The method was validated and met the acceptance criteria of the guide used. Matrix effect was observed due to the inefficiency of reduction of humic substances when present in high concentration in the samples. The use of the internal standard for each paraben, however, corrected the deviations caused by the matrix effect allowing ng L-1 limits.

Comparison of this study results with countries such as China,6565 Huang, C.; Wu, L. H.; Liu, G. Q.; Shi, L.; Guo, Y.; Arch. Environ. Contam. Toxicol. 2018, 75, 224.

66 Lu, J.; Mao, H.; Li, H.; Wang, Q.; Yang, Z.; Sci. Total Environ. 2017, 605-606, 1064.
-6767 Peng, X.; Yu, Y.; Tang, C.; Tan, J.; Huang, Q.; Wang, Z.; Sci. Total Environ. 2008, 397, 158. India,77 Ramaswamy, B. R.; Shanmugam, G.; Velu, G.; Rengarajan, B.; Larsson, D. G. J.; J. Hazard. Mater. 2011, 186, 1586. Portugal,6868 Jonkers, N.; Sousa, A.; Galante-Oliveira, S.; Barroso, C. M.; Kohler, H. P. E.; Giger, W.; Environ. Sci. Pollut. Res. 2010, 17, 834. Switzerland,88 Jonkers, N.; Kohler, H. P. E.; Dammshäuser, A.; Giger, W.; Environ. Pollut. 2009, 157, 714. UK6969 Kasprzyk-Hordern, B.; Dinsdale, R. M.; Guwy, A. J.; Water Res. 2009, 43, 363. and other Brazilian1616 Pompei, C. M. E.; Campos, L. C.; da Silva, B. F.; Fogo, J. C.; Vieira, E. M.; Chemosphere 2019, 226, 210. reservoir was evaluated. Different samples cleanup procedures were described for the reported results varying from liquid-liquid extraction to offline and automated SPE for further analysis by GC-MS or LC-MS/MS. The parabens concentrations ranged from less than the LOQ up to 18,300 ng L-1 for methylparaben, 1,240 ng L-1 for ethylparaben, 5,940 ng L-1 for propylparaben and 4,470 ng L-1 for butylparaben. The literature results highlight China and Switzerland as the countries from which highest concentrations for these parabens were found. The explanation for these results may be the high demographic density of the collected sites of the samples. The collection site in China6565 Huang, C.; Wu, L. H.; Liu, G. Q.; Shi, L.; Guo, Y.; Arch. Environ. Contam. Toxicol. 2018, 75, 224.

66 Lu, J.; Mao, H.; Li, H.; Wang, Q.; Yang, Z.; Sci. Total Environ. 2017, 605-606, 1064.
-6767 Peng, X.; Yu, Y.; Tang, C.; Tan, J.; Huang, Q.; Wang, Z.; Sci. Total Environ. 2008, 397, 158. was from a region of 12 million people, with only 70% of the treated sewage; Switzerland,77 Ramaswamy, B. R.; Shanmugam, G.; Velu, G.; Rengarajan, B.; Larsson, D. G. J.; J. Hazard. Mater. 2011, 186, 1586. although a region with a smaller number of residents, had collections in sites where only 10-21% of the sewage is treated and around 160 thousand people lives. A study1616 Pompei, C. M. E.; Campos, L. C.; da Silva, B. F.; Fogo, J. C.; Vieira, E. M.; Chemosphere 2019, 226, 210. reported 0.62 µg L-1 of MP from Lobo reservoir in Itirapina, São Paulo State a value much higher than the one found for urban water in this study and region. No quantification results were observed on South Wales rivers, UK,1616 Pompei, C. M. E.; Campos, L. C.; da Silva, B. F.; Fogo, J. C.; Vieira, E. M.; Chemosphere 2019, 226, 210.,6565 Huang, C.; Wu, L. H.; Liu, G. Q.; Shi, L.; Guo, Y.; Arch. Environ. Contam. Toxicol. 2018, 75, 224. possibly because the lowest LOQ of their method was 0.2 µg L-1, over 65 times higher than that presented in our study and likewise their wastewater passed by activated sludge treatment that was reported with higher efficiency in the removal of parabens (> 99% efficiency).1616 Pompei, C. M. E.; Campos, L. C.; da Silva, B. F.; Fogo, J. C.; Vieira, E. M.; Chemosphere 2019, 226, 210.,6565 Huang, C.; Wu, L. H.; Liu, G. Q.; Shi, L.; Guo, Y.; Arch. Environ. Contam. Toxicol. 2018, 75, 224. Our finds showed positive results of MP at 20.0 ng L-1 in sample site 3, MP at 18.0 ng L-1 and BP at 4.1 ng L-1 in sample site 4 reporting closer results to Portugal6868 Jonkers, N.; Sousa, A.; Galante-Oliveira, S.; Barroso, C. M.; Kohler, H. P. E.; Giger, W.; Environ. Sci. Pollut. Res. 2010, 17, 834. that presented results of 17 ng L-1 for MP and 27 ng L-1 for PP.

