Abstract
Introduction The dental implant rehabilitation protocol is a well-established treatment option for edentulous patients with a high success rate. However, there are factors that can lead to difficulty in continuation and even complete loss of rehabilitation. Microgaps are spaces found between the implant and prosthetic abutment that are caused by the limit of precision in the manufacturing of the implant. This space can cause micromovements and bacterial microleakage that can compromise the long-term useful life of the implant.
Objective Measure and evaluate microgaps in angled and straight implants using scanning electron microscopy. To carry out the study, a total of 30 SIN® osseointegrated implants were used, 15 of which were straight and 15 angled.
Material and method For the comparative analysis of these components, scanning electron microscopy was used, carried out by a properly calibrated and experienced researcher.
Result It was observed that the microgap measurements of both components are in accordance with the clinically acceptable values presented in the literature, however the microgap values of the angled components were considerably higher compared to the straight components.
Conclusion Although the values presented corroborate the data presented in the literature, additional studies are necessary for a more comprehensive and in-depth understanding of the relationship between the microgaps of the morse cone system components.
Descriptors: Dental implants; dental implant-abutment design; dentistry
Resumo
Introdução O protocolo de reabilitação com implantes dentários é uma opção de tratamento bem estabelecida para pacientes desdentados com alta taxa de sucesso. No entanto, existem fatores que podem levar à dificuldade de continuidade e até mesmo à perda completa da reabilitação. Microgaps são espaços encontrados entre o implante e o pilar protético que são causados pelo limite de precisão na fabricação do implante. Esse espaço pode causar micromovimentos e microinfiltração bacteriana que podem comprometer a vida útil do implante a longo prazo.
Objetivo Medir e avaliar microgaps em implantes angulados e retos usando microscopia eletrônica de varredura. Para a realização do estudo, foram utilizados um total de 30 implantes osseointegrados SIN®, sendo 15 retos e 15 angulados.
Material e método Para a análise comparativa desses componentes, foi utilizada a microscopia eletrônica de varredura, realizada por pesquisador devidamente calibrado e experiente.
Resultado Observou-se que as medidas de microgaps de ambos os componentes estão de acordo com os valores clinicamente aceitáveis apresentados na literatura, porém os valores de microgaps dos componentes angulados foram consideravelmente maiores em comparação aos componentes retos.
Conclusão Embora os valores apresentados corroborem os dados apresentados na literatura, estudos adicionais são necessários para uma compreensão mais abrangente e aprofundada da relação entre os microgaps dos componentes do sistema cone morse.
Descritores: Implantes dentários; projeto do implante dentário-pivô; odontologia
INTRODUCTION
Dental implant rehabilitation has become a widely accepted and highly successful treatment option for edentulous patients1-5. As a result, various types of implants and therapeutic protocols have been developed. However, several factors can lead to the failure of the rehabilitation protocol. These include surgical trauma, peri-implant diseases, microleakage, variation in implant anatomy in the crest area, patient age, genetics, systemic conditions, and oral hygiene6-8.
Due to its ability to protect the implant from unwanted loads during the bone consolidation phase, the two-piece implant protocol (implant and prosthetic abutment) is widely used. However, the implant and abutment cannot be perfectly aligned due to limitations in precision during their production9. As a result, a gap is created between the prosthetic abutment and the implant (implant-abutment interface). This gap can be a source of micromovements and bacterial microleakage, allowing microorganisms to pass freely between the oral cavity and the internal cavity of the implant1,2,4-6,8-13. The presence of bacteria is closely related to inflammatory processes that can lead to mucositis, peri-implantitis, and eventually bone loss1-14.
The health of the soft tissues and the size of the implant-abutment interface strictly depend on the material of the prosthetic abutment, as well as its design, surface topography, and preparation8. Currently, most implants are made of titanium and its alloys8-15, and according to Liu, Yang9, the microgap in titanium implants is considerably smaller than in zirconia implants.
In different Morse cone implant systems, the degree of conicity and the connection area vary, which are primarily responsible for differences in bacterial penetration. The applied torque value is also important. Generally, a large connection area results in a small degree of conicity, and a high torque value translates to a low level of bacterial microleakage9.
