Inter-session confidence of balance assessment with TOBtrainerMR

Abstracts

INTRODUCTION: A new device for assessing and training dynamic balance, the TOBtrainerMR, has been recently developed. OBJECTIVE: To asses the inter-session reliability of balance assessment using the TOBtrainerMR. The study has also the purpose of giving error scores so that true changes in performance can be identified. METHODS: Thirty healthy sedentary subjects (nine males, 21 females, age = 27.9 ± 2.9 years) participated in this study. The TOBtrainerMR was used to assess balance in the medial-lateral plane with eyes open (MLEO) and with eyes closed (MLEC), and in the anterior-posterior plane with eyes open (APEO) and with eyes closed (APEC). Subjects were instructed to be in a double-limb standing position with the hands at their sides trying to maintain platform stability. All the data were registered with the TOBT software. RESULTS: The samples were correlated, t-Tests assessed, learning effects and Intraclass Correlation Coefficients assessed in their reliability. Standard Errors of Measurement and Smallest Detectable Differences were calculated to assess measurement error. No significant differences between the first and the second session measurements were found. The Intraclass Correlation Coefficients, the Standard Errors of Measurement and the Smallest Detectable Differences for the four modalities measured ranged from 0.71 to 0.83, 0.32º to 0.80º and 0.90º to 2.22º, respectively. CONCLUSIONS: Inter-session reliability for balance assessment using the TOBtrainerMR was good. Future researchers have now reference data to evaluate whether differences in two different scores are real or due to measurement error and what changes are needed in a subject's score to be sure that a real change has occurred.

balance; postural control; Intraclass Correlation Coefficient


INTRODUÇÃO: Um novo dispositivo para a avaliação e equilíbrio dinâmico de formação, o TOBtrainerMR, foi recentemente desenvolvido. OBJETIVO: Avaliar a confiabilidade entre sessões de avaliação de equilíbrio usando o TOBtrainerMR. O estudo também tem a finalidade de dar contagens de erro para que verdadeiras mudanças no desempenho possam ser identificadas. MÉTODOS: Trinta indivíduos sedentários saudáveis (nove homens, 21 mulheres com idade de 27,9 ± 2,9 anos) participaram deste estudo. O TOBtrainerMR foi usado para avaliar o equilíbrio no plano medial-lateral com os olhos abertos (MLOA) e com os olhos fechados (MLOF) e no plano anteroposterior com os olhos abertos (APOA) e com os olhos fechados (APOF). Os indivíduos foram instruídos para estar em posição ereta com braços ao longo do corpo e mãos lateralmente tentando manter a estabilidade da plataforma. Todos os dados foram registrados com o software TOBT. RESULTADOS: As amostras foram relacionadas, testes t avaliados, efeitos de aprendizagem e coeficientes de correlação intracIasse avaliados na sua confiabilidade. Erros padrão de medição e menores diferenças detectáveis foram calculadas para avaliar o erro de medição. Não foram encontradas diferenças significativas de medição entre a primeira e a segunda sessão. Os coeficientes de correlação intraclasse, os erros padrão de medição e as menores diferenças detectáveis para as quatro modalidades medidas variavam de 0,71 0,83, 0,32º a 0,80º e 0,90º a 2,22º, respectivamente. CONCLUSÕES: A confiabilidade entre sessões para avaliação de equilíbrio utilizando o TOBtrainerMR foi boa. Futuros investigadores têm agora dados de referência para avaliar se as diferenças de duas diferentes pontuações são reais ou se devem ao erro de medição e que mudanças são necessárias na pontuação do sujeito para se ter certeza de que ocorreu uma mudança real.

equilíbrio; controle postural; confiabilidade


ORIGINAL ARTICLE

OTHER

Inter-session confidence of balance assessment with TOBtrainerMR

Claudio Oyarzo MauricioI; Mercedes Schmitt RungueI; Roberto LarraguibelII; Daniel Rojano OrtegaIII; Francisco José Berral de la RosaIII

IFinis Terrae University, Santiago, Chile

IIARTOficio, Santiago, Chile

IIIPablo de Olavide University, Seville, Spain

Mailing address

ABSTRACT

INTRODUCTION: A new device for assessing and training dynamic balance, the TOBtrainerMR, has been recently developed.