Conclusions

Herein we report a method for quantification of methylparaben, ethylparaben, propylparaben and butylparaben in urban waters (São Carlos, SP, Brazil) by direct sample injection using a SPE-LC-MS/MS system. Quantification limits in the order of ng L-1 with good accuracy and precision were achieved using internal standard for each paraben.

Methylparaben was found at collection sites 2, 3 and 4 with values from lower than LOQ up to 20.0 ng L-1. Butylparaben was found only at site 4 at 4.1 ng L-1.

The main features of the reported method are analysis run of 9.5 min including the cleanup cycles and columns conditioning, lower collection volumes of urban water and solvent usage. The SPE cartridge used showed robustness allowing over 500 sample injections (800 µL each) with good chromatographic performance.

Acknowledgments

We acknowledge the Agency for the Improvement of Higher Education Personnel (CAPES) (finance code 001) for the financial support to the graduate program. The National Council for Scientific and Technological Development (CNPq) is also acknowledged for the grant 3025557/2018-0. The equipment and technical support provided by Agilent Brazil is also acknowledged.

References

  • 1
    Garner, N.; Siol, A.; Eilks, I.; Chem. Action 2014, 103, 38, available at https://25593d2a-1813-4818-aca3-d56dd316e2c9.filesusr.com/ugd/7faa15_9e074968b5ee427a8cdca17b0a0aaa87.pdf, accessed in December 2020.
    » https://25593d2a-1813-4818-aca3-d56dd316e2c9.filesusr.com/ugd/7faa15_9e074968b5ee427a8cdca17b0a0aaa87.pdf
  • 2
    Yoom, H.; Shin, J.; Ra, J.; Son, H.; Ryu, D.; Kim, C.; Lee, Y.; Sci. Total Environ. 2018, 634, 677.
  • 3
    Saunders, L. J.; Mazumder, A.; Lowe, C. J.; Environ. Toxicol. Chem. 2016, 35, 919.
  • 4
    Daughton, C. G.; Ternes, T. A.; Environ. Health Perspect 1999, 107, 907.
  • 5
    Błędzka, D.; Gromadzińska, J.; Wąsowicz, W.; Environ. Int. 2014, 67, 27.
  • 6
    Brausch, J. M.; Rand, G. M.; Chemosphere 2011, 82, 1518.
  • 7
    Ramaswamy, B. R.; Shanmugam, G.; Velu, G.; Rengarajan, B.; Larsson, D. G. J.; J. Hazard. Mater. 2011, 186, 1586.
  • 8
    Jonkers, N.; Kohler, H. P. E.; Dammshäuser, A.; Giger, W.; Environ. Pollut. 2009, 157, 714.
  • 9
    Kang, S.; Kim, S.; Park, J.; Kim, H. J.; Lee, J.; Choi, G.; Choi, S.; Kim, S.; Kim, S. Y.; Moon, H. B.; Kim, S.; Kho, Y. L.; Choi, K.; Sci. Total Environ. 2013, 461-462, 214.
  • 10
    Cabaleiro, N.; De la Calle, I.; Bendicho, C.; Lavilla, I.; TrAC, Trends Anal. Chem. 2014, 57, 34.
  • 11
    Li, X. J.; Tang, X.; Zhou, X. H.; Chen, M. S.; Dai, J. L.; Xiong, W. M.; Zou, Z. X.; Liu, H. J.; Lv, S. Y.; Lin, H.; Lian, M.; Appl. Mech. Mater. 2013, 333-335, 1993.
  • 12
    Terasaki, M.; Takemura, Y.; Makino, M.; Environ. Chem. Lett. 2012, 10, 401.
  • 13