The most recommended technique for evaluating and measuring the implant-abutment interface is scanning electron microscopy (SEM)6. This technique provides high-resolution images with great depth of field of the object while maintaining a fixed and predetermined position6. Studies assessing microgaps in SIN implants are scarce in the literature. Therefore, the aim of this work was to evaluate and measure microgaps at the interface using SEM.
MATERIAL AND METHOD
This laboratory study evaluated the interface between the dental implant and the prosthetic abutment in order to measure the microgaps. It was conducted at the Department of Dentistry at the Federal University of Sergipe (UFS, Brazil), in collaboration with the Department of Physics and Materials Engineering.
Sample Characterization
A total of 30 osseointegrated SIN® implants (São Paulo, Brazil) were used, made of commercially pure titanium (c. p Ti), conforming to the NBR ISO 5832 standard16, sized 3.8 x 11.5 mm with a Morse Cone prosthetic system. The samples were subdivided into two groups: 15 were coupled with straight abutments of AIMP 4003C-H 4.0 mm, and 15 were coupled with angled abutments of AIAM 4003C-H 4.0 mm and 3.0 mm.
Sample Analysis
For the comparative analysis of the microgaps in the samples, scanning electron microscopy (SEM) (JSM-6510 LV, JEOL USA, Inc) with an acceleration voltage of 5 kV was used. Measurements were taken by a single, properly calibrated, and experienced researcher. SEM magnifications of 4,000 times were used, with captures obtained at the micrometer (μm) scale. Five areas (A1-A5) around the circumference between the component and the implant were analyzed. For each area, five measurements were taken, resulting in 150 measurements tabulated in Excel 16.0 (Microsoft, Washington, USA).
Statistical Analysis
The Shapiro-Wilk normality test was applied, and for comparison between groups, the Mann-Whitney test for independent variables was used. The significance level for all tests was set at 5% (p ≤ 0.05). Data were analyzed using Bioestat 5.0 software17.
RESULT
After the microscopy analyses, the microgaps were measured in micrometers and tabulated as shown in Figure 1. Figure 2 demonstrates a comparison of the size of microgaps from the t-test results, showing that the angled components have larger microgap values than the straight components. The sample group of angled components (CA) and the sample group of straight components (CR) were analyzed separately and comparatively.
SEM analysis of the straight prosthetic component in different regions demonstrating the microgap between prosthetic abutment and implant in a magnificence of 4000 times.
Table 1 shows the results of the Shapiro-Wilk test and presents the mean microgap values separated by groups (A1 to A5), their standard deviation, and the p-value. It is notable that the mean values at the ends of both components are higher than their average values, but the discrepancy is more pronounced in the straight components. Regarding standard deviation values, a similar pattern to the mean microgaps is observed; however, in angled components, the standard deviation values are more standardized regardless of the region.
Table 2 shows the results of the normality Lilliefors test p-value and presents that the p-values are <0.05, then the data follow a normal distribution. Table 3 shows the results of the Mann-Whitney test for independent samples, which present different p-values for each sample. The p-value of the one-tailed test suggests a significant difference between the samples, indicating that one tends to be larger or smaller than the other. The p-value of the two-tailed test is not small enough to indicate a significant difference in general between the sample distributions.
DISCUSSION
The long-term success of dental implant rehabilitation demonstrates the effectiveness of this treatment option1-5. However, risks such as material manufacturing defects, surgical errors, and anatomical variations can impede the success of the rehabilitation6-8. Therefore, the connection between the implant and the prosthetic abutment should be as closely aligned as possible to avoid large gaps, thereby preventing micromovements and bacterial microleakage1-2,4-6,8-13. In this context, the images produced in our study using SEM reveal the presence of minimal microgaps in SIN implants, both in straight and angled components.
In the Morse Cone connection, fixation and stability depend not only on the fixation screw but also on the friction between the conical parts and the surface of the component18. Emphasizing the importance of the absence of microgaps at the implant/component interface, our study presents the quantification of existing microgaps and their variations within the same manufacturer.