OBJECTIVE: To asses the inter-session reliability of balance assessment using the TOBtrainerMR. The study has also the purpose of giving error scores so that true changes in performance can be identified.

METHODS: Thirty healthy sedentary subjects (nine males, 21 females, age = 27.9 ± 2.9 years) participated in this study. The TOBtrainerMR was used to assess balance in the medial-lateral plane with eyes open (MLEO) and with eyes closed (MLEC), and in the anterior-posterior plane with eyes open (APEO) and with eyes closed (APEC). Subjects were instructed to be in a double-limb standing position with the hands at their sides trying to maintain platform stability. All the data were registered with the TOBT software.

RESULTS: The samples were correlated, t-Tests assessed, learning effects and Intraclass Correlation Coefficients assessed in their reliability. Standard Errors of Measurement and Smallest Detectable Differences were calculated to assess measurement error. No significant differences between the first and the second session measurements were found. The Intraclass Correlation Coefficients, the Standard Errors of Measurement and the Smallest Detectable Differences for the four modalities measured ranged from 0.71 to 0.83, 0.32� to 0.80� and 0.90� to 2.22�, respectively.

CONCLUSIONS: Inter-session reliability for balance assessment using the TOBtrainerMR was good. Future researchers have now reference data to evaluate whether differences in two different scores are real or due to measurement error and what changes are needed in a subject's score to be sure that a real change has occurred.

Keywords: balance, postural control, Intraclass Correlation Coefficient.

INTRODUCTION

Balance is a basic human need which allows keeping the body weight center on the support basis1-3. Balance control is generally defined as a complex motor ability which involves the integration of sensory responses, elaboration of a response of the central nervous system and the motor capacity to perform this response so that the posture goals can be reached1,3. Balance disorders in the elderly have been widely studied1-8. Moreover, there are many studies about balance in the active and sports population which establish relations with injury prevention and recovery9-12.

Statically speaking, balance may be defined as the ability to keep a support basis with minimum movement, while dynamically speaking, it can be the ability to perform a task during the maintenance of a stable position13,14. Many tests have been developed to assess static and dynamic balance. Some of them are especially used with the elderly and have proved to be useful in prevention of falls. Examples of these tests are the Romberg test, the Tinetti test, the Stand Up and Run test, Functional reach test and the Bestest test15-20. Some others, such as the Cartwheel test12,16 are used with a more active population. There are also more sophisticated systems which assess balance quantitative and qualitatively, such as the AMTI AccuSway Plus Force Platform, the NeuroCom EquiTest, the Biodex Stability test, the Trainer 2000 Kinestetic Ability and the Balance Master20-26.

Dynamic balance conditions have been investigated with the use of some of these systems, such as the Biodex Stability System, generating instability through the instrument itself13. In this flow of thinking, a new instrument for dynamics balance assessment and training, the TOBtrainerMR, has been developed and patented by Pablo de Olavide at the University of Seville, Spain. The TOBtrainerMR is at least three times less costly than the majority of the instruments previously mentioned here, besides being easy to be carried (it weighs only 7 kg).

The aim of the present study is to assess inter-session confidence of balance assessment using the TOBtrainerMR. The evaluator's participation limited to providing instructions to the subjects, which makes the evaluation almost independent from the evaluator. Moreover, the authors have the aim to offer error scores so that the actual performance alterations can be identified.