    Madureira, T. V.; Barreiro, J. C.; Rocha, M. J.; Cass, Q. B.; Tiritan, M. E.; J. Chromatogr. A 2009, 1216, 7033.
  • 14
    Zhang, D.; Gersberg, R. M.; Ng, W. J.; Tan, S. K.; Environ. Pollut. 2014, 184, 620.
  • 15
    Soni, M. G.; Carabin, I. G.; Burdock, G. A.; Food Chem. Toxicol. 2005, 43, 985.
  • 16
    Pompei, C. M. E.; Campos, L. C.; da Silva, B. F.; Fogo, J. C.; Vieira, E. M.; Chemosphere 2019, 226, 210.
  • 17
    Marta-Sanchez, A. V.; Caldas, S. S.; Schneider, A.; Cardoso, S. M. V. S.; Primel, E. G.; Environ. Sci. Pollut. Res. 2018, 25, 14460.
  • 18
    Chen, Y.; Deng, P.; Xie, P.; Shang, R.; Wang, Z.; Wang, S.; Chemosphere 2017, 168, 1628.
  • 19
    Kang, H. M.; Kim, M. S.; Hwang, U. K.; Jeong, C. B.; Lee, J. S.; Chemosphere 2019, 226, 388.
  • 20
    Ribeiro, C.; Ribeiro, A. R.; Maia, A. S.; Gonçalves, V. M. F.; Tiritan, M. E.; Crit. Rev. Anal. Chem. 2014, 44, 142.
  • 21
    Caban, M.; Lis, E.; Kumirska, J.; Stepnowski, P.; Sci. Total Environ. 2015, 538, 402.
  • 22
    Liu, Z.-h; Lu, G.-n; Yin, H.; Dang, Z.; Littier, H.; Liu, Y.; TrAC, Trends Anal. Chem. 2015, 64, 149.
  • 23
    Hodgeson, J. W.; Collins, J.; Barth, R. E.; Method 552: Determination of Haloacetic Acids in Drinking Water by Liquid-Liquid Extraction, Derivatization, and Gas Chromatography with Electron Capture Detection; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1990, available at https://www.o2si.com/docs/epa-method-552.pdf, accessed in December 2020.
    » https://www.o2si.com/docs/epa-method-552.pdf
  • 24
    Munch, J. W.; Method 525.2: Determination of Organic Compounds in Drinking Water by Liquid-Solid Extraction and Capillary Column Gas Chromatography/Mass Spectrometry; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.epa.gov/sites/production/files/2015-06/documents/epa-551.1.pdf, accessed in December 2020.
    » https://www.epa.gov/sites/production/files/2015-06/documents/epa-551.1.pdf
  • 25
    Munch, J. W.; Hautman, D. P.; Method 551.1: Determination of Chlorination Disinfection Byproducts, Chlorinated Solvents, and Halogenated Pesticides/Herbicides in Drinking Water by Liquid-Liquid Extraction and Gas Chromatography with Electron-Capture Detection; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.epa.gov/sites/production/files/2015-06/documents/epa-525.2.pdf, accessed in December 2020.
    » https://www.epa.gov/sites/production/files/2015-06/documents/epa-525.2.pdf
  • 26
    Munch, J. W.; Method 515.2: Determination of Chlorinated Acids in Water Using Liquid-Solid Extraction and Gas Chromatography with an Electron Capture Detector; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.o2si.com/docs/epa-method-515.2.pdf, accessed in December 2020.
    » https://www.o2si.com/docs/epa-method-515.2.pdf
  • 27