This study demonstrates higher average microgap values in the uppermost region (A1) and the lowermost region (A5) for both straight and angled components, with a greater discrepancy between the values at the extremities (A1 and A5) and the central regions (A2, A3, and A4) in straight components. These values contradict the findings of Duraisamy et al.19, who observed smaller values in the upper region and larger values in the lower region, which could be explained by differences in implant designs and brands. Lopes et al.20, in their study of Neodent® implants, reported average values similar to the straight components in our research. However, the maximum values in our study are relatively higher, regardless of the component type. On the other hand, the average values reported by Costa et al.6 differ from those presented in our study, with differences in implant design and brands likely being a determining factor for this discrepancy.
After analyzing the statistical data, it can be observed that the microgaps present in this study are within clinically acceptable standards for the joining of components. Aspects such as the precision and stability of the connection between components and implants have been the focus of research by manufacturers aiming to improve the quality of mechanical parts through enhanced machining processes, resulting in greater precision. Additionally, investments in materials that support or minimize screw loosening have been reported in the literature18.
Mohammadi et al.11 state that when the abutment is connected to the implant, gaps between the components are inevitable and can become potential sites for bacterial microleakage, which may lead to the infiltration of inflammatory cells. This inflammatory process around the implant can cause peri-implantitis and even bone loss, ultimately affecting the long-term success of the rehabilitation21,22. However, studies by Kowalski et al.8, Duraisamy et al.19, Jemt, Book23, and Solá-Ruíz et al.24 demonstrate that microgaps of less than 10 micrometers do not have harmful effects on either soft or hard tissues. Comparing with our study, it can be concluded that the average microgap values are satisfactory since they are below 10 micrometers. Additionally, due to the stability of the Morse cone system, oxidation between the components is observed in the gap region, similar to cold welding, which sometimes acts as an effective physical seal, preventing bacterial proliferation6.
When comparing the average values between these components, a notable disparity is observed in the central regions (A2, A3, and A4), where the values for angled components are significantly higher. This discrepancy can be attributed to the unique design characteristics of these components. Conversely, the average values at the extremities (A1 and A5) between the straight and angled components are similar. Additionally, we identified that some of the analyzed regions did not exhibit parametric distribution, suggesting the possibility of irregularities that could affect the implant adaptation.
In the two-piece implant system, although micromovement in conical connections decreases due to precise manufacturing of the implant and abutment, the current production process cannot eliminate micromovement entirely. A limitation of this study include that the microgap was evaluated in the absence of loading mechanics. Future studies should assess the microgap under the combined application of dynamic mechanical testing and fatigue, where micromovements produced by chewing could lead to significant micro-adaptations between the abutment and implant that are important for clinical practice.
CONCLUSION
In light of the above, we can conclude that microgaps are present in both straight and angled components, and they are within clinically acceptable limits. The results show higher microgap values in the angled component group, suggesting conformational changes related to manufacturing. Additional studies are needed for a more comprehensive and in-depth understanding of the relationship between microgaps in Morse cone system components.
ACKNOWLEDGEMENTS
This research utilized facilities at the Multi-User Nanotechnology Center of UFS (CMNano-UFS), a National Multi-User Research Center supported by the Funding Authority for Studies and Projects (FINEP). We thank the technical support provided by the CMNano-UFS teams during the experiments of proposal 085/2023. The provision of implant samples was granted by SINGULAR Implants®. This work was supported by the Coordination for the Improvement of Higher Education Personnel-Brazil (CAPES) - Funding Code 001.
-
How to cite:
Lobo YL, Mendes GN, Santana LAM, Floresta LG, Marqueti AC, Takeshita WM, et al. Comparative analysis of microgaps in angled and straight components: a laboratory study. Rev Odontol UNESP. 2024;53:e20240024. https://doi.org/10.1590/1807-2577.02424
REFERENCES
-
1 Cascos R, Celemín-Viñuela A, Mory-Rubiños N, Gómez-Polo C, Ortega R, Agustín-Panadero R, et al. Influence of the use of transepithelial abutments vs. titanium base abutments on microgap formation at the dental implant–abutment interface: an in vitro study. Materials (Basel). 2023;16(19):6532. http://doi.org/10.3390/ma16196532 PMid:37834669.
» http://doi.org/10.3390/ma16196532 -
2 Carlos LV, Carlos NC, Sm Karina L, Sunil BK, Carlos PE, Olga LG. Comparative study of bacterial microfiltration in the implant-abutment interface, with straight and conical internal connections, in vitro. Clin Exp Dent Res. 2021 Dec;7(6):1014-24. http://doi.org/10.1002/cre2.439 PMid:34151544.