METHODS

Subjects

Thirty healthy subjects (nine men, 21 women, age = 27.9 ± 2.9 years) with no involvement in physical activity over two times per week and absence of history of having been considered elite athletes, participated in this study. None of them have reported lower extremity or spine injury in the last previous six months. None of them have suffered central nervous, visual or vestibular systems disorder or have ingested any drug which could alter balance. All subjects provided written consent for participation in the study, according to the Declaration of Helsinki Revised in 2008.

Instruments

The TOBtrainerMR is a 40-cm fiberglass platform with 12 levels of unidirectional inclination, which can be medial-lateral um (ML) or anteroposterior um (AP), depending on the subject's position on the platform. It has three radius of curvature (25 cm, 15 cm and 10 cm) for increase of the difficulty level and can also evaluate balance in any position.

The TOBtrainerMR has an accelerometer (ADXL105) which may be connected to two microcomputers, PIC16F877A or PIC16F876A. The first has a sampling rate of 20 Hz, is used without a computer, has two modalities (assessment and training) and presents the information on the instrument's display (figure 1). The second one is only used for assessment, has a sampling rate of 40 Hz and needs a computer attached to it. Both micro-controllers can be used with a visual reference to aid the subject improve his/her balance.


Testing procedure

The test protocol consisted of two sessions separated by a time interval of three weeks. Each session consisted of three 20-second tests for each modality, using a radius of curvature of 15 cm (medium difficulty level). The four modalities were medial-lateral with eyes open (MLEO), medial-lateral with eyes closed (MLEC), anteroposterior with eyes open (APEO) and anteroposterior with eyes closed (APOF). The tests were performed in the following order: MLEO, MLEC, APEO and APEC.

All tests were performed in a closed room with good light. There were no visual or sound distractions. The subjects were then instructed to remain at standing position on the two limbs with hands along the body trying to maintain stability on the platform. Therefore, they were helped by a visual reference which consisted of a light signal which remained in the middle of the screen when the platform was at the horizontal plane. In case the subjects were not able to maintain bilateral position on the platform, the attempt was discarded.

Before the beginning of the tests, the protocol was explained to all subjects and they performed a 20-second familiarization test immediately before the test of each modality.

A PIC16F876 micro controller was used so that the sampling rate was set at 40 Hz. All data were recorded with the TOBT software. The mean value of the angle bias was admitted for each test and the best result out of the three used tests for each modality was applied. A 60-second interval was given between tests, while a five-minute one was given between modalities.

Statistical analysis

Statistical analysis used the SPSS program for Windows, version 17.0. The means and standard deviations of all variables have been measured. The Kolmogorov-Smirnov test was applied to verify normality of all variables, and as this condition was always verified, a t test for paired samples was used for comparison of the first and the second measurement with the aim to discard any systematic differences due to the learning effects. The significance level of the test was set at p < 0.05. A Two-way intraclass correlation coefficient model (2,1) (ICC) was used to evaluate confidence. The following criteria were established for the ICC values27:

- Bad: ICC < 0.40;

- Average: 0.40 < ICC < 0.70;

- Good: 0.70 < ICC < 0.90;

- Excellent: ICC > 0.90.

From the ICC values, we calculated the standard error measurement (SEM) and the smallest detectable difference (SDD) through the following formulas28:

Where SD is standard deviation of the scores of all subjects.

RESULTS

Table 1 presents the mean and standard deviation for the four modalities in both sessions. The t tests of related samples did not reveal differences between measurements from the first and second sessions (p < 0.01), evidencing hence that there were no differences between both sessions due to learning effects. The ICC values for confidence assessment between sessions 1 and 2 ranged between 0.71 (for MLEC) and 0.83 (for MLEO). Table 2 presents the ICC, SEM and SDD values for all the measured variables.