    Munch, J. W.; Method 508: Determination of Chlorinated Pesticides in Water by Gas Chromatography with An Electron Capture Detector; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at https://www.o2si.com/docs/epa-method-508.pdf, accessed in December 2020.
    » https://www.o2si.com/docs/epa-method-508.pdf
  • 28
    Munch, J. W.; Method 506: Determination of Phthalate and Adipate Esters in Drinking Water by Liquid-Liquid Extraction or Liquid-Solid Extraction and Gas Chromatography with Photoionization Detection; Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 1995, available at http://www.cromlab.es/Articulos/Metodos/EPA/500/506.pdf, accessed in December 2020.
    » http://www.cromlab.es/Articulos/Metodos/EPA/500/506.pdf
  • 29
    Majors, E. R. In Sample Preparation Fundamentals for Chromatography; Agilent: Delaware, USA, 2013, available at https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf, accessed in December 2020.
    » https://www.agilent.com/cs/library/primers/Public/5991-3326EN_SPHB.pdf
  • 30
    Raynie, E. D.; LCGC North America 2016, 34, 174, available at https://www.chromatographyonline.com/view/trends-sample-preparation-4, accessed in December 2020.
    » https://www.chromatographyonline.com/view/trends-sample-preparation-4
  • 31
    Chirila, E. S.; Draghici, C.; Dobrinas, S. In Chemicals as Intentional and Accidental Global Environmental Threats; Simeonov, L.; Chirila, E. eds.; Springer: Brasov, 2006, p. 7.
  • 32
    Petrovic, M.; Farré, M.; de Alda, M. L.; Perez, S.; Postigo, C.; Köck, M.; Radjenovic, J.; Gros, M.; Barcelo, D.; J. Chromatogr. A 2010, 1217, 4004.
  • 33
    Ramos, L.; J. Chromatogr. A 2012, 1221, 84.
  • 34
    Poole, C. F.; Gunatilleka, A. D.; Sethuraman, R.; J. Chromatogr. A 2000, 885, 17.
  • 35
    Hennio, M.-C.; J. Chromatogr. A 1999, 856, 3.
  • 36
    Gilart, N.; Borrull, F.; Fontanals, N.; Marcé, R. M.; Trends Environ. Anal. Chem. 2014, 1, e8.
  • 37
    Fumes, B. H.; Silva, M. R.; Andrade, F. N.; Nazario, C. E. D.; Lanças, F. M.; TrAC, Trends Anal. Chem. 2015, 71, 9.
  • 38
    Poole, C. F.; TrAC, Trends Anal. Chem. 2003, 22, 362.
  • 39
    Barreiro, J. C.; Vanzolini, K. L.; Madureira, T. V.; Tiritan, M. E.; Cass, Q. B.; Talanta 2010, 82, 384.
  • 40
    Valls-Cantenys, C.; Scheurer, M.; Iglesias, M.; Sacher, F.; Brauch, H. J.; Salvadó, V.; Anal. Bioanal. Chem. 2016, 408, 6189.
  • 41
    Cerqueira, M. B. R.; Caldas, S. S.; Primel, E. G.; J. Chromatogr. A 2014, 1336, 10.
  • 42
    Bylda, C.; Thiele, R.; Kobold, U.; Volmer, D. A; Analyst 2014, 139, 2265.
  • 43
    Lopes, B. R.; Barreiro, J. C.; Cass, Q. B.; J. Pharm. Biomed. Anal. 2016, 130, 318.
  • 44
    Barciela-Alonso, M. C.; Otero-Lavandeira, N.; Bermejo-Barrera, P.; Microchem. J. 2017, 132, 233.
  • 45
    Cassiano, N. M.; Barreiro, J. C.; Cass, Q. B.; J. Braz. Chem. Soc. 2014, 25, 9.
  • 46