» http://doi.org/10.1002/cre2.439 -
3 Baseri M, Radmand F, Hamedi R, Yousefi M, Kafil HS. Immunological aspects of dental implant rejection. BioMed Res Int. 2020 Dec;2020:7279509. http://doi.org/10.1155/2020/7279509 PMid:33376734.
» http://doi.org/10.1155/2020/7279509 -
4 Khajavi A, Mohseni S, Peymani A, Amjadi M. In vitro bacterial leakage at the implant-abutment connection of two dental implant systems with internal connection. Front Dent. 2020 Dec;17:32. http://doi.org/10.18502/fid.v17i32.5196 PMid:36042810.
» http://doi.org/10.18502/fid.v17i32.5196 -
5 Vélez J, Peláez J, López-Suárez C, Agustín-Panadero R, Tobar C, Suárez MJ. Influence of implant connection, abutment design and screw insertion torque on implant-abutment Misfit. J Clin Med. 2020 Jul 24;9(8):2365. https://doi.org/10.3390/jcm9082365 PMid: 32722131.
» https://doi.org/10.3390/jcm9082365 -
6 Costa MB, Ferreira LF, Takeshita WM, Marqueti AC, Trento CL. Evaluation of the interface between the straight prosthetic component and the Morse Cone-type internal connection of the dental implant in cross section by scanning electron microscopy. Rev Odontol UNESP. 2020;49:e20200033. http://doi.org/10.1590/1807-2577.03320
» http://doi.org/10.1590/1807-2577.03320 - 7 Ertem SY, Gungormus M, Ozdogan MS, Orhan M. A microbiological assessment of peri-implant sites and implant-abutment interfaces in diabetic and healthy individuals. Int J Clin Exp Med. 2020;13(1):208-15.
-
8 Kowalski J, Puszkarz AK, Radwanski M, Sokolowski J, Cichomski M, Bourgi R, et al. Micro-CT evaluation of microgaps at implant-abutment connection. Materials (Basel). 2023 Jun;16(12):4491. http://doi.org/10.3390/ma16124491 PMid:37374674.
» http://doi.org/10.3390/ma16124491 -
9 Liu Y, Wang J. Influences of microgap and micromotion of implant-abutment interface on marginal bone loss around implant neck. Arch Oral Biol. 2017 Nov;83:153-60. http://doi.org/10.1016/j.archoralbio.2017.07.022 PMid:28780384.
» http://doi.org/10.1016/j.archoralbio.2017.07.022 -
10 Mao Z, Beuer F, Wu D, Zhu Q, Yassine J, Schwitalla A, et al. Microleakage along the implant-abutment interface: a systematic review and meta-analysis of in vitro studies. Int J Implant Dent. 2023 Sep;9(1):34. http://doi.org/10.1186/s40729-023-00494-y PMid:37733145.
» http://doi.org/10.1186/s40729-023-00494-y -
11 Mohammadi F, Hajmousaei M, Vaziri N, Arshad M. Bacterial leakage at implant-abutment interface with different intermediate materials. J Oral Implantol. 2019 Dec;45(6):451-5. http://doi.org/10.1563/aaid-joi-D-18-00313 PMid:31580765.
» http://doi.org/10.1563/aaid-joi-D-18-00313 -
12 Molinero-Mourelle P, Cascos-Sanchez R, Yilmaz B, Lam WYH, Pow EHN, Del Río Highsmith J, et al. Effect of fabrication technique on the microgap of CAD/CAM cobalt-chrome and zirconia abutments on a conical connection implant: an in vitro study. Materials (Basel). 2021 Apr;14(9):2348. http://doi.org/10.3390/ma14092348 PMid:33946477.
» http://doi.org/10.3390/ma14092348 -
13 Vinhas AS, Aroso C, Salazar F, López-Jarana P, Ríos-Santos JV, Herrero-Climent M. Review of the mechanical behavior of different implant-abutment connections. Int J Environ Res Public Health. 2020 Nov;17(22):8685. http://doi.org/10.3390/ijerph17228685 PMid:33238476.