DISCUSSION

Concerning our dynamics balance assessment, the ICC values found for MLEO, MLEC, APEO and APEC ranged from 0.71 and 0.83, which means that confidence after three weeks of interval was good. Such results are according to the ones found by Mattacola et al.24, who obtained an ICC of 0.78 for the dynamics protocol of two limbs with eyes open and of 0.84 for the dynamic protocol of two limbs with eyes closed using the Chattecx Balance System. The SEM describes within which variation the actual score of a subject will lie as a result of an impossibility of confidence of the assessment, The SDD is an index which shows the necessary difference between scores separated of a subject so that the difference in these scores is considered real. The ICC is without an unit, but the SEM and the SDD present the same units of measure under interest, and we believe that these are more useful than the ICC itself. An important limitation of our study is that the quantity of tested conditions was very low. Balance assessment should also consider the way these balance strategies alter with support and sensory conditions, as well as limitations of the task. Moreover, these results were obtained from a healthy and young population and it is not clear whether the results would have been similar with an older population or another with balance disorders. Another possible limitation is the choice of the radius of curvature for the adjustment of the level of difficulty. Our subjects were not involved in physical activity more frequent than twice a week; therefore, it would have been possible to find higher ICC values using a radius of curvature of 25 cm. Some authors 29,30 reported good confidence when used a steady stability level with the Biodex, but Pereira et al.25 state that the decreasing stability protocol could have been a better option for subjects who are engaged in sports, since it is difficult to choose which stability level better represent its reality. The use of a similar protocol with the TOBTrainerMR could also be the best way to assess balance in sports populations.

CONCLUSION

Although there is certain variability in the balance assessment using the TOBtrainerMR, the ICC values found in our study present good confidence. Since there are no previous studies with the TOBtrainerMR, there are no SEM and SDD values either. Future researchers would have, from now on, reference data to assess within which variation an actual score of a subject will lie and whether differences in two scores are real or due to an error in measurement.