    Munch, J. W.; Method 528: Determination of Phenols in Drinking Water by Solid Phase Extraction and Capillary Column Gas Chromatography/Mass Spectrometry (GC/MS); Environmental Monitoring Systems Laboratory Office of Research and Development U.S. Environmental Protection Agency: Cincinnati, Ohio, 2000, available at https://www.epa.gov/sites/production/files/2015-09/documents/m_528.pdf, accessed in December 2020.
    » https://www.epa.gov/sites/production/files/2015-09/documents/m_528.pdf
  • 47
    Moghimi, A.; Hosseini, S. A.; Far, A. M.; Orient. J. Chem. 2015, 31, 263.
  • 48
    Gołębiowski, M.; Stepnowski, P.; Leszczyńska, D.; Chem. Pap. 2017, 71, 831.
  • 49
    Buszewski, B.; Szultka, M.; Crit. Rev. Anal. Chem. 2012, 42, 198.
  • 50
    Liska, I.; J. Chromatogr. A 1993, 655, 163.
  • 51
    Tlili, I.; Caria, G.; Ouddane, B.; Ghorbel-Abid, I.; Ternane, R.; Trabelsi-Ayadi, M.; Net, S.; Sci. Total Environ. 2016, 563-564, 424.
  • 52
    Kuklenyik, Z.; Calafat, A. M.; Barr, J. R.; Pirkle, J. L.; J. Sep. Sci. 2011, 34, 3606.
  • 53
    Heuett, N. V.; Ramirez, C. E.; Fernandez, A.; Gardinali, P. R.; Sci. Total Environ. 2015, 511, 319.
  • 54
    Fernández-Ramos, C.; Šatínský, D.; Solich, P.; Talanta 2014, 129, 579.
  • 55
    García-Galán, M. J.; Díaz-Cruz, M. S.; Barceló, D.; Talanta 2010, 81, 355.
  • 56
    Bones, J.; Thomas, K.; Nesterenko, P. N.; Paull, B.; Talanta 2006, 70, 1117.
  • 57
    Cass, Q. B.; Cassiano, N. M. In Cromatografia Líquida: Novas Tendências e Aplicações; Elsevier: Rio de Janeiro, Brazil, 2015.
  • 58
    MassHunter Optimizer, version B07; Agilent Inc., Santa Clara, CA, USA, 2014.
  • 59
    Smith, G.; Bioanalysis 2012, 4, 865.
  • 60
    González-Mariño, I.; Quintana, J. B.; Rodríguez, I.; Cela, R. S.; Rapid Commun. Mass Spectrom 2009, 23, 1756.
  • 61
    Carder, K. L.; Steward, R. G.; Harvey, G. R.; Ortner, P. B.; Limnol. Oceanogr. 1989, 34, 68.
  • 62
    Weigel, S.; Kallenborn, R.; Hühnerfuss, H.; J. Chromatogr. A 2004, 1023, 183.
  • 63
    Guenu, S.; Hennion, M. C.; J. Chromatogr. A 1996, 737, 15.
  • 64
    Lin, W. C.; Chen, H. C.; Ding, W. H.; J. Chromatogr. A 2005, 1065, 279.
  • 65
    Huang, C.; Wu, L. H.; Liu, G. Q.; Shi, L.; Guo, Y.; Arch. Environ. Contam. Toxicol. 2018, 75, 224.
  • 66
    Lu, J.; Mao, H.; Li, H.; Wang, Q.; Yang, Z.; Sci. Total Environ. 2017, 605-606, 1064.
  • 67
    Peng, X.; Yu, Y.; Tang, C.; Tan, J.; Huang, Q.; Wang, Z.; Sci. Total Environ. 2008, 397, 158.
  • 68
    Jonkers, N.; Sousa, A.; Galante-Oliveira, S.; Barroso, C. M.; Kohler, H. P. E.; Giger, W.; Environ. Sci. Pollut. Res. 2010, 17, 834.
  • 69
    Kasprzyk-Hordern, B.; Dinsdale, R. M.; Guwy, A. J.; Water Res. 2009, 43, 363.

Publication Dates

  • Publication in this collection
    30 Apr 2021
  • Date of issue
    May 2021

History

  • Received
    08 Sept 2020
  • Accepted
    21 Dec 2020
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br