» http://doi.org/10.3390/ijerph17228685 -
14 Gehrke P, Burg S, Peters U, Beikler T, Fischer C, Rupp F, et al. Bacterial translocation and microgap formation at a novel conical indexed implant abutment system for single crowns. Clin Oral Investig. 2022 Feb;26(2):1375-89. http://doi.org/10.1007/s00784-021-04112-2 PMid:34401947.
» http://doi.org/10.1007/s00784-021-04112-2 -
15 Lauritano D, Moreo G, Lucchese A, Viganoni C, Limongelli L, Carinci F. The impact of implant-abutment connection on clinical outcomes and microbial colonization: a narrative review. Materials (Basel). 2020 Mar;13(5):1131. http://doi.org/10.3390/ma13051131 PMid:32138368.
» http://doi.org/10.3390/ma13051131 - 16 Associação Brasileira de Normas Técnicas – ABNT. NBR ISO 5832: Implantes para cirurgia - Materiais metálicos. Rio de Janeiro: ABNT; 2008.
- 17 Ayres M, Ayres M Jr, Ayres DL. BioEstat: aplicações estatísticas nas áreas das ciências biomédicas. 5. ed. Belém: Sociedade Civil Mamirauá; 2007.
- 18 da Silva-Neto JP, Nóbilo MA, Penatti MP, Simamoto PC Jr, das Neves FD. Influence of methodologic aspects on the results of implant-abutment interface microleakage tests: a critical review of in vitro studies. Int J Oral Maxillofac Implants. 2012 Jul-Aug;27(4):793-800. PMid:22848880.
-
19 Duraisamy R, Krishnan CS, Ramasubramanian H, Sampathkumar J, Mariappan S, Navarasampatti Sivaprakasam A. Compatibility of nonoriginal abutments with implants: evaluation of microgap at the implant-abutment interface, with original and nonoriginal abutments. Implant Dent. 2019 Jun;28(3):289-95. http://doi.org/10.1097/ID.0000000000000885 PMid:31124826.
» http://doi.org/10.1097/ID.0000000000000885 -
20 Lopes PA, Carreiro AFP, Nascimento RM, Vahey BR, Henriques B, Souza JCM. Physicochemical and microscopic characterization of implant-abutment joints. Eur J Dent. 2018 Jan-Mar;12(1):100-4. http://doi.org/10.4103/ejd.ejd_3_17 PMid:29657532.
» http://doi.org/10.4103/ejd.ejd_3_17 -
21 Hermann JS, Schoolfield JD, Schenk RK, Buser D, Cochran DL. Influence of the size of the microgap on crestal bone changes around titanium implants. A histometric evaluation of unloaded non-submerged implants in the canine mandible. J Periodontol. 2001 Oct;72(10):1372-83. http://doi.org/10.1902/jop.2001.72.10.1372 PMid:11699479.
» http://doi.org/10.1902/jop.2001.72.10.1372 -
22 Adell R, Lekholm U, Rockler B, Brånemark PI, Lindhe J, Eriksson B, et al. Marginal tissue reactions at osseointegrated titanium fixtures (I). A 3-year longitudinal prospective study. Int J Oral Maxillofac Surg. 1986 Feb;15(1):39-52. http://doi.org/10.1016/S0300-9785(86)80010-2 PMid:3083005.
» http://doi.org/10.1016/S0300-9785(86)80010-2 - 23 Jemt T, Book K. Prosthesis misfit and marginal bone loss in edentulous implant patients. Int J Oral Maxillofac Implants. 1996 Sep-Oct;11(5):620-5. PMid:8908860.
-
24 Solá-Ruíz MF, Selva-Otaolaurruchi E, Senent-Vicente G, González-de-Cossio I, Amigó-Borrás V. Accuracy combining different brands of implants and abutments. Med Oral Patol Oral Cir Bucal. 2013 Mar;18(2):e332-6. http://doi.org/10.4317/medoral.18137 PMid:23229250.
» http://doi.org/10.4317/medoral.18137
Publication Dates
-
Publication in this collection
22 Nov 2024 -
Date of issue
2024
History
-
Received
16 Sept 2024 -
Accepted
02 Oct 2024