REFERENCES

  • 1. Horak F. Clinical assessment of balance disorders. Gait Posture 1997;6:76-84.
  • 2. Peterka R, Loughlin P. Dynamic Regulation of Sensorimotor Integration in Human Postural Control. J Neurophysiol 2004;91:410-23.
  • 3. Winter DA. Human balance and posture control during standing and walking. Gait Posture 1995;3:193-214.
  • 4. Castillo A, D'Andréa J, Camanho, G. Evaluating the Center of Gravity of Dislocations in Soccer Players With and Without Reconstruction of the Anterior Cruciate Ligament using a Balance Platform. Clinics 2009;64:163-70.
  • 5. Gribble PA, Hertel J, Denegar CR, Buckley WE. The Effects of Fatigue and Chronic Ankle Instability on Dynamic Postural Control. J Athl Train 2004;39:321-9.
  • 6. Luoto BM. One-footed and externally disturbed two-footed postural control in patients with chronic low back pain and healthy control subjects. Spine 1998;23:2081-9.
  • 7. Mok NW, Brauer SG, Hodges PW. Hip strategy for balance control in quiet standing is reduced in people with low back pain. Spine 2004;11:107-12.
  • 8. Prioli A, Freitas Júnior P, Barela J. Physical Activity and Postural Control in the Elderly: Coupling between Visual Information and Body Sway. Gerontology 2005;51:145-8.
  • 9. Emery CA, Meeuwisse WH. The effectiveness of a neuromuscular prevention strategy to reduce injuries in youth soccer: a cluster-randomised controlledtrial. Br J Sports Med 2010;44:555-62.
  • 10. Mcguide T, Keene J. The Effect of a Balance Training Program on the Risk of Ankle Sprains in High School Athletes. Am J Sports Med 2006;34:1003-111.
  • 11. Pasanen K, Parkkari J, Pasanen M, Hiiloskorpi H, Mäkinen T, Järvinen M, et al. Neuromuscular training and the risk of leg injuries in female floorball players: cluster randomised controlled study. Br J Sports Med 2008;42:802-6.
  • 12. Plisky PJ, Rauh MJ, Kaminski TW, Underwood FB. Star Excursion Balance Test as a Predictor of Lower Extremity Injury in High School Basketball Players. J Orthop Sports Phys Ther 2006;36:911-9.
  • 13. Ricotti L. Static and dynamic balance in young athletes. J Hum Sport Exerc 2011;6:616-28.
  • 14. Winter DA, Patla AE, Frank JS. Assessment of balance control in humans. Med Prog Technol 1990;16:31-51.
  • 15. Duncan PW, Weiner DK, Chandler J, Studenski S. Functional Reach: a new clinical measure of balance. J Gerontol 1990;45:192-7.
  • 16. Gribble P, Hertel J. Considerations for Normalizing Measures of the Star Excursion Balance Test. Meas Phys Educ Exerc Sci 2003;7:89-100.
  • 17. Horak FB, Wrisley DM, Frank J. The Balance Evaluation System Test (BESTest) to differentiate balance deficits. Phys Ther 2009:89:484-98.
  • 18. Tinetti ME. Performance-oriented assessment of mobility problems in elderly patients. J Am Geriatr Soc 1986;24:119-26.
  • 19. Whitney J, Lord S, Close J. Streamling assessment and intervention in a falls clinic using the Timed Up and Go Test and Physiological Profile Assessment. Age Ageing 2005;34:567-71.
  • 20. Yim-Chiplins P, Talbot L. Defining and Measuring Balance in Adults. Biol Res Nurs 2000;1:321-31.
  • 21. Arnold BL, Schmitz RJ. Examination of Balance Measures Produces by Biodex Stability System. J Athl Train 1998;33:323-7.
  • 22. Artuso A, Garozzo A, Contucci A, Frenguelli A, Di Girolamo S. Role of dynamic posturography (Equitest) in the identification of feigned balance disturbances. Acta Otorhinolaryngol Ital 2004;24:8-12.
  • 23. Ekdahl C, Jarnlo GB, Andersson SI. Standing balance in healthy subjects. Evaluation of a quantitative test battery on a force platform. Scand J Rehabil Med 1989;21:187-95.
  • 24. Mattacola CG, Lebsack DA, Perrin DH. Intertester Reliability of Assessing Postural Sway Using the Chattecx Balance System. J Athl Train 1995;30:237-42.
  • 25. Pereira HM, De Campos TF, Santos MB, Cardoso JR, Garcia MC, Cohen M. Influence of knee position on the postural stability index registered by the Biodex Stability System. Gait Posture 2008;28:668-72.
  • 26. Walker C, Brouwer B, Gulham E. Use of Visual Feedback in Retraining Balance Following acute Stroke. Phys Ther 2000;80:886-95.
  • 27. Munro AG, Herrington LC. Between-session reliability of the star excursion balance test. Phys Ther Sport 2010;11:128-32.
  • 28. Weir JP. Quantifying Test-Retest Reliability Using the Intraclass Correlation Coefficient and the SEM. J Strength Cond Res 2005;19:231-40.
  • 29. Cachupe W, Shifflett B, Kahanov L, Wulghalter E. Reliability of biodex balance system measures. Meas Phys Educ Exerc Sci 2001;9:240-52.
  • 30. Pincivero D, Lephart SM, Henry TJ. Learning effects and reliability of the biodex stability system. J Athl Train 1995;30:S35.
  • Correspondência:

    Daniel Rojano Ortega
    Universidade Pablo de Olavide
    Sevilha, Espanha
    c/ María Fulmen
    Edificio Los Granados, bloque 2 1° E
    41019-Sevilla, Espanha
    E-mail:
  • Publication Dates

    • Publication in this collection
      10 Dec 2013
    • Date of issue
      Oct 2013

    History

    • Received
      06 June 2012
    • Accepted
      07 Dec 2012
    Sociedade Brasileira de Medicina do Exercício e do Esporte Av. Brigadeiro Luís Antônio, 278, 6º and., 01318-901 São Paulo SP, Tel.: +55 11 3106-7544, Fax: +55 11 3106-8611 - São Paulo - SP - Brazil
    E-mail: atharbme@uol.com